RIGHTEST DIABETES MANAGEMENT SOFTWARE AND ADAPTER (EXISTING CLEARED K113007),RIGHTEST NEW ADAPTERS-3,GE DIABETES MANAGEM
Applicant
Bionime Coporation
Product Code
NBW · Clinical Chemistry
Decision Date
Apr 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The Bionime Rightest Diabetes Management System is an over-the-counter software system for use by Health Care Professionals and Patients with diabetes as an aid for managing diabetes. User(s) can transfer blood glucose readings from the following glucose meters: Rightest Blood Glucose Monitoring System GM550 and GM250 with Rightest PC Link Adaptor (GP550), Rightest Blood Glucose Monitoring System GM700 with Rightest PC Link Adaptor (GP550) or Rightest Bluetooth PC Link (GP700), Rightest Blood Glucose Monitoring System GM650 with Rightest PC Link Adaptor for GM650 (GP650), and Rightest Blood Glucose Monitoring System GM300 with Rightest PC Link Adaptor for GM300 (GP300) to a personal computer for the purpose of viewing, analyzing and printing the blood glucose readings, as well as to backup and to recover users' profile and data. The Bionime Rightest Diabetes Management System is not intended to provide treatment decisions, nor should it substitute professional opinion. All medical diagnoses and treatment plans should be performed by a licensed healthcare professional.
Device Story
System facilitates transfer of blood glucose readings from compatible Bionime/GE meters to PC via USB or wireless (Bluetooth) PC link adapters. Software (Rightest/GE GP200) enables data visualization (tables, trend charts, pie charts) and report printing for diabetes management. Used by patients and healthcare professionals in home or clinical settings. Output aids in monitoring diabetes control; does not provide treatment decisions. Benefits include improved data organization and tracking of glucose trends over time.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing (data transmission accuracy, memory rollover synchronization) and a user performance evaluation (n=45) demonstrating that laypersons can successfully connect adapters and import data to the software. Readability assessment (Flesch-Kincaid) confirmed manuals are appropriate for 8th-grade reading level.
Technological Characteristics
System comprises glucose meters (electrochemical dehydrogenase sensor), PC link adapters (USB/Bluetooth), and management software. Meters use FAD-GDH or Glucose Oxidase reagents. Connectivity via USB or Bluetooth. Software runs on Windows (XP, Vista, 7).
Indications for Use
Indicated for patients with diabetes and healthcare professionals to manage diabetes by transferring, viewing, analyzing, and printing blood glucose data from compatible Bionime/GE meters. Not for treatment decisions, diagnosis, or screening.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K113007 — BIONIME DIABETES MANAGEMENT SYSTEM VI.0 · Bionime Corporation · Feb 27, 2012
K123089 — FREESTYLE AUTO-ASSIST SOFTWARE VERSON 2.0 WITH MODEL PRT24238-001 · Abbott Laboratories · Apr 5, 2013
Submission Summary (Full Text)
{0}
Bionime Corporation
510(K) SUMMARY
APR 17 2014
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: k133522
## 1. Submitter's Identification:
BIONIME CORPORATION
No.100, Sec. 2, Daqing St., South Dist.,
Taichung City 40242, Taiwan (R.O.C.)
c/o IVDD Regulatory Consultant
2700 La Paz Road, Suite 266B
Mission Viejo, CA 92691
Contact Person: Feng-Yu Lee
Phone Number: 1-949-951-3046
FAX Number: 1-949-951-2042
Date Summary Prepared: April 15, 2014
## 2. Name of the Device:
Rightest Blood Glucose Monitoring System GM700
Rightest Blood Glucose Monitoring System GM650
Rightest Blood Glucose Monitoring System GM300
GE300 Talking Blood Glucose Monitoring System
GE200 Blood Glucose Monitoring System
Bionime Rightest Diabetes Management System
GE Diabetes Management System
## 3. Common or Usual Name:
| Device Name | Product Code | Classification | Regulation | Panel |
| --- | --- | --- | --- | --- |
| Glucose Test System | NBW: Blood Glucose Test System, Over-the-Counter | Class II | 21 CFR § 862.1345 | Clinical Chemistry (75) |
| Glucose Test System | LFR: Glucose Dehydrogenase, Glucose | Class II | 21 CFR § 862.1345 | Clinical Chemistry (75) |
| Glucose Test System | CGA: Glucose Oxidase, Glucose | Class II | 21 CFR § 862.1345 | Clinical Chemistry (75) |
| Calculator/Data Processing Module | JQP: Calculator/ Data Processing Module for | Class I | 21 CFR § 862.2100 | Clinical Chemistry |
8 - 1
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{1}
Bionime Corporation
| for Clinical Use | Clinical Use | | | (75) |
| --- | --- | --- | --- | --- |
## 4. Device Description:
4.1 Rightest Blood Glucose Monitoring System GM700 consists of following devices: Rightest Blood Glucose Meter GM700, Rightest Blood Glucose Test Strip GS700, Rightest Control Solution GC700, lancing device and sterile lancets. The Rightest Blood Glucose Meter GM700, when used with the Rightest Blood Glucose Test Strips GS700, quantitatively measures glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System GM700 is verified by the Rightest Control Solution GC700.
4.2 Rightest Blood Glucose Monitoring System GM650 consists of following devices: Rightest Blood Glucose Meter GM650, Rightest Blood Glucose Test Strip GS650, Rightest Control Solution GC650, lancing device and sterile lancets. The Rightest Blood Glucose Meter GM650, when used with the Rightest Blood Glucose Test Strips GS650, quantitatively measures glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System GM650 is verified by the Rightest Control Solution GC650.
4.3 Rightest Blood Glucose Monitoring System GM300 consists of following devices: Rightest Blood Glucose Meter GM300, Rightest Blood Glucose Test Strip GS300, Rightest Control Solution GC300, lancing device and sterile lancets. The Rightest Blood Glucose Meter GM300, when used with the Rightest Blood Glucose Test Strips GS300, quantitatively measures glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System GM300 is verified by the Rightest Control Solution GC300.
4.4 GE200 Blood Glucose Monitoring System consists of following devices: GE200 Blood Glucose Meter, GE200 Blood Glucose Test Strip, GE200 Control Solution, lancing device and sterile lancets. The GE200 Blood Glucose Meter, when used with the GE200 Blood Glucose Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance of the GE200 Blood Glucose Monitoring System is verified by the GE200 Control Solution.
4.5 GE300 Talking Blood Glucose Monitoring System consists of following devices: GE300 Talking Blood Glucose Meter, GE300 Talking Blood Glucose Test Strip, GE300 Series Control Solution, lancing device and sterile lancets. The GE300 Talking Blood Glucose Meter, when used with the GE300 Talking Blood Glucose Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance of the GE300 Talking Blood Glucose Monitoring System is verified by the GE300 Series Control Solution.
The following devices have already been cleared. The purpose of this submission is to add additional compatible meters for use with these systems.
8 - 2
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{2}
Bionime Corporation
4.6 The Bionime Rightest Diabetes Management System allows the transfer of blood glucose readings from a compatible Rightest Glucose meter to a PC via PC link adapter (through USB and wireless connection).
The Data analysis features enable the user(s) to view and analyze blood glucose readings from different meal time periods. Other features including data tables, trend charts, pie charts, and printed reports are available for viewing and analyzing these readings within the different time slots.
The revised system includes: 1) Installation CD with setup files for Rightest GP200 Diabetes Management Software and Rightest PC Link Adapter Driver, 2) One or multiple Rightest Adapters: GP550, GP700, GP650 and GP300
Compatible Meters include: GM250, GM550, GM700, GM650 and GM300.
4.7 The GE Diabetes Management System allows the transfer of blood glucose readings from a compatible GE Glucose meter to a PC via PC link adapter (through USB and wireless connection).
The Data analysis features enable the user(s) to view and analyze blood glucose readings from different meal time periods. Other features including data tables, trend charts, pie charts, and printed reports are available for viewing and analyzing these readings within the different time slots.
The revised system includes: 1) Installation CD with setup files for GE GP200 Diabetes Management Software and GE PC Link Adapter Driver, 2) One or multiple GE Adapters: GE GP550, GE GP700 and GE GP300.
Compatible Meters include: GE100, GE200 and GE300.
## 5. Intended Use/Indications for Use:
The Rightest and GE Diabetes Management System Software has been previously cleared by FDA under k113007 and k113007/A001, respectively.
### a. Bionime Rightest Diabetes Management System
The Bionime Rightest Diabetes Management System is an over-the-counter software system for use by Health Care Professionals and Patients with diabetes as an aid for managing diabetes. User(s) can transfer blood glucose readings from the following glucose meters: Rightest Blood Glucose Monitoring System GM550 and GM250 with Rightest PC Link Adaptor (GP550), Rightest Blood Glucose Monitoring System GM700 with Rightest PC Link Adaptor (GP550) or Rightest Bluetooth PC Link (GP700), Rightest Blood Glucose Monitoring System GM650 with Rightest PC Link Adaptor for GM650 (GP650), and Rightest Blood Glucose Monitoring System GM300 with Rightest PC Link Adaptor for GM300 (GP300) to a personal computer
8 - 3
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{3}
Bionime Corporation
for the purpose of viewing, analyzing and printing the blood glucose readings, as well as to backup and to recover users' profile and data.
The Bionime Rightest Diabetes Management System is not intended to provide treatment decisions, nor should it substitute professional opinion. All medical diagnoses and treatment plans should be performed by a licensed healthcare professional.
## b. GE Diabetes Management System
The GE Diabetes Management System is an over-the-counter software system for use by Health Care Professionals and Patients with diabetes as an aid for managing diabetes. User(s) can transfer blood glucose readings from the following glucose meters: GE100 Blood Glucose Monitoring System with GE PC Link Adaptor (GE GP550), GE300 Talking Blood Glucose Monitoring System with GE PC Link Adaptor for GE300 (GE GP300) or GE200 Blood Glucose Monitoring System with GE PC Link Adaptor (GE GP5500 and GE Bluetooth PC Link (GE GP700) to a personal computer for the purpose of viewing, analyzing and printing the blood glucose readings, as well as to backup and to recover users' profile and data.
The GE Diabetes Management System is not intended to provide treatment decisions, nor should it substitute professional opinion. All medical diagnoses and treatment plans should be performed by a licensed health care professional.
## c. Rightest Blood Glucose Monitoring System GM700:
The Rightest Blood Glucose Monitoring System GM700 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700 is intended to be used by a single person and should not be shared.
The Rightest Blood Glucose Monitoring System GM700 is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM700 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
The Rightest Blood Glucose Test Strips GS700 are for use with the Rightest Blood Glucose Meter GM700 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{4}
Bionime Corporation
## d. Rightest Blood Glucose Monitoring System GM650:
The Rightest Blood Glucose Monitoring System GM650 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM650 is intended to be used by a single person and should not be shared.
The Rightest Blood Glucose Monitoring System GM650 is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM650 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
The Rightest Blood Glucose Test Strips GS650 are for use with the Rightest Blood Glucose Meter GM650 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.
## e. Rightest Blood Glucose Monitoring System GM300:
The Rightest Blood Glucose Monitoring System GM300 is intended for in vitro diagnostic use (outside of body). It is indicated to be used by professional healthcare personnel or diabetes at home to measure the glucose concentration for aiding diabetes management. The glucose concentration is measured with quantitative capillary whole blood from the fingertip, palm, and forearm by using Rightest Blood Glucose Monitoring System GM300. This device is not intended for testing neonate blood samples.
Special conditions for use statement(s): Rightest System provides plasma equivalent results.
## f. GE200 Blood Glucose Monitoring System:
The GE200 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE200 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
The GE200 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE200 Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
8 - 5
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{5}
Bionime Corporation
The GE200 Blood Glucose Test Strips are for use with the GE200 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.
## g. GE300 Talking Blood Glucose Monitoring System:
The GE300 Talking Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE300 Talking Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
The GE300 Talking Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE300 Talking Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
The GE300 Blood Glucose Test Strip is for use with the GE300 Talking Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.
## 6. Predicate Device Information:
The Rightest Blood Glucose Monitoring System and GE Blood Glucose Monitoring System are substantially equivalent to the predicate device noted below.
Name: Rightest Blood Glucose Monitoring System GM700
Device Company: Bionime Corporation
510(K) Number: K110737 and K123008
Name: Rightest Blood Glucose Monitoring System GM650
Device Company: Bionime Corporation
510(K) Number: K120423 and K123008
Name: Rightest Blood Glucose Monitoring System GM300
Device Company: Bionime Corporation
510(K) Number: K062567
Name: GE200 Blood Glucose Monitoring System
Device Company: Bionime Corporation
510(K) Number: K123008
Name: GE300 Talking Blood Glucose Monitoring System
Device Company: Bionime Corporation
8 - 6
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{6}
Bionime Corporation
510(K) Number: K120423 and K123008
The Rightest Diabetes Management System and the GE Diabetes Management System are substantially equivalent to the predicate device noted below.
Name: Bionime Diabetes Management System
Device Company: Bionime Corporation
510(K) Number: K113007
Name: GE Diabetes Management System
Device Company: Bionime Corporation
510(K) Number: K113007/A001
# 7. Comparison to Predicate Devices:
Specification Comparison
| Item | Subject Device | | Predicate Device | |
| --- | --- | --- | --- | --- |
| | Bionime Rightest Diabetes Management System | GE Diabetes Management System | Bionime Diabetes Management System (BDMS) | GE Diabetes Management System |
| Similarities | | | | |
| Intended use/Indication(s) for use | Same | | BDMS is intended for use by Health Care professionals and patients with diabetes. When it is installed on a personal computer (PC), a database can be used to store multiple users’ profiles and manage their blood glucose readings. People who use BDMS must login to their own profile. BDMS enables users to view and organize the blood glucose readings which are uploaded from Rightest Glucose meter(s). Users can backup and recover their profiles and print blood glucose data and report from their PC. | |
| COM Ports scan | Same | | Automatic scan for connected ports | |
| Multiple patients | Same | | Can list data for multiple patients | |
| Data list/ Summary report | Same | | List all readings inside selected time frame | |
| Target periods | Same | | Daily time periods can be changed | |
| Average week | Same | | 7-day report | |
| Average Day | Same | | 1-day report | |
| Password | Same | | Does not have password protection on the software | |
| Providers | Same | | Does not have providers but does allow for multiple databases | |
| Reports and charts | Same | | Data List, Standard Day report, Trend report, Pie charts | |
8 - 7
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{7}
# Bionime Corporation
| Change Meter settings | Same | Does not allow change to meter settings through the Diabetes Management System |
| --- | --- | --- |
| Operating System | Same | Windows XP Professional™
Windows Vista Professional™
Windows 7 Professional™ |
| Associate a meter to a person | Same | Does not associate a meter to a person |
| Standard week report | Same | No standard week report |
Additional Adapters and compatible meters to be added under this application (Rightest Diabetes Management System):
Table 1
| Item | Subject Pairs (Meter / Adapter) | | Predicate (Meter / Adapter) | |
| --- | --- | --- | --- | --- |
| | Meter | Adapter | Meter | Adapter |
| New pair 1 | Rightest GM700 | GP550 | Rightest GM700 (K110737/K12300 8) | GP550 |
| Similarities/Differences | | | | |
| Operation | The meter connected to the adapter by plug in and transmit data to computer. | | Same | |
| Connection | Mini USB port | Plug in to meter USB Port | Same | Same |
| Function | A device of quantitative glucose measuring system. | A device to transmit data from meter. | Same | Same |
| Feature | Up to 1000 data stored | Cable | Same | Same |
| Appearance | A meter with a display window, two side buttons, one main button, and a test strip port. | A connector to meter and computer | Same | Same |
| Minimum Sample Volume | 0.75 μL | None Applicable | Same | None Applicable |
| Test Time | 5 seconds | None Applicable | Same | None Applicable |
| Measuring Range | 20-600 mg/dL | None Applicable | Same | None Applicable |
| Operating Relative Humidity Range | 10-90% | None Applicable | Same | None Applicable |
| Memory Capacity | 1000 blood glucose test results with date and time | None Applicable | Same | None Applicable |
| Coding | Auto Coding | None Applicable | Same | None Applicable |
| Meter Storage Conditions | 14 ~140 □ (-10 ~ 60□) | None Applicable | Same | None Applicable |
| Test Strip Storage Conditions | 39 ~86 □ (4 ~ 30□), <90 % relative humidity | None Applicable | Same | None Applicable |
| The unit of measurement data | Fix on mg/dL | None Applicable | Same | None Applicable |
8 - 8
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{8}
Bionime Corporation
| Sample | Capillary whole blood | None Applicable | Same | None Applicable |
| --- | --- | --- | --- | --- |
| Interference | Uric acid ≥ 16 mg/dL
Xylose ≥ 10 mg/dL
Ascorbic acid ≥ 3 mg/dL
Dopamine HCl ≥ 1.25 mg/dL
L-Dopa ≥ 2 mg/dL | None Applicable | Same | None Applicable |
| Voice Function | No | None Applicable | Same | None Applicable |
| Strip Reagent | 1.FAD-glucose Dehydrogenase (FAD-GDH) 12.1%
2.Potassium ferricyanide 48.5%
3.Non-reactive ingredients 39.4% | None Applicable | Same | None Applicable |
| Measurement Technology | Dehydrogenase Electrochemical Sensor | None Applicable | Same | None Applicable |
| Operating Temperature Range | 43 ~111 °C (6 ~ 44 °C) | None Applicable | Same | None Applicable |
| Compatible to DMS | Compatible to Bionime Rightest Diabetes Management System | Compatible to Bionime Rightest Diabetes Management System | None Applicable | Same |
Table 2
| Item | Subject Pairs (Meter / Adapter) | | Predicate (Meter / Adapter) | |
| --- | --- | --- | --- | --- |
| | Meter | Adapter | Meter | Adapter |
| New pair 2 | Rightest GM700 | GP700 (Bluetooth) | Rightest GM700 (K110737/K12300 8) | GP550 |
| Similarities/Differences | | | | |
| Operation | The meter connected to the adapter by plug in and transmit data to computer. | | Same | |
| Connection | Mini USB port | Plug in to meter USB Port | Same | Same |
| Function | A device of quantitative glucose measuring system. | A device to transmit data from meter. | Same | Same |
| Feature | Up to 1000 data stored | Wireless | Same | Cable |
| Appearance | A meter with a display window, two side buttons, one main button, and a test strip port. | A connector to meter. | Same | A connector to meter and computer |
| Minimum Sample Volume | 0.75 μL | None Applicable | Same | None Applicable |
| Test Time | 5 seconds | None Applicable | Same | None Applicable |
8 - 9
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{9}
Bionime Corporation
| Measuring Range | 20-600 mg/dL | None Applicable | Same | None Applicable |
| --- | --- | --- | --- | --- |
| Operating Relative Humidity Range | 10-90% | None Applicable | Same | None Applicable |
| Memory Capacity | 1000 blood glucose test results with date and time | None Applicable | Same | None Applicable |
| Coding | Auto Coding | None Applicable | Same | None Applicable |
| Meter Storage Conditions | 14 ~140 °F (-10 ~ 60°C) | None Applicable | Same | None Applicable |
| Test Strip Storage Conditions | 39 ~86 °F (4 ~ 30°C), < 90% relative humidity | None Applicable | Same | None Applicable |
| The unit of measurement data | Fix on mg/dL | None Applicable | Same | None Applicable |
| Sample | Capillary whole blood | None Applicable | Same | None Applicable |
| Interference | Uric acid ≥ 16 mg/dL
Xylose ≥ 10 mg/dL
Ascorbic acid ≥ 3 mg/dL
Dopamine HCl ≥ 1.25 mg/dL
L-Dopa ≥ 2 mg/dL | None Applicable | Same | None Applicable |
| Voice Function | No | None Applicable | Same | None Applicable |
| Strip Reagent | 1.FAD-glucose Dehydrogenase (FAD-GDH) 12.1%
2.Potassium ferricyanide 48.5%
3.Non-reactive ingredients 39.4% | None Applicable | Same | None Applicable |
| Measurement Technology | Dehydrogenase Electrochemical Sensor | None Applicable | Same | None Applicable |
| Operating Temperature Range | 43 ~111 °F (6 ~ 44°C) | None Applicable | Same | None Applicable |
| Compatible to DMS | Compatible to Bionime Rightest Diabetes Management System | Compatible to Bionime Rightest Diabetes Management System | None Applicable | Same |
Table 3
| Item | Subject Pairs (Meter / Adapter) | | Predicate (Meter / Adapter) | |
| --- | --- | --- | --- | --- |
| | Meter | Adapter | Meter | Adapter |
| New pair 3 | Rightest GM650 | GP650 (USB) | Rightest GM650 (K120423/K123008) | GP550 |
| Similarities/Differences | | | | |
| Operation | The meter connected to the adapter by plug in | | Same | |
8 - 10
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{10}
Bionime Corporation
| | and transmit data to computer. | | | |
| --- | --- | --- | --- | --- |
| Feature | Up to 500 data stored | Cable | Same | Same |
| Function | A device of quantitative glucose measuring system. | A device to transmit data from meter. | Same | Same |
| Appearance | A meter with a display window, two side buttons, one main button, and a test strip port. | A connector to meter and computer | Same | Same |
| Connection | Far-red sensor | Plug in | Same | Same |
| Minimum Sample Volume | 0.75 μL | None Applicable | Same | None Applicable |
| Test Time | 5 seconds | None Applicable | Same | None Applicable |
| Measuring Range | 20-600 mg/dL | None Applicable | Same | None Applicable |
| Operating Relative Humidity Range | 10-90% | None Applicable | Same | None Applicable |
| Memory Capacity | 500 blood glucose test results with date and time | None Applicable | Same | None Applicable |
| Coding | Auto Coding | None Applicable | Same | None Applicable |
| Meter Storage Conditions | 14~140 °F (-10~60°C) | None Applicable | Same | None Applicable |
| Test Strip Storage Conditions | 39~86 °F (4~30°C), < 90% relative humidity | None Applicable | Same | None Applicable |
| The unit of measurement data | Fix on mg/dL | None Applicable | Same | None Applicable |
| Sample | Capillary whole blood | None Applicable | Same | None Applicable |
| Interference | Uric acid ≥ 16 mg/dL
Xylose ≥ 10 mg/dL
Ascorbic acid ≥ 3 mg/dL
Dopamine HCl ≥ 1.25 mg/dL
L-Dopa ≥ 2 mg/dL | None Applicable | Same | None Applicable |
| Voice Function | Yes | None Applicable | Same | None Applicable |
| Strip Reagent | 1.FAD-glucose Dehydrogenase (FAD-GDH) 12.1%
2.Potassium ferricyanide 48.5%
3.Non-reactive ingredients 39.4% | None Applicable | Same | None Applicable |
| Measurement Technology | Dehydrogenase Electrochemical Sensor | None Applicable | Same | None Applicable |
| Operating Temperature Range | 43~111 °F (6~44°C) | None Applicable | Same | None Applicable |
| Compatible to DMS | Compatible to Bionime Rightest Diabetes Management System | Compatible to Bionime Rightest Diabetes Management System | None Applicable | Same |
8 - 11
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{11}
Bionime Corporation
Table 4
| Item | Subject Pairs (Meter / Adapter) | | Predicate (Meter / Adapter) | |
| --- | --- | --- | --- | --- |
| | Meter | Adapter | Meter | Adapter |
| New pair 4 | Rightest GM300 | GP300 (USB) | Rightest GM300, (K062567) | GP550 |
| Similarities/Differences | | | | |
| Operation | The meter connected to the adapter by plug in and transmit data to computer. | | Same | |
| Function | A device of quantitative glucose measuring system. | A device to transmit data from meter. | Same | Same |
| Appearance | A meter with a display window, two side buttons, one main button, and a test strip port. | A connector to meter and computer | Same | Same |
| Connection | Code key base | Plug in | Same | Same |
| Feature | Up to 300 data stored | Cable | Same | Same |
| Minimum Sample Volume | 1.4 μL | None Applicable | Same | None Applicable |
| Test Time | 8 seconds | None Applicable | Same | None Applicable |
| Measuring Range | 20-600 mg/dL | None Applicable | Same | None Applicable |
| Operating Relative Humidity Range | 10-90% | None Applicable | Same | None Applicable |
| Memory Capacity | 300 blood glucose test results with date and time | None Applicable | Same | None Applicable |
| Coding | Code Key | None Applicable | Same | None Applicable |
| Meter Storage Conditions | 14~140 °F (-10~60°C) | None Applicable | Same | None Applicable |
| Test Strip Storage Conditions | 39~86 °F (4~30°C), < 90% relative humidity | None Applicable | Same | None Applicable |
| The unit of measurement data | Fix on mg/dL | None Applicable | Same | None Applicable |
| Sample | Capillary whole blood | None Applicable | Same | None Applicable |
| Interference | Uric acid ≥ 9 mg/dL
Methyldopa ≥ 1.5 mg/dL
Cholesterol ≥ 250 mg/dL
L-Dopa ≥ 1.5 mg/dL | None Applicable | Same | None Applicable |
| Voice Function | No | None Applicable | Same | None Applicable |
| Strip Reagent | 1. Glucose Oxidase 8.5%
2. Potassium ferricyanide | None Applicable | Same | None Applicable |
8 - 12
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{12}
Bionime Corporation
Additional Adapters and compatible meters to be added under this application (GE Diabetes Management System):
Table 5
| Item | Subject Pairs (Meter / Adapter) | | Predicate (Meter / Adapter) | |
| --- | --- | --- | --- | --- |
| | Meter | Adapter | Meter | Adapter |
| New pair 1 | GE200 | GE GP550 (USB) | GE200 (K123008) | GE GP550 |
| Similarities/Differences | | | | |
| Operation | The meter connected to the adapter by plug in and transmit data to computer. | | Same | |
| Connection | Mini USB port | Plug in | Same | Same |
| Function | A device of quantitative glucose measuring system. | A device to transmit data from meter. | Same | Same |
| Feature | Up to 1000 data stored | Cable | Same | Same |
| Appearance | A meter with a display window, two side buttons, one main button, and a test strip port. | A connector to meter and computer | Same | Same |
| Minimum Sample Volume | 0.75 μL | None Applicable | Same | None Applicable |
| Test Time | 5 seconds | None Applicable | Same | None Applicable |
| Measuring Range | 20-600 mg/dL | None Applicable | Same | None Applicable |
| Operating Relative Humidity Range | 10-90% | None Applicable | Same | None Applicable |
| Memory Capacity | 1000 blood glucose test results with date and time | None Applicable | Same | None Applicable |
| Coding | Auto Coding | None Applicable | Same | None Applicable |
| Meter Storage Conditions | 14~140 °F (-10~60°C) | None Applicable | Same | None Applicable |
| Test Strip Storage Conditions | 39~86 °F (4~30°C), < 90% relative humidity | None Applicable | Same | None Applicable |
| The unit of measurement | Fix on mg/dL | None Applicable | Same | None Applicable |
8 - 13
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{13}
Bionime Corporation
| data | | | | |
| --- | --- | --- | --- | --- |
| Sample | Capillary whole blood | None Applicable | Same | None Applicable |
| Interference | Uric acid ≥ 16 mg/dL
Xylose ≥ 10 mg/dL
Ascorbic acid ≥ 3 mg/dL
Dopamine HCl ≥ 1.25 mg/dL
L-Dopa ≥ 2 mg/dL | None Applicable | Same | None Applicable |
| Voice Function | No | None Applicable | Same | None Applicable |
| Strip Reagent | 1.FAD-glucose Dehydrogenase (FAD-GDH) 12.1%
2.Potassium ferricyanide 48.5%
3.Non-reactive ingredients 39.4% | None Applicable | Same | None Applicable |
| Measurement Technology | Dehydrogenase Electrochemical Sensor | None Applicable | Same | None Applicable |
| Operating Temperature Range | 43 ~111 °F (6 ~ 44°C) | None Applicable | Same | None Applicable |
| Compatible to DMS | Compatible to GE Diabetes Management System | Compatible to GE Diabetes Management System | None Applicable | Same |
Table 6
| Item | Subject Pairs (Meter / Adapter) | | Predicate (Meter / Adapter) | |
| --- | --- | --- | --- | --- |
| | Meter | Adapter | Meter | Adapter |
| New pair 2 | GE200 | GE GP700 (Bluetooth) | GE200 (K123008) | GE GP550 |
| Similarities/Differences | | | | |
| Operation | The meter connected to the adapter by plug in and transmit data to computer. | | Same | |
| Connection | Mini USB port | Plug in | Same | Same |
| Function | A device of quantitative glucose measuring system. | A device to transmit data from meter. | Same | Same |
| Feature | Up to 1000 data stored | Wireless | Same | Cable |
| Appearance | A meter with a display window, two side buttons, one main button, and a test strip port. | A connector to meter. | Same | A connector to meter and computer |
| Minimum Sample Volume | 0.75 μL | None Applicable | Same | None Applicable |
| Test Time | 5 seconds | None Applicable | Same | None Applicable |
| Measuring Range | 20-600 mg/dL | None Applicable | Same | None Applicable |
8 - 14
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{14}
Bionime Corporation
| Operating Relative Humidity Range | 10-90% | None Applicable | Same | None Applicable |
| --- | --- | --- | --- | --- |
| Memory Capacity | 1000 blood glucose test results with date and time | None Applicable | Same | None Applicable |
| Coding | Auto Coding | None Applicable | Same | None Applicable |
| Meter Storage Conditions | 14 ~140 °F (-10 ~ 60°C ) | None Applicable | Same | None Applicable |
| Test Strip Storage Conditions | 39 ~86 °F (4 ~ 30°C ), < 90% relative humidity | None Applicable | Same | None Applicable |
| The unit of measurement data | Fix on mg/dL | None Applicable | Same | None Applicable |
| Sample | Capillary whole blood | None Applicable | Same | None Applicable |
| Interference | Uric acid ≥ 16 mg/dL
Xylose ≥ 10 mg/dL
Ascorbic acid ≥ 3 mg/dL
Dopamine HCl ≥ 1.25 mg/dL
L-Dopa ≥ 2 mg/dL | None Applicable | Same | None Applicable |
| Voice Function | No | None Applicable | Same | None Applicable |
| Strip Reagent | 1.FAD-glucose Dehydrogenase (FAD-GDH) 12.1%
2.Potassium ferricyanide 48.5%
3.Non-reactive ingredients 39.4% | None Applicable | Same | None Applicable |
| Measurement Technology | Dehydrogenase Electrochemical Sensor | None Applicable | Same | None Applicable |
| Operating Temperature Range | 43 ~111 °F (6 ~ 44°C ) | None Applicable | Same | None Applicable |
| Compatible to DMS | Compatible to GE Diabetes Management System | Compatible to GE Diabetes Management System | None Applicable | Same |
8 - 15
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{15}
Bionime Corporation
Table 7:
| Item | Subject Pairs (Meter / Adapter) | | Predicate (Meter / Adapter) | |
| --- | --- | --- | --- | --- |
| | Meter | Adapter | Meter | Adapter |
| New pair 3 | GE300 | GE GP300 (USB) | GE300 (K120423/K123008) | GE GP550 |
| Similarities/Differences | | | | |
| Operation | The meter connected to the adapter by plug in and transmit data to computer. | | Same | |
| Feature | Up to 500 data stored | Cable | Same | Same |
| Function | A device of quantitative glucose measuring system. | A device to transmit data from meter. | Same | Same |
| Appearance | A meter with a display window, two side buttons, one main button, and a test strip port. | A connector to meter and computer | Same | Same |
| Connection | Far-red sensor | Plug in | Same | Same |
| Minimum Sample Volume | 0.75 μL | None Applicable | Same | None Applicable |
| Test Time | 5 seconds | None Applicable | Same | None Applicable |
| Measuring Range | 20-600 mg/dL | None Applicable | Same | None Applicable |
| Operating Relative Humidity Range | 10-90% | None Applicable | Same | None Applicable |
| Memory Capacity | 500 blood glucose test results with date and time | None Applicable | Same | None Applicable |
| Coding | Auto Coding | None Applicable | Same | None Applicable |
| Meter Storage Conditions | 14~140 °F (-10~60°C) | None Applicable | Same | None Applicable |
| Test Strip Storage Conditions | 39~86 °F (4~30°C), < 90% relative humidity | None Applicable | Same | None Applicable |
| The unit of measurement data | Fix on mg/dL | None Applicable | Same | None Applicable |
| Sample | Capillary whole blood | None Applicable | Same | None Applicable |
| Interference | Uric acid ≥ 16 mg/dL
Xylose ≥ 10 mg/dL
Ascorbic acid ≥ 3 mg/dL
Dopamine HCl ≥ 1.25 mg/dL
L-Dopa ≥ 2 mg/dL | None Applicable | Same | None Applicable |
| Voice Function | Yes | None Applicable | Same | None Applicable |
| Strip Reagent | 1.FAD-glucose Dehydrogenase (FAD-GDH) 12.1% | None Applicable | Same | None Applicable |
8 - 16
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{16}
Bionime Corporation
| | 2.Potassium ferricyanide
48.5%
3.Non-reactive ingredients
39.4% | | | |
| --- | --- | --- | --- | --- |
| Measurement Technology | Dehydrogenase
Electrochemical Sensor | None Applicable | Same | None Applicable |
| Operating Temperature Range | 43 ~111 °F (6 ~ 44°C) | None Applicable | Same | None Applicable |
| Compatible to DMS | Compatible to GE
Diabetes Management System | Compatible to GE
Diabetes Management System | None Applicable | Same |
## 8. Discussion of Non-Clinical and Clinical Tests Performed for Determination of Substantial Equivalence:
Verification and validation of test results were evaluated to establish the performance, functionality and reliability of the Rightest Diabetes Management System, and the GE Diabetes Management System when used with the following systems: Rightest Blood Glucose Monitoring System GM700, Rightest Blood Glucose Monitoring System GM650, Rightest Blood Glucose Monitoring System GM300, GE300 Talking Blood Glucose Monitoring System, and GE200 Blood Glucose Monitoring System
## Data Transmission Memory Rollover Synchronization Verification
Data transmission accuracy and memory data rollover synchronization were tested at each meter’s full data memory capacity to ensure successful upload and storage into software database of the Rightest/GE Diabetes Management System. Software test used 3 meters per adapter.
| Meter/Adapter | Expected Test Result | Result |
| --- | --- | --- |
| Rightest GM700 Meter/GP550 Adapter (USB) | Full set of data (1000 records) transferred from meter to software was accurately downloaded
Meter memory data rollover accurately transmitted to Rightest DMS with additional new data | Pass |
| Rightest GM700 Meter/GP700 Adapter (Bluetooth) | Full set of data (1000 records) transferred from meter to software was accurately downloaded
Meter memory data rollover accurately transmitted to Rightest DMS with additional new data | Pass |
| Rightest GM650 Meter/GP650 Adapter (USB) | Full set of data (500 records) transferred from meter to software was accurately downloaded | Pass |
8 - 17
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{17}
Bionime Corporation
| | Meter memory data rollover accurately transmitted to Rightest DMS with additional new data | |
| --- | --- | --- |
| Rightest GM300 Meter/ GP300 Adapter (USB) | Full set of data (300 records) transferred from meter to software was accurately downloaded | Pass |
| | Meter memory data rollover accurately transmitted to Rightest DMS with additional new data | |
| GE200 Meter/ GEGP550 (USB) | Full set of data (1000 records) transferred from meter to software was accurately downloaded | Pass |
| | Meter memory data rollover accurately transmitted to GE DMS with additional new data | |
| GE200 Meter/ GEGP700 (Bluetooth) | Full set of data (1000 records) transferred from meter to software was accurately downloaded | Pass |
| | Meter memory data rollover accurately transmitted to GE DMS with additional new data | |
| GE300 Meter/ GEGP300 (USB) | Full set of data (500 records) transferred from meter to software was accurately downloaded | Pass |
| | Meter memory data rollover accurately transmitted to GE DMS with additional new data | |
## User Performance Evaluation:
The User (Layperson) Performance Evaluation of PC Link Adapters study included 45 participants of varying age, sex, background, education level, and work experience. The study was conducted from October to December 2013.
The study demonstrates that English-speaking/reading laypersons are able to follow user instructions for the PC link adapters and Software user manual to successfully connect PC Link Adapters and paired meter to PC installed and import data to GP200 Diabetes Management Software.
8 - 18
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{18}
Bionime Corporation
## Electromagnetic Compatibility (EMC):
To evaluate the compliance of safety requirements for electrical equipment – PC Adapters (GP300, GP650, and GP700), the testing and compliance reports were performed by SGS Taiwan Electronics & Communication Laboratory.
The EMC reports confirmed that each PC Adapter met performance criteria indicated by the standards.
## Readability Assessment:
The Readability Assessment Tests were performed using the Flesch-Kincaid Grade Level Score to evaluate the readability of the PC Adapter Guides and PC Adapter User Manuals for both Rightest BDMS and GE DMS Software.
The obtained results demonstrate that PC Adapter Guides and PC Adapter User Manuals for both Rightest BDMS and GE DMS Software each received a Flesch-Kincaid Grade Level that indicates each text is expected to be understood by an average student in the eighth grade. The assessments ranged from 6.2 to 7.4 and therefore each text meets the criteria.
## 9. Conclusion:
Results of performance evaluation of the Rightest Diabetes Management System and the GE Diabetes Management System with addition of compatible blood glucose meters and their respective PC Link Adapters demonstrate that the subject devices are substantially equivalent to the predicate devices.
8 - 19
510k k133522, Rightest and GE DMS with PC Link Adapters, 04/15/14
{19}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Document Control Center – WO66-G609
Silver Spring, MD 20993-0002
April 17, 2014
BIONIME COPORATION
C/O FENG-YU LEE
2700 LA PAZ ROAD
SUITE 266B
MISSION VIEJO CA 92691
Re: K133522
Trade/Device Name: Rightest Blood Glucose Monitoring System GM700
Rightest Blood Glucose Monitoring System GM650
Rightest Blood Glucose Monitoring System GM300
GE300 Talking Blood Glucose Monitoring System
GE200 Blood Glucose Monitoring System
Bionime Rightest Diabetes Management System
GE Diabetes Management System
Regulation Number: 21 CFR 862.1345
Regulation Name: Glucose test system
Regulatory Class: II
Product Code: NBW, LFR, CGA, JQP
Dated: March 03, 2014
Received: March 6, 2014
Dear Feng-Yu Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21
{20}
Page 2—Feng-Lu Lee
CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Courtney H. Lias -S
Courtney H. Lias, Ph.D.
Director
Division of Chemistry and Toxicology Devices
Office of In Vitro Diagnostics
and Radiological Health
Center for Devices and Radiological Health
Enclosure
{21}
| DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use** | Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page. |
| --- | --- |
510(k) Number (if known)<br/>
k133522
Device Name<br/>
GE Diabetes Management System
Indications for Use (Describe)<br/>
The GE Diabetes Management System is an over-the-counter software system for use by Health Care Professionals and Patients with diabetes as an aid for managing diabetes. User(s) can transfer blood glucose readings from the following glucose meters: GE100 Blood Glucose Monitoring System with GE PC Link Adaptor (GE GP550), GE300 Talking Blood Glucose Monitoring System with GE PC Link Adaptor for GE300 (GE GP300) and GE200 Blood Glucose Monitoring System with GE PC Link Adaptor (GE GP550) or GE Bluetooth PC Link (GE GP700) to a personal computer for the purpose of viewing, analyzing and printing the blood glucose readings, as well as to backup and to recover users' profile and data.
The GE Diabetes Management System is not intended to provide treatment decisions, nor should it substitute professional opinion. All medical diagnoses and treatment plans should be performed by a licensed health care professional.
Type of Use (Select one or both, as applicable)<br/>
☑ Prescription Use (Part 21 CFR 801 Subpart D)<br/>
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY<br/>
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Katherine Serrano -S
FORM FDA 3881 (1/14)
Page 1 of 2
PSC Publishing Services (301) 441-6740
{22}
| DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use** | Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page. |
| --- | --- |
510(k) Number (if known)<br/>
k133522
Device Name<br/>
Rightest Blood Glucose Monitoring System GM300
Indications for Use (Describe)<br/>
The Rightest Blood Glucose Monitoring System GM300 is intended for in vitro diagnostic use (outside of body). It is indicated to be used by professional healthcare personnel or diabetes at home to measure the glucose concentration for aiding diabetes management. The glucose concentration is measured with quantitative capillary whole blood from the fingertip, palm, and forearm by using Rightest Blood Glucose Monitoring System GM300. This device is not intended for testing neonate blood samples.
Special conditions for use statement(s): Rightest System provides plasma equivalent results.
Type of Use (Select one or both, as applicable)<br/>
☑ Prescription Use (Part 21 CFR 801 Subpart D)<br/>
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY<br/>
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Katherine Serrano -S
FORM FDA 3881 (1/14)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{23}
| DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use** | Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page. |
| --- | --- |
510(k) Number (if known)<br/>k133522
Device Name<br/>Rightest Blood Glucose Monitoring System GM700
Indications for Use (Describe)<br/>The Rightest Blood Glucose Monitoring System GM700 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700 is intended to be used by a single person and should not be shared.
The Rightest Blood Glucose Monitoring System GM700 is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM700 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
The Rightest Blood Glucose Test Strips GS700 are for use with the Rightest Blood Glucose Meter GM700 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.
Type of Use (Select one or both, as applicable)<br/>☐ Prescription Use (Part 21 CFR 801 Subpart D)<br/>☑ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY<br/>Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Katherine Serrano -S
FORM FDA 3881 (1/14) Page 1 of 2 PTC Publishing Services (301) 443-6740 EF
{24}
| DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use** | Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page. |
| --- | --- |
510(k) Number (if known)<br/>
k133522
Device Name<br/>
Bionime Rightest Diabetes Management System
Indications for Use (Describe)
The Bionime Rightest Diabetes Management System is an over-the-counter software system for use by Health Care Professionals and Patients with diabetes as an aid for managing diabetes. User(s) can transfer blood glucose readings from the following glucose meters: Rightest Blood Glucose Monitoring System GM550 and GM250 with Rightest PC Link Adaptor (GP550), Rightest Blood Glucose Monitoring System GM700 with Rightest PC Link Adaptor (GP550) or Rightest Bluetooth PC Link (GP700), Rightest Blood Glucose Monitoring System GM650 with Rightest PC Link Adaptor for GM650 (GP650), and Rightest Blood Glucose Monitoring System GM300 with Rightest PC Link Adaptor for GM300 (GP300) to a personal computer for the purpose of viewing, analyzing and printing the blood glucose readings, as well as to backup and to recover users' profile and data.
The Bionime Rightest Diabetes Management System is not intended to provide treatment decisions, nor should it substitute professional opinion. All medical diagnoses and treatment plans should be performed by a licensed healthcare professional.
Type of Use (Select one or both, as applicable)<br/>
☑ Prescription Use (Part 21 CFR 801 Subpart D)<br/>
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY<br/>
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Katherine Serrano -S
FORM FDA 3881 (1/14) Page 1 of 2 PHC Publishing Services (301) 441-6740 EF
{25}
| DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use** | Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page. |
| --- | --- |
510(k) Number (if known)<br/>
k133522
**Device Name**<br/>
GE200 Blood Glucose Monitoring System
**Indications for Use (Describe)**
The GE200 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE200 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
The GE200 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE200 Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
The GE200 Blood Glucose Test Strips are for use with the GE200 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.
Type of Use (Select one or both, as applicable)<br/>
☐ Prescription Use (Part 21 CFR 801 Subpart D)<br/>
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY<br/>
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Katherine Serrano -S
FORM FDA 3881 (1/14) Page 1 of 2
PX Publishing Services (910) 411-6749 EF
{26}
| DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use** | Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page. |
| --- | --- |
510(k) Number (if known)<br/>
k133522
Device Name<br/>
GE300 Talking Blood Glucose Monitoring System
Indications for Use (Describe)<br/>
The GE300 Talking Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE300 Talking Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
The GE300 Talking Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE300 Talking Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
The GE300 Blood Glucose Test Strip is for use with the GE300 Talking Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.
Type of Use (Select one or both, as applicable)<br/>
☐ Prescription Use (Part 21 CFR 801 Subpart D)<br/>
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY<br/>
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Katherine Serrano -S
FORM FDA 3881 (1/14)
Page 1 of 2
PPC Publishing Services (301) 443-6740
{27}
| DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use** | Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page. |
| --- | --- |
510(k) Number (if known)<br/>k133522
Device Name<br/>Rightest Blood Glucose Monitoring System GM650
Indications for Use (Describe)<br/>The Rightest Blood Glucose Monitoring System GM650 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM650 is intended to be used by a single person and should not be shared.
The Rightest Blood Glucose Monitoring System GM650 is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM650 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
The Rightest Blood Glucose Test Strips GS650 are for use with the Rightest Blood Glucose Meter GM650 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.
Type of Use (Select one or both, as applicable)<br/>☐ Prescription Use (Part 21 CFR 801 Subpart D)<br/>☑ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY<br/>Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Katherine Serrano -S
FORM FDA 3881 (1/14)
Page 1 of 2
PSC Publishing Services (301) 443-6740
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.