The Destiknee™- Total Knee System is indicated for patients with severe knee pain and disability due to: . Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis . Correction of functional deformities. . . Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy. . Moderate valgus, varus, or flexion trauma. . . Knee fractures untreatable by other methods. . The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.
Device Story
Destiknee Total Knee System; cemented, single-use orthopedic prosthesis. Components: femoral component (Co-Cr-Mo alloy), tibial component (metal-backed or all-poly), and all-poly patella component. Used in orthopedic surgery to replace articulating surfaces of distal femur and proximal tibia. Measured resection technique. Surgeon implants device to restore joint function, alleviate pain, and correct deformities. Benefits: improved mobility and pain relief for patients with severe joint degeneration or trauma.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included endurance properties (cyclical testing), stability, surface stress, and kinematic range of motion.
Technological Characteristics
Materials: Cobalt-Chromium-Molybdenum alloy (ASTM F-75) and UHMWPE (ASTM F648). Designs: Posterior Stabilized (PS) and Cruciate Retaining (CR). Configurations: Modular (metal-backed) or non-modular (all-poly) tibial components. Cemented application. Single-use. No software or electronic components.
Indications for Use
Indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, functional deformities, post-traumatic loss of knee joint contour (patellofemoral erosion/dysfunction/prior patellectomy), moderate valgus/varus/flexion trauma, or knee fractures untreatable by other methods.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2016
Meril Healthcare Private Limited % Jon Ward President and CEO AJW Technology Consultants, Inc. 445 Apollo Beach Blvd. Apollo Beach, Florida 33572
Re: K160771
Trade/Device Name: Destiknee Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: November 14, 2016 Received: November 14, 2016
Dear Jon Ward:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 4. INDICATIONS FOR USE STATEMENT
### Indications for Use
510(k) Number (if known): K160771
## Device Name: Destiknee™- Total Knee System
Indications for Use: The Destiknee™- Total Knee System is indicated for patients with severe knee pain and disability due to:
- . Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis
- Correction of functional deformities. .
- . Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
- Moderate valgus, varus, or flexion trauma. .
- . Knee fractures untreatable by other methods.
The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.
Prescription Use: X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use: (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IFNECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Meril" in a sans-serif font. The letters are white against a light blue background. There is a small yellow dot above the "i" in "Meril," which replaces the typical dot.
## 5. 510(k) SUMMARY
## 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 92.
#### 5.1 Applicant:
Meril Healthcare Private Limited First Floor, H1-H3, Meril Park, Survey No. 135/2/B & 174/2, Muktanand Marg, Chala, Vapi - 396 191, Gujarat, INDIA
#### 5.2 Contact Person:
Umesh Sharma Deputy General Manager (DGM) Meril Healthcare Private Limited H1-H3, Meril Park, Survey No. 135/2/B & 174/2, Muktanand Marg, Chala, Vapi - 396 191, Gujarat, INDIA Phone: +91 - 2603063811 Mobile: +91 - 9904805571 Fax: +91 260-3052125 E mail: umesh.sharma@merillife.com
#### 5.3 Regulatory Correspondent:
Jon Ward President and CEO AJW Technology Consultants, Inc. 445 Apollo Beach Blvd. Apollo Beach, FL 33572 Phone: 813-645-2855 x104 Fax: 813-645-2856 Email: wardjp(@ajwtech.com
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Meril" in a sans-serif font on a blue background. The "i" in Meril is dotted with two circles, one yellow and one white. The word is centered in the image and takes up most of the space.
ર્સ્ડ
| 5.4 | Date prepared: |
|-----|----------------|
|-----|----------------|
14Dec2016
| Device information: | |
|----------------------|--------------------------------------------------------------------------------------------------|
| Proprietary Name: | Destiknee™ Total Knee System |
| Common / Usual Name: | Total Knee Replacement |
| Regulation name: | Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis. |
| Review Panel | Orthopaedic |
| Regulation Number: | 21 CFR 888.3560 |
| Product Code: | JWH |
| Device Class: | Class II |
#### 5.6 Predicate Device
- a. The Freedom® Total Knee System (K082019, K090411, K091280)
#### 5.7 Device Description:
The Destiknee™ Total Knee System consists of following components.
- . Femoral Component
- . Tibial Component which is available in following two designs
- o Metal Backed Tibial Component which consists of Tibial Base Plate and Tibial Articular Surface (viz. Tibial Liner)
- All Poly Tibial Component (viz. All Poly) and o
- All Poly Patella Component .
Each of these components are described below.
#### 1. Femoral Component
Femoral Component is fabricated from Cobalt-Chromium-Molybdenum alloy and is intended for cemented application to replace the articulating surface of the distal femur in a measured resection technique.
{5}------------------------------------------------
# Mer
The Femoral Component is available in two designs viz. Posterior Stabilized (PS) and Cruciate Retaining (CR).Each design is classified into Left and Right configurations. Each Left and Right configuration is available in eight different sizes.
#### 2. Tibial Component
Tibial Component is available in following two designs:
- a) Metal Backed (modular/two pieces)
- b) All Poly (non-modular/single piece)
Metal backed design of Tibial component is modular two piece design viz. Tibial Base Plate and Tibial Articular Surface (viz. Tibial Liner).
The Tibial Base Plate is fabricated from Cobalt-Chromium-Molybdenum alloy and is intended for use with the Tibial Liner for cemented application to replace the articulating surface of the proximal tibia in a measured resection.
Tibial Liner is fabricated from industry standard Ultra High Molecular Weight Polyethylene (UHMWPE) and is intended for use with the Tibial Base Plate to replace the articulating surface of the proximal tibia in a measured resection. It has a locking mechanism on the peripheral edges of the distal surface to lock into the Tibial Base Plate. The Tibial Liner is available in two designs viz. Posterior Stabilized (PS) and Cruciate Retaining (CR).Each PS and CR designs are available in 14 configurations. Each configuration is further available in four different thickness viz. 9mm, 11mm, 14mm and 17mm.
All poly design is a non modular single piece design viz. All Poly Tibial Component (viz. All Poly)fabricated from industry standard Ultra High Molecular Weight Polyethylene (UHMWPE) and is intended for cemented application to replace the articulating surface of the proximal tibia in a measured resection. All Poly is available in two designs viz. Posterior Stabilized (PS) and Cruciate Retaining (CR).
Each PS and CR designs are available in 12 configurations. Each configuration is available in five different thickness viz. 9mm, 11mm, 14mm, 17mm and 20mm.
{6}------------------------------------------------
All Poly Patella Component is fabricated from industry standard Ultra High Molecular Weight Polyethylene (UHMWPE). It is available in a symmetric single radius design. The symmetrical design is available in 4 diameters.
All materials used in Destiknee™ Total Knee System meet ASTM standards for implants, viz. ASTM F648 "Standard Specification for Ultra High Molecular Weight Polyethylene Powder and Fabricated Form for Surgical Implants" and ASTM F-75 "Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting alloy for Surgical Implants (UNS R30075)".
#### 5.8 Indications for Use:
The Destiknee™ Total Knee System is indicated for patients with severe knee pain and disability due to:
- . Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities. .
- Post-traumatic loss of knee joint contour, particularly when there is patellofemoral . erosion, dysfunction, or prior patellectomy.
- Moderate valgus, varus, or flexion trauma. .
- Knee fractures untreatable by other methods. .
The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.
#### Comparison of Technological characteristics: રું. તે જેવી સાંતર તાલુકામાં આવેલું એક ગામનાં મુખ્યત્વે ખાતે પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખે
The Destiknee™ Total Knee system is identical to the predicate device with respect to intended use, device design, materials, and method of sterilization.
All the necessary performance tests to determine the material and mechanical characteristics have been performed on predicate device viz. Freedom® Total Knee System.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Meril" in a sans-serif font against a blue background. The letters are white, except for a small yellow dot above the "i" and a white dot to the right of the yellow dot. The overall design is simple and clean, with a focus on readability and a modern aesthetic.
## 5.10 Non clinical Performance data:
Destiknee™ Total Knee System is identical to the predicate device viz. Freedom® Total Knee System with respect to intended use, device design, materials, and method of sterilization.
All necessary performance tests to determine the material and mechanical characteristics have been performed on identical device. Performance testing includes: endurance properties (cyclical testing), stability, surface stress, kinematic range of motion.
There have been no adverse event reports associated with the Destiknee™M Total Knee System and no recalls have been initiated.
## 5.11 Conclusion
The Destiknee™ Total Knee System is identical to predicate device with respect to intended use, device design, materials and method of sterilization. Therefore, Destiknee™ Total Knee System is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.