The PrimeWire PRESTIGE® Plus Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.
Device Story
Steerable 0.014" guide wire with pressure transducer mounted 3 cm proximal to tip; used to measure intravascular blood pressure. Operates by connecting to SmartMap®, s5™ Family, or ComboMap® systems to provide hemodynamic data. Used in clinical settings (e.g., cath lab) by physicians during diagnostic or interventional procedures. Enhancements include hydrophilic coating for improved torqueability and lubricity compared to predicate. Output displayed on compatible systems to assist clinicians in diagnosing and treating blood vessel disease.
Clinical Evidence
Bench testing only. Performance verified via design verification plan including biocompatibility (ISO 10993-1), particulate testing (USP 788), tensile strength, and torqueability. No clinical data presented.
Technological Characteristics
0.014" diameter guide wire; 185 cm or 300 cm lengths; pressure transducer 3 cm from tip. Features hydrophilic coating. Biocompatibility per ISO 10993-1. Connectivity via preconnected cable to Volcano console systems. Sterilization method leveraged from predicate.
Indications for Use
Indicated for patients undergoing diagnostic angiography or interventional procedures requiring blood pressure measurement in coronary or peripheral vessels.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
K111395
Volcano Corporation July 25, 2011
PrimeWire PRESTIGE®Plus Pressure Guide Wire Special 510(k)
#### 510 (K) Summary
AUG - 1 2011
# PrimeWire PRESTIGE® Plus Pressure Guide Wire
| Date Prepared: | July 7, 2011 | |
|----------------------|-----------------------------------------------------------------------------------|-------------|
| Submitted by: | Volcano Corporation<br>3661 Valley Centre Dr.<br>Suite 200<br>San Diego, CA 92130 | |
| Contact person: | Marilyn Pourazar<br>Senior Director of Regulatory Affairs | |
| Phone number: | (858) 720-4116 | |
| Facsimile number: | (858) 720-0612 | |
| Device Name: | PrimeWire PRESTIGE® Plus Pressure Guide Wire | |
| Classification name: | > 870.1330 Catheter guide wire | Class<br>II |
| | > 870.2870 Catheter tip pressure transducer | II |
#### Predicate Device:
The Volcano PrimeWire PRESTIGE® Plus Pressure Guide Wire is substantially equivalent to the following:
| 510(k) Number | Product Name | Clearance Date |
|---------------|-----------------------------------------|----------------|
| K100930 | PrimeWire PRESTIGE® Pressure Guide Wire | April 30, 2010 |
#### Device Description:
The PrimeWire PRESTIGE® Plus pressure guide wire is a steerable guide wire with a pressure transducer mounted 3 cm proximal to the tip. The PrimeWire PRESTIGE® Plus guide wire measures pressure when used with the SmartMap®, s5™ Family, and ComboMap® systems. The PrimeWire PRESTIGE® Plus guide wire has a diameter of 0.014" (0.36 mm) and is available in lengths of 185 cm or 300 cm, and also in straight or pre-shaped tips. The PrimeWire PRESTIGE® Plus guide wire is packaged preconnected to the connector with a torque device to facilitate navigation through the vasculature.
l
{1}------------------------------------------------
#### Intended Use:
The PrimeWire PRESTIGE® Plus Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.
### Device Technological Characteristics and Comparison to Predicate Device:
The Volcano Corporation PrimeWire PRESTIGE® Plus Pressure Guide Wire Device is substantially equivalent to the predicate device, PrimeWire PRESTIGE® Pressure Guide Wire.
The PrimeWire PRESTIGE® Plus Pressure Guide Wire Device uses the same fundamental scientific technology and has the same intended use as that of the predicate device.
#### Performance Data:
Applicable testing was performed in accordance with the Design Verification Plan including a Risk Analysis addressing the impact of enhancements to the device and components. The test results indicate the PrimeWire PRESTIGE® Plus Pressure Guide Wire is comparable to the predicate device.
| Design Input Description | Design Change | Required Testing,<br>Leveraged or<br>Justification |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Working Length - product shall come<br>in 185cm and 300 cm lengths | No change from predicate device. | Testing |
| Flexible Distal Length - length of<br>flexible section shall be ~ 30 cm in<br>length | No change from predicate device. | Testing |
| Distal Radiopaque Length - length<br>of distal tip shall be 3.0 cm | No change from predicate device. | Testing |
| Design Input Description | Design Change | Required Testing,<br>Leveraged or<br>Justification |
| Maximum outer diameter of .0145" | No change from predicate device. | Justification –<br>100% verified in<br>manufacturing |
| The distal 1.0 cm of the wire shall be<br>equivalent stiffness to the predicate<br>device tip. | No change from predicate device. | Testing |
| Better torqueability than predicate<br>device in distal or tortuous anatomy. | Replaced silicon-based coating with<br>hydrophilic coating | Testing |
| Maintain lubricity within the body<br>throughout procedure (approx. 60<br>minutes). | Replaced silicon-based coating with<br>hydrophilic coating | Testing |
| Particulate generation in simulated use<br>shall meet USP 788: Maximum of 3000<br>particles $\ge$ 10 microns; maximum of<br>300 particles $\ge$ 25 microns. | Replaced silicon-based coating with<br>hydrophilic coating | Testing |
| Sensor Housing to Core Tensile shall<br>be $\ge$ 1.0 pounds. | No change from predicate device. | Testing |
| Tip Tensile $\ge$ 1.0 pounds. | No change from predicate device. | Testing |
| Turns to Failure $\ge$ 10 turns | No change from predicate device. | Testing |
| Wire connector must withstand 10<br>insertions in connector. | No change from predicate device. | Testing |
| Sensor shall have same accuracy as<br>predicate device | No change from predicate device. | Leveraged from<br>previously cleared<br>predicate device |
| Sensor drift shall not exceed 5 mmHG<br>in 10 minutes | No change from predicate device. | Leveraged from<br>previously cleared<br>predicate device and<br>100% verified in<br>manufacturing |
| Subjecting PLUS to clinically relevant<br>tortuosity shall not cause erratic output | No change from predicate device. | |
| 3 year shelf life | Replaced silicon-based coating with<br>hydrophilic coating | Testing |
| Biocompatibility | Replaced silicon-based coating with<br>hydrophilic coating | Testing |
| Packaging | No change from predicate device. | Leveraged from<br>previously cleared<br>predicate device |
| Sterilization | No change from predicate device. | Leveraged from<br>previously cleared<br>predicate device |
## Summary of Performance Characteristics
#### K111395 2
{2}------------------------------------------------
{3}------------------------------------------------
# Biocompatibility Studies:
The biocompatibility of the PrimeWire PRESTIGE® Plus was verified in accordance with ISO 10993-1, Biological Evaluation of Medical Devices and FDA Memorandum # G95-1. Test results confirmed biocompatibility of the pressure wire was tested as an external communicating, circulating blood contact, limited exposure (<24 hrs) device.
The following table summarizes biocompatibility test results on the following PrimeWire PRESTIGE® Plus materials:
| Test Description | Results |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Cytotoxicity Study using the Colony Assay Extraction Method | Pass |
| ISO Intracutaneous Study (Irritation Test), Extract 0.9% NaCl | Pass |
| ISO Intracutaneous Study (Irritation Test), Extract Sesame Oil | Pass |
| ISO Systemic Toxicity Study, Sesame Oil | Pass |
| ISO Systemic Toxicity Study, 0.9% NaCl | Pass |
| ISO Maximum Sensitization Study, Extract Sesame Oil | Pass |
| ISO Maximum Sensitization Study, Extract 0.9% NaCl | Pass |
| Pyrogen Study, Material Mediated, 0.9% NaCl | Pass |
| ASTM Partial Thromboplastin Time - Plasma Extract | Pass |
| ASTM Hemolysis CMF PBS Extract | Pass |
| C3a Complement Activation (Hemocompatibility), Normal<br>Human Serum Extract | Pass |
| SC5b-9 Complement Activation (Hemocompatibility), Human<br>Serum Extract | Pass |
| In Vivo Thromboresistance Study, Peripheral Vessel / Jugular<br>Vein (vessel to be determined based on device evaluation)<br>(Note: the control device will be identified in the NAMSA<br>protocol before test execution) | Pass |
| Preliminary Extraction, Japanese MHLW | Pass |
| Exaggerated Extraction Method 1 or 2, Japanese MHLW (test<br>type depends on preliminary extraction results) | Pass |
| Maximum Sensitization Study, Method 1 or 2, MHLW (test<br>type depends on preliminary extraction results) | Pass |
Image /page/3/Picture/6 description: The image contains a sequence of digits and a letter. The sequence starts with the letter 'K', followed by four instances of the digit '1', and then the digits '395'. There is also a small superscript '4' to the right of the number.
{4}------------------------------------------------
Volcano Corporation July 25, 2011
# Conclusion:
The Volcano PrimeWire PRESTIGE® Plus Pressure Guide Wire Device has the same performance specifications, fundamental scientific technology and intended use as that of the predicate device, PrimeWire PRESTIGE® Pressure Guide Wire. The design enhancements to the device do not raise any new questions regarding safety or efficacy. The performance data and a declaration of conformity with design controls support a determination of continuing substantial equivalence of the modified device to the predicate device.
\$\qquad K11139\overline{5}^{5}\$
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three wavy lines representing the eagle's body. The eagle is facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Volcano Corporation c/o Ms. Marilyn Pourazar Senior Director, Regulatory Affairs 3661 Valley Centre Drive Suite 200 San Diego, CA 92130
AUG - 1 2011
Re: K111395 Trade/Device Name: PrimeWire PRESTIGE® Plus Pressure Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter, Percutaneous Regulatory Class: Class II Product Code: DQX Dated: July 12, 2011 Received: July 13, 2011
Dear Ms. Pourazar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr re and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractly and i warranties. We remind you, however, that device labeling must be truthful and not much you
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements if the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Ms. Marilyn Pourazar
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersQffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
K
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K11395
Device Name: PrimeWire PRESTIGE® Plus Pressure Guide Wire Device
Indications for Use:
The PrimeWire PRESTIGE® Plus Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
---
(Division Sign-Off)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Numbe
Page 1 of
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.