Internal Joint Stabilizer - Elbow

K153208 · Skeletal Dynamics, LLC · OZI · Dec 31, 2015 · Orthopedic

Device Facts

Record IDK153208
Device NameInternal Joint Stabilizer - Elbow
ApplicantSkeletal Dynamics, LLC
Product CodeOZI · Orthopedic
Decision DateDec 31, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Internal Joint Stabilizer – Elbow is intended to provide temporary stabilization of the elbow joint after trauma or chronic elbow dislocation.

Device Story

Internal elbow fixator provides temporary joint stabilization post-trauma or chronic dislocation. Device comprises base plate assembly (titanium alloy/cobalt chrome), connecting rod, boom, locking joints, and cobalt chrome axis pin. Implanted via surgical procedure using specialized instrumentation (trajectory guides, K-wires, drills). Device bridges anconeous muscle; axis pin inserted into humerus rotation axis; base plate attached to ulna via compression screws. Intended for removal after 6-8 weeks once ligaments heal. Used in clinical/surgical settings by orthopedic surgeons. Provides mechanical stability equivalent to external hinged fixators; facilitates functional recovery; reduces need for external fixation hardware.

Clinical Evidence

Prospective, multi-center, non-randomized cohort study (n=26). Primary endpoints: Broberg Morrey Functional Rating (fair or better) and absence of recurrent dislocation at 6 months post-explant. Results: 23/24 (95.83%) achieved functional success; 23/24 (95.83%) had no recurrent dislocations. Safety objective met; device-related adverse events (n=1 axis pin back out) comparable to retrospective external fixator cohort.

Technological Characteristics

Materials: Titanium alloy and cobalt chrome. Components: Base plate assembly, connecting rod, boom, locking joints, axis pin, compression screws. Mechanical fixation principle. Provided non-sterile; user-facility sterilization required. Instrumentation includes trajectory guides, K-wires, depth gauges, drills, counter-torque tool, T-10 driver.

Indications for Use

Indicated for patients requiring temporary stabilization of the elbow joint following trauma or chronic elbow dislocation.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top of the seal in a circular fashion. The seal is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 31, 2015 Skeletal Dynamics, LLC % John Smith, M.D., J.D. Regulatory Counsel Hogan Lovells US LLP 555 Thirteenth Street, N.W. Washington, District of Columbia 20004 Re: K153208 Trade/Device Name: Internal Joint Stabilizer - Elbow Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: OZI, LXT Dated: November 4, 2015 Received: November 5, 2015 Dear Dr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K153208 Device Name Internal Joint Stabilizer - Elbow Indications for Use (Describe) The Internal Joint Stabilizer – Elbow is intended to provide temporary stabilization of the trauma or chronic elbow dislocation. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary #### Submitter: Skeletal Dynamics, LLC 8905 SW 87 Avenue, Suite 201 Miami, FL 33176 Phone: (305) 596-7585 Facsimile: (305) 596-7591 Contact person: Ana M. Escagedo Date Prepared: December 30, 2015 Establishment Registration Number: 3006742481 ### Name and Classification: | Name: | Internal Joint Stabilizer – Elbow | |-----------------|-------------------------------------------------------------------------------------------| | Common Name: | Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component,<br>Metal Composite | | Classification: | 21 C.F.R. § 888.3030 | | Product Code: | OZI; LXT | | Class: | II | #### Predicate Device: Howmedica (now Stryker) Dynamic Joint Distractor II (K002923) ### Device Description - Technological Characteristics: The Skeletal Dynamics Internal Joint Stabilizer – Elbow is an internal elbow fixator, with three principal components: - A Base Plate assembly consisting of a low profile Base Plate, contoured to fit within . the flat surface of ulna at the olecranon process; a Connecting Rod, a Boom and two Locking Joints designed to bridge over the anconeous muscle. The Connecting Rod and Boom are adjustable by means of the Locking Joints to accommodate different patient anatomies. The Base Plate assembly is made of titanium alloy and cobalt chrome. - A cobalt chrome Axis Pin threads through the Boom of the Base Plate assembly and ● is inserted in the axis of rotation of the humerus, and - . Titanium compression screws to attach the Base Plate Assembly to the ulna. Specialized instrumentation includes Axis Trajectory Guides (available in 3 sizes), K- Wire Guide, K-Wires, Depth Gauges, Drills, a Counter-torque tool, and a T-10 driver and quick connect handle. The Internal Joint Stabilizer - Elbow is intended to be removed once the elbow ligaments have healed, usually within 6 - 8 weeks after implantation. {4}------------------------------------------------ The system is provided non-sterile and is sterilized in the user facility. ## Intended Use / Indications for Use The Internal Joint Stabilizer – Elbow is intended to provide temporary stabilization of the elbow joint after trauma or chronic elbow dislocation. ## Summary of Substantial Equivalence: The substantial equivalence of the Internal Joint Stabilizer – Elbow to the predicate device is demonstrated by the same intended use, and similar indications for use, materials, design (fundamental scientific technology), performance, sterility, and packaging and does not present any new or different issues of safety and effectiveness. ## Performance Data: Preclinical and clinical analysis and testing demonstrated that the Internal Joint Stabilizer – Elbow is substantially equivalent to the predicate device. Mechanical testing, which established equivalency, included the testing described below: - Fatigue Evaluate loosening with repetitive physiological loading. . - Stability - Evaluate IJS-E joint in comparison to Dynamic Joint Distractor II. - Component Strength Evaluate grip, torsional and bending strength of the joint . components in loading. - . Cadaveric Testing - Demonstrate that the IJS can be consistently implanted and removed in accordance with the surgical technique, and evaluate whether there is soft tissue impingement when used. ## Clinical Study: Clinical testing was also performed in a prospective, multi-center, non-randomized, cohort study, conducted at 6 sites in the US in which 26 subjects were enrolled (26 consented and screened). The purpose of the Internal Joint Stabilizer - Elbow study was to demonstrate safety and effectiveness of the Internal Joint Stabilizer - Elbow. The primary efficacy objective was to confirm that the IJS-E provides temporary stabilization of the elbow joint and allows functional recovery after trauma or chronic dislocation as effectively as hinged external fixation. The primary safety objective was to confirm that the IJS-E does not contribute to a greater number of device related adverse events than hinged external fixation. Subject outcomes were compared to a retrospective cohort of subjects who underwent external hinged fixation with FDA cleared external fixators at a major metropolitan hospital in the US. A subject was an efficacy success if at six months post explant: the Broberg Morrey Functional Rating was fair or better; and no recurrent dislocation occurred during use of the Internal Joint Stabilizer – Elbow or after its removal. Safety success examined device related adverse events. Twenty-four (24) out of 26 subjects completed the study; two subjects were lost to follow up. Out of the subjects who completed the study, 23 of the 24 (95.83%) had a Broberg Morrey Functional Rating of Fair or better (95% Cl: 78.87%, 99.89%). Out of the 24 subjects who {5}------------------------------------------------ completed the study, 23 subjects did not have recurrent dislocations (95% Cl: 78.87%, 99.89%). The primary safety objective was also met since the Internal Joint Stabilizer – Elbow did not contribute to a greater number of device related adverse events than the hinged external fixator studied in the retrospective data review. The only device related Adverse Event experienced by the test subject was back out of an Axis Pin (n=1). Other Adverse Events related to the study experienced by study participants included malreduction (N=1). The Internal Joint Stabilizer - Elbow study demonstrates that the IJS-E will perform equivalently to currently marketed external joint fixation devices in providing temporary stabilization of the elbow joint after trauma or chronic elbow dislocation. ### Conclusion: Based on the testing described above evaluating the differences between the subject and predicate device, Skeletal Dynamics concludes that Internal Joint Stabilizer - Elbow is substantially equivalent to the predicate device identified in this premarket notification.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...