K142598 · Opsens, Inc. · DQX · Jun 12, 2015 · Cardiovascular
Device Facts
Record ID
K142598
Device Name
Opto Wire and OptoMonitor System
Applicant
Opsens, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Jun 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/ or other any interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease.
Device Story
OptoWire and OptoMonitor System measures intravascular blood pressure and fractional flow reserve (FFR). System comprises sterile, single-use fiberoptic pressure-sensing guidewire (OptoWire) and reusable electronic signal processing/display unit (OptoMonitor). Used in clinical settings (e.g., cath lab) by physicians during angiography/interventional procedures. OptoWire transmits pressure signals via fiber optics to OptoMonitor; monitor processes signals to display real-time pressure curves and numerical values (mean aortic/distal pressure, Pd/Pa ratio). Physician uses output to assess hemodynamic significance of vessel lesions, guiding treatment decisions. Benefits include accurate, real-time physiological assessment of blood vessel disease.
Clinical Evidence
No clinical data. Evidence consists of bench testing (mechanical, functional, electrical safety per IEC 60601-1, EMC per IEC 60601-1-2) and animal studies characterizing in-vivo handling, compatibility, and pressure measurement accuracy compared to predicate devices.
Technological Characteristics
Guidewire: 0.014" diameter, stainless steel/nitinol shaft, Teflon/silicone coating, fiberoptic pressure sensor. Monitor: LCD touchscreen, electrical isolation (Class 2), operating range 70-106 kPa, pressure range -30 to 300 mmHg. Connectivity: fiberoptic sensor to monitor. Sterilization: Ethylene oxide (ISO 11135-1). Software: IEC 62304 compliant.
Indications for Use
Indicated for patients requiring intravascular pressure measurement in coronary or peripheral blood vessels during diagnostic angiography or interventional procedures to obtain hemodynamic information, such as fractional flow reserve, for diagnosis and treatment of blood vessel disease.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
PrimeWire Prestige Plus Pressure Guide Wire (K111395)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2015
Opsens. Inc. % Pamela Papineau President Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, Massachusetts 01432
Re: K142598
Trade/Device Name: OptoWire and OptoMonitor System Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX, DXO Dated: May 13, 2015 Received: May 14, 2015
Dear Pamela Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
A. Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142598
#### Device Name OptoWire and OptoMonitor System
#### Indications for Use (Describe)
To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/ or other any interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size: 20px;">☑</span> </div> |
|----------------------------------------------|------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span style="font-size: 20px;">☐</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 – 510(k) Summary
### General Information
| Owner's Name: | Opsens, Inc. |
|-------------------------|---------------------------------------------------------------------------------------------|
| Address: | 2014 rue Cyrille Duquet, #125<br>Quebec, QC G1N 4N6 Canada |
| Telephone: | 418-682-9996 |
| Fax Number: | 418-682-9939 |
| Contact Person: | Pamela Papineau, RAC (US, EU, CAN)<br>Delphi Medical Device Consulting, Inc. |
| Address: | 5 Whitcomb Avenue<br>Ayer, MA 01432 |
| Telephone Number: | (978) 772-3552 |
| Fax Number: | (978) 796-5460 |
| Subject Device: | |
| Trade Name: | OptoWire and OptoMonitor System |
| Common Name: | Intravascular Pressure Monitoring System |
| Product Code: | DXQ / DXO |
| FDA Regulation: | 21 CFR 870.1330 - Catheter Guide Wire<br>21 CFR 870.2870 – Catheter Tip Pressure Transducer |
| Device Classification: | Class II |
| Predicate Devices: | |
| Product Name: | PrimeWire Prestige Plus Pressure Guide Wire (Volcano Corp.) |
| Common Name: | Pressure Guide Wire |
| Product Code: | DQX |
| FDA Regulation: | 21 CFR 870.1330 – Catheter Guide Wire |
| Device Classification: | Class II |
| Premarket Notification: | K111395 |
| Product Name: | ComboMap Pressure and Flow System (Volcano Corp.) |
ComboMap Pressure and Flow System (Volcano Corp.) Intravascular Pressure Monitoring System OBJ 21 CFR 870.1200 – Diagnostic Intravascular Catheter Class II K041134
## Indications for Use:
Premarket Notification:
Common Name:
FDA Regulation: Device Classification:
Product Code:
To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or other any interventional procedures. Blood pressure
{4}------------------------------------------------
measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease.
## Device Description:
The OptoWire and OptoMonitor System includes a pressure sensing guidewire (OptoWire), electronic signal processing and display units (OptoMonitor) that process signals received from the OptoWire to display intravascular blood pressure and fractional flow reserve (FFR) values, and various connection cables. The OptoWire and OptoWire cable are sterile, single-use devices; the remaining system components are reusable.
## Substantial Equivalence:
The OptoWire and OptoMonitor System is substantially equivalent to the Volcano Corp. PrimeWire Prestige Plus Pressure Sensing Guidewire (K111395) and the Volcano Corp. ComboMap Pressure and Flow System (K041134). Substantial equivalence is based on indications for use, technological characteristics, device materials, performance characteristics; physical and functional test results, and conformity with standards. A substantial equivalence summary table is provided at the end of this 510(k) Summary.
## Biocompatibility Testing:
Testing has been performed on the OptoWire device in accordance with the ISO 10993 family of standards to confirm that all patient-contacting materials are biocompatible. This testing included cytotoxicity, intracutaneous toxicity, acute systemic toxicity, sensitization, hemocompatibility and pyrogenicity.
#### Sterilization, Packaging and Shelf Llife:
The OptoWire is sterilized to a 106 SAL using an ethylene oxide process that has been validated in accordance with ISO 11135-1. The remaining system components are non-sterile. Testing has been performed to confirm that all packaging (sterile barrier and outer packaging) meets predetermined acceptance criteria with respect to sterile package seal strength and seal integrity, and that the packaging adequately protects all system components throughout transportation and storage. The sterile product shelf life has been validated through package and product testing of unaged devices as well as devices that have been held at an elevated temperature to simulate the labelled shelf life. Samples will also be tested after real time aging.
#### Software:
The OptoMonitor contains software that has been developed, documented and validated in accordance with industry standards (IEC 62304 - Medical device software - Software life cycle processes) and FDA guidance (General Principles of Software Validation; Final Guidance for FDA and Industry). Complete software design, risk and validation documents have been provided in accordance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Software validation has demonstrated that the system software operates in accordance with design specifications and the system operates as intended when used for its intended use.
{5}------------------------------------------------
# Electrical Safety and EMC Testing:
The OptoMonitor has been tested in accordance with IEC 60601-1 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance) to demonstrate electrical safety; testing in accordance with IEC 60601-1-2 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests) has confirmed electromagnetic compatibility.
# Performance Testing - Bench:
Bench testing has been performed to confirm that the OptoWire and OptoMonitor System meets predetermined acceptance criteria and is substantially equivalent to the predicate devices. The following is a summary of the performance (bench) testing presented in this 510(k):
| Component | Testing | Purpose | Met Acceptance<br>Criteria? | |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| | | | | |
| | OptoWire | | | |
| Mechanical /<br>Functional<br>Verification | • Tip Testing<br>• Wire<br>Flexibility/Support<br>• Tensile Test<br>• Torque Tests<br>• Turn to Failure<br>• Fatigue Test<br>• Coating Testing<br>• Compatibility Test<br>• Surface Inspection<br>• Corrosion Test<br>• Dimensional testing | • Sensor Sensitivity to<br>Bending<br>• Pressure and Sensor<br>Testing<br>• Optical Contrast<br>• Optical Sensor Testing<br>• Ambient Pressure<br>Range<br>• Signal Loss<br>• Connection/<br>Disconnection Testing<br>• Age testing | • Functionality<br>and<br>Performance<br>• Safety<br>• Usability | Yes. All internal<br>design criteria were<br>successfully met.<br>Subject device<br>compared favorably<br>to predicate devices<br>tested. |
| OptoMonitor | | | | |
| • SCU<br>Hardware<br>• DU<br>Hardware<br>• HCU<br>Hardware<br>• System | • visual & design<br>• specs & environment<br>• single fault conditions<br>• electrical performance | • Age Testing<br>• Transportation Testing<br>• Cleaning | • Functionality<br>and<br>Performance<br>• Reliability<br>• Safety<br>• Usability<br>• Environmental | Yes. All internal<br>design criteria were<br>successfully met. |
## Performance Testing - Animal:
Animal studies have been performed to characterize the in-vivo performance of the OptoWire and OptoMonitor System, as compared to the predicate device and other currently marketed devices with the same indications for use. Animal testing included assessment of the system handling characteristics, compatibility with other devices, and pressure measurement capabilities.
## Conclusion:
The OptoWire and OptoMonitor System has been demonstrated to be substantially equivalent to the predicate devices.
{6}------------------------------------------------
| | OptoWire and OptoMonitor System<br>(current submission; Opsens, Inc.) | PrimeWire Prestige Plus<br>Pressure Guide Wire<br>(K111395; Volcano Corp.) | ComboMap Pressure and Flow System<br>(K041134; Volcano Corp.) |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Common/Usual Name | Intravascular Pressure Guide Wire<br>and Monitoring System | Intravascular Pressure Guide Wire | Intravascular Pressure Monitoring System |
| Device Class | Class II | Class II | Class II |
| Product Code / Regulation | DQX / 21 CFR 870.1330<br>DXO / 21 CFR 870.2870 | DQX / 21 CFR 870.1330 | OBJ / 21 CFR 870.2870 |
| Regulation Name | Pressure Guide Wire<br>Diagnostic Intravascular Monitor | Pressure Guide Wire | Diagnostic Intravascular Monitor |
| Indications for Use | To measure pressure in blood vessels,<br>including both coronary and peripheral<br>vessels, during diagnostic angiography<br>and/or any interventional procedures.<br>Blood pressure measurements provide<br>hemodynamic information, such as<br>fractional flow reserve, for the diagnosis<br>and treatment of blood vessel disease. | To measure pressure in blood vessels,<br>including both coronary and peripheral<br>vessels, during diagnostic angiography<br>and/or any interventional procedures.<br>Blood pressure measurements provide<br>hemodynamic information for the diagnosis<br>and treatment of blood vessel disease.<br>Intended for use with signal processing<br>system such as Volcano ComboMap. | A multi-mode system intended for use in<br>all blood vessels, including coronary and<br>peripheral arteries, to measure<br>intravascular blood pressure and/or blood<br>flow velocities during diagnostic<br>angiography and/or interventional<br>procedures.<br>Intended for use with pressure sensing<br>guidewire such as Volcano PrimeWire. |
| System Components | Sterile, disposable guidewire<br>Reusable signal processor / monitor<br>Embedded software<br>Connecting cables | Sterile, disposable guidewire | Reusable signal processor / monitor<br>Embedded software<br>Connecting cables |
| System Capabilities | Measurement of intravascular blood<br>pressure including FFR. | Measurement of intravascular blood<br>pressure and flow including FFR<br>(when used with pressure/flow system) | Measurement of intravascular blood<br>pressure and flow including FFR<br>(when used with pressure wire) |
| Prescription Use | Rx Only | Rx Only | Rx Only |
| Pressure Sensing & Signal<br>Transmission Technology | Fiberoptic sensor & fiber bundle embedded<br>in guidewire | Hard wired strain gauge embedded in<br>guidewire | Hard wired strain gauge embedded in<br>compatible guidewire |
| | OptoWire and OptoMonitor System<br>(current submission; Opsens, Inc.) | PrimeWire Prestige Plus<br>Pressure Guide Wire<br>(K111395; Volcano Corp.) | ComboMap Pressure and Flow System<br>(K041134; Volcano Corp.) |
| Sterile, Single Use Patient<br>Contact Component? | Yes – OptoWire | Yes – PrimeWire Prestige Plus | No – computer-controlled pressure and<br>flow monitoring instrument intended for<br>use with compatible pressure guidewire |
| FFR Capability? | Yes | Yes | Yes |
| FFR Viewing? | Yes | N/A | Yes |
| Basis for FFR Determination | Simultaneous acquisition of 2 pressure<br>values: distal pressure from sensor<br>embedded in OptoWire; aortic pressure<br>from external pressure transducer | Simultaneous acquisition of 2 pressure<br>values: distal pressure from sensor<br>embedded in PrimeWire; aortic pressure<br>from external pressure transducer | Simultaneous acquisition of 2 pressure<br>values: distal pressure from sensor<br>embedded in compatible guidewire; aortic<br>pressure from external pressure transducer |
| Operating Temperature<br>(Monitor) | 15°C to 30°C | N/A | 16°C to 32°C |
| Transport Temperature<br>(Monitor) | -25°C to 60°C | N/A | -20°C to 60°C |
| Operating Relative Humidity<br>(Monitor) | 10% to 85% non-condensing | N/A | 30% to 75% |
| Storage Temperature<br>(Monitor) | Room Temperature | N/A | -20°C to 60°C |
| Operating Pressure | 70 to 106 kPa | N/A | 50 to 106 kPa |
| Pressure Range | -30 to 300 mmHg | -30 to 300 mmHg | -30 to 300 mmHg |
| Pressure Accuracy | +/- 1 mmHg plus +/- 1% of reading<br>(pressure range -30 to 50 mmHg) or +/- 3%<br>of reading (pressure range 50 to 300<br>mmHg) | Unknown | +/- 1 mmHg plus +/- 1% of reading<br>(pressure range -30 to 50 mmHg) or +/-<br>3% of reading (pressure range 50 to 300<br>mmHg) |
| Thermal Zero Shift | <0.3 mmHg/deg C | <0.3 mmHg/deg C | N/A |
| Zero Drift | <1 mmHg/h | Unknown | N/A |
| Electrical Isolation | Class 2 (double isolation) | N/A | Class 1 |
| User Interface | Touchscreen | N/A | Touchscreen, Remote Control |
| Auto-zeroing | Yes | N/A | Yes |
| | OptoWire and OptoMonitor System<br>(current submission; Opsens, Inc.) | PrimeWire Prestige Plus<br>Pressure Guide Wire<br>(K111395; Volcano Corp.) | ComboMap Pressure and Flow System<br>(K041134; Volcano Corp.) |
| Real Time Curves | Aortic instantaneous pressure, aortic mean pressure, distal instantaneous pressure,<br>distal mean pressure | N/A | Aortic instantaneous pressure, aortic mean pressure, distal instantaneous pressure,<br>distal mean pressure, Pa/Pd trend curve |
| Real Time Numerical Values | Mean aortic pressure, mean distal pressure,<br>mean Pd/mean Pa | N/A | Mean aortic pressure, mean distal pressure,<br>mean Pd/mean Pa, heartbeat |
| Minimum Pd/Pa Cursor<br>(Detection of FFR Locus) | Yes | N/A | Yes |
| Recording Values | Instantaneous Pa, Pd and Pd/Pa; mean Pa;<br>mean Pd; mean Pd/mean Pa | N/A | Instantaneous Pa, Pd and Pd/Pa; mean Pa;<br>mean Pd; mean Pd/mean Pa |
| Display Monitor | LCD | N/A | LCD |
| Aortic Input | High Level (100 mmHg/V) | N/A | High Level (100 mmHg/V)<br>Low Level (5uV/V/mmHg) |
| Aortic Output | No | N/A | Low Level (5uV/V/mmHg) |
| Distal Input | OptoWire (optical) | N/A | Compatible pressure wire (PrimeWire)<br>(electrical) |
| Distal Output | Low Level (5uV/V/mmHg) | N/A | Low Level (5uV/V/mmHg) |
| Guidewire OD | 0.014" | 0.014" | N/A |
| Guidewire Length | 175 cm | 185 cm, 300 cm | N/A |
| Guidewire Shaft Material | Stainless Steel; Nitinol | Stainless Steel; SS | N/A |
| Guidewire Coating | Teflon; Silicone | Teflon; GlyDx Hydrophilic coating | N/A |
| Guidewire Tip Configuration | Straight | Straight, pre-shaped "J" | N/A |
| Guidewire Tip Length | 3.5 cm | 3.0 cm | N/A |
| Radiopaque Tip? | Yes | Yes | N/A |
| Pressure Sensor Location | 3.5 cm from distal tip | 3.0 cm from distal tip | N/A |
# Substantial Equivalence Comparison
{7}------------------------------------------------
{8}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.