COALITION SPACER

K083389 · Globus Medical, Inc. · OVE · Mar 26, 2009 · Orthopedic

Device Facts

Record IDK083389
Device NameCOALITION SPACER
ApplicantGlobus Medical, Inc.
Product CodeOVE · Orthopedic
Decision DateMar 26, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The COALITION™ Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION™ Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant.

Device Story

COALITION™ Spacer is a stand-alone cervical interbody fusion device; provides structural stability following discectomy. Device features radiolucent polymer body with titanium alloy or tantalum markers; superior/inferior surface protrusions grip vertebral endplates to resist expulsion. Titanium alloy screws inserted through anterior portion into adjacent vertebral bodies for bony fixation. Implant filled with autogenous bone graft. Used in clinical setting by surgeons for cervical spine fusion. Benefits include stabilization of spinal segment in DDD patients.

Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and performance characteristics consistent with the Class II Special Controls Guidance Document for Intervertebral Fusion Devices.

Technological Characteristics

Materials: Radiolucent polymer, titanium alloy, tantalum (ASTM F2026, F136, F1295, F560). Design: Interbody fusion spacer with integrated screw fixation. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-T1) at one level, defined as discogenic pain with degeneration confirmed by history and radiographic studies, who have failed at least six weeks of nonoperative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K083389 Page\$\frac{1}{1}\$ # 510(k) SUMMARY: COALITION™ SPACER MAR 2 6 2009 - Company: Globus Medical Inc. 2560 General Armistead Ave. Audubon. PA 19403 (610) 415-9000 Kelly J. Baker, Ph.D Contact: Director, Clinical Affairs & Regulatory - Device Name: COALITION™ Spacer Classification: Product Code ODP. Class II. 21 CFR §888.3080 Intervertebral body fusion device. - PATRIOT™ Predicate(s): Cervical Spacer (Colonial™ ACDF Spacer) K072991 and other legally marketed devices. #### Device Description: The COALITION™ Spacer is a stand-alone cervical interbody fusion device used to provide structural stability in skeletally mature individuals following discectorny. The spacers are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to aid in expulsion resistance. Screws are inserted through the anterior titanium portion of the implant into adjacent vertebral bodies for bony fixation. The spacer is to be filled with autogenous bone graft material. The COALITION™ Spacer is made from radiolucent polymer, with titanium alloy or tantalum markers, as specified in ASTM F2026, F136, F1295, and F560. The anterior portion of the implant and the mating screws are manufactured from titanium alloy, as specified in ASTM F136 and F1295. ### Intended Use: The COALITION™ Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION™ Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant. ## Basis for Substantial Equivalence: The COALITION™ Spacer has been evaluated in accordance with the "Class II Special Controls Guidance Document: Intervertebral Fusion Device", June 12, 2007 and have been found to meet the criteria set forth in the quidance document in terms of indications, design, and performance. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is written around the perimeter of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Globus Medical, Inc. % Kelly J. Baker, Ph.D. Director, Clinical Affairs & Regulatory 2560 General Armistead Avenue Audubon, Pennsylvania 19403 SEP 12 2011 Re: K083389 Trade/Device Name: COALITION™ Spacer Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: March 23, 2009 Received: March 24, 2009 Dear Dr. Baker: This letter corrects our substantially equivalent letter of March 26, 2009. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other {2}------------------------------------------------ Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Sincerely yours, Mark A. Milken Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Indications for Use Statement X083389 510(k) Number: Device Name: COALITION™ Spacer ## Indications: The COALITION™ Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION™ Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant. OR Prescription Use × (Per 21 CFR §801.109) Over-The-Counter Use (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) D. Thm Division Division of General, Restorative, and Neurological D 510(k) Number K68339
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...