OptiFix Fixation System - Straight 5mm X 39 cm - 30 absorbable fasteners, OptiFix Fixation System - Straight 5mm x 39 cm - 15 absorbable fasteners
K142873 · C.R. Bard, Inc. · GDW · Mar 12, 2015 · General, Plastic Surgery
Device Facts
Record ID
K142873
Device Name
OptiFix Fixation System - Straight 5mm X 39 cm - 30 absorbable fasteners, OptiFix Fixation System - Straight 5mm x 39 cm - 15 absorbable fasteners
Applicant
C.R. Bard, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Mar 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OptiFix™ Absorbable Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.
Device Story
OptiFix™ Absorbable Fixation System is a sterile, single-use device for soft tissue approximation and mesh fixation in open or laparoscopic hernia repair. The system consists of a pistol-shaped deployment tool and preloaded absorbable fasteners made of Poly (D, L)-lactide. The clinician operates the device by inserting the 39 cm shaft through a 5mm trocar and using a resistive trigger to mechanically guide and launch fasteners into tissue. The fasteners feature retention heads to secure the mesh. The device is intended for use by surgeons in clinical settings. The system provides a mechanical method for tissue fixation, replacing manual suturing or other fixation methods, potentially reducing procedure time and improving mesh stability. Safety and performance are supported by bench testing, including actuation torque, fastener deployment, and resorption profile analysis, demonstrating equivalence to the predicate device.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including actuation torque, fastener deployment, fastener gap height, ball burst testing, mesh compatibility, and resorption profile. Biocompatibility was evaluated per ISO 10993-1, with fastener biocompatibility supported by data from the reference device (SorbaFix, K082396).
Technological Characteristics
Single-use, sterile (gamma irradiation 25-40 kGy) fixation system. Fasteners: Poly (D, L)-lactide, injection-molded, 6.7 mm length. Deployment device: 39 cm shaft, pistol-grip handle, resistive trigger mechanism. Mechanical actuation principle. Tissue contact: <24 hours for deployment device; permanent implant for fasteners. No software or electronic components.
Indications for Use
Indicated for soft tissue approximation and surgical mesh fixation during open or laparoscopic procedures, including hernia repair.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 12, 2015
Davol Incorporated Ms. Christine Lloyd Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886
Re: K142873
Trade/Device Name: OptiFix™ Absorbable Fixation System Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: February 6, 2015 Received: February 9, 2015
Dear Ms. Lloyd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142873
Device Name OptiFix™ Absorbable Fixation System
Indications for Use (Describe)
The OptiFix™ Absorbable Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 11pt;"> <span style="color: black;"> <span style="font-weight: bold;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 11pt;"> <span style="color: black;"> <span style="font-weight: bold;">□</span> Over-The-Counter Use (21 CFR 801 Sub </span> </span> </span> |
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
Davol Inc. 100 Crossings Boulevard Warwick, RI 02886
| Contact Person: | Christine Lloyd |
|-----------------|-------------------------------|
| | Regulatory Affairs Specialist |
| Phone: | (401) 825-8447 |
| Fax: | (401) 825-8753 |
| E-mail: | Christine.Lloyd@crbard.com |
Date Prepared: March 12 2015
#### II. DEVICE
| Name of Device: | OptiFix™ Absorbable Fixation System [product ordering codes 0113126 (deployment device contains 30 fasteners) and 0113127 (deployment device contains 15 fasteners)] |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Staple Implantable/Implantable Staple |
| Classification Name: | Implantable staple (21 CFR §878.4750) |
| Regulatory Class: | II |
| Product Code: | GDW |
#### III. PREDICATE DEVICE
OptiFixTM Absorbable Fastener System, K132134; marketed by Davol, Inc., This predicate has not been subject to a design-related recall.
The reference device Davol Absorbable Fastem - SorbaFix™ (K082396) is used in this submission.
#### IV. DEVICE DESCRIPTION
The OptiFix™ Absorbable Fixation System is a sterile (via gamma) single use product that is comprised of a deployment component and an absorbable fastener component. Two product ordering codes are to be packaged for distribution; each contains the same ergonomically designed deployment device. The variation will be the preloaded fasteners; either 15 or 30. The shaft of the OptiFix™ Absorbable Fixation System is 39 cm in length and is designed for use with 5mm trocars. The fasteners are designed with retention features and are manufactured from Poly (D, L)-lactide.
#### SECTION 7
{4}------------------------------------------------
#### V. INDICATIONS FOR USE
The OptiFix™ Absorbable Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.
The Indications for Use statement for the subject device is identical to the predicate device. Both the subject and predicate devices have the same intended use for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair using the mechanical actuation (trigger) force to guide the fastener and launch the implant is the principle for both the proposed and predicate devices.
At a high level, the proposed and predicate devices are based on the following same elements:
- Deployment component mechanical component used to guide the fastener into tissue ●
- Use of a mechanical component for positioning and launching the implant ●
- User-controlled mechanical trigger to guide the fastener (implant) ●
- Fastener component - absorbable component based on the same material used to fixate surgical mesh to tissue
At a high level, the following differences exist between the proposed and predicate devices:
- The proposed deployment device has a resistive trigger and the predicate has a rotating ● trigger.
- The proposed fastener component has slight differences in geometry than the predicate.
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Performance standards
No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
{5}------------------------------------------------
# Biocompatibility testing
The biocompatibility evaluation for the OptiFix™ Absorbable Fixation System was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
Testing for the fastener included Cytotoxicity and the complete biocompatibility profile for the OptiFix fastener has been supported by full biocompatibility testing on the reference device, SorbaFix (K082396) as listed in the following. The fastener materials between the OptiFix and the reference device have been demonstrated to be substantially equivalent.
- Sensitization ●
- Intracutaneous Reactivity
- Single-dose Systemic Toxicity ●
- Rabbit Pyrogenicity
- 14-Day Repeat Dose Intravenous Toxicity
- Genotoxicity ●
- Chronic Toxicity, 13 weeks
- . Implantation. 52 weeks
Testing for the new materials within the deployment device included Cytotoxicity, Sensitization, Intracutaneous Reactivity, and Single-dose Systemic Toxicity.
The deployment component of the proposed OptiFix™ Absorbable Fixation System is determined to be tissue contacting for duration of less than 24 hours, while the absorbable fasteners are determined to be permanent implants.
# Testing
The following non-clinical tests were completed for the proposed and predicate device. The OptiFix™ Absorbable Fixation System passed all the test requirements and demonstrated substantial equivalence to the test results of the predicate device.
- Performance and Functional testing of the OptiFix™ Absorbable Fixation System-● deployment device and fastener
- Actuation (trigger) torque .
- . Fastener deployment
- . Fastener gap height
- . Ball burst testing
- Mesh compatibility testing of the OptiFix™ Absorbable Fixation System
- Resorption profile of the fastener
All samples tested met the established acceptance criteria.
#### SECTION 7
{6}------------------------------------------------
# Animal Study
Animal studies were not performed on the proposed device for this submission. Based on identical materials and comparable mechanical strengths measured at the bench level, the in vivo safety and performance of the proposed device was evaluated via the animal and histological studies conducted on the reference device, the SorbaFix Fixation Device.
## Clinical Studies
Clinical studies were not performed for the submission of this device nor were clinical studies performed for the predicate device.
| Device Features | Proposed device<br>OptiFix™ | Predicate device<br>OptiFix™ (K132134) |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Soft tissue fixation | Soft tissue fixation |
| Indication For use | Indicated for the approximation<br>of soft tissue and fixation of<br>surgical mesh to tissues during<br>open or laparoscopic surgical<br>procedures, such as hernia repair. | Indicated for the approximation of<br>soft tissue and fixation of surgical<br>mesh to tissues during open or<br>laparoscopic surgical procedures,<br>such as hernia repair. |
| Fastener Material | Poly (D,L) Lactide | Poly (D,L) Lactide |
| Fastener Violet Dye | D & C Violet No. 2<br>conforms to 21 CFR §74.3602 | D & C Violet No. 2<br>conforms to 21 CFR §74.3602 |
| Fastener Body<br>Contact | Long term implant (>30 days)<br>contacting tissue and/or bone | Long term implant (>30 days)<br>contacting tissue and/or bone |
| Fastener<br>Shape/Design | Push Tack with retention features<br>on end | Push Tack with retention features<br>on end |
| Fastener Dimensions | 6.7 mm overall fastener length<br>Fastener head: 3.7 mm length; 3.5<br>mm thickness | 6.7 mm overall fastener length<br>Fastener head: 3.7 mm length; 3.5<br>mm thickness |
| Fastener<br>Manufacturing<br>Method | Injection-Molded | Injection-Molded |
| Fastener Absorption<br>Time | 360 days | 360 days |
| Fastener Quantity<br>per Device | 15 & 30 fasteners | 15 & 30 fasteners |
# Table 7-1: Device Substantial Equivalence – General Characteristics
#### Premarket Notification for OptiFix™ Absorbable Fixation System
{7}------------------------------------------------
| Device Features | Proposed device<br>OptiFix™ | Predicate device<br>OptiFix™ (K132134) |
|----------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Fastener Delivery<br>System | Push - Impact tube pushes<br>fasteners forward over a straight<br>guidewire | Push – Impact tube pushes<br>fasteners forward over a straight<br>guidewire |
| Deployment Device<br>Handle design | Pistol/Gun shape | Pistol/Gun shape |
| Deployment Device<br>Shaft Nominal<br>Length | 39 cm in length | 39 cm in length |
| Deployment Device<br>Trigger | Resistive trigger | Rotating trigger |
| Deployment Device<br>Trigger Stroke | Swift and full | Slow and progressive |
| Deployment Device<br>Guidewire | Moveable | Stationary |
| Device Sterilization | Gamma Irradiation<br>(25 - 40 kGy) | Gamma Irradiation<br>(25 - 40 kGy) |
## VIII. CONCLUSIONS
The OptiFix™ Absorbable Fixation System is substantially equivalent to the legally marketed predicate device for the following reasons:
- A) The same intended use and indications for use as the predicate device.
- B) Both devices use a similar fixation technology to deliver the fasteners by compressing a trigger.
- C) Both devices house similar synthetic absorbable fastener with retention head design.
- D) Identical materials with long history of biocompatible use in medical instrumentation and implantation.
- E) Identical technological characteristics to the predicate devices such as: Fastener dimensions trigger handle, penetration depth, and shaft length.
- F) Same principle of operation.
The comparative analysis as well as the bench and preclinical testing results demonstrated that the OptiFix™ Absorbable Fixation System should perform as safe, and as effective, as the predicate device that is currently marketed for the same intended use.
#### SECTION 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.