GLUCOFACTS EXPRESS DIABETES MANAGEMENT SOFTWARE

K082486 · Bayer Healthcare, LLC · NBW · Oct 8, 2008 · Clinical Chemistry

Device Facts

Record IDK082486
Device NameGLUCOFACTS EXPRESS DIABETES MANAGEMENT SOFTWARE
ApplicantBayer Healthcare, LLC
Product CodeNBW · Clinical Chemistry
Decision DateOct 8, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

GLUCOFACTS® Express Diabetes Management Software is an over-the-counter software program for use by health care professionals and patients with diabetes for transferring blood glucose data from a blood glucose meter to a personal computer for the purpose of viewing and printing reports that display blood glucose readings from Bayer's CONTOUR®, CONTOUR® TS, BREEZE®, and BREEZE®2 blood glucose meters.

Device Story

Software application transfers blood glucose results, time, date, and markers from Bayer blood glucose meters (CONTOUR, CONTOUR TS, BREEZE, BREEZE2) to a personal computer via serial or USB cable. Used by patients at home or healthcare professionals in clinics. Transforms raw meter data into five visual formats: electronic logbook, glucose trend by date, daily trend (standard day), weekly trend (standard week), and summary charts (histogram/pie chart). Enables printing of reports. Assists users in visualizing glucose patterns to support diabetes management decisions.

Clinical Evidence

Performance assessed in a study with 51 participants (43 lay users, 8 healthcare professionals). Study demonstrated ease of use and understandability of results.

Technological Characteristics

Software-based diabetes management system. Architecture: Java-based. Connectivity: PC-based with automatic meter detection and web-based update capability. Features: glucose range/alarm configuration, automated report generation, simplified UI/UX.

Indications for Use

Indicated for patients with diabetes and healthcare professionals to transfer, review, and analyze blood glucose data from specific Bayer meters to support diabetes management. Suitable for home and clinical use.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## Bayer HealthCare Diabetes Care OCT 0 8 2008 Image /page/0/Picture/8 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" written twice, once horizontally and once vertically, intersecting at the "Y". The text is enclosed in a circle. The logo is in black and white. #### 510(k) SUMMARY #### GLUCOFACTS® Express Data Management Software This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | The assigned 510(k) number is | K082484 | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Prepared: | August 25, 2008 | | Submitter: | Bayer HealthCare Diabetes Care | | Address: | 430 South Beiger Street<br>Mishawaka, IN 46544<br>Phone (574) 256-3441; FAX (574) 256-3519 | | Contact: | Roger Sonnenburg, Sr. Manager, Regulatory Affairs | | Device: | Trade/Proprietary Name: GLUCOFACTS® Express Diabetes<br>Management Software | | Common/Usual Name: | Diabetes data management software program. | | Classification: | Division of Clinical Laboratory Devices<br>Panel - Clinical Chemistry and Toxicology<br>Classification Code - 75 NBW, JQP | | Predicate Device: | Ascensia® Glucofacts® Data Management System,<br>k024234 | | Device Description: | This software application allows the transfer of blood<br>glucose results, along with time, date, and certain data<br>markers, from a Bayer blood glucose meter to a personal<br>computer through the use of a serial or USB cable. Data<br>analysis includes allowing the home-user or healthcare<br>professional to view the data in five different ways:<br>1. Electronic logbook where all of the data can be seen<br>2. Glucose trend of the results by date<br>3. Daily blood glucose trend (standard day)<br>4. Weekly blood glucose trend (standard week)<br>5. Summary chart (histogram or pie chart) | | Intended Use: | GLUCOFACTS® Express Diabetes Management Software is an<br>over-the-counter software program for use by health care | {1}------------------------------------------------ # | | professionals and patients with diabetes for transferring<br>blood glucose data from a blood glucose meter to a personal<br>computer for the purpose of viewing and printing reports that<br>display blood glucose readings from Bayer's CONTOUR®,<br>CONTOUR® TS, BREEZE®, and BREEZE®2 blood glucose<br>meters. | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Technological Characteristics: | There were no changes to the fundamental scientific<br>technology. | | Comparison to<br>Predicate device: | GLUCOFACTS® Express Diabetes Management Software is<br>similar in function to the predicate device, Ascensia<br>Glucofacts Data Management System, k024234, but has<br>been updated to make it compatible with the most recent<br>Bayer blood glucose meters and to improve the ease of use. | | Assessment of<br>Performance: | The performance was assessed in a study that included 51<br>people - 43 lay users and eight healthcare professionals.<br>The study showed that the program is easy to use and the<br>results are understandable by the users. | | Conclusion: | The results of the verification and validation studies of the<br>GLUCOFACTS® Express Diabetes Management Software<br>demonstrated that the product is safe and effective in the<br>hands of lay users and healthcare professionals. The<br>product is substantially equivalent to the predicate device. | {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) of the United States. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 OCT 0 8 2008 Bayer Healthcare, LLC c/o Mr. Roger Sonnenburg 430 South Beiger St. Mishawaka, IN 46544 Re: k082486 Trade/Device Name: Glucofacts Express Diabetes Management Software Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, JQP Dated: September 16, 2008 Received: September 17, 2008 Dear Mr. Sonnenburg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Féan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: GLUCOFACTS® Express Diabetes Management Software Indications For Use: GLUCOFACTS® Express Diabetes Management Software is an overthe-counter software program for use by persons with diabetes in the home and by healthcare professionals in healthcare facilities to assist with transferring blood glucose data from a blood glucose monitor to a personal computer to allow reviewing and analyzing the data to support effective diabetes management. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use × (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (OIVD) Division Sign-Off Page 1 of ____________________________________________________________________________________________________________________________________________________________________ Office of In Vitm Diagnostic Device Evaluation and Safety 5103) K082486
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...