K141580 · Tso3, Inc. · FRC · Dec 17, 2014 · General Hospital
Device Facts
Record ID
K141580
Device Name
STERIZONE SELF-CONTAINED BIOLOGICAL INDICATOR
Applicant
Tso3, Inc.
Product Code
FRC · General Hospital
Decision Date
Dec 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The STERIZONE® BI+ Self-contained Biological Indicator (SCBI) is intended for routine monitoring of the STERIZONE® VP4 Sterilizer, which offers a single pre-set sterilization cycle ("Cycle 1"). The SCBI should only be used in a Test Pack configuration to monitor Cycle 1. The SCBI placed within the STERIZONE® VP4 Test Pack monitors exposure to both vaporized hydrogen peroxide (H2O2 or VHP) and ozone (O3) which are both used in the STERIZONE® VP4 Sterilizer. The STERIZONE® VP4 Test Pack is intended to have equivalent to greater resistance than worst case devices and loads in any load configuration.
Device Story
Device is a self-contained biological indicator (SCBI) and test pack for monitoring STERIZONE® VP4 Sterilizer. Input: exposure to vaporized hydrogen peroxide and ozone during sterilization cycle. SCBI contains Geobacillus stearothermophilus spores on stainless steel carrier. After processing, SCBI is retrieved and incubated in dry-bath (55-60°C) for 18 hours. Output: visual color change (purple to yellow) indicates bacterial growth; no color change (remains purple) indicates successful sterilization (spore kill). Used in clinical settings by healthcare staff. Provides objective verification of sterilization efficacy; ensures process achieved conditions to kill 6-log spore population. Test pack includes syringe, plunger, and diffusion restrictor to simulate worst-case load resistance.
Clinical Evidence
Bench testing only. Studies validated viable spore population (>=1x10^6), growth inhibition, 18-hour incubation time, media recovery, stability, and positive controls. Test pack performance confirmed equivalent or greater resistance than worst-case loads and full half-cycle in STERIZONE® VP4 Sterilizer.
Technological Characteristics
SCBI: 316 stainless steel carrier, Geobacillus stearothermophilus spores, soybean digest media with bromocresol purple pH indicator. Test Pack: 10 mL Becton-Dickinson plastic syringe, 18-gauge (1.02 mm) PTFE cannula (50.8 mm long) as diffusion restrictor. Single-use, disposable. Incubation: 55-60°C dry-bath. Sterilization agents: H2O2 and ozone.
Indications for Use
Indicated for routine monitoring of the STERIZONE® VP4 Sterilizer Cycle 1. Used by healthcare personnel to assess spore kill of Geobacillus stearothermophilus (1x10^6 spores) on stainless steel carriers following sterilization processing.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2014
TSO3 Inc. C/O Thomas Richards, Ph.D Consultant IM3, Inc. 512F NE 81st Street, Suite 101 Vancouver, WA 98665
Re: K141580
> Trade/Device Name: Sterizone® BI+ Self-Contained Biological Indicator, Sterizone VP4 Test Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: November 14, 2014 Received: November 17, 2014
Dear Dr. Richards.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin Keith Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141580
Device Name STERIZONE® BI+ Self-contained Biological Indicator STERIZONE® VP4 Test Pack
#### Indications for Use (Describe)
The STERIZONE® BI+ Self-contained Biological Indicator (SCBI) is intended for routine monitoring of the STERIZONE® VP4 Sterilizer, which offers a single pre-set sterilization cycle ("Cycle 1"). The SCBI should only be used in a Test Pack configuration to monitor Cycle 1. The SCBI placed within the STERIZONE® VP4 Test Pack monitors exposure to both vaporized hydrogen peroxide (H2O2 or VHP) and ozone (O3) which are both used in the STERIZONE® VP4 Sterilizer. The STERIZONE® VP4 Test Pack is intended to have equivalent to greater resistance than worst case devices and loads in any load configuration.
STERIZONE® VP4 Test Pack is a device composed of the STERIZONE® BI+ Self-contained Biological Indicator (TSO3 product code 42602, including crusher), a 10 mL syringe and its plunger, and a diffusion restrictor (sold in the form of a kit - TSO3 product code: 44020). An external STERIZONE® CI+ Chemical Indicator (TSO3 product code 43810) is also added to allow differentiating processed from unprocessed test packs. All components of the Test Pack are single-use, disposable items.
The STERIZONE® VP4 Test Pack is constructed by first inserting the STERIZONE® BI+ Selfcontained Biological Indicator inside the syringe, with the SCBI cap facing to the Luer-lock of the syringe. The plunger is then inserted to the 10 mL mark of the syringe. The diffusion restrictor is screwed to the Luer-lock of the syringe. The chemical indicator is then inserted in the opening between the plunger and the syringe.
The test pack is placed within the load on the upper shelf of the STERIZONE® VP4 Sterilizer loading rack.
After processing, the SCBI is retrieved from the test pack. The SCBI is intended to provide users with a means to assess spore kill by the STERIZONE® VP4 Sterilizer. A "no growth" result from the SCBI after 18 hours of incubation indicates that the process achieved the conditions necessary to kill at least 1 × 10° viable spores of Geobacillus stearothermophilus (6 logs) on the SCBI inoculated stainless steel carrier.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
$$\mathbf{^{\hat{G}}}$$
## 5.1. Applicant's Name and Address and submission date
## Applicant's Name and Address
TSO3 Inc., 2505, avenue Dalton, Québec (Quebec) Canada G1P 3S5
## TSO3 Contact Person, Telephone, FAX
Marc Chaunet, Director of Regulatory Affairs Tel : (418) 651-0003 ext.242 FAX : (418) 653-5726 E-mail: mchaunet(@tso3.com
### U.S. Contact
Contact: Thomas Richards, Ph.D. of IM3, Inc. Phone: 503-415-0250 Email: tomami20x@gmail.com
#### Submission Date
November 13, 2014
### 5.2. Name of the device
### Trade Name
STERIZONE® BI+ Self-contained Biological Indicator STERIZONE® VP4 Test Pack
### Common Name
Biological Indicator (Test Pack)
### Classification Name (if known)
Indicator, Biological Sterilization Process
### Regulatory Class
Class II under Sterilization Process Indicator in 21 CFR 880.2800 (b) by the General Hospital and Personal Use Devices Panel. Product code: FRC
#### 5.3. Legally Marketed Equivalent Device Name(s)
STERRAD® CYCLESURE® 24 Biological Indicator (K123017)
STERRAD® 100NX DUO Cycle Test Pack (K111391)
Verify® V24 Self-contained Biological Indicator (K090514)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a chemical structure diagram. The diagram depicts a molecule with the letters 'TS' on the left side, followed by 'O' with a subscript of '3' on the right side. The oxygen atoms are connected to the sulfur atom, forming a tetrahedral arrangement.
#### 5.4. Description of device
The STERIZONE® BI+ Self-contained Biological Indicator (TSO3 product code 42602) consists of at least 10° Geobacillus stearothermophilus viable spores, known to be the reference microorganism for the STERIZONE® VP4 Sterilizer sterilization process, grouped on a stainless steel carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. The presence of G. stearothermophilus spores is detected by a visual color change (media turns yellow). The yellow color change indicates bacterial growth. No change of color indicates that the process achieved the conditions necessary to kill at least 1 × 10 viable spores of G. stearothermophilus (6 logs) on the SCBI inoculated stainless steel carrier. The final readout of a negative result (media remains purple) is made after 18 hours of incubation when using a dry-bath type incubator.
The STERIZONE® VP4 Test Pack is a device composed of the STERIZONE® BI+ Selfcontained Biological Indicator (TSO3 product code 42602, including crusher), a 10 mL syringe and its plunger, and a diffusion restrictor (sold in the form of a kit - TSO3 product code: 44020). A STERIZONE® CI+ Chemical Indicator (TSO3 product code 43810) is also added, external to the syringe, to allow differentiating processed test packs. All components of the Test Pack are single-use, disposable items.
### 5.5. Statement of Intended use
The STERIZONE® BI+ Self-contained Biological Indicator (SCBI) is intended for routine monitoring of the STERIZONE® VP4 Sterilizer, which offers a single pre-set sterilization cycle ("Cycle 1"). The SCBI should only be used in a Test Pack configuration to monitor Cycle 1. The SCBI placed within the STERIZONE® VP4 Test Pack monitors exposure to both vaporized hydrogen peroxide (H2O2 or VHP) and ozone (O3) which are both used in the STERIZONE® VP4 Sterilizer The STERIZONE® VP4 Test Pack is intended to have equivalent to greater resistance than worst case devices and loads in any load configuration.
{5}------------------------------------------------
## SECTION 5 - 510(k) SUMMARY
Image /page/5/Picture/1 description: The image shows a chemical structure diagram. The diagram depicts a molecule with the letters 'T' and 'S' on the left side, followed by 'O' with a subscript '3' on the right side. The 'O' atoms are connected to the 'S' atom, forming a tetrahedral arrangement.
The STERIZONE® VP4 Test Pack is a device composed of the STERIZONE® BI+ Selfcontained Biological Indicator (TSO3 product code 42602, including crusher), a 10 mL syringe and its plunger, and a diffusion restrictor (sold in the form of a kit - TSO3 product code: 44020). A STERIZONE® CI+ Chemical Indicator (TSO3 product code 43810) is also added, external to the syringe, to allow differentiating processed test packs. All components of the Test Pack are single-use, disposable items.
The STERIZONE® VP4 Test Pack is constructed by first inserting the STERIZONE® BI+ Selfcontained Biological Indicator inside the syringe, with the SCBI cap facing to the Luer-lock of the syringe. The plunger is then inserted to the 10 mL mark of the syringe. The diffusion restrictor is screwed to the Luer-lock of the syringe. The chemical indicator is then inserted in the opening between the plunger and the syringe.
The test pack is then placed within the load on the upper shelf of the STERIZONE® VP4 Sterilizer loading rack.
After processing, the SCBI is retrieved from the test pack. The STERIZONE® BI+ Selfcontained Biological Indicator (SCBI) is intended to provide users with a means to assess spore kill by the STERIZONE® VP4 Sterilizer. A "no growth" result from the STERIZONE® BI+ Self-contained Biological Indicator after 18 hours of incubation indicates that the process achieved the conditions necessary to kill at least 1 × 106 viable spores of Geobacillus stearothermophilus (6 logs) on the SCBI inoculated stainless steel carrier.
### 5.6. Substantial equivalence
The indicator organism, spore population and physical construction of the STERIZONE® BI+ Self-contained Biological Indicator are similar to the predicate devices for the proposed device (Table 1).
{6}------------------------------------------------
Image /page/6/Picture/6 description: The image shows a chemical structure diagram. The diagram depicts a molecule with the formula TSO3. The letters 'TS' are prominently displayed on the left side of the diagram, followed by the letter 'O' and the number '3' as a subscript, indicating the presence of three oxygen atoms. The oxygen atoms are arranged in a triangular configuration on the right side of the diagram.
The performance characteristics and intended use of the STERIZONE® BI+ Self-contained Biological Indicator are the same as for the STERRAD® CYCLESURE® 24 Self-contained Biological Indicator (K123017) and to the Verify® V24 Self-contained Biological Indicator (K090514) (Table 1).
Table 1. Comparison between the intended use and claims for the STERIZONE® BI+ Self-contained Biological Indicator, and the STERRAD® CYCLESURE® 24 Self-contained Biological Indicator and to the Verify® V24 Self-contained Biological Indicator
| Features | STERIZONE® BI+ Self-<br>contained Biological<br>Indicator | STERRAD® CycleSure®<br>24 Biological Indicator<br>(K123017) | Verify® V24 Self-<br>contained Biological<br>Indicator (K090514) |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Picture of the<br>device | Image: STERIZONE® BI+ Self-contained Biological Indicator | Image: STERRAD® CycleSure® 24 Biological Indicator (K123017) | Image: Verify® V24 Self-contained Biological Indicator (K090514) |
| Type of BI | Self-contained BI | Self-contained BI | Self-contained BI |
| Organism:<br>Spore species<br>& Strain | Geobacillus<br>stearothermophilus<br>ATCC 7953 | Geobacillus<br>stearothermophilus<br>ATCC 7953 | Geobacillus<br>stearothermophilus<br>ATCC 7953 |
| Viable spore<br>population | Equal or greater than 1×106 | Equal or greater than 1×106 | Equal or greater than 1×106 |
| Intended Use | Sterilization process<br>indicator | Sterilization process<br>indicator | Sterilization process<br>indicator |
| Indications for<br>Use | The STERIZONE® BI+<br>Self-contained Biological<br>Indicator is intended to<br>provide users with a means<br>to assess spore kill by the<br>STERIZONE® VP4<br>Sterilizer. | The STERRAD®<br>CycleSure® Biological<br>Indicator is intended to be<br>used as a standard method<br>for frequent monitoring of<br>the STERRAD® Sterilizer<br>cycles. | The Verify® V24 Self-<br>contained Biological<br>Indicator is intended to be<br>used as a standard method<br>for frequent monitoring of<br>the Amsco® V-PRO™ Low<br>Temperature Sterilization<br>System. |
| Intended Use<br>• Method of<br>sterilization | STERIZONE® VP4<br>Sterilizer | STERRAD® Sterilization<br>System<br>All types of STERRAD®<br>Sterilization Cycles, e.g:<br>STERRAD® 100S,<br>STERRAD® 50<br>STERRAD® NX™<br>STERRAD® 100NX™ | Amsco® V-PRO™ Low<br>Temperature Sterilization<br>System |
| •<br>Primary | Hydrogen peroxide | Hydrogen peroxide | Hydrogen peroxide |
| Features | STERIZONE® BI+ Self-<br>contained Biological<br>Indicator | STERRAD® CycleSure®<br>24 Biological Indicator<br>(K123017) | Verify® V24 Self-<br>contained Biological<br>Indicator (K090514) |
| sterilizing<br>agent | | | |
| Resistance<br>characteristics:<br>• D-value | $ΔP_{VS280}$ = 0.65 Torr<br>Time = 4.3 seconds<br>Dose = 0.39 mg/L | Specific to the utilized<br>resistometer and claimed<br>sterilization system | Specific to the utilized<br>resistometer and claimed<br>sterilization system |
| • Survival-kill<br>window | | | |
| | $ΔP_{VS280}$ = ≥ 1.28 and ≤ 11.0<br>Torr<br>Time = ≥ 6.1 and ≤ 102.5<br>seconds<br>Dose = ≥ 0.69 and ≤ 15.61<br>mg/L | | |
| Culture<br>Conditions | Crushable "onion skin"<br>glass containing a<br>proprietary formulated<br>soybean digest base with a<br>bromocresol purple pH<br>indicator. | Crushable glass containing a<br>proprietary formulated<br>soybean digest base with a<br>pH indicator. | Crushable glass containing a<br>proprietary formulated<br>soybean digest base with a<br>pH indicator. |
| Carrier<br>materials | Disc made of polished 316<br>Stainless Steel (non-porous<br>carrier type) | Non-cellulosic dry spore<br>strip made of fiberglass | Bottom of polypropylene<br>vial |
| Incubation<br>temperature | 55 - 60 °C | 55 - 60 °C | 55 - 60 °C |
| Incubation<br>time | 18 hours | 24 hours | 24 hours |
| • Inoculated<br>carrier<br>Primary<br>Pack | Capsule: Flexible<br>polypropylene vial to hold<br>both dry spore disc and the<br>ampoule medium.<br>Cap: White polypropylene | Capsule: Flexible polymeric<br>vial to hold both dry spore<br>strip and the ampoule<br>medium.<br>Cap: White polypropylene | Capsule: Flexible<br>polypropylene vial directly<br>inoculated with spores and<br>containing the ampoule<br>medium. |
| | cap. The cap filter is Tyvek®<br>1073B non-woven<br>polyethylene | cap. The cap filter is made<br>of non-woven polyethylene. | Cap: Orange polymeric cap.<br>The cap filter is made of<br>non-woven polyethylene. |
| Storage<br>Conditions | 15-30°C (59-86°F) | 2-25°C (35-77°F)<br>under dry conditions | 21-25°C (70-77°F)<br>40-60 % RH |
| Labeling | Instructions for use | Instructions for use | Instructions for use |
| Features | STERIZONE® BI+ Self-<br>contained Biological<br>Indicator | STERRAD® CycleSure®<br>24 Biological Indicator<br>(K123017) | Verify® V24 Self-<br>contained Biological<br>Indicator (K090514) |
| | Carton label | Carton label | Carton label |
| | Shipping label | Shipping label | Shipping label |
| Accessories | Activator (BI crusher) | Activator (BI crusher) | Activator (BI crusher) |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a chemical structure diagram. The diagram depicts a molecule with the letters 'TS' on the left, followed by a central sulfur atom bonded to three oxygen atoms, represented by 'O3'. The 'O' atoms are arranged in a triangular configuration around the sulfur atom.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a chemical structure diagram. The diagram depicts a molecule with the letters 'TS' on the left, followed by a central atom 'O' connected to three other 'O' atoms, represented as O3. The 'O' atoms are drawn with a lighter shade of gray compared to the 'TS' letters, indicating a different element or group within the molecule.
Table 2 summarizes the specific device configuration for rendering each of the aforementioned biological indicators into a corresponding "test pack" or a resistant challenge to sterilization. Each version of a test pack is constructed of a biological indicator, a chemical indicator, and a diffusion restrictor intended to increase resistance. The diffusion restrictor varies from a sealed pouch (STERIS® Verify – K090514) to a vial with a cap having a defined orifice (STERRAD® 100NX DUO Cycle Test Pack – K111391). The STERIZONE® VP4 Test Pack has equivalent to greater resistance than the worst case devices and loads in any load configuration, and is designed to be more resistant than the full half-cycle, including exposure to hydrogen peroxide and ozone. The predicate STERRAD Test Pack is claimed to be at least as resistant to the sterilization process as the biological model developed for the DUO Cycle.
Table 2. Comparison of the STERIZONE® VP4 Test Pack with the STERRAD® 100NX DUO Cycle Test Pack and the STERIS® Verify® V24 "Test Pack"
| Features | STERIZONE® VP4<br>Test Pack | STERRAD® 100NX DUO<br>Cycle Test Pack<br>(K111391) | STERIS® Verify® V24<br>"Test Pack"<br>(K090514) |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Product<br>description | A STERIZONE® BI+ Self-<br>contained BI, and a syringe<br>with diffusion restrictor for<br>holding the BI during the<br>sterilization cycle, along<br>with a process chemical<br>indicator. | A STERRAD® CycleSure®<br>Self-contained BI, and a<br>STERRAD® NX® Test<br>Vial with cap for holding<br>the BI during the<br>sterilization cycle. | A STERIS Verify® Self-<br>contained BI, a Verify®<br>chemical indicator strip, and<br>a low-temperature<br>sterilization pouch1 |
| Picture of the<br>device | Image: STERIZONE VP4 Test Pack | Image: STERRAD 100NX DUO Cycle Test Pack | Image: STERIS Verify V24 "Test Pack" |
| Indications for<br>Use | The STERIZONE® VP4<br>Test Pack is used for routine<br>monitoring of the | The STERRAD® 100NX<br>DUO Cycle Test Pack is<br>used for routine | The Verify® V24 Self-<br>contained Biological<br>Indicator is intended to be |
| Features | STERIZONE® VP4<br>Test Pack | STERRAD® 100NX DUO<br>Cycle Test Pack<br>(K111391) | STERIS® Verify® V24<br>"Test Pack"<br>(K090514) |
| | STERIZONE® VP4<br>Sterilizer cycle and for the<br>performance validation of<br>the STERIZONE® VP4<br>Sterilizer system using<br>hospital-defined loads. | monitoring of the<br>STERRAD 100NX DUO<br>Sterilization Cycle and is<br>also used for the periodic<br>testing of a STERRAD<br>100NX System DUO<br>cycle, using hospital-<br>defined loads containing<br>devices that do not exceed<br>claims of the cycle. The<br>STERRAD 100NX DUO<br>Cycle Test Pack consists<br>of a STERRAD<br>CYCLESURE 24<br>Biological Indicator, vial<br>and cap to hold the BI. | used as a standard method<br>for frequent monitoring of<br>the Amsco® V-PRO™ Low<br>Temperature Sterilization<br>System. |
| Intended Use<br>Method of<br>sterilization | STERIZONE® VP4<br>Sterilizer | STERRAD® Sterilization<br>System | Amsco® V-PRO™ Low<br>Temperature Sterilization<br>System |
| Sterilizing<br>agent(s) | Hydrogen peroxide / ozone | Hydrogen peroxide | Hydrogen peroxide |
| Biological<br>challenge | STERIZONE® BI+ Self-<br>contained Biological<br>Indicator - Min. 1×106 | STERRAD® CycleSure®<br>24 Self-contained<br>Biological Indicator - Min.<br>1×106 | STERIS Verify® V24 Self-<br>contained Biological<br>Indicator - Min. 1×106 |
| | Geobacillus<br>stearothermophilus ATCC<br>7953 spores | Geobacillus<br>stearothermophilus ATCC<br>7953 spores | Geobacillus<br>stearothermophilus ATCC<br>7953 spores |
| Holder | Syringe and plunger-<br>10 mL Becton-Dickinson<br>(B-D) plastic syringe with<br>Luer-lock connector and its<br>plunger | STERRAD® NX® Test<br>Vial<br>9.1 mL (0.558 in3)<br>polyethylene vial | STERIS Low Temperature<br>Sterilization Pouch |
| Diffusion<br>restrictor | 18 gauge (1.02 mm<br>diameter),<br>2-inch (50.8 mm) long<br>polytetrafluoroethylene<br>(PTFE) cannula with Luer-<br>lok™ attachment | Polyethylene vial cap with<br>a single orifice of 1.4-1.55<br>mm diameter & 2.7 mm<br>length (i.e., thickness of<br>cap) | Sealed Pouch |
| Resistance<br>characteristics | Demonstrated to have<br>equivalent to greater<br>resistance than the worst<br>case devices and loads in<br>any load configuration. | The Test Pack is at least as<br>resistant to the sterilization<br>process as the biological<br>model developed for the<br>DUO Cycle. | Specific to the claimed<br>sterilization system |
| Features | STERIZONE® VP4<br>Test Pack | STERRAD® 100NX DUO<br>Cycle Test Pack<br>(K111391) | STERIS® Verify® V24<br>"Test Pack"<br>(K090514) |
| | Demonstrated to be more<br>resistant than the full half-cycle, including exposure to<br>hydrogen peroxide and<br>ozone. | | |
<sup>1</sup> Use of a pouch can increase resistance so as to create a test pack.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a chemical structure with the letters 'T', 'S', and 'O' arranged in a specific configuration. The 'T' is on the left, followed by 'S', and then 'O' with a subscript '3'. The 'O' and subscript '3' are drawn with gray lines, while the 'T' and 'S' are drawn with black lines. The image appears to represent a chemical formula or abbreviation, possibly related to an organic compound or a chemical reaction.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a chemical structure diagram. The diagram depicts a molecule with the letters 'TS' on the left, followed by a central sulfur atom connected to three oxygen atoms (O3) on the right. The oxygen atoms are arranged in a triangular configuration around the sulfur atom.
## Conclusion:
The STERIZONE® BI+ Self-contained Biological Indicator and STERIZONE® VP4 Test Pack are substantially equivalent to the STERRAD® CYCLESURE® 24 Self-contained Biological Indicator (K123017), to the STERRAD® 100NX DUO Cycle Test Pack (K111391), and to the Verify® V24 Self-contained Biological Indicator (K090514) with respect to intended use, indications for use, and critical technological characteristics. Overall, the subject device and predicate devices have identical intended use, and general indications for use, although the STERIZONE® VP4 Test Pack monitors exposure to both hydrogen peroxide and ozone whereas the predicate devices monitor only exposure to hydrogen peroxide. Additionally, all devices have the same general technological characteristics, and the same operating principles. Performance testing demonstrates that the STERIZONE® BI+ Self-contained Biological Indicator and STERIZONE® VP4 Test Pack are substantially equivalent to the identified predicate devices. Minor differences in technology between the subject devices and predicates do not raise new questions of safety and effectiveness when the devices are used as labeled.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows a chemical structure. The structure is labeled with the letters "TSO" on the left side. To the right of the letters, there is a chemical structure with three oxygen atoms. The number 3 is written as a subscript to the right of the last oxygen atom.
## 5.7. Assessment of performances data
#### Summary of nonclinical performance tests 5.7.1.
The studies were conducted to confirm that the performance characteristics of the STERIZONE® BI+ Self-contained Biological Indicator are similar to the predicate devices (3).
## Table 3. Summary of nonclinical tests performed to demonstrate Safety and Effectiveness
| | Performance Requirements for Effectiveness | Results |
|-------------------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Viable population assay | Passed<br>Within specification |
| 2 | Growth inhibition by carrier and pack materials | Passed<br>No inhibition induced |
| 3 | Reduced incubation time validation | Passed<br>18 hours using a dry-bath type incubator<br>adjusted to 55 – 60 °C |
| 4 | Effect of sterilization process on recovery media | Passed<br>No effect |
| 5 | Stability of biological read | Passed<br>Stable for 7 days |
| 6 | Positive Controls | Passed<br>Viability demonstrated |
| 7 | Stability (shelf life) evaluation | Passed<br>Ongoing stability evaluation |
| 8 | BI validation in the STERIZONE® VP4 process | Passed |
| 9 | Test Pack performance evaluation in the<br>STERIZONE® VP4 Sterilizer process | Passed<br>Demonstrated to have equivalent to greater<br>resistance than the worst case devices and<br>loads in any load configuration<br><br>Demonstrated to be more resistant than the <i>full</i><br>half-cycle, including exposure to hydrogen<br>peroxide and ozone |
| Performance Requirements for Safety | | |
| 1 | Safe for use | Passed<br>No safety issue |
#### Overall Performance Conclusions 5.7.2.
Performance tests demonstrate that the STERIZONE® BI+ Self-contained Biological Indicator and STERIZONE® VP4 Test Pack are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.