STRATAFIX Spiral MONOCRYL Knotless Tissue Control Device, STRATAFIX Spiral MONOCRYL Plus Knotless Tissue Control Device
K151200 · ETHICON, Inc. · GAM · Aug 13, 2015 · General, Plastic Surgery
Device Facts
Record ID
K151200
Device Name
STRATAFIX Spiral MONOCRYL Knotless Tissue Control Device, STRATAFIX Spiral MONOCRYL Plus Knotless Tissue Control Device
Applicant
ETHICON, Inc.
Product Code
GAM · General, Plastic Surgery
Decision Date
Aug 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Devices and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Devices are indicated for general soft tissue approximation where use of an absorbable suture is appropriate.
Device Story
STRATAFIX™ Spiral MONOCRYL™ and Plus Knotless Tissue Control Devices are monofilament, synthetic absorbable surgical devices; composed of glycolide and ε-caprolactone copolymer. Devices feature unidirectional barbs cut into the suture core to anchor tissue without surgical knots; include a surgical needle on one end and a fixation loop at the other. Plus version incorporates antibacterial agent IRGACARE® MP (triclosan). Used by surgeons in clinical settings for soft tissue approximation. Barbed design allows tissue approximation without knot-tying; provides mechanical anchoring. Benefits include simplified closure and reduced need for manual knotting. Devices are single-use and sterile.
Clinical Evidence
No clinical data. Evidence consists of non-clinical laboratory performance testing, including suture diameter, tensile strength, needle attachment, biocompatibility (intramuscular/subcutaneous implantation, irritation, pyrogenicity), and in-vitro/in-vivo performance (breaking strength retention, wound holding strength, wound healing, bacterial colonization). Testing confirms conformance to USP Monograph for absorbable surgical sutures and FDA Class II Special Controls Guidance.
Technological Characteristics
Materials: Copolymer of glycolide and ε-caprolactone (poliglecaprone 25). Plus version includes triclosan (max 2360 µg/m). Design: Monofilament, unidirectional barbs, variable loop. Sizes: 2-0 to 4-0. Sterilization: Ethylene Oxide. Single-use. Mechanical anchoring principle.
Indications for Use
Indicated for general soft tissue approximation in patients where use of an absorbable suture is appropriate.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
MONOCRYL Plus Antibacterial (poliglecaprone 25) Suture (K050845)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, facing right. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2015
Ethicon Incorporated a Johnson & Johnson company Ms. Donna Marshall Manager, Regulatory Affairs Route 22 West, P.O. Box 151 Somerville, New Jersey 08876
Re: K151200
Trade/Device Name: STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device and STRATAFIX™ Spiral MONOCRYLTM Plus Knotless Tissue Control Device Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/l-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: July 9, 2015 Received: July 10, 2015
Dear Ms. Marshall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{2}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours.
## David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151200
#### Device Name
STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device
#### Indications for Use (Describe)
STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device are indicated for soft tissue approximation where use of an absorbable suture is appropriate.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Traditional 510(k)
STRATAFIX™ Spiral MONOCRYL™ and
STRATAFIX™ Spiral MONOCRYL™ PLUS
Knotless Tissue Control Device Ethicon, Inc.
Image /page/3/Picture/1 description: The image features the logo for ETHICON, INC. The text "ETHICON, INC." is displayed in a bold, red font. Below the company name, there is a smaller tagline that reads "a Johnson & Johnson company" in a cursive font.
# 510(k) Summary
| Submitter: | Ethicon Inc. a Johnson & Johnson company<br>P.O. Box 151<br>Route 22 West<br>Somerville, NJ 08876-0151<br>USA |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Donna Marshall<br>Manager, Regulatory Affairs<br>Ethicon, Inc. a Johnson & Johnson company<br>Ph: (908) 541-3990<br>Fax: (908) 218-2595<br>e-mail: dmarsha2@its.jnj.com |
| Date Prepared: | August 12, 2015 |
| Device Trade Name: | STRATAFIX™ Spiral MONOCRYL™<br>Knotless Tissue Control Device<br>and<br>STRATAFIX™ Spiral MONOCRYL™ Plus<br>Knotless Tissue Control Device |
| Device Common Name: | Suture, Surgical, Absorbable, Polyglcolic Acid |
| Class: | II |
| Classification Name: | Suture, Absorbable, Synthetic, Polyglcolic Acid<br>(21 CFR 878.4493) |
| Product Code: | GAM |
| Predicate Device | 510(k) Number |
|-------------------------------------------------------------------------|---------------|
| MONOCRYLTM Plus Antibacterial (poliglecaprone 25) Suture | K050845 |
| QuillTM MONODERMTM Knotless Tissue Closure Device, Variable Loop Design | K141778 |
| | K123836 |
| | K123409 |
{5}------------------------------------------------
Traditional 510(k) STRATAFIX™ Spiral MONOCRYL™ and STRATAFIX™ Spiral MONOCRYL™ PLUS Knotless Tissue Control Device Ethicon. Inc.
### Device Description:
The STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device is a monofilament, synthetic absorbable device prepared from a copolymer of glycolide and s-caprolactone. Poliglecaprone 25 copolymer has been found to be nonpyrogenic and elicits only a slight tissue reaction during absorption.
The STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device, Variable Loop Design consists of barbed suture material, armed with a surgical needle on one end and a fixation loop at the opposite end. The STRATAFIX™ Spiral MONOCRYL™ Device is designed to anchor with a closed loop at one end and a unidirectional barbed section on the other end. The STRATAFIX™ Spiral MONOCRYL™ Device barbs are oriented in one direction to allow tissue approximation without the need to tie surgical knots.
While the formation of barbs in the STRATAFIX™ Spiral MONOCRYL™ Device reduces the tensile strength relative to non-barbed suture material of the same size, tying of knots in nonbarbed suture materials also reduces their effective strengths. For this reason, the strength of the STRATAFIX™ Spiral MONOCRYL™ Device can be compared to USP knot strength of nonbarbed sutures. USP designations for diameter are used to describe the STRATAFIX™ Spiral MONOCRYL™ Device suture material after barbing, except for minor variation in suture diameter with a maximum overage of 0.1 mm.
The STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device is an antibacterial monofilament, synthetic absorbable device prepared from a copolymer of glycolide and s-caprolactone. The device contains IRGACARE®* MP (triclosan), a broad spectrum antibacterial agent, at no more than 2360 µg/m. Poliglecaprone 25 copolymer has been found to be nonpyrogenic and elicits only a slight tissue reaction during absorption.
The STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device. Variable Loop Design consists of barbed suture material, armed with a surgical needle on one end and a fixation loop at the opposite end. The STRATAFIX™ Spiral MONOCRYL™ Plus Device is designed to anchor with a closed loop at one end and a unidirectional barbed section on the other end. The STRATAFIX™ Spiral MONOCRYL™ Plus Device barbs are oriented in one direction to allow tissue approximation without the need to tie surgical knots.
While the formation of barbs in the STRATAFIX™ Spiral MONOCRYL™ Plus Device reduces the tensile strength relative to non-barbed suture material of the same size, tying of knots in nonbarbed suture materials also reduces their effective strengths. For this reason, the strength of the STRATAFIX™ Spiral MONOCRYL™ Plus Device can be compared to USP knot strength of non-barbed sutures. USP designations for diameter are used to describe the STRATAFIX™ Spiral MONOCRYL™ Plus Device suture material after barbing, except for minor variation in suture diameter with a maximum overage of 0.1 mm.
{6}------------------------------------------------
### Indications for Use:
STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Devices and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Devices are indicated for general soft tissue approximation where use of an absorbable suture is appropriate.
### Performance Data:
Non-clinical laboratory performance testing was performed demonstrating that STRATAFIX™ Spiral MONOCRYLTM Knotless Tissue Control Devices and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Devices conforms to the current USP Monograph for absorbable surgical sutures, except for diameter. This testing was performed in accordance with FDA's Class II Special Controls Guidance Document: Surgical Sutures, Issued June 3. 2003. In addition, bench and animal testing was provided showing that the device performed as intended and as claimed.
Below you will find a list of non-clinical performance data completed for the STRATAFIX™ Spiral MONOCRYLTM Knotless Tissue Control Devices and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Devices:
| Bench Testing | Biocompatiblity | In-vitro/In-vivo |
|-------------------|----------------------------------------|-----------------------------|
| Suture Diameter | Intramuscular Implantation | Breaking Strength Retention |
| Tensile Strength | Subcutaneous Implantation | Wound Holding Strength |
| Needle Attachment | Irritation (Intracutaneous Reactivity) | Wound Healing |
| | Pyrogenicity | Bacterial Colonization |
## Substantial Equivalence:
STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Devices and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Devices have the same intended use and similar indications for use as the predicate devices. The technological differences between the subject devices, STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device and the predicate devices raise no new questions of safety or effectiveness. STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Devices and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Devices met all testing criteria to demonstrate substantial equivalence to the predicates devices. The following table compares the subject devices to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
{7}------------------------------------------------
Traditional 510(k) STRATAFIX™ Spiral MONOCRYL™ and STRATAFIX™ Spiral MONOCRYL™ PLUS Knotless Tissue Control Device Ethicon, Inc.
## Comparison Table:
### Comparison of Subject Device Characteristics to Predicate Devices
| Characteristic | Subject Device:<br>STRATAFIXTM Spiral<br>MONOCRYLTM Knotless<br>Tissue Control Device | Subject Device: STRATAFIXTM<br>Spiral MONOCRYLTM Plus Knotless<br>Tissue Control Device | Predicate Device:<br>MONOCRYL Plus<br>Antibacterial Sutures<br>(K050845) | Predicate Device:<br>QuillTM Knotless Tissue<br>Closure Device<br>(Variable Loop Design)<br>Comprised of<br>MONODERMTM<br>(PGA-PCL)<br>(K141778, K123836, K123409) |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | STRATAFIXTM Spiral<br>MONOCRYLTM Knotless<br>Tissue Control Device is<br>indicated for use in soft tissue<br>approximation where the use<br>of absorbable sutures is<br>appropriate. | STRATAFIXTM Spiral<br>MONOCRYLTM Plus Knotless Tissue<br>Control Device is indicated for use in<br>soft tissue approximation where the use<br>of absorbable sutures is appropriate. | MONOCRYL Plus<br>Antibacterial sutures are<br>indicated for use in general<br>soft tissue approximation<br>and/or ligation, but not for use<br>in cardiovascular or<br>neurological tissues,<br>microsurgery or ophthalmic<br>surgery. | QuillTM MONODERM device is<br>indicated for use in soft tissue<br>approximation where use of<br>absorbable sutures is<br>appropriate. |
| Technological means for<br>achieving intended use | Monofilament strand to<br>approximate tissue anchored<br>without the need to tie surgical<br>knots. | Monofilament strand to approximate<br>tissue anchored without the need to tie<br>surgical knots. | Monofilament strand to<br>approximate tissue anchored<br>with surgical knots. | Monofilament strand to<br>approximate tissue anchored<br>without the need to tie surgical<br>knots. |
| Absorbable/<br>Suture Material | MONOCRYL is composed of<br>a copolymer of glycolide and<br>ε-caprolactone | MONOCRYL is composed of a<br>copolymer of glycolide and ε-<br>caprolactone | MONOCRYL is composed of<br>a copolymer of glycolide and<br>ε-caprolactone | MonodermTM (PGA-PCL) –<br>composed of a copolymer of<br>glycolide and e-caprolactone |
| Suture Design | Monofilament strand. Single<br>armed, unidirectional barbs<br>made by cutting into the core<br>of the suture strand. Contains<br>adjustable loop. | Monofilament strand. Single armed,<br>unidirectional barbs made by cutting<br>into the core of the suture strand.<br>Contains adjustable loop. | Monofilament strand | Uni-directional barbs along the<br>long axis of the monofilament |
| Colorant (if dyed) | Undyed | Undyed | Undyed or Dyed (D&C Violet<br>No.2) | Undyed or Violet (D&C Violet<br>No.2) |
| USP Suture Size | 2-0 through 4-0 | 2-0 through 4-0 | 0, 1, 2-0 through 6-0 | 0, 2-0 through 5-0 |
| Tensile Strength | Straight Tensile meets USP | Straight Tensile meets USP Knot | Meets USP requirement for | Straight Tensile meets USP Knot |
| | Knot Tensile requirements for<br>Synthetic Absorbable Sutures<br>of the same size equivalent | Tensile requirements for Synthetic<br>Absorbable Sutures of the same size<br>equivalent | Synthetic Absorbable Sutures | Tensile requirements for<br>Synthetic Absorbable Sutures of<br>one size smaller equivalent |
| Needle Pull-Off | Meets USP requirements | Meets USP requirements | Meets USP requirements | Meets USP requirements |
| Antibacterial Agent | Not Applicable | Irgacare MP (Triclosan) | Irgacare MP (Triclosan) | Not Applicable |
| Maximum Level of<br>Antibacterial Agent | Not Applicable | < 2360 µg/m | < 2360 µg/m | Not Applicable |
| Approximate % Breaking<br>Strength Retention (BSR) | 62% @ 7 days<br>27% @ 14 days | 62% @ 7 days<br>27% @ 14 days | 50 – 60% @ 7 days<br>20 – 30% @ 14 days<br>(undyed) | 62% @ 7 days<br>27% @ 14 days |
| Absorption Profile | Essentially absorbed by<br>91days post-implantation | Essentially absorbed by 91days post-<br>implantation | Essentially complete between<br>91 and 119 days | Essentially complete between 90<br>and 120 days |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Single Use? | Yes | Yes | Yes | Yes |
| Packaging Configuration | Plastic tray with paper lid<br>(contains the suture) placed<br>within foil pouch (sterile<br>barrier). Sterile foil pouches<br>are packaged in a sales carton. | Plastic tray with paper lid or paper<br>folder (contains the suture) placed<br>within foil pouch (sterile barrier).<br>Sterile foil pouches are packaged in a<br>sales carton. | Paper folder or plastic tray<br>within a foil pouch. Pouches<br>placed within sales carton | Device wound onto inner<br>support card, within a Single<br>Barrier Foil Pouch |
{8}------------------------------------------------
{9}------------------------------------------------
Traditional 510(k) STRATAFIX™ Spiral MONOCRYL™ and STRATAFIX™ Spiral MONOCRYL™ PLUS
Knotless Tissue Control Device Ethicon, Inc.
### Conclusion:
Based on the intended use, technological characteristics, safety and performance testing STRATAFIX™ Spiral MONOCRYL™ Knotless Tissue Control Device and STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device have been shown to be appropriate for its intended use and are considered to be substantially equivalent to the predicate devices.
* Trademark
IRGACARE®* MP (triclosan) "Registered Trademark of BASF Group"
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.