K142218 · Globus Medical, Inc. · ODP · Mar 31, 2015 · Orthopedic
Device Facts
Record ID
K142218
Device Name
Coalition AGX Plate and Coalition AGX Spacer
Applicant
Globus Medical, Inc.
Product Code
ODP · Orthopedic
Decision Date
Mar 31, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The COALITION AGX™ Plate is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis. The COALITION AGX™ Spacer is an interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-TI) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The COALITION AGX™ Spacer is to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation such as the COALITION AGX™ Plate, ASSURE®, PROVIDENCE™, VIP®, XTEND®, or UNIFY™ Anterior Cervical Plate Systems. When used with the COALITION AGX™ Plate, the assembly takes on the indications for use of the COALITION AGX™ Spacer, with the COALITION AGXTM Plate acting as the supplemental fixation.
Device Story
COALITION AGX™ system comprises anterior cervical plates and interbody fusion spacers; used for spinal stabilization and fusion. Spacers are inserted into the intervertebral space following discectomy; filled with autogenous bone graft to promote fusion. Plates provide supplemental anterior screw fixation to the cervical spine. Protrusions on spacer surfaces grip vertebral endplates to resist expulsion. System used by surgeons in clinical settings. Provides structural stability; facilitates fusion; addresses discogenic pain and spinal deformities. Performance verified through mechanical testing and cadaveric studies.
Clinical Evidence
No clinical data. Evidence consists of bench testing (static and dynamic compression, torsion, shear, expulsion, subsidence, and plate pull-off) conducted per ASTM F1717, F2077, and F2267, and biomechanical cadaveric testing to validate performance and substantial equivalence.
Technological Characteristics
Plates and screws manufactured from titanium alloy (ASTM F136, F1295, F1472). Spacers manufactured from radiolucent PEEK polymer (ASTM F2026) with titanium alloy or tantalum markers (ASTM F136, F1295, F1472, F560). System is a mechanical implant; no software or energy source.
Indications for Use
Indicated for skeletally mature patients with cervical spine (C2-T1) degenerative disc disease (DDD) at one level, or cervical spine (C2-C7) trauma, tumors, deformity, pseudarthrosis, failed fusion, spondylolisthesis, or spinal stenosis. Requires 6 weeks of non-operative treatment for DDD. Contraindications include skeletal immaturity.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, suggesting a sense of community and support. The profiles are positioned within a circular frame, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" encircling the figure. The text is written in a simple, sans-serif font, and the overall design is clean and professional.
April 9, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Globus Medical, Incorporated Kelly J. Baker, Ph.D. Senior Vice President, Regulatory and Clinical Affairs 2560 General Armistead Avenue Audubon, Pennsylvania 19403
Re: K142218
Trade/Device Name: COALITION AGX™ Plate and COALITION AGX™ Spacer Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, OVE, KWQ Dated: March 31, 2015 Received: March 31, 2015
Dear Dr. Baker:
This letter corrects our substantially equivalent letter of March 31, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
Page 2 - Kelly J. Baker, Ph. D.
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142218
Device Name
COALITION AGX™ Plate and COALITION AGX™ Spacer
#### Indications for Use (Describe)
The COALITION AGX™ Plate is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.
The COALITION AGX™ Spacer is an interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-TI) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The COALITION AGX™ Spacer is to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation such as the COALITION AGX™ Plate, ASSURE®, PROVIDENCE™, VIP®, XTEND®, or UNIFY™ Anterior Cervical Plate Systems. When used with the COALITION AGX™ Plate, the assembly takes on the indications for use of the COALITION AGX™ Spacer, with the COALITION AGXTM Plate acting as the supplemental fixation.
Type of Use (Select one or both, as applicable)
| <div> <span> 区 Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------|
| <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
==============================================================================================================================================================================
{3}------------------------------------------------
# 510(k) SUMMARY: COALITION AGX™ System
| Company: | Globus Medical Inc.<br>2560 General Armistead Avenue<br>Audubon, PA 19403<br>(610) 930-1800 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Kelly Baker, Ph.D.<br>Senior Vice President, Regulatory and Clinical Affairs |
| Date Prepared: | March 23, 2015 |
| Device Name: | COALITION AGXTM Plate and COALITION AGXTM Spacer |
| Classification: | 21CFR §888.3080 Intervertebral Body Fusion Devices<br>21CFR §888.3080 Intervertebral Fusion Device with<br>Integrated Fixation, Cervical<br>Product Codes: ODP, OVE<br>Regulatory Class: II; Panel Code: 87<br>21CFR §888.3060 Spinal Intervertebral Body Fixation<br>Orthosis<br>Product Code: KWQ<br>Regulatory Class: II; Panel Code: 87 |
| Primary Predicate: | SUSTAIN® Radiolucent Medium (K130478) |
| Additional<br>Predicates: | VIP® Anterior Cervical Plate System (K081391)<br>InterPlateTM C (K092070)<br>IN:C2TM Spinal Fixation System (K122630)<br>PEEK Prevail® Cervical Interbody Device (K073285)<br>PATRIOT® Colonial® ACDF Spacer (K072991)<br>COALITION® Spacer (K083389) |
#### Purpose:
The purpose of this submission is to request clearance for the COALITION AGX™ Plate and COALITION AGX™ Spacer.
#### Device Description:
The COALITION AGX™ Plate is an anterior, cervical fixation device available in various heights and widths to fit the anatomical needs of a wide variety of patients. The plates are made from titanium alloy, as specified in ASTM F136, F1295, and F1472. The screws for use with the COALITION AGX™ Plates are manufactured from titanium alloy, as specified in ASTM F136 and F1295.
{4}------------------------------------------------
COALITION AGX™ Spacers are anterior cervical interbody fusion devices used to provide structural stability in skeletally mature individuals following discectorny. The spacers are available in various heights and footprints to fit the anatomical needs of a wide variety of patients. These devices are to be filled with autogenous bone graft material. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to resist expulsion. The COALITION AGX™ Spacers are manufactured from radiolucent PEEK polymer. with titanium alloy or tantalum markers, as specified in ASTM. F2026, F136. F1295, F1472, and F560.
#### Indications for Use:
The COALITION AGX™ Plate is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies). trauma (including fractures). tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.
The COALITION AGX™ Spacer is an interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The COALITION AGX™ Spacer is to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the COALITION AGX™ Plate, ASSURE®, PROVIDENCE™, VIP®, XTEND®, or UNIFY™ Anterior Cervical Plate Systems. When used with the COALITION AGX™ Plate, the assembly takes on the indications for use of the COALITION AGX™ Spacer, with the COALITION AGX™ Plate acting as the supplemental fixation.
## Performance Data:
Mechanical testing for the COALITION AGX™ Plate (static and dynamic compression bend, and static torsion), for the COALITION AGX™ Spacer (static and dynamic compression, static and dynamic compression-shear, static and dynamic torsion, and subsidence), and for the COALITION AGX™ Plate and Spacer (static compression, compression-shear, torsion, expulsion, subsidence, and plate pull-off) was conducted in accordance with the Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s (May 3, 2004) and ASTM F1717, F2077 and F2267, as applicable. Design validation/cadaver testing was conducted to ensure the COALITION AGX™ System performance is acceptable for its intended use and to ensure substantial equivalence to the predicate(s). Biomechanical cadaveric testing was also conducted. Performance and comparative analysis data demonstrate substantial equivalence to the predicate device(s).
{5}------------------------------------------------
K142218 Page 3 of 3
# Basis for Substantial Equivalence:
COALITION AGX™ implants are similar to the predicate devices with respect to technical characteristics, performance, design, materials, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices. COALITION AGX™ implants are as safe, as effective, and perform as well as or better than predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.