ZipE Knotless Tissue Repair and Attachment Devices
K162429 · Ziptek, LLC · MAI · Mar 13, 2017 · Orthopedic
Device Facts
Record ID
K162429
Device Name
ZipE Knotless Tissue Repair and Attachment Devices
Applicant
Ziptek, LLC
Product Code
MAI · Orthopedic
Decision Date
Mar 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ZIPE® Knotless Tissue Repair and Attachment Devices are intended for fixation of tissue to bone and tissue to tissue. This product is intended for the following indications: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair. Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair. Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar and Radial Collateral Ligament Reconstruction. Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction.
Device Story
Implantable soft tissue fixation system; comprises resorbable button disc (suture capture), titanium screw or biocomposite bone anchor (plug), UHMWPE braided #2 suture with pre-tied half-hitched protuberances, nitinol wire capture holder, and nitinol wire suture shuttle. Pre-assembled on stainless steel insertion driver with TPE tip protector. Operated by surgeon in clinical setting. Principle: anchor inserted into bone; suture capture pushed down driver over knotted suture; capture locks against suture protuberances, preventing reverse movement; secures tissue to bone or tissue to tissue. Benefits: knotless fixation, simplified surgical technique.
Clinical Evidence
No clinical data provided. Substantial equivalence based on bench testing, including static and cyclic pullout testing, insertion torque testing, degradation testing, and comprehensive biocompatibility and sterilization validation.
Technological Characteristics
Materials: Titanium or biocomposite bone anchor, UHMWPE suture, PLA/caprolactone copolymer capture disc. Components: Suture capture, bone anchor, braided suture, nitinol wire holders, stainless steel insertion driver. Principle: Mechanical locking via suture capture disc over knotted suture protuberances. Sterilization: Ethylene Oxide (EO).
Indications for Use
Indicated for fixation of tissue to bone and tissue to tissue in shoulder, foot/ankle, knee, hand/wrist, and elbow orthopedic procedures requiring soft tissue repair or reconstruction.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device (K151200)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and clean, and it is easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 13, 2017
Ziptek, LLC % Janice Hogan Partner Hogan Lovells Us LLP 1835 Market Street 29th Floor Philadelphia, Pennsylvania 19103
Re: K162429 Trade/Device Name: ZipE® Knotless Tissue Repair and Attachment Devices Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI, KGS Dated: February 10, 2017 Received: February 10, 2017
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K162429
Device Name
ZipE® Knotless Tissue Repair and Attachment Devices
Indications for Use (Describe)
The ZIPE® Knotless Tissue Repair and Attachment Devices are intended for fixation of tissue to tissue.
This product is intended for the following indications:
- Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceos Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- . Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction. Metatarsal Ligament Repair
- Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar and Radial Collateral Ligament Reconstruction
- Elbow: Biceps Tendon Reattachment. Tennis Elbow Repair. Ulnar or Radial Collateral Ligament Reconstruction o
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
Over-The-Counter Use (21 CFR 801
Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug
- Administration
- Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
## 510(k) SUMMARY
## ZIPE® Knotless Tissue Repair and Attachment Devices
| Preparation Date: | February 10, 2017 |
|--------------------|-----------------------------------------------------------------|
| Applicant/Sponsor: | Ziptek, LLC<br>1250 S Tamiami Trl Ste 303<br>Sarasota, FL 34239 |
| Contact Person: | Dr. William F. Bennett<br>(941) 953-5509 |
| Trade Name: | ZipE® Knotless Tissue Repair and Attachment Device |
| Common Name: | Soft Tissue Fixation Device |
#### Classification Name:
- MBI (888.3040) Fastener, fixation, nondegradable, soft tissue ●
- MAI (888.3030) Fastener, fixation, biodegradable, soft tissue
- KGS (878.4930) Retention device, suture
#### Predicate Devices:
| Predicate | 510(k) Number | Device | Manufacturer |
|-----------------------------|---------------|---------------------------------------------------------------|------------------------|
| Primary (metallic anchor) | K974847 | Arthrex Parachute Corkscrew Suture<br>Anchor | Arthrex, Inc. |
| Additional | K071704 | Sleeve and Button Soft Tissue Devices | Biomet Sports Medicine |
| Primary (resorbable anchor) | K071177 | Arthrex Biocomposite Suture Anchors | Arthrex, Inc. |
| Reference (material) | K983843 | Arthrex Bio-Button | Arthrex, Inc. |
| Reference (material) | K151200 | STRATAFIX™ Spiral MONOCRYL™<br>Knotless Tissue Control Device | Ethicon, Inc. |
#### Device Description:
The ZipE® Knotless Tissue Repair and Attachment Devices are implantable soft tissue fixation and suture retention devices comprising of a resorbable pledge('button disc ("suture capture"), a titanium screw or biocomposite bone anchor ("plug"), an ultra-high polyethylene (UHMWPE) braided #2 suture having pre-tied half hitched protuberances along its length, a nitinol wire capture holder and a nitinol wire suture shuttle. All components are preloaded onto a stainless steel insertion driver having a TPE tip protector and a silicone O-ring encircling the diameter of the handle to hold the system in place. The implantable elements of the system are the bone anchor, suture and capture; the other components are part of the driver insertion system.
The capture is designed to interact with the knotted suture by locking with the protuberances on the suture so that once advanced across the protuberance, it cannot move backward. Once the anchor is inserted into the bone, the
{4}------------------------------------------------
pushed down with the stainless steel driver, traveling forward over the knotted suture, preventing movement in the reverse direction, and thereby locking the repair tissue to the underlying tissue.
The ZipE® Knotless Tissue Repair and Attachment Devices come with all components pre-assembled into the insertion driver and ready to be operated by the user. Products may be sold separately.
## Indications for Use:
The ZIPE® Knotless Tissue Repair and Attachment Devices are intended for fixation of tissue to bone and tissue to tissue.
This product is intended for the following indications:
- Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-● Clavicular Separation Repair, Deltoid Repair, Capsulolabral Reconstruction
- . Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valqus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair
- . Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair
- . Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar and Radial Collateral Ligament Reconstruction
- Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction
## Non-clinical Testing:
ZIPTEK LLC performed extensive non-clinical testing for the ZIPE® Knotless Tissue Repair and Attachment Devices, including:
| Bench testing | Static pullout testing<br>Cyclic pullout testing<br>Degradation testing<br>Insert Torque testing |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>testing | Systemic toxicity<br>Reverse mutation<br>Intracutaneous injection<br>Kligman maximization<br>Micronucleus assay<br>Lymphoma mutagenesis<br>Intramuscular implantation<br>Pyrogen test<br>Systemic injection test<br>MEM elution test<br>Neutral red uptake test |
{5}------------------------------------------------
| Sterilization validation | Bioburden<br>EO residual testing<br>Endotoxin (LAL) Validation (*Endotoxin testing is conducted on every batch of the device as part of the release test) |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Package and<br>transportation testing | Transportation validation<br>Packaging testing<br>Accelerated aging |
## Clinical Testing:
None provided as a basis for substantial equivalence.
## Technological Characteristics:
The ZipE® Knotless Tissue Repair and Attachment Devices and the primary predicate Arthrex Parachute Corkscrew IK9748471 are similar in design, have the same intended use and similar indications, technological characteristics and principles of operation. The technological difference between them relies on the locking mechanism of the suture by the suture (button disc). The Sleeve and Button Anchor [K071704] also uses a button disc as a suture retention device as part of the anchoring system. The device anchor plug and the primary predicate Arthrex Biocomposite Anchors [K071177] are material and have the same intended use, similar indications, technological characteristics and principles of operation. The most significant technological difference relies on the eyelet design and suture attachment mechanisms. Some predicate devices have an eyelet and some do not; therefore, this does not represent a new design feature compared to other previously cleared suture anchors. The suture-capture material is made of a PLA/caprolactone is utilized as a copolymer with other suture materials, an example, the STRATAFIX Knotless Tissue Control Device [K151200]. The Arthrex Bio-Button [K983843] and the Arthrex Parachute Corkscrew [K974847] have a surgical button/capture made from the PLA family of materials.
## Substantial Equivalence Summary:
The ZipE® Knotless Tissue Repair and Attachment Devices is substantially equivalent to predicate devices, which have the same intended use and similar technological characteristics. Minor differences between ZipE® Knotless Tissue Repair and Attachment Devices and the predicate devices do not raise different questions concerning safety and effectiveness and do not adversely impact the performance, function, or intended use of the device, as confirmed by performance testing. Thus, the ZipE® Knotless Tissue Repair and Attachment Devices are substantially equivalent to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.