Ziostation2 is an image processing application software available for installation onto customer owned hardware. This application software can be networked to provide for sharing of resources. This application software receives medical images from modalities (mage scanning devices such as CT) or image archives such as PACS through network or media and provides for the viewing, quantification, manipulation, communication, printing, and management of medical images. This application software is intended for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which Ziostation2 is being offered.
Device Story
Ziostation2 is a DICOM-based image processing software installed on customer-owned hardware. It ingests medical images from modalities (CT, MRI, Ultrasound, etc.) or PACS. The software provides tools for viewing, quantification, manipulation, and management of these images. It is operated by trained medical professionals in clinical settings to assist in the analysis of anatomy, physiology, and pathology. The software includes specialized modules for cardiac analysis (coronary, RV, TAVR), perfusion analysis (CT/SPECT), lesion analysis, liver analysis, and MR tractography. Clinicians use the output to supplement standard radiological interpretation; the clinician retains ultimate responsibility for diagnosis. The device benefits patients by providing advanced visualization and quantitative tools to aid in clinical decision-making and monitoring of disease progression or treatment response.
Clinical Evidence
Bench testing only. The device underwent integration testing, verification testing, regression testing, and Beta validation. Risk management processes were applied to address potential hazards. No clinical data was required or provided.
Technological Characteristics
Software-only device; no physical materials. Operates as a DICOM image management and processing system. Features include image fusion, noise reduction, motion coherence, automated vessel labeling, HU range color coding, ROI measurement, polar map generation (stress/rest/reversibility), and auto-segmentation. Connectivity via network or media for PACS/modality integration.
Indications for Use
Indicated for trained medical professionals to supplement radiological image interpretation. Supports viewing, quantification, manipulation, communication, printing, and management of medical images from modalities (CT, MRI, Ultrasound, Digital X-ray, Angiography, PET, SPECT, NM, SC, Mammography, Radiofluoroscopic, RT). Contraindicated for primary interpretation of lossy compressed mammographic or digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 4, 2015
Qi Imaging, LLC % Mr. Richard Ball Director, Regulatory & Quality Affairs 1301 Shoreway Road. Suite 325 BELMONT CA 94002
Re: K151212 Trade/Device Name: Ziostation2 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 8, 2015 Received: October 14, 2015
Dear Mr. Ball:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151212
Device Name Ziostation2
Indications for Use (Describe)
Ziostation2 is an image processing application software available for installation onto customer owned hardware. This application software can be networked to provide for sharing of resources.
This application software receives medical images from modalities (mage scanning devices such as CT) or image archives such as PACS through network or media and provides for the viewing, quantification, manipulation, communication, printing, and management of medical images.
This application software is intended for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which Ziostation2 is being offered.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Ziostation2 Qi Imaging, LLC
This 510(k) summary is prepared in accordance with 21 CFR 807.92.
#### General Information
| Trade Name | Ziostation2 |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Common Name<br>Classification Name | Picture Archiving and Communications System (PACS)<br>System, Image Processing, Radiological (21 CFR § 892.2050 - LLZ) |
| Applicant: | Qi Imaging, LLC<br>1301 Shoreway Road, Suite 325<br>Belmont, CA 94002<br>Tel 650-413-1364<br>Fax 650-596-7319 |
| Contact | Richard Ball<br>Sr. Director, Regulatory and Quality Affairs |
#### Intended Use
- . Ziostation (hereinafter "this application software") is an image processing application software available for installation onto customer-owned hardware. This application software can be networked to provide for sharing of resources.
- This application software receives medical images from modalities (medical image scanning devices such as CT) or image archives such as PACS through network or media and provides for the viewing, quantification, manipulation, communication, printing, and management of medical images.
- . This application software is intended for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images.
- . Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which Ziostation2 is being offered.
Note: The clinician retains the ultimate responsibility for making the proper diagnosis based on standard radiological practices and visual comparison of the separate, unprocessed images. Ziostation2 is a tool to be used in support of those standard practices and visual comparisons.
{4}------------------------------------------------
## Predicate Devices
Predicate devices to Ziostation2 Software Tool
| Device Name | 510(k)<br>Number |
|---------------------------------------------------------------------------|------------------|
| Ziostation<br>(Multiple Protocol Enhancements) | K070205 |
| Ziostation with<br>Cardiac Function Analysis and Calcium Scoring Software | K083446 |
| CARDIQ Express v2.0 | K073138 |
| EBW NM 2.0 | K111336 |
| PET VCAR | K063324 |
| CT Perfusion 4 | K052839 |
| Vitrea CT Myocardial Analysis | K112531 |
| Syngo Volume Perfusion, CT Body | K092013 |
| AW Volume Share 5 | K110834 |
| Synapse 3D Liver Analysis | K110186 |
| Signa HDxT HD | K121676 |
| Achieva R2 1.5T/3.0T | K110151 |
| SureSubtraction Ortho CSSO-001A | K130960 |
| CardEP | K031261 |
| Vitrea CT Transcatheter Aortic Valve Replacement Planning | K122578 |
#### Device Description
ZIOSTATION2 is a basic DICOM image management system to further aid clinicians in their analysis of anatomy, physiology and pathology. Universal functions such as data retrieval, storage, management, querying and listing, and output are handled by the basic Ziostation2 software. Various imaging tools and techniques can be invoked to process images from the following image types: CT, MRI, Ultrasound, Digital X-ray X-ray Angiography, PET, SPECT, NM, SC, Mammography, X-ray Radiofluoroscopic image, RT Image.
#### New Features – Substantial Equivalence Discussion
| | Image Viewing/Tool Enhancements – Ref. Ziostation K070205 and Ziostation w/ Cardiac Function |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Analysis K083446 as predicate devices | |
| Capability | Substantial Equivalence |
| Fusion Viewer | K070205 Fusion Viewer is a workflow enhancement and is substantially equivalent to the current Ziostation ability to fuse images. |
| PhyZiodynamics | K070205 PhyZiodynamics reconstructs data with the addition of noise reduction and motion coherence to improve image quality and is substantially equivalent to the current Ziostation data reconstruction capability. |
{5}------------------------------------------------
| CT Coronary<br>Analysis | K070205 (Secondary K073138) Improvements to CT Coronary include<br>automatic vessel labeling and color coding of HU ranges of specified areas<br>for easier visualization of suspected areas of interest. This tool with its<br>expanded capabilities is substantially equivalent to the current Ziostation<br>coronary analysis tool. |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CT Colon Analysis | K070205 (Secondary K110834) The added capability to automatically<br>measure and report on user-specified ROI's is a workflow enhancement.<br>This capability is substantially equivalent to the current Ziostation CT<br>Colon tool. |
| CT/SPECT Cardiac<br>Fusion | K070205 (Secondary K111336) The functionality being added to the<br>already cleared CT/SPECT Cardiac Fusion ability of the current Ziostation<br>device is to provide data and polar maps of stress, rest, reversibility and<br>washout. This workflow enhancement is substantially equivalent to the<br>current Ziostation CT/SPECT Cardiac Fusion tool. |
| SUV Measurement | K070205 (Secondary K063324) Uses current multi-exam comparison<br>capability of PET images to visualize and monitor disease progression or<br>response to treatment or therapy. This workflow enhancement is<br>substantially equivalent to the current Ziostation PET imaging tool. |
| CT Subtraction | K070205 (Secondary K130960) CT Subtraction improves the current<br>capability of contrast enhancement visualization. This improvement is<br>substantially equivalent to the current Ziostation contrast enhancement<br>capability. |
| EP Planning | K070205 (Secondary K031261) Using currently available display and<br>measurement tools, EP Planning allows the physician to concentrate on<br>certain specific areas of the heart. This tool is substantially equivalent to<br>the current Ziostation display and measurement capability. |
| TAVR | K070205 (Secondary K122578) Using currently available display and<br>measurement tools, TAVR allows the physician to concentrate on certain<br>specific areas of the heart, including aortic valve and surrounding<br>structures. This tool is substantially equivalent to the current Ziostation<br>display and measurement capability. |
| CT Cardiac RV | K083446 (Secondary K110834) Using currently available display and<br>measurement tools, CT Cardiac RV adds the capability to calculate and<br>display various functional parameters of the right ventricle. This added<br>capability is substantially equivalent to the current Ziostation capability to<br>provide functional ventricular information. |
## Additional Image Processing Functionalities
| CT Perfusion Analysis | |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CT Perfusion<br>Analysis | K052839 Both the Qi Imaging feature and the GE predicate feature are<br>image analysis software applications that allow evaluation of dynamic CT<br>data following the injection of a compact bolus of contrast material,<br>generating information with regard to changes in image intensity over<br>time. Both features allow assessment of the type and extent of various<br>perfusion related parameters. The Qi Imaging CT Perfusion feature is<br>substantially equivalent to GE CT Perfusion 4. |
| CT Volume<br>Perfusion | K052839 Both the Qi Imaging feature and the GE predicate feature are<br>image analysis software applications that allow evaluation of dynamic CT<br>data following the injection of a compact bolus of contrast material,<br>generating information with regard to changes in image intensity over |
{6}------------------------------------------------
| | time. Both features allow assessment of the type and extent of various<br>perfusion related parameters in user-specified planes of orientation. The Qi<br>Imaging CT Volume Perfusion feature is substantially equivalent to GE<br>CT Perfusion 4. |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CT Myocardial<br>Perfusion | K112531 Both the Qi Imaging feature and the Vital Images feature<br>provide color-coded polar maps of the contrasted myocardium based on<br>the calculation of transmural perfusion ratios. The Qi Imaging CT<br>Myocardial Perfusion feature is substantially equivalent to Vital Images<br>Vitrea CT Myocardial Analysis |
| CT Dynamic<br>Myocardial Perfusion | K092013 Both the Qi Imaging feature and the Siemens feature provide 3-<br>dimensional calculation of Blood Flow, Blood Volume and other<br>functional parameters by displaying color coded maps of individual<br>parameters on user-specified regions of the myocardium. The Qi Imaging<br>CT Dynamic Myocardial Perfusion feature is substantially equivalent to<br>Siemens syngo Volume Perfusion – CT Body |
| CT Lesion Analysis | K110834 Both the Qi Imaging feature and the GE feature allow extraction<br>and measurement of lesions, enabling observation of changes over time.<br>The Qi Imaging CT Lesion Analysis feature is substantially equivalent to<br>GE AW Volume Share 5 |
| CT Liver Analysis | K110186 Both the Qi Imaging CT Liver feature and the Fuji Film feature<br>allow auto-segmentation/calculation of liver features such as liver<br>structure, volumes, tumors and manual definition of separation plane<br>proposals. The Qi Imaging CT Liver Analysis feature is substantially<br>equivalent to Fuji Film Synapse 3D Liver Analysis |
| MR Tractography | K121676 and K110151 Qi Imaging MR Tractography feature, the GE<br>device and the Philips device all present DTI Tractography data showing<br>colored tracts of various anatomical fibers of interest overlaid onto the<br>surrounding anatomy. The Qi Imaging MR Tractography feature is<br>substantially equivalent to the comparable software features of both GE<br>Signa HDXT and Philips Achieva R41.5T & 3.0T. |
# Materials
This software tool consists entirely of software. No materials are contained in this product.
# Testing Summary
The ZIOSTATION2 software package successfully completed integration testing/verification testing prior to Beta validation. Regression testing was also performed on all functionality present on Ziostation. Software Beta testing/validation was successfully completed prior to final testing and release. In addition, potential hazards have been addressed by the Qi Imaging Risk Management process.
# Summary of Substantial Equivalence
ZIOSTATION2 is substantially equivalent in intended use and function to the composite of predicate devices identified above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.