SORBAFIX (TM) ABSORBABLE FIXATION SYSTEM, PERMAFIX (TM) FIXATION SYSTEM

K111153 · C.R. Bard, Inc. · GDW · May 24, 2011 · General, Plastic Surgery

Device Facts

Record IDK111153
Device NameSORBAFIX (TM) ABSORBABLE FIXATION SYSTEM, PERMAFIX (TM) FIXATION SYSTEM
ApplicantC.R. Bard, Inc.
Product CodeGDW · General, Plastic Surgery
Decision DateMay 24, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K111153 · May 24, 2011SORBAFIX (TM) ABSORBABLE FIXATION SYSTEM, PERMAFIX (TM) FIXATION SYSTEMC.R. Bard, Inc.Medical Device Reports (MDRs)Medical device reports from clinical practice were investigated to identify a specific patient incidence, resulting in the expansion of the contraindication statement in the device labeling as a remedial action.MDR; Postmarket surveillance; Labeling change

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Internal investigation of medical device reports; Retrospective review of medical device reportsPatients undergoing procedures with the device; Sample Size: One patientNot applicable for this studySafety/Adverse event identification

Indications for Use

The SorbaFix Absorbable Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. The PermaFix Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

Device Story

Sterile, single-use fixation systems (SorbaFix and PermaFix) deliver synthetic fasteners to approximate soft tissue and secure surgical mesh. SorbaFix uses absorbable poly(D, L)-lactide fasteners; PermaFix uses non-absorbable acetal copolymer fasteners. Devices feature 36 cm shafts (5 mm diameter) with piloting tips, compatible with most 5 mm trocars. Operation involves manual trigger-actuated delivery of fasteners. A mechanical fastener gauge on the ergonomic handle provides visual feedback on remaining fastener count. Used by surgeons in OR settings for hernia repair or similar procedures. Modifications include improved drive mechanism (clutch relocation) and addition of fastener gauge. Benefits include secure mesh fixation and tissue approximation.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence supported by design verification and validation testing demonstrating performance equivalent to predicate devices.

Technological Characteristics

Sterile, single-use, manual mechanical fixation system. SorbaFix fasteners: poly(D, L)-lactide (absorbable). PermaFix fasteners: acetal copolymer (non-absorbable). Delivery system: stainless steel components. Shaft length: 36 cm; outer diameter: 5 mm. Mechanical fastener gauge included. Biocompatibility per ISO 10993.

Indications for Use

Indicated for soft tissue approximation and surgical mesh fixation during open or laparoscopic procedures (e.g., hernia repair).

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS MAY 2 4 2011 This 510(k) Summary is provided per the requirements of section 807.92(c). ### Submitter Information | Submitter's Name: | Davol, Inc., Subsidiary of C. R. Bard, Inc. | |----------------------|-----------------------------------------------------------------------------------------------------------------------------| | Address: | 100 Crossings Boulevard<br>Warwick, RI 02886 | | Contact Person: | Zheng Liu<br>Regulatory Affairs Project Manager<br>Tel: (401) 825-8671<br>Fax: (401) 825-8765<br>Email: jane.liu@crbard.com | | Device Name | | | Trade Name: | SorbaFix™ Absorbable Fixation System<br>PermaFix™ Fixation System | | Device Common Name: | Staple, Implantable | | Classification Code: | Class II, §878.4750, Product Code GDW | ## Predicate Device Names - K082396 Davol Absorbable Fastener System, cleared on January 02, 2009 - K092483 PermaFix Fixation System, cleared on December 18, 2009 # Device Description The proposed SorbaFix Absorbable Fixation System is a sterile single use device that delivers either 15 or 30 synthetic absorbable fasteners. The fasteners are dyed with D & C Violet No. 2 in accordance with 21 CFR 74.1602. The shaft of the SorbaFix Absorbable Fixation System is 36 cm in length, including a piloting tip. The fasteners are 6.7 mm in length and are manufactured from Poly (D, L)-lactide. The fixation instrument shafts have an outer diameter of 5 mm, and may be used in open procedures or with most 5 mm trocars in laparoscopic procedures. The device includes a fastener gauge located on the back of the handpiece for user convenience only. The gauge will move right to left as the fasteners are deployed and indicates the approximate level of fasteners remaining in the device. The proposed PermaFix Fixation System is a sterile single use device that delivers PREMARKET NOTIFICATION FOR THE SORBAFIX™ ABSORBABLE FIXATION SYSTEM AND PERMAFIXTM FIXATION SYSTEM > -CONFIDENTIAL-SECTION 8 {1}------------------------------------------------ - have the same implantable fastener design and similar delivery system design: - have the same materials as the predicate device: - have the same shelf life; - use the same sterilization method; and - are packaged and sterilized using the same materials and processes. The contraindication statement was expanded to the proposed SorbaFix and PermaFix IFU's as a remedial action taken following the internal investigation of the medical device reports from one incidence on one patient during one procedure. The proposed SorbaFix Absorbable Fixation System and PermaFix Fixation System are made from materials identical to their respective predicates (SorbaFix K082396 and PermaFix K092483). The SorbaFix fasteners are composed of an absorbable polymer material, i.e. poly (D, L) lactide, dyed with D&C Violet No. 2 in accordance with 21 CFR 74.1602. The PermaFix fasteners are manufactured from a non-absorbable acetal copolymer material. The other patient contacting components in the delivery system are all made from the same stainless steel as that of the predicates. Full biocompatibility testing was conducted on both the predicate SorbaFix and PermaFix fasteners, the long term implants, in accordance with the ISO 10993 series requirements. The results of these tests were submitted under K082396 (Davol Absorbable Fastener System) and K092483 (PermaFix Fixation System) respectively. The biocompatibility data has ensured that the devices are safe for their intended use. No further biocompatibility testing is necessary because the proposed devices incorporate the same materials as their respective predicates. Other modifications covered under this Special 510(k) include: - labeling changes to remove the word "laparoscopic" from the 36 cm shaft devices for clarity purposes; - Modification of component dimensional specifications in the delivery system by removing the location of the clutch component from the end of the drive train towards the trigger to improve the drive mechanism: - Addition of a fastener gauge to the back of the ergonomic handle for user convenience. Tables summarizing the similarities and differences between the proposed devices and their respective predicates are provided in Section 15.0. # Substantial Equivalence: Design verification and validation testing were conducted and demonstrated that the proposed SorbaFix and PermaFix met the requirements set up in the product performance specifications for their respective predicate devices (SorbaFix K082396 and PREMARKET NOTIFICATION FOR THE SORBAFIX™ ABSORBABLE FIXATION SYSTEM AND PERMAFIXTM FIXATION SYSTEM > -CONFIDENTIAL-SECTION 8 {2}------------------------------------------------ PermaFix K092483). A summary of the design control activities are provided in Section 18.0 of this Premarket Notification. In summary, based on the performance data and the substantial equivalence comparison tables (reference Section 15.0), Davol Inc. believes that the proposed SorbaFix Absorbable Fixation System and PermaFix Fixation System are substantially equivalent to their respective predicates (SorbaFix K082396 and PermaFix K092483). PREMARKET NOTIFICATION FOR THE SORBAFIX™ ABSORBABLE FIXATION SYSTEM AND PERMAFIXTM FIXATION SYSTEM > -CONFIDENTIAL-SECTION 8 {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 C. R. Bard, Inc. % Davol, Inc. Ms. Zheng Liu Regulatory Affairs Project Manager 100 Crossings Boulevard Warwick, Rhode Island 02886 MAY 2 4 2011 Re: K111153 Trade/Device Name: SorbaFix™ and PermaFix™ Absorbable Fixation Systems Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: April 22, 2011 Received: April 25, 2011 Dear Ms. Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Zheng Liu Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510(k) Number (if known): SorbaFix™ Absorbable Fixation System Device Name: The SorbaFix Absorbable Fixation System is indicated for Indications for Use: the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. PermaFix™ Fixation System Device Name: The PermaFix Fixation System is indicated for the Indications for Use: approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Kacor for MXM (Division Sign Off) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K111153
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...