K152869 · Tearscience, Inc. · HKI · Jan 14, 2016 · Ophthalmic
Device Facts
Record ID
K152869
Device Name
LipiView II Ocular Surface Interferometer
Applicant
Tearscience, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Jan 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
The LipiView II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of: · Specular (interferometric) observations of the tear film. Using these images, LipiView® II measures the absolute thickness of the tear film lipid layer. • Meibomian glands under near-infrared (NIR) illumination · The ocular surface and eyelids under white illumination All of these image types can be photographically documented and visually monitored.
Device Story
Bench-top ophthalmic imaging device; captures digital images/video of tear film, meibomian glands, and ocular surface. Inputs: specular reflection (white light interferometry) for tear film; NIR illumination for meibomian glands; white light for ocular surface/eyelids. Operates via physician-controlled touchscreen interface; includes motion stage, camera, and handheld NIR lid everter. Outputs: visual images/video on display, printed reports, or exported files. Used in-office by physicians to monitor ocular surface health; does not provide diagnosis; no treatment/life-sustaining function. Benefits: enables visualization and documentation of ocular structures to assist clinical assessment.
Clinical Evidence
Bench testing only. Performance testing included usability testing, software validation, and verification of compliance with ISO 15004-2 (light hazard), ANSI/IEC 60601 (electrical safety/EMC), and UL 94V-1 (flammability). No clinical data required.
Technological Characteristics
Bench-top ophthalmic camera; Windows-based OS; touchscreen GUI. Materials: flame retardant per UL 94V-1. Illumination: Class I white LED, NIR LED. Connectivity: USB, wireless network. Sterilization: alcohol disinfection for patient contact parts (chin/forehead rest). Standards: ISO 15004-2, ANSI/IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for adult patients; physician use for capturing, archiving, manipulating, and storing digital images of tear film lipid layer thickness, meibomian glands (via NIR), and ocular surface/eyelids (via white light).
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a wing-like shape above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 14, 2016
Tearscience, Inc. % Dr. Christy Coleman Vice President, Clinical, Quality and Regulatory Affairs 5151 McCrimmon Pkwy., Suite 250 Morrisville, North Carolina 27650
Re: K152869
Trade/Device Name: Lipiview II Ocular Surface Interferometer Regulation Number: 21 CFR 886.1120 Regulation Name: Opthalmic Camera Regulatory Class: Class II Product Code: HKI, HJO Dated: December 8, 2015 Received: December 9, 2015
Dear Dr. Coleman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152869
Device Name
LipiView® II Ocular Surface Interferometer
Indications for Use (Describe)
The LipiView II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of:
· Specular (interferometric) observations of the tear film. Using these images, LipiView® II measures the absolute thickness of the tear film lipid layer.
• Meibomian glands under near-infrared (NIR) illumination
· The ocular surface and eyelids under white illumination
All of these image types can be photographically documented and visually monitored.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for TearScience. The logo features a stylized blue eye on the left, followed by the word "TearScience" in a blue, slightly italicized font. A registered trademark symbol is present to the right of the word.
# 510(k) SUMMARY
| PREPARATION DATE: | September 29, 2015 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | TearScience, Inc.<br>5151 McCrimmon Parkway, Suite 250<br>Morrisville, NC 27560 |
| | Tel: (919) 467-4007<br>Fax: (919) 467-3300 |
| CONTACT PERSON: | Christy Coleman, OD, MPH<br>Vice President, Clinical, Quality and Regulatory Affairs |
| DEVICE TRADE NAME: | LipiView® II Ocular Surface Interferometer |
| COMMON NAME: | Ophthalmic Imaging Device |
| CLASSIFICATION NAME: | Ophthalmic Camera |
| DEVICE CLASSIFICATION: | Class II; 21 CFR 886.1120 and 21 CFR 886.1850 |
| PRODUCT CODE: | HKI, HJO |
| PREDICATE DEVICES: | PRIMARY: LipiView® Ocular Surface Interferometer<br>Class II under 21 CFR 886.1120 and 21 CFR 886.1850;<br>Product Code HKI, HJO; Applicant: TearScience, Inc.;<br>Originally cleared under K091935 on October 23, 2009;<br>Expanded Indications for Use cleared under K122481 on<br>December 31, 2012<br><br>SECONDARY: Optomed Smartscope M5 Digital Camera<br>Class II under 21 CFR 886.1120; Product Code HKI;<br>Applicant: Optomed Oy.;<br>Cleared under K132186 on December 18. 2013 |
#### DEVICE DESCRIPTION:
The LipiView® II Ocular Surface Interferometer is a bench-top device used as an ophthalmic camera for imaging the lipid layer of the tear film, meibomian glands, ocular surface and eyelids. There are three imaging modes: Lipid, Gland and Ocular. LipiView II contains a computer system and electronics, two position chin and forehead rest, camera and lens, motion stage, illuminator and touch screen display. LipiView II includes pushbutton controls and handheld near-infrared lid everter for gland imaging.
{4}------------------------------------------------
In the Lipid Imaging mode, LipiView II operates on the principle of white light interferometry and provides an interferometry color assessment of the tear film by specular reflection. The computer system captures a video image file that is recorded over time since the interference pattern changes as the tear film is distributed across the cornea during blinking. The video image of the ocular surface may be viewed on the computer screen display, in a printed report, or captured on video and exported to USB-attached storage or a mapped network drive.
In the Gland Imaging Mode, Lipi View II uses near infrared (NIR) illumination and an NIR-sensitive camera mounted on a motion control system to image the meibomian glands. The tissue between the meibomian glands and the surface of the eyelid are transparent to NIR light. The glands reflect NIR wavelengths, allowing them to be imaged.
In the Ocular Imaging mode, LipiView II captures high-resolution still images or video to record relevant findings on the ocular surface and/or eyelids. The images can be taken using room light only or the device's built-in white light.
The LipiView II Interferometer does not provide a diagnosis. The device has no treatment or no life-sustaining functions, and presents no potential for serious risk to the health, welfare or safety of patients.
## INTENDED USE:
The LipiView® II Ocular Surface Interferometer is intended to image the eye and surrounding area. The LipiView II Interferometer is a prescription device for use by a physician during an in-office exam.
Indications for Use: The LipiView II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of:
- . Specular (interferometric) observations of the tear film. Using these images, Lipi View II measures the absolute thickness of the tear film lipid layer.
- Meibomian glands under near-infrared (NIR) illumination
- . The ocular surface and eyelids under white illumination
All of these image types can be photographically documented and visually monitored.
As an ophthalmic camera, LipiView II has the same intended use to image the eye and surrounding area as the predicate devices, LipiView Interferometer (K091935, K122481) and Optomed Smartscope M5 Digital Camera (K132186). The Indications for Use for LipiView II are similar to the predicate LipiView device, except that the parts of the eye and surrounding area being imaged have been expanded to include the meibomian glands, ocular surface and eyelids. The expanded Indications for Use do not alter the intended use of the LipiView II Interferometer. LipiView II does not provide a diagnosis, and imaging other parts of the eye does not affect the safety and effectiveness of the device.
{5}------------------------------------------------
### TECHNOLOGICAL CHARACTERISTICS:
The LipiView II Interferometer has the same fundamental scientific technology as the predicate devices. As summarized below, most of the technological characteristics of the LipiView II Interferometer remain unchanged from the predicate LipiView device. In addition, LipiView II and Optomed Smartscope M5 with optics module ES2 use similar visible and infrared LED illumination to image the eye. A comparison of the technology between LipiView II and each of the predicate devices is summarized below.
Comparison to LipiView Interferometer: The predicate LipiView device and LipiView II have the same principle of operation as an ophthalmic camera to image the eye and for specular observations of the tear film using white light interferometry. Both bench-top devices have similar overall dimensions and components, including computer system and electronics, chin and forehead rest, camera and lens, motion stage, illuminator and touch screen display. Both devices may be used with off-the-shelf USB accessories and operate wirelessly with networks.
Similar design features include: adjustable forehead and chin rest with the same patient contact material and disinfection method; digital video camera adjusted using a motion stage; white light Class I LED illuminator; and a visible LED fixation light. Both devices comply with ISO15004-2 standard for light hazard protection. Both devices have an AC power source in compliance with ANSI/IEC 60601 standards for electrical safety and electromagnetic compatibility (EMC). Both devices have flame retardant materials per UL 94V-1 standard. Both devices have ambient temperature of parts of device held by operator or accessible to patient in compliance with ISO 15004-1 standard.
The software on both devices runs on a Microsoft Windows-based operating system with a touchscreen display graphical user interface. Similar software design features include: password-protected user login: patient entry into a database: real-time video display to acquire and view images; touchscreen user controls for camera and video playback; image acquisition process with storage of lossless AVI format video images; printed or saved reports; and export of images to USB external media or electronic medical record system via wireless network. In addition, both devices use the same tear film lipid layer image processing for analysis and measurement of the lipid layer thickness.
Different design features include new hardware components in LipiView II: pushbutton controls, a two-position adjustable chin rest, a high definition camera, surface NIR illumination, and a handheld NIR lid everter. These design changes are to support imaging other parts of the eye and surrounding area, such as the meibomian glands. The primary software changes in LipiView II are an update to the Windows-based operating system, new graphical user interface appearance; and new meibomian gland imaging and ocular imaging functionality. Other minor hardware and software changes relative to the predicate LipiView device were made for user convenience or product enhancement.
Comparison to Optomed M5 Digital Camera: The Optomed M5 Digital Camera with ES2 optics module and LipiView II have the same principle of operation as an ophthalmic camera for digital imaging of the eye. Both devices have a camera, lens, visible and infrared LED illumination and visible LED fixation light. Both devices
{6}------------------------------------------------
comply with ISO15004-2:2007 standard for light hazard protection. Both devices have a power source in compliance with ANSI/IEC 60601-1-1 and 60601-1-2 standards for electrical safety and EMC. Also, both devices have display systems to view images and both provide image storage capability on external media.
LipiView II is a bench-top device with a computer system, software and a touchscreen display graphical user interface. Conversely, Optomed Smartscope M5 with optic module ES2 consists of a camera handset, attachable ophthalmic lens, cradle for charging and image transfer to a PC by USB connection. LipiView II has an integrated chin rest to support the patient's head, whereas. Optomed Smartscope utilizes an adapter to a slit lamp biomicroscope, which has support for the patient's head.
#### PERFORMANCE TESTING:
Performance testing demonstrates that the design changes in LipiView II do not raise new questions of safety and effectiveness, and LipiView II is substantially equivalent to the predicate devices.
- Usability testing and software validation demonstrate that LipiView II is as safe O and as effective as the predicate devices in the intended use to image the eye and surrounding area, including meibomian glands, ocular surface and eyelids.
- As for the predicate devices, LipiView II meets the standards for light hazard o protection, electrical safety and electromagnetic compatibility and material flammability. Therefore, the use of near infrared illumination does not raise new safety or effectiveness questions.
- The new hardware features in LipiView II, including two-position chin rest, high o definition camera, NIR illumination and handheld NIR lid everter, meet all design requirements for safety and performance, and therefore, do not adversely affect safety and effectiveness as compared to the predicate device.
## CONCLUSIONS:
The LipiView II Ocular Surface Interferometer has the same intended use and the same fundamental scientific technology as the predicate devices. The expanded Indications for Use for LipiView II relative to the predicate devices do not alter the intended use and do not affect safety and effectiveness of the device. Performance testing demonstrates the LipiView II Interferometer is substantially equivalent in intended use and technological characteristics to the predicate devices. The technology differences do not raise new questions of safety and effectiveness, and do not adversely affect safety and effectiveness of the device. Based on overall design verification and validation and bench performance testing, LipiView II is at least as safe and effective as the predicate devices.
{7}------------------------------------------------
| TABLE 5-1. PREDICATE DEVICE COMPARISON TABLE | | | |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison Feature | LipiView II Interferometer | Predicate LipiView Interferometer<br>(K091935 and K122481) | Optomed Smartscope M5<br>Digital Camera (K132186) |
| Intended Use | Intended to take photographs of the eye and the<br>surrounding area | Intended to take photographs of the eye and the<br>surrounding area | Intended to take photographs of<br>the eye and the surrounding area |
| Indications for Use | The LipiView II Ocular Surface Interferometer is<br>an ophthalmic imaging device intended for use by<br>a physician in adult patients to capture, archive,<br>manipulate and store digital images of:<br>• Specular (interferometric) observations of the tear<br>film. Using these images, LipiView II measures<br>the absolute thickness of the tear film lipid layer.<br>• Meibomian glands under near-infrared (NIR)<br>illumination.<br>• The ocular surface and eyelids under white<br>illumination.<br>All of these image types can be photographically<br>documented and visually monitored. | The LipiView Ocular Surface Interferometer is an<br>ophthalmic imaging device that is intended for use<br>by a physician in adult patients to capture, archive,<br>manipulate, and store digital images of specular<br>(interferometric) observations of the tear film,<br>which can be visually monitored and<br>photographically documented. Using these<br>images, the LipiView® Interferometer measures<br>the absolute thickness of the tear film lipid layer. | Optomed Smartscope M5 with<br>optics module ES2 is intended to<br>capture images and video of the<br>surface of the human eye and<br>surrounding areas. |
| Rx only/OTC | Prescription Device | Prescription Device | Prescription Device |
| Method of Operation<br>Ophthalmic Imaging | Ophthalmic camera with digital imaging | Ophthalmic camera with digital imaging | Ophthalmic camera with digital<br>imaging |
| Method of Operation<br>Specular Observation<br>of Tear Film | Tear film images and lipid layer thickness<br>measurement based on specular reflection using<br>white light interferometry | Tear film images and lipid layer thickness<br>measurement based on specular reflection using<br>white light interferometry | None |
| Illumination Source | • Visible White Class I LED illumination<br>• Visible LED fixation light<br>• Near-Infrared LED illumination and<br>transillumination of the eyelid | • Visible White Class I LED illumination<br>• Visible LED fixation light | • Visible and Infrared LED<br>illumination<br>• Red LED Fixation Light<br>• White and Cobalt Blue Light |
| Illumination Exposure | Complies with ISO 15004-2:2007 Group 2 | Complies with ISO 15004-2:2007 Group 1 | Complies with ISO 15004-2:2007<br>Group 2 instrument for safety |
| Parameters | instrument for safety | instrument for safety | |
| Brightness Control | No brightness control adjustment by user | No brightness control adjustment by user | Unknown |
| Material Flammability | Materials near light source comply with UL 94V-1 | Materials near light source comply with UL 94V-1 | Unknown |
| Maximum Temperature of<br>Held or Accessible Parts | Ambient temperature of parts of device held by<br>operator or accessible to patient | Ambient temperature of parts of device held by<br>operator or accessible to patient | Unknown |
| TABLE 5-1. PREDICATE DEVICE COMPARISON TABLE | | | |
| Comparison Feature | LipiView II Interferometer | Predicate LipiView Interferometer<br>(K091935 and K122481) | Optomed Smartscope M5<br>Digital Camera (K132186) |
| Data Collection and/or<br>Display Systems | • Touchscreen display to view images<br>• Stores images on computer and external media<br>• Support for off-the-shelf USB printer and media<br>that are compatible with Windows 7 and USB2.0 | • Touchscreen display to view images<br>• Stores images on computer and external media<br>• Support for off-the-shelf USB printer and media<br>compatible with Windows XP and USB1.0 | • Active matrix color TFT LCD<br>• Stores images on external<br>media (flash memory card)<br>• USB1.1 terminal compatible<br>with Windows XP/Vista/7 |
| Electrical Safety and<br>Electromagnetic<br>Compatibility | • AC-powered<br>• Complies with ANSI 60601-1:2005 (3rd edition)<br>and IEC 60601-1-2: 2007 (3rd edition) | • AC-powered<br>• Complies with IEC 60601-1:1995 (2nd edition)<br>and IEC 60601-1-2: 2001 (2nd edition) | • Rechargeable NiMH Battery<br>with charging unit<br>• Complies with IEC 60601-<br>1:2005 (3rd edition) and IEC<br>60601-1-2: 2007 (3rd edition) |
| Chin/Forehead Rest<br>Support | Integrated chin rest and forehead support is<br>disinfected with alcohol prior to use and storage | Integrated chin rest and forehead support is<br>disinfected with alcohol prior to use and storage | May be used with slit lamp<br>adapter for chin rest support |
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.