K131164 · Siemens Medical Solutions USA, Inc. · IYN · Dec 24, 2013 · Radiology
Device Facts
Record ID
K131164
Device Name
ACUSON
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Dec 24, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The S2000 and S3000 ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult, Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, transesophageal, transvaginal, peripheral vessel, musculo-skeletal (conventional), musculoskeletal (superficial) and neonatal cardiac) and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the oction to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography: Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging". The Acuson Acunav Ultrasound Catheter is intended for intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.
Device Story
Multi-purpose mobile diagnostic ultrasound systems; acquire harmonic ultrasound echo data; display in B-Mode, PW/CW/Color/Amplitude Doppler, Harmonic, and 3D/4D imaging. Operated by clinicians in clinical settings. Includes VTQ (Virtual Touch Quantification) for shear wave elasticity imaging; uses focused push beams to induce tissue displacement; measures shear wave velocity to assess tissue stiffness. Includes Arterial Health Package (AHP) for Intima Media Thickness measurement. Output displayed on flat panel monitor; used adjunctively by physicians for clinical diagnosis. Benefits include non-invasive visualization of anatomy/physiology and quantitative assessment of tissue elasticity and vascular health.
Clinical Evidence
Bench testing only. No clinical data submitted. Devices evaluated for acoustic output, biocompatibility, and electrical/mechanical safety per standards (IEC 60601-2-37, ISO 10993-1, etc.).
Technological Characteristics
Mobile ultrasound system; B-mode, Doppler, 3D/4D imaging. VTQ shear wave elasticity imaging uses focused push beams (mechanical index 1.7) and detection beams. Patient contact materials tested to ISO 10993-1. Connectivity via standard ultrasound interfaces. Software-controlled; complies with UL 60601-1, IEC 60601-2-37, and AIUM/NEMA UD-2/UD-3 standards.
Indications for Use
Indicated for fetal, abdominal, intraoperative, pediatric, small parts, transcranial, OB/GYN, cardiac, pelvic, neonatal/adult cephalic, vascular, musculoskeletal, and peripheral vascular imaging in adult and pediatric patients. Includes intra-cardiac and intra-luminal visualization via Acunav catheter.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
KB1164
page 1 of 7
S2000 and S3000 VTQ 510(k) Submission
DEC 2 4 2013
# 510(k) Summary Prepared April 17, 2013
- Siemens Medical Solutions, Inc., 1. Sponsor: Ultrasound Division 685 East Middlefield Road Mountain View, California 94043
| Contact Person: | Shelly Pearce |
|-----------------|---------------------------|
| | Telephone: (650) 694-5988 |
| | Fax: (650) 694-5580 |
- 2. Device Name: Acuson S2000 and S3000 Diagnostic Ultrasound Systems
- Common Name: Diagnostic Ultrasound System
# Classification:
Regulatory Class: ll Review Category: Tier II Classification Panel: Radiology
| Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN |
|------------------------------------------|---------------|---------------------|
| Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 | Product Code 90-IYO |
| Diagnostic Ultrasound Transducer | FR # 892.1570 | Product Code 90-ITX |
| Diagnostic Ultrasound Catheter | FR # 870.1200 | Product Code OBJ |
# 3. Legally Marketed Predicate Devices
The Acuson S2000 and S3000 Ultrasound Systems are substantially equivalent to the company's own S2000 and S3000 Ultrasound Systems and Supersonic Imagine Shearwave Elastography (K111674, K122825, K112255, K121329)
# 4. Device Description:
The S2000 and S3000 Ultrasound Systems are multi-purpose mobile, software controlled diagnostic ultrasound systems with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, or Harmonic Imaging and 3D/4D Imaging on a Flat Panel Display. It is substantially equivalent to the S2000 (K111674) and S3000 system (K122825) which are legally marketed devices.
# 5. Intended Use
The S2000 and S3000 ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
{2}------------------------------------------------
| Siemens Medical Solutions, Inc. | S2000 and S3000 VTQ |
|---------------------------------|---------------------|
| Ultrasound Division | 510(k) Submission |
The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".
The Acuson Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.
| | Feature / Characteristic | Submission<br>Device<br>S2000 | Submission<br>Device<br>S3000 | Acuson<br>S3000<br>K122825 | Acuson<br>S2000<br>K111674 | Supersonic<br>Imagine<br>K112255 /<br>K121329 |
|---------------------------------------------------|------------------------------------------|-------------------------------|-------------------------------|----------------------------|-----------------------------------------------|-----------------------------------------------|
| | Indications for Use: | | | | | |
| ■ | Fetal | √ | √ | √ | √ | |
| ■ | Abdominal | √ | √ | √ | √ | √ |
| ■ | Intraoperative abdominal and<br>vascular | √ | √ | √ | √ | |
| ■ | Intraoperative neurological | -- | -- | -- | -- | -- |
| ■ | Pediatric | √ | √ | √ | √ | √ |
| ■ | Small Organ | √ | √ | √ | √ | √ |
| ■ | Neonatal cephalic | √ | √ | √ | √ | |
| ■ | Adult Cephalic | √ | √ | √ | √ | |
| ■ | Cardiac | √ | √ | √ | √ | |
| ■ | Trans-esophageal | √ | √ | √ | √ | |
| ■ | Transrectal | √ | √ | √ | √ | √ |
| ■ | Transvaginal | √ | √ | √ | √ | √ |
| ■ | Peripheral vessel | √ | √ | √ | √ | √ |
| ■ | Laparoscopic | -- | -- | -- | -- | -- |
| ■ | Musculo-skeletal (conventional) | √ | √ | √ | √ | √ |
| ■ | Musculo-skeletal (superficial) | √ | √ | √ | √ | √ |
| | Center Frequencies Supported: | | | | | |
| ■ | 2.0 MHz | √ | √ | √ | √ | √ |
| ■ | 3.0 MHz | √ | √ | √ | √ | √ |
| ■ | 3.2 MHz | √ | √ | √ | √ | √ |
| Feature / Characteristic | Submission<br>Device<br>S2000 | Submission<br>Device<br>S3000 | Acuson<br>S3000<br>K122825 | Acuson<br>S2000<br>K111674 | Supersonic<br>Imagine<br>K112255/<br>K121329 | |
| • 3.3 MHz | √ | √ | √ | √ | √ | |
| • 4.2 MHz | √ | √ | √ | √ | √ | |
| • 4.4 MHz | √ | √ | √ | √ | √ | |
| • 4.8 MHz | √ | √ | √ | √ | √ | |
| • 5.0 MHz | √ | √ | √ | √ | √ | |
| • 5.2 MHz | √ | √ | √ | √ | √ | |
| • 6.0 MHz | √ | √ | √ | √ | √ | |
| • 6.5 MHz | √ | √ | √ | √ | √ | |
| • 6.9 MHz | √ | √ | √ | √ | √ | |
| • 9.5 MHz | √ | √ | √ | √ | √ | |
| • 10.0 MHz | √ | √ | √ | √ | √ | |
| Modes: | | | | | | |
| • B | √ | √ | √ | √ | √ | |
| • Parallel processing in B mode | √ | √ | √ | √ | √ | |
| • M | √ | √ | √ | √ | | |
| • PWD (Pulsed Wave Doppler) | √ | √ | √ | √ | √ | |
| • CWD (Continuous Wave Doppler) | √ | √ | √ | √ | √ | |
| • D (Color Doppler) | √ | √ | √ | √ | √ | |
| • Amplitude Doppler | √ | √ | √ | √ | √ | |
| • Combined (BMDC) | √ | √ | √ | √ | √ | |
| Features: | | | | | | |
| • Quad processing in color | √ | √ | √ | √ | | |
| • Native™ tissue harmonic imaging | √ | √ | √ | √ | | |
| • SieScape™ panoramic imaging | √ | √ | √ | √ | | |
| • Color SieScape™ panoramic<br>imaging | √ | √ | √ | √ | | |
| • 3-Scape™ real-time 3D imaging | √ | √ | √ | √ | | |
| • fourSight™ 4D transducer<br>technology | √ | √ | √ | √ | | |
| • TEQ™ ultrasound technology | √ | √ | √ | √ | | |
| • Extend imaging technology | √ | √ | √ | √ | | |
| • Cardiac Imaging<br>physiological signal display | √ | √ | √ | √ | | |
| • syngo ® Auto OB measurements | √ | √ | √ | √ | | |
| • Advanced SieClear™ spatial<br>compounding | √ | √ | √ | √ | | |
| • STIC (Fetal Heart Imaging) | √ | √ | √ | √ | | |
| • Amnioscopic rendering | √ | √ | √ | √ | | |
| • Cadence contrast agent imaging | √ | √ | √ | √ | | |
| • Clarify™ vascular enhancement<br>technology | √ | √ | √ | √ | | |
| • eSie™ Touch elasticity imaging | √ | √ | √ | √ | √ | |
| • syngo ® Auto Left heart | √ | √ | √ | √ | | |
| • syngo ® Velocity Vector Imaging | √ | √ | √ | √ | | |
| • Semi Auto-segmentation (eSie<br>Calc) | √ | √ | √ | √ | | |
| Feature / Characteristic | Submission<br>Device<br>S2000 | Submission<br>Device<br>S3000 | Acuson<br>S3000<br>K122825 | Acuson<br>S2000<br>K111674 | Supersonic<br>Imagine<br>K112255 /<br>K121329 | |
| Custom Tissue Imaging / Speed of<br>Sound | √ | √ | √ | √ | | |
| AHP | √ | √ | √ | √ | | |
| VTQ | √ | √ | | | √ | |
| 18L6HD Transducer | √ | √ | √ | √ | | |
| 6C1HD Transducer | √ | √ | √ | √ | | |
| 8C3HD Transducer | | √ | √ | | | |
| Monitor: FPD | √ | √ | √ | √ | √ | |
| Output Display Standard (Track 3) | √ | √ | √ | √ | | |
| Patient Contact Materials | Tested to ISO<br>10993-1 | Tested to ISO<br>10993-1 | Tested to ISO<br>10993-1 | Tested to ISO<br>10993-1 | Tested to ISO<br>10993-1 | |
| UL 60601-1 Certified | √ | √ | √ | √ | √ | |
#### 6. Summary of Technological Characteristics - New Device Compared to Predicate
S2000/3000 510(k) Submission
Page 8 of 64
{3}------------------------------------------------
# S2000 and S3000 VTQ 510(k) Submission
:
.
S2000/3000 510(k) Submission
Page 9 of 64 .
.
{4}------------------------------------------------
# S2000 and S3000 VTQ 510(k) Submission
| Shear Wave Elasticity Imaging Software | VTQ | Supersonic<br>Imagine<br>Aixplorer | Comments |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparable Exam Types | Abdomen,<br>Breast,<br>Thyroid | Abdomen,<br>Breast,<br>Thyroid | |
| Transducer for same Exam Types | 6C1 HD,<br>4C1, 4V1,<br>9L4 | SC6-1<br>SL15-4 | |
| Operating Mode | Single Frame | Low Frame<br>Rate Real<br>Time | Note 1 |
| Cool Down period before live imaging<br>resumes | Yes (variable.<br>~1-3 sec) | No | Note 1 |
| Push Pulse sequence | Dual Push<br>Beams: one<br>on each side<br>of<br>measurement<br>ROI | Low (variable<br>with FOV<br>size)<br>Reference<br>Figure 13 in<br>Supersonic<br>White Paper | Note 1 |
| Multiple Push Pulse Focal Zones | No (Single<br>focal position<br>tracks with<br>ROI location<br>in depth) | "Mach Cone" | In both cases, radiation force uniformity is<br>maintained with depth by optimizing push beam<br>focus at region of interest. "Mach Cone" is the<br>Supersonic term for compound transmit focusing<br>"Supersonic Imagine has implemented a two<br>dimensional image display representing shear<br>wave speed or Young's modulus. The region of<br>interest size is adjustable within limits. A potential<br>advantage is the ability to visualize variability of<br>the elasticity parameter over a wider area of<br>interest, however this is visual reference without<br>quantification of variance. Outside the USA, the<br>ability to get a mean value is provided for the<br>larger area or with reduced measurement ROI |
| Two Dimensional Display with Elasticity<br>Region of Interest | No | Yes | size, a localized value. This feature is not cleared<br>in the USA. VTQ provides a mean shear wave |
| | | | speed over a predefined region of interest that is<br>not adjustable. For the 4C1, 6C1 HD and 4V1<br>abdomen transducers, the ROI size is 10mm x<br>5mm. For the 9L4 transducer, the ROI size is 5 x<br>5 mm. Multiple measurements can be made over<br>larger areas and statistics for Mean, Median.<br>Standard Deviation and IQR (Interquartile range)<br>are provided for analyzing multiple<br>measurements. There is no evidence to suggest<br>one method is more accurate than the other,<br>therefore we believe they are equivalent in clinical<br>use." |
| Localized Quantitative Shear Wave<br>Velocity measurement | Yes | Yes -<br>Predicate not<br>cleared in the<br>U.S.A. | |
| Localized Quantitative Young's Modulus<br>measurement | No | Yes -<br>Predicate not<br>cleared in the<br>U.S.A. | Supersonic Imagine has implemented the shear<br>wave velocity scale in the color bar which<br>indicates relative shear wave speed in the image<br>and has been cleared in the U.S. Young's<br>modulus is not cleared. Siemens VTQ is<br>equivalent to Supersonic except Siemens is<br>requesting clearance for point measurements in<br>addition to displaying shear wave velocity on the<br>color bar. When Young's modulus is displayed<br>(kPa), assumptions are made regarding tissue<br>density and viscosity that may not be correct in a<br>wide range of biological tissues and is therefore<br>an indirect measurement. Shear wave velocity is<br>a direct measurement. |
| Localized Shear Wave Velocity<br>measurement area | 6C1 HD,<br>4C1, 4V1:<br>10mm<br>(height) by 6<br>mm (width)<br>9L4: 5mm<br>(height) by 6<br>mm (width) | Yes -<br>Predicate not<br>cleared in the<br>U.S.A. | With VTQ, shear wave velocity estimates must<br>pass a confidence interval check (≥0.8) to display<br>a value, otherwise an "x.xx" result is displayed. |
| Shear Wave Propagation "Movie" (clip) | No | Yes | The Supersonic shear wave propagation movie<br>provides information specific to their<br>implementation of shear wave imaging, as noted<br>in Note 1. |
| Shear Wave Velocity Measurement Range | 0.5-5.5 m/sec<br>(6C1 HD,<br>4C1, 4V1)<br>0.5-8.4 m/sec<br>(9L4) | SC6-1: 0-10<br>m/sec<br>(default 0-3.2<br>m/sec)<br>SL15-4: 0-10<br>m/sec | Siemens has chosen a lower velocity cutoff of 0.5<br>m/s to avoid potential erroneous shear wave<br>velocity estimates. In in-vivo studies, shear wave<br>velocities below 0.5 m/s have not been<br>encountered so there is no impact on clinical<br>efficacy. For abdomen (liver) applications, in<br>several in-vivo studies the maximum recorded<br>values did not exceed 4.4 m/sec. |
| Color Coded Shear Wave Velocity Display | No | Yes | |
| Adjustable Maximum and Minimum<br>Velocity Scale | No | Yes | |
| In Color Code, Red is 'stiff' and blue is<br>'soft' | N/A | Yes | |
| Color map is transparent over B mode<br>display | N/A | Yes | |
| Other system features: | | | |
| B Mode Imaging | Yes | Yes | |
| Pulse Wave Doppler Imaging | Yes | Yes | |
| Color Flow Doppler Imaging | Yes | Yes | |
| Spatial Compound Imaging | Yes | Yes | |
| Speckle Reduction Image Processing | Yes | Yes | |
| Tissue Harmonic Imaging | Yes | Yes | |
S2000/3000 510(k) Submission
Page 10 of 64
{5}------------------------------------------------
# S2000 and S3000 VTQ 510(k) Submission
.
S2000/3000 510(k) Submission
Page 11 of 64
{6}------------------------------------------------
#### S2000 and S3000 VTQ 510(k) Submission
# Note 1:
There are differences in push pulse sequencing and shear wave detection between the predicate device (Supersonc Aixplorer) and Siemens VTQ. Siemens uses focused push beams on each side of the measurement ROI and a fixed number of closely spaced focused detection beams resulting in higher signal to noise ratio in shear wave imaging than the predicate device. A reduced push beam mechanical index is used on the Aixplorer as compared to the constant value of 1.7 with VTQ, which allows pseudo-real time operation on the Aixplorer of approximately 1 frame per second. The user is required to compound the image at the region of interest over several seconds with the Aixplorer to allow the image to "build up" over several frames to improve signal to noise ratio. VTQ is a single frame image acquisition that allows higher signal to noise ratio without image compounding and generates a more consistent shear wave velocity estimation based on our experience in in-vivo studies.
# 7. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence.
The devices have been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform with applicable medical device safety standards. The system complies with the following voluntary standards:
- UL 60601-1, Safety Requirements for Medical Equipment .
- IEC 60601-2-37 Diagnostic Ultrasound Safety Standards l
- CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment ●
- AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical . Acoustic Output Indices on Diagnostic Ultrasound Equipment
- . AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound
- 93/42/EEC Medical Devices Directive .
- Safety and EMC Requirements for Medical Equipment .
- · EN/IEC 60601-1
- · EN/IEC 60601-1-1
- EN/IEC 60601-1-2 .
- I IEC 1157 Declaration of Acoustic Power
- . ISO 10993-1 Biocompatibility
Cleared patient contact materials, electrical and mechanical safety are unchanged.
# 8. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence.
The S2000, S3000 and Aixplorer use the same technology and principles as existing devices, clinical data is not required.
# 9. Summary
Intended uses and other key features are consistent with traditional clinical practice and FDA guidelines. The design and development process of the manufacturer conforms with 21 CFR 820 Quality System Regulation and ISO 13485:2003 quality system standards. The product is
S2000/3000 510(k) Submission
{7}------------------------------------------------
designed to conform with applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore it is the opinion of Siemens Medical that the devices are substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
S2000/3000 510(k) Submission
Page 13 of 64
{8}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W066-G609 Silver Spring, MD 20993-0002
December 24, 2013
Siemens Medical Solutions USA, Inc. c/o Ms. Shelly Pearce Regulatory Affairs 1230 Shorebird Way MOUNTAIN VIEW CA 94043
Re: K131164
Trade/Device Name: Acuson S2000 and S3000 Diagnostic Ultrasound Systems Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, OBJ Dated: November 19, 2013 Received: November 22, 2013
Dear Ms. Pearce:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This determination of substantial equivalence applies to the following transducers intended for use with the Acuson S2000 Diagnostic Ultrasound Systems, as described in your premarket notification:
# Transducer Model Number
CW2 9L4 Linear Array 6C2 Curved Array 4VI Phased Array 7CF2 Curved Array 8V3 Phased Array EV8C4 V7M TEE
CWS 14L5 Multi-D Array 4C1 Curved Array 10V4 Phased Array 9EVF4 Curved Array 4V Ic Phased Array 8C3HD Curved Array ACUNAV 8F
EC9-4 Curved Array 4PI Phased Array 6CIHD Curved Array 14L5 SP Linear Array V5Ms Multiplane TEE ୧୮3 18L6 HD Linear Array ACUNAV 10F
{9}------------------------------------------------
Page 2-Ms. Pearce
..............................................................................................................................................................................
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 1 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/Medica|Devices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Samung Hi. Mesgo
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
#### 1.3 Indications for Use
K131164 A. 510(k) Number (if known):
Device Name: S2000 and S3000 Diagnostic Ultrasound Systems
#### Indications for Use:
The S2000 and S3000 ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult, Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, transesophageal, transvaginal, peripheral vessel, musculo-skeletal (conventional), musculoskeletal (superficial) and neonatal cardiac) and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the oction to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subcinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography: Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".
The Acuson Acunav Ultrasound Catheter is intended for intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.
Prescription Use × (Part 21CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
currence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Semung M. Mase
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health (OIR)
510(k).
Page 1 of
S2000/3000 510(k) Submission
Page 17 of 64
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.