OptiFix AT Absorbable Fixation System with Articulating Technology - 30 Fasteners, OptiFix AT Absorbable Fixation System with Articulating Technology - 15 Fasteners
K170278 · C.R. Bard, Inc. · GDW · May 31, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170278
Device Name
OptiFix AT Absorbable Fixation System with Articulating Technology - 30 Fasteners, OptiFix AT Absorbable Fixation System with Articulating Technology - 15 Fasteners
Applicant
C.R. Bard, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
May 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OptiFix AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.
Device Story
OptiFix AT is a sterile, single-use, mechanical fixation system for hernia repair; used in open or laparoscopic surgery by surgeons. Device consists of a pistol-grip deployment handle and a shaft containing preloaded absorbable fasteners (15 or 30 count). Key features include 360-degree handle rotation and an articulating shaft tip to improve access to surgical sites. Operation involves manual mechanical actuation to drive fasteners through mesh into tissue. Fasteners are composed of Poly(L-lactide-co-glycolide) and provide long-term fixation before absorbing. Device facilitates mesh placement, potentially reducing surgical time and improving fixation precision in confined anatomical spaces.
Clinical Evidence
No clinical studies performed. Safety and feasibility supported by non-clinical bench testing (actuation torque, deployment, fastener gap height, ball burst, mesh compatibility, in-vitro degradation) and an animal study comparing the device to the secondary predicate (AbsorbaTack). Animal study evaluated mesh contracture, fastener seating, tissue in-growth, and inflammatory response. Biocompatibility testing conducted per ISO 10993-1.
Technological Characteristics
Mechanical deployment system for absorbable fasteners. Fasteners made of Poly(L-lactide-co-glycolide) dyed with D & C Violet No. 2. Shaft length 37cm, compatible with 5mm trocars. Features 360-degree handle rotation and articulating shaft tip. Sterilized via gamma irradiation. No electrical or software components.
Indications for Use
Indicated for soft tissue approximation and surgical mesh fixation during open or laparoscopic procedures, including hernia repair, in patients requiring such surgical intervention.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Medtronic Covidien AbsorbaTack Absorbable Fastener System (K123109)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 31, 2017
C.R. Bard Incorporated Ms. Katherine Earle Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886
Re: K170278
Trade/Device Name: Optifix AT Absorbable Fixation System With Articulating Technology - 30 Fasteners, Optifix AT Absorbable Fixation System With Articulating Technology - 15 Fasteners
Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: April 28, 2017 Received: May 1, 2017
Dear Ms. Earle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{2}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170278
#### Device Name
OptiFix AT Absorbable Fixation System with Articulating Technology - 30 Fasteners, OptiFix AT Absorbable Fixation System with Articulating Technology - 15 Fasteners
#### Indications for Use (Describe)
The OptiFix AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <span></span> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
I. SUBMITTER Davol Inc. 100 Crossings Boulevard Warwick, RI 02886
| Contact Person: | Katherine Earle |
|-----------------|-------------------------------|
| | Regulatory Affairs Specialist |
| Phone: | (401) 825-8450 |
| Fax: | (401) 825-8765 |
| E-mail: | |
katherine.earle@crbard.com
Date Prepared: May 30, 2017
#### II. DEVICE
| 510(k) Number: | K170278 |
|-----------------------|----------------------------------------------------------------------------------|
| Name of Device: | OptiFix <sup>TM</sup> AT Absorbable Fixation System with Articulating Technology |
| Common or Usual Name: | Implantable Staple |
| Classification Name: | Implantable Staple (21 CFR §878.4750) |
| Regulatory Class: | II |
| Product Code: | GDW |
#### III. PREDICATE DEVICE
The primary predicate device for this submission is the OptiFix" Absorbable Fixation System (K142873), cleared March 12, 2015; marketed by Davol, Inc. The primary predicate, hereinafter referred to as "the predicate" throughout the remainder of the submission, has not been subject to a design-related recall.
The secondary predicate device for this submission is the Medtronic Covidien AbsorbaTack™ Absorbable Fastener System (K123109), cleared January 27, 2012.
{5}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The OptiFix™ AT Absorbable Fixation System with Articulating Technology, hereinafter referred to as "OptiFix ""AT", is a sterile (via gamma) single use device that is comprised of a deployment component and an absorbable fastener component. Two different product ordering codes are to be packaged for distribution; each contains the same ergonomically designed deployment device. The variation will be the number of preloaded fasteners; either 15 or 30. The deployment shaft of the OptiFix™ AT device is 37cm in length and designed for use with 5mm trocars. The tip of the shaft can be articulated and the handle of the device can be rotated 360 degrees to facilitate access for fixation during surgery. The fasteners are designed with retention features and are manufactured from Poly (L-lactide-co-glycolide) and are dyed with D & C Violet No. 2 (<0.15% by weight).
# V. INDICATIONS FOR USE
The OptiFix™ AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.
The Indications for Use statement for the subject device is identical to that of the primary predicate device (OptiFix "-K142873), and similar to that of the secondary predicate device (AbsorbaTack"-K123109). The proposed and predicate devices are all intended for the fixation of surgical mesh/prosthetic to tissue during open or laparoscopic surgical procedures, such as hernia repair.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The fixation of surgical mesh/prosthetic to tissue during open or laparoscopic surgical procedures, such as hernia repair, using a mechanical actuation force to guide and launch the fastener implant is the operating principle for the proposed and predicate devices.
The proposed and predicate devices are based on the following similar elements:
- Deployment component - mechanical component used to position and launch the fastener (implant) into tissue
- Fastener component -absorbable component composed of well-characterized absorbable polymer, used to fixate surgical mesh to tissue.
- O The fastener shape/design and dimensions of the subject device are similar to that of the primary predicate device (OptiFix" -K142873).
- The fastener material for the subject device (Poly(1-lactide-co-glycolide)) is o identical to that of the secondary predicate device (AbsorbaTack" -K123109).
{6}------------------------------------------------
The following differences exist between the proposed and predicate devices:
- The handle configuration of the proposed device can be rotated 360° and the deployment ● shaft articulated to facilitate fixation, unlike the predicate devices which have no articulation in the deployment shaft and a static handle.
- The deployment shaft of the proposed device is shorter in length (37 cm) compared to the deployment shaft of the primary predicate device (39cm).
- The proposed absorbable fastener has minor geometric differences and is composed of a different material (Poly(L-lactide-co-glycolide)) than the primary predicate absorbable fastener (Poly(D,L)-lactide).
## VII. PERFORMANCE DATA
The following performance data is provided in support of the substantial equivalence determination.
#### Performance standards
No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
## Biocompatibility testing
The biocompatibility evaluation for the OptiFix™ AT device was conducted in accordance with FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". June 16, 2016; and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
The proposed device is supported by the biocompatibility testing listed below.
- Cytotoxicity
- Sensitization ●
- Intracutaneous (Irritation)
- Systemic Acute Toxicity
- Sub-Chronic Toxicity (13 weeks)
- Chronic Toxicity (26 weeks)
- Genotoxicity - Mouse Peripheral Blood Micronucleus Study
- Genotoxicity - Bacterial Reverse Mutation Study
- Genotoxicity - Mouse Lymphoma Assay
- Implantation - 4, 8, 12, 26, 39, 52, 77 weeks
- Pyrogenicity
All samples tested met the acceptance criteria and the proposed OptiFix™ AT device is biocompatible for its intended use.
{7}------------------------------------------------
## Electrical safety and electromagnetic compatibility (EMC)
The proposed OptiFix™ AT device is not an electro-mechanical medical device nor is it a medical system, therefore this section does not apply.
# Software Verification and Validation Testing
The proposed OptiFix " AT device does not contain software; therefore this section does not apply.
# Product Testing
The following non-clinical tests were completed for the proposed and primary predicate device. The proposed OptiFix™ AT device passed all the test requirements and demonstrated substantial equivalence to the test results of the primary predicate device.
- · Performance and Functional testing of the OptiFix™ AT device
- Actuation (trigger) Torque ●
- Fastener Deployment ●
- Fastener Gap Height
- Ball Burst Testing
- · Mesh compatibility testing of the OptiFix " AT device
- · Resorption Profile of the fastener
- In-Vitro Degradation ●
- In-Vitro to In-Vivo Correlation
All samples tested met the established acceptance criteria.
# Animal Study
The safety and feasibility of the OptiFix™ AT device was evaluated by animal and histological evaluation of the tissue in the treatment fixated areas. The evaluation used the secondary predicate, AbsorbaTack™ (K123109), as a control since the fastener is composed of the same material, Poly (L-lactide-co-glycolide) in an 82%/18% molar ratio, and has the same intended use as the OptiFix" AT device. The study demonstrated that the OptiFix AT device can safely, approximate soft tissue and fixate surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. The following endpoints were evaluated:
- Mesh contracture
- · Fastener seating properties
- · Tissue in-growth (T-peel analysis)
- · Host inflammatory/fibrotic response
# Clinical Studies
Clinical studies were not performed for the submission of this proposed device nor were clinical studies performed for the primary predicate device (OptiFix-K142873).
{8}------------------------------------------------
## VIII. CONCLUSIONS
The proposed OptiFix™ AT device is substantially equivalent to the legally marketed primary predicate device for the following reasons:
- A) The same intended use as the primary predicate device.
- B) Similar technological characteristics as the primary predicate device such as: Fastener shape/design, trigger handle, penetration depth, and shaft length. The differences in technology do not adversely impact the safety and performance of the device.
- C) Identical principle of operation to deliver fasteners perpendicular to the tissue plane.
The comparative analysis, see Table 5-1 below, as well as the bench and preclinical testing results demonstrate that the proposed OptiFix™ AT device performs as intended and is substantially equivalent to the primary predicate device that is currently marketed for the same intended use.
| Device Features | Proposed device<br>OptiFix™ AT | Predicate device<br>OptiFix™ (K142873) |
|------------------------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Identical to predicate | Soft tissue fixation |
| Indication For<br>Use | Identical to predicate | Indicated for the approximation of<br>soft tissue and fixation of surgical<br>mesh to tissues during open or<br>laparoscopic surgical procedures,<br>such as hernia repair. |
| Fastener Material | Resorbable Polymer<br>Poly(L-lactide-co-glycolide)<br>(PLG) | Resorbable Polymer<br>Poly (D,L) Lactide |
| Fastener Violet<br>Dye | Identical to predicate | D & C Violet No. 2<br>(<0.15% by weight)<br>conforms to 21 CFR §74.1602 |
| Fastener Body<br>Contact | Identical to predicate | Long term implant (>30 days)<br>contacting tissue and/or bone |
| Fastener<br>Shape/Design | Push Tack with retention features<br>on end; nested tip<br>Image: [purple fastener] | Push Tack with retention features<br>on end<br>Image: [purple fastener] |
| Device Features | Proposed device<br>OptiFix™ AT | Predicate device<br>OptiFix™ (K142873) |
| Fastener<br>Dimensions | 7.1 mm overall fastener length<br>Fastener head: 3.5 mm diameter | 6.7 mm overall fastener length<br>Fastener head: 3.7 mm diameter |
| Fastener<br>Manufacturing<br>Method | Identical to predicate | Injection-Molded |
| Fastener<br>Absorption Time | Identical to predicate | 360 days |
| Fastener Quantity<br>per Device | Identical to predicate | 15 & 30 fasteners |
| Fastener Delivery<br>System | Identical to predicate | Push – Impact tube pushes<br>fasteners forward over a<br>guidewire |
| Deployment<br>Component<br>Handle design | Pistol/Gun shape with 360 degree<br>rotation option | Pistol/Gun shape with no rotation |
| Deployment<br>Component Shaft | 37 cm in length, 6 cm of<br>articulation | 39 cm in length, no articulation |
| Deployment<br>Component<br>Configuration | Image: OptiFix AT Deployment Component Configuration | Image: OptiFix (K142873) Deployment Component Configuration |
| Deployment<br>Component Body<br>Contact | Identical to predicate | Transient use - Less than 24 hour<br>duration |
| Device<br>Sterilization | Identical to predicate | Gamma Irradiation<br>(25 - 40 kGy) |
| Device<br>Packaging | Identical to predicate | Device is placed into an SBS<br>cardboard insert which is inserted<br>into a foil pouch, sealed and then<br>placed into an SBS paperboard<br>printed shelf carton. |
Table 5-1: Device Substantial Equivalence - General Characteristics
{9}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.