K181892 · Ziosoft, Inc. · LLZ · Oct 9, 2018 · Radiology
Device Facts
Record ID
K181892
Device Name
Ziostation2
Applicant
Ziosoft, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Indications for Use
This application is image processing application software available for installation onto customer-owned hardware. This application software can be networked to provide for sharing of resources. This application software receives medical image data (CT, MRI, Ultrasound, Digital X-ray, X-ray Angiography, PET, SPECT, Nuclear Medicine, Secondary Capture, Mammography, X-ray Radiofluoroscopic image, and RT Image) from modalities or image archives such as PACS through network or media, and provides for the viewing, quantification, manipulation, communication, printing and management of medical images. This application software is intended only for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which this application is being offered.
Device Story
Ziostation2 is a DICOM-based image management and processing software suite installed on customer-owned hardware. It ingests medical images from modalities (CT, MRI, Ultrasound, etc.) or PACS via network/media. Clinicians use the software to view, quantify, manipulate, and manage these images to aid in the analysis of anatomy, physiology, and pathology. New functionalities include Goddard score classification for LAA analysis, lung lobe extraction, advanced MPR batch processing, virtual bronchoscopy (route planning), CT lung resection planning, and MR myocardial T1 mapping. The software provides post-processing analysis tools; it does not perform image scanning or communicate with real-time interventional devices like bronchoscopes. Healthcare providers use the output to supplement radiological interpretation and surgical planning. The device benefits patients by providing advanced visualization and quantitative analysis to support clinical decision-making.
Clinical Evidence
Bench testing only. The device underwent integration testing, verification testing, and regression testing on all functionalities. Potential hazards were addressed through the Ziosoft Risk Management process. No clinical data was required or provided.
Technological Characteristics
Software-only image processing application. Operates on customer-owned hardware. Supports DICOM standard for image data. Features include MPR, segmentation (lung lobes, bronchi, lesions), quantification (Goddard score, T1 mapping), and virtual navigation. Connectivity via network or media. No physical materials or energy sources.
Indications for Use
Indicated for trained medical professionals to view, quantify, manipulate, communicate, print, and manage medical images (CT, MRI, Ultrasound, Digital X-ray, X-ray Angiography, PET, SPECT, Nuclear Medicine, Secondary Capture, Mammography, X-ray Radiofluoroscopic, RT Image) to supplement radiological interpretation. Contraindicated for primary interpretation of lossy compressed mammographic images and digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
ACHIEVA, INTERA AND PANORAMA 1.0T, RELEASE 2.5 (K063559)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.
October 9, 2018
Ziosoft, Inc. % Richard Ball Regulatory Consultant Ziosoft USA Inc. 1301 Shoreway Road, Suite 325 BELMONT, CA 94002
Re: K181892
Trade/Device Name: Ziostation2 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ, JAK, LNH Dated: June 29, 2018 Received: July 13, 2018
Dear Richard Ball:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hole 2. Nils
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181892
Device Name Ziostation2
#### Indications for Use (Describe)
• This application is image processing application software available for installation onto customer-owned hardware. This application software can be networked to provide for sharing of resources.
• This application software receives medical image data (CT, MRI, Ultrasound, Digital X-ray, X-ray Angiography, PET, SPECT, Nuclear Medicine, Secondary Capture, Mammography, X-ray Radiofluoroscopic image, and RT Image) from modalities or image archives such as PACS through network or media, and provides for the viewing, quantification, manipulation, communication, printing and management of medical images.
•This application software is intended only for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images.
· Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which this application is being offered.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| | <span> <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Ziostation2 Ziosoft, Inc.
This 510(k) summary is prepared in accordance with 21 CFR 807.92.
### General Information
| Trade Name | Ziostation2 |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Common Name<br>Classification Name | Picture archiving and communications system<br>System, Image Processing, Radiological (21 CFR § 892.2050 - LLZ) |
| Applicant: | Ziosoft, Inc.<br>MitaKokusai Bldg., 1-4-28 Mita,<br>Minato-ku, Tokyo 108-0073, Japan |
| Contact | Richard Ball<br>Regulatory Consultant<br>Telephone: +1 (650) 413 1300<br>E-mail:<br>Dball@ziosoftinc.com |
### Intended Use
- This application is image processing application software available for installation onto customer-owned hardware. This application software can be networked to provide for sharing of resources.
- . This application software receives medical image data (CT, MRI, Ultrasound, Digital X-ray, X-ray Angiography, PET, SPECT, Nuclear Medicine, Secondary Capture, Mammography, X-ray Radiofluoroscopic image, and RT Image) from modalities or image archives such as PACS through network or media, and provides for the viewing, quantification, manipulation, communication, printing and management of medical images.
- . This application software is intended only for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images.
- . Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which this application is being offered.
### Predicate Devices
Predicate devices to Ziostation2 Software Tool
{4}------------------------------------------------
| Device Name | 510(k)<br>Number |
|---------------------------------------------------------------------------|------------------|
| Ziostation2 | K151212 |
| CT PULMONARY ANALYSIS | K130552 |
| SYNAPSE 3D LUNG AND ABDOMEN ANALYSIS | K130542 |
| LUNGPOINT PLANNING AND VIRTUAL BRONCHOSCOPIC<br>NAVIGATION (VBN) SOFTWARE | K112051 |
| IQQA-BODYIMAGING SOFTWARE | K141745 |
| SYNGO MR B17 | K082427 |
| MR-CT VVA | K140587 |
| ACHIEVA, INTERA AND PANORAMA 1.0T, RELEASE 2.5 | K063559 |
The already-cleared Ziostation2 (K151212) is the primary predicate device.
# Device Description
Ziostation2 is a basic DICOM image management system to further aid clinicians in their analysis of anatomy, physiology and pathology. Universal functions such as data retrieval, storage, management, querying and listing, and output are handled by the basic Ziostation2 software. Various imaging tools and techniques can be invoked to process images from the following image types: CT, MRI, Ultrasound, Digital X-ray, X-ray Angiography, PET, SPECT, Nuclear Medicine, Secondary Capture, Mammography, X-ray Radiofluoroscopic image, RT Image.
# Materials
This software tool consists entirely of software. No materials are contained in this product.
# Testing Summary
The Ziostation2 software package successfully completed integration testing/verification testing prior to validation. Regression testing was also performed on all functionality present on Ziostation2 prior to release. In addition, potential hazards have been addressed by the Ziosoft Risk Management process.
# Summary of Substantial Equivalence
# Similarities and Differences
# 1) Ziostation2 with already cleared aspects
With respect to the already-cleared intended use of Ziostation2, no significant functional differences exist between Ziostation2 and SE K151212. Certain improvements in workflow and usability have been implemented. The function of each tool being improved or added and the difference between it and the current Ziosoft, Inc. Ziostation2 and/or predicate device are discussed in the following items.
{5}------------------------------------------------
# 2) LAA analysis (Goddard)
LAA analysis (Goddard) is adding Goddard score classification to Ziostation2. The feature of Goddard score classification is legally marketed in SYNAPSE 3D LUNG AND ABDOMEN ANALYSIS K130542. Other features are previously cleared in CT PULMONARY ANALYSIS K130552 including LAA analysis.
### 3) Lung Lobes
Lung Lobes is adding lung lobes extraction to Ziostation2. The feature of lung lobes extraction is legally marketed in SYNAPSE 3D LUNG AND ABDOMEN ANALYSIS K130542. Other features are previously cleared in CT PULMONARY ANALYSIS K130552.
### 4) Advanced MPR Batch
Advanced MPR Batch is a workflow enhancement. No significant functional differences exist between SE K151212.
### 5) Virtual Bronchoscopy
Virtual Bronchoscopy is a new functionality added to Ziostation2. The feature of showing the route to the desired target and showing Virtual Bronchoscopy is legally marketed in SYNAPSE 3D LUNG AND ABDOMEN ANALYSIS K112051 and LUNGPOINT PLANNING AND VIRTUAL BRONCHOSCOPIC NAVIGATION (VBN) SOFTWARE. Other features are previously cleared in SE K151212 and CT PULMONARY ANALYSIS K130552. Virtual Bronchoscopy does not communicate with real Bronchoscope.
# 6) CT Lung Resection Planning
CT Lung Resection Planning is a new functionality added to Ziostation2. The feature of extracting bronchi and lung lobes, extracting lesions and providing the proposal for lung resection are legally marketed in IQQA-BODYIMAGING SOFTWARE K141745 and SYNAPSE 3D LUNG AND ABDOMEN ANALYSIS K130542. Some other lung analysis features are previously cleared in CT PULMONARY ANALYSIS K130552.
# 7) MR Myocardial T1 Mapping
MR Myocardial T1 Mapping is a new functionality added to Ziostation2, providing analysis of myocardium MR T1 mapping which is legally marketed in SYNGO MR B17 K082427, MR-CT VVA K140587 and ACHIEVA, INTERA AND PANORAMA 1.0T, RELEASE 2.5 K063559.
MR Myocardial T1 Mapping provides post process analysis and does not provide image scanning.
### 8) Conclusion
With respect to the tools and features offered by Ziosoft, no significant functional differences exist between Ziostation2 software and those portions of the legally marketed SE devices that perform the same functions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.