IN:C2 SPINAL FIXATION SYSTEM

K122630 · Spine Smith Partners L.P. · OVE · Dec 5, 2012 · Orthopedic

Device Facts

Record IDK122630
Device NameIN:C2 SPINAL FIXATION SYSTEM
ApplicantSpine Smith Partners L.P.
Product CodeOVE · Orthopedic
Decision DateDec 5, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The IN:C2 Spinal Fixation System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. IN:C2 is a stand-alone device intended to be used with an anterior cover plate and a minimum two provided bone screws angled both cephalad and caudal with a minimum of one screw into each vertebral body. The implants are to be used with autogenous bone graft and implanted via an open, anterior approach. Patients should have at least six weeks of non-operative treatment prior to treatment with an intervertebral cage.

Device Story

IN:C2 Spinal Fixation System is a stand-alone intervertebral body fusion device used during spinal fusion to stabilize the spine while bony fusion develops. The device consists of a 'U' shaped PEEK block with titanium markers for radiological positioning, an anterior cover plate assembly, and two titanium bone screws. It is implanted via an open anterior approach by a surgeon. The cover plate and screws provide supplemental fixation, with screws angled cephalad and caudal into vertebral bodies. The device is used with autogenous bone graft. It benefits patients by providing structural support and stabilization for cervical disc disease.

Clinical Evidence

No clinical data. Evidence consists of bench testing including static axial compression, subsidence testing, and pushout testing per ASTM standards (F2077-03, F2267-04, F1877-05).

Technological Characteristics

Materials: PEEK block, titanium markers, titanium bone screws. Design: 'U' shaped intervertebral cage with anterior cover plate assembly. Connectivity: None. Sterilization: Not specified. Software: None. Standards: ASTM F2077-03, ASTM F2267-04, ASTM F1877-05.

Indications for Use

Indicated for anterior cervical interbody fusion in skeletally mature patients with cervical disc disease (intractable radiculopathy/myelopathy, herniated disc, or osteophyte formation with nerve root/spinal cord compression) at one level (C2-C3 to C7-T1). Requires 6 weeks prior non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### SpineSmith IN:C2 Spinal Fixation Svstem #### 5 2012 510(k) Summary of Safety and Effectiveness DEC EC 5 2012 SUBMITTED BY SpineSmith Partners, LLP 93 Red River Austin, TX 78701 3006404071 ESTABLISHMENT REGISTRATION NUMBER CONTACT PERSON Clifton (Chris) Naivar Director - Quality and Regulatory Affairs Phone: 512-637-2068 Fax: 512-637-2096 Email: cnaivar@spinesmithusa.com SUBMISSION PREPARED BY Clifton (Chris) Naivar Director - Quality and Regulatory Affairs Phone: 512-637-2068 DATE PREPARED September 30, 2012 OVE - intervertebral fusion device with integrated CLASSIFICATION fixation, cervical REGULATION NUMBER 21 CFR 888.3080 Intervertebral Body Fusion Device COMMON NAME SpineSmith IN:C2 Spinal Fixation System PROPRIETARY NAME #### IDENTIFICATION OF PREDICATE DEVICES: The SpineSmith IN:C2 System was determined to be substantially equivalent to the previously cleared Cimplicity System (K080971) as well as the Coalition Spacer (K083389) and the InterPlate C-Ti (K092070). #### DEVICE DESCRIPTION: SpineSmith Partners LLP developed the IN:C2 Spinal Fixation System to be used during spinal fusion. IN:C2 serves to stabilize the spine while bony fusion develops. The IN:C2 System consists of a 'U' shaped PEEK block in multiple footprint configurations, heights and lordosis angles. The PEEK implants contain a titanium marker intended to verify position radiologically. The IN:C2 is a stand-alone system, intended for use with its cover plate assembly and two titanium bone screws provided. The IN:C2 implant is intended to be implanted via an open anterior approach. {1}------------------------------------------------ The anterior cover plate assembly attaches to the anterior most portion of the device. and includes housing features for placement of two bone screws angled cephalad and caudal. The cover plate assembly and integrated screws are supplemental fixation. #### INDICATIONS: The IN:C2 Spinal Fixation System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. IN:C2 is a stand-alone device intended to be used with an anterior cover plate and a minimum two provided bone screws angled both cephalad and caudal with a minimum of one screw into each vertebral body. The implants are to be used with autoqenous bone graft and implanted via an open, anterior approach. Patients should have at least six weeks of non-operative treatment prior to treatment with an intervertebral cage. #### MECHANICAL TESTING: The following non-clinical tests were conducted: - ASTM F2077-03, Test Methods for Intervertebral Body Fusion Devices. ● - ASTM F2267-04. Standard Test Method for Measuring Load Induced . Subsidence of an Intervertebral Body Fusion Device Under Static Axial Compression. - Draft Document Submitted to ASTM F-04.25.02.02. Static Pushout Test . Method for Intervertebral Body Fusion Devices. - ASTM F1877-05 Standard Practice for Characterization of Particles . #### CONCLUSIONS: The subject and predicate device share the same intended use, primary implant design and material of manufacture. The non-clinical mechanical test results demonstrate that any minor differences do not impact device performance as compared to the predicates and demonstrate that the IN:C2 System is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines extending from its base, enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 December 5, 2012 Spine Smith Partners, L.P. % Clifton Naivar Director, Quality and Regulatory Affairs 93 Red River Austin, Texas 78701 Re: K122630 Trade/Device Name: IN:C2 Spinal Fixation System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE Dated: November 2, 2012 Received: November 5, 2012 #### Dear Mr. Naivar We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Clifton Naivar forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. # Sincerely vours. Mark N. Méikerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### INDICATIONS FOR USE #### 510(k) Number (if known): K122630 Device Name: #### IN:C2 Spinal Fixation System Indications for Use: The IN:C2 Spinal Fixation System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with hemiated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. IN:C2 is a stand-alone device intended to be used with an anterior cover plate and a minimum two provided bone screws angled both cephalad and caudal with a minimum of one screw into each vertebral body. The implants are to be used with autogenous bone graft and implanted via an open, anterior approach. Patients should have at least six weeks of non-operative treatment prior to treatment with an intervertebral cage. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ## Caroline Rhim -S (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: 122630
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...