The SuperSonic Imagine AIXPLORER® ultrasound system and transducer are intended for general purpose pulse echo ultrasound imaging and Doppler fluid flow analysis of the human body.
Device Story
Cart-based ultrasound system for non-invasive diagnostic imaging; utilizes linear, curved, micro-convex, and motorized linear array transducers. Inputs: acoustic signals processed via scan converter; outputs: B-Mode, Color Flow, Pulsed Wave Doppler, Harmonic Imaging, Amplitude Doppler, 3D imaging, and ShearWave™ elastography displayed on LCD monitor. Operated by clinicians in clinical settings. System allows image capture, measurement, and report generation. Facilitates clinical evaluation and diagnosis through enhanced image quality and elastography; supports biopsy guidance. Benefits include non-invasive diagnostic assessment of tissue and fluid flow.
Clinical Evidence
Bench testing only. Performance verified through compliance with electrical safety (IEC 60601-1, 60601-1-1, 60601-1-2, 60601-2-37), acoustic output standards (NEMA UD-2, UD-3), and biocompatibility (ISO 10993-1). Additional verification testing performed for Panoramic and High PRF measurement features. All results met pre-determined specifications.
Technological Characteristics
Cart-based ultrasound system; LCD monitor; scan converter; linear, curved, micro-convex, and motorized linear array transducers. Complies with IEC 60601-1 (safety), IEC 60601-2-37 (ultrasound safety), and NEMA UD-3 (acoustic output). Biocompatible materials per ISO 10993-1. Modes: B-Mode, Color Flow, Pulsed Wave Doppler, Harmonic Imaging, Amplitude Doppler, 3D, and ShearWave™ elastography.
Indications for Use
Indicated for use in Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans-vaginal, and Neonatal Cephalic applications. Provides anatomical structure measurements for these same applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K121329
#### AUG 2 4 2012 510(k) Summary of Safety and Effectiveness
This summary of safety and effectiveness information is submitted in accordance with 21 CFR §807.92.
### 1) Submitter's name, address, telephone number, contact person
Submitted by: SuperSonic Imagine, S.A. Les Jardins de la Duranne - Bât. E & F 510, rue René Descartes 13857 Aix-en-Provence Cedex France Telephone: 011 33 442 99 24 24
Distributed by: SuperSonic Imagine, Inc. 11714 North Creek Parkway N Suite 150 Bothell, WA 98011 North America Telephone: +1(425) 686 6380
Corresponding Official:
Jacques Souquet Chief Executive Officer Telephone: 011 33 442 99 24 35
Date: 2012/08/21
### 2) Name of the device, including the trade or proprietary name if applicable. the common or usual name, and the classification name, if known:
Common/Usual Name: Diagnostic Ultrasound System with Accessories Proprietary Name: Aixplorer®
Classification:
Regulatory Class: II
| Classification Name: | 21 CFR Section | Product Code |
|------------------------------------------|----------------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
### 3) Substantially Equivalent/Predicate Devices
AIXPLORER® Ultrasound Imaging System (K102041), cleared on 10/13/2010 Siemens Acuson S2000TM Diagnostic Ultrasound System (K072786), cleared on 11/13/2007 Philips iU22 Ultrasound System (K093563), cleared on 02/01/2010
{1}------------------------------------------------
### 4) Description of Device
The SuperSonic Imagine AIXPLORER® system is a cart based ultrasound imaging system used to perform non-invasive diagnostic general purpose ultrasound imaging studies. The system contains a scan converter and can be coupled to a variety of linear, curved, micro-convex, and motorized linear array transducers to produce images, which are displayed on a LCD monitor. An adiustable control panel with integrated touch screen allows the user to perform an ultrasound exam quickly and efficiently in accordance with ALARA principles. The system also allows the user to perform measurements, capture images to digital memory or to an external device (such as a printer), and review diagnostic studies in the form of a report. The system functions in a manner identical to the predicate devices for the imaging modes: B-Mode, Color Flow, Pulsed Wave Doppler, Harmonic Imaging, Amplitude Doppler, 3D imaging and for ShearWave™ elastography.
## 5) Intended Use
The SuperSonic Imagine AIXPLORER® ultrasound system and transducer are intended for general purpose pulse echo ultrasound imaging and Doppler fluid flow analysis of the human body.
## 6) Indication for Use
The SuperSonic Imagine AIXPLORER® ultrasound system is indicated for use in the following applications: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans-vaginal and Neonatal Cephalic.
The system also provides the ability to measure anatomical structures (Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Peripheral Vascular, GYN, Pelvic, Pediatric, Urology, Transrectal, Trans-vaginal, Neonatal Cephalic).
## 7) Safety Considerations
As a Track 3 ultrasound device, the SuperSonic Imagine AIXPLORER® ultrasound system is designed to comply with the "Standard For Real Time Display Of Thermal And Mechanical Acoustic Output Indices On Diagnostic Ultrasound Equipment" AIUM/NEMA 2004a published by the National Electrical Manufacturers Association as UD -3. With respect to limits on acoustic outputs. the SuperSonic Imagine AIXPLORER® ultrasound system complies with the FDA quideline limits set in the September 9, 2008, 510(k) diagnostic ultrasound guidance.
With regard to general safety, the SuperSonic Imagine AIXPLORER® ultrasound system scanner is designed to comply with IEC 60101 - 1 (2005) Medical Electrical Equipment. Part 1: General Requirements for Basic Safety and Essential Performance, and IEC 60601 - 2-37 (2007): Particular Requirements For The Safety Of Ultrasonic Medical Diagnostic And Monitoring Equipment.
| Mechanical Index | 1.9 (Maximum) |
|------------------|-------------------|
| TIS/TIB | 0.1 - 4.0 (Range) |
| ISPTA (d) | 720 mW/cm2 |
| ISPPA (d) | 0 - 700 W/cm2 |
The device's acoustic output limits are:
The limits are the same as predicate Track 3 devices. These considerations apply to all modes the system offers.
### 8) Comparison to Predicate Devices
The SuperSonic Imagine AIXPLORER® system and transducers are substantially equivalent to the predicate devices with regard to intended use, imaging capabilities, safety and effectiveness.
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis. .
- The systems have the same clinical indications for use.
{2}------------------------------------------------
- . The systems have the same B-Mode (grayscale imaging) and Doppler capabilities.
- . The systems have similar capability in terms of harmonic imaging, spatial compound imaging, elastography imaging and other image post-processing features to improve the image quality and aid in clinical evaluation and diagnosis.
- . The transducers are similar in materials, manufacture and clinical capability.
- . The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- . The systems have acoustic power levels which are below the applicable FDA limits.
- . The systems have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- . The systems have been found to be manufactured in compliance with approved electrical and physical safety standards.
#### Nonclinical Performance Data 9)
Non-clinical testing was conducted per the following standards to support a determination of substantial equivalence to the predicate devices.
| Reference Standard | Tests Performed |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 3rd Edition | All applicable electrical, basic safety and essential<br>performance tests. |
| UL 60601-1 1st Edition | All applicable electrical, basic safety and essential<br>performance tests specific to the U.S.A. |
| IEC 60601-1-1 2nd Edition | All applicable tests pertaining to Medical Electrical<br>Systems. |
| IEC 60601-1-2 3rd Edition | All applicable testing pertaining to electromagnetic<br>compatibility. |
| IEC 60601-2-37 2nd Edition | All applicable testing pertaining to the particular<br>requirements for the safety of ultrasonic medical<br>diagnostic and monitoring equipment. |
| NEMA UD 2 (Rev. 3) | All tests applicable in order to demonstrate<br>compliance with the "Accoustic Output<br>Measurement Standard for Diagnostic Ultrasound<br>Equipment". |
| NEMA UD 3 (Rev. 2) | All tests applicable in order to demonstrate<br>compliance with the "Standard For Real Time<br>Display Of Thermal And Mechanical Acoustic<br>Output Indices On Diagnostic Ultrasound<br>Equipment". |
| ISO 10993-1 | Applicable biocompatibility tests per FDA 510(k)<br>Memorandum - #G95-1 - per the appropriate<br>device category. |
In addition to the referenced standards testing, SuperSonic Imagine conducted the following performance tests with respect to the Panoramic and High PRF features:
| Panoramic Measurement Verification Tests |
|------------------------------------------|
| High PRF Measurement Verification Tests |
The above testing confirmed that the Aixplorer System performs according to the stated intended use. All data fell within pre-determined product specifications and external standard requirements. Results of nonclinical testing confirmed the substantial equivalence of the Aixplorer System to the predicate device(s).
{3}------------------------------------------------
## 10) Conclusion
The documentation provided demonstrates that:
- 1) The system and transducers are substantially equivalent to the predicate devices.
- 2) There are no new questions of safety and effectiveness concerning the SuperSonic Imagine AIXPLORER® ultrasound system and transducers.
- 3) The ultrasound device has been scientifically evaluated and has been demonstrated to be at least as safe and effective as the predicate devices cited in item 3. The system's acoustic power levels are below the applicable FDA limits.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
### Food and Drug Administration
10903 New Hampshire Avenue Silver Spring, MD 20993
AUG 2 4 2012
Jacques Souquet, Ph.D. Chief Executive Officer SUPERSONIC imagine Les Jardins de la Duranne 510, rue Rene Descartes - Bat. E et F 13 857 AIX-EN-PROVENCE CEDEX FRANCE
Re: K121329
Trade/Device Name: AIXPLORER® Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: July 2, 2012 Received: July 2, 2012
Dear Dr. Souquet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the AIXPLORER® Ultrasound System, as described in your premarket notification:
Transducer Model Number
SL15-4 (1D Linear Array) SC6-1 (Curved Array) SE12-3 (Endocavitary)
SLV 16-5 (Motorized Linear) SL10-2 (Linear) SMC12-3 (Micro-Curved)
{5}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely Yours
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{6}------------------------------------------------
510(k) number (if known):
Device Name: AIXPLORER® Ultrasound System
Device Name: Alixi EURER® Online of the malysis of the human body as follows:
#### Indications for Use:
Indications for Use.
The SuperSonic Imagine AIXPLORER® ultrasound system is indicated for use in the following The SuperSonic Tinagile Anxi CONEN® Citaceae Chical Musculoskeletal, Vasculoskeleta, Vascular, Peripheral
applications: Abdominal, Small Organs, Musculoskeletal, Superficial applications: Abdominal, Omail Organo) metal, Trans-vaginal and Neonatal Cephalic.
The system also provides the ability to measure anatomical structures (Abdominal, Small Organs, The System also "provides" the "admy" to "htoudro" anatomical CSYN, Pelvic, Pediatric, Urology, Transrectal, Trans-vaginal, Neonatal Cephalic).
Prescription Use __XX_______ OR Over-The-Counter Use _
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 11
(Division Sign-Off)
Division of Radiological Devices
DIVD
Stok 8/2/3/29
{7}------------------------------------------------
510(k) number (if known):
Device Name: AIXPLORER® Ultrasound System
Device Name: AIXPLORER® Olirasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as files
| Clinical Application | | Mode of Operation | | | | | | | |
|---------------------------|---------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|-------------------------|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | | |
| Other | Abdominal (including urolology) | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, 10, N, 9 | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, 10, N, 9 | |
| | Small Organ (Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, 10, N, 9 | |
| | Neonatal Cephalic | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 9 | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8 | |
| | Trans-vaginal | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8 | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, 10, N, 9 | |
| | Musculo-skeletal (Superficial) | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, 10, N, 9 | |
| | Intravascular, | | | | | | | | |
| | GYN | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, 10 | |
| | Pelvic | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, 10 | |
| | Other (Specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, 10, N, 9 | |
| | Other (Specify) | P | P | | | P | P, 1, 2, 3, N, 4 | P, 5, 6, 7, 8, N, 9 | |
N = new indication; P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6: Spatial Compounding
o: Hannomo millionio
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
(Division Sign-Off)
Division of Radiological Devices
10: 3D Imaging
111
510k
OR Over-The-Counter Use Prescription Use __XX_ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{8}------------------------------------------------
510(k) Number (if known):
Device Name: SL15-4 transducer (1D Linear Array Transducer)
Device Name: SE 15-4 transdaoor (1D Einear Ana) 1 flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|-------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging & | Fetal | | | | | | | |
| Other | Abdominal | P | | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, 9 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, 9 |
| | Small Organ (for example Breast,<br>Thyroid, Testicle, Prostate, Penis) | P | | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, 9 |
| | Neonatal Cephalic | N | | N | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 9 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, 9 |
| | Musculo-skeletal (Superficial) | P | | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, 9 |
| | Intravascular | | | | | | | |
| | GYN | | | | | | | |
| | Pelvic | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | P | | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, 9 |
| Vessel | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
Auditional Comments:
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
2: Combined modes include: B+ ShearWave™
3: Combined modes include: B+ Pulsed Wave
4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
# 6: Spatial Compounding
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
sion Sian-Off
Division of Radiological Devices
10: 3D Imaging
(D)v
510
OR Over-The-Counter Use Prescription Use __XX__ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Page 3 of 8
{9}------------------------------------------------
510(k) Number (if known):
Device Name: SC6-1 transducer (curved array transducer)
Device Name: SOO- Fitansador (our rou and) « and flow analysis of the human body as follows:
| | Clinical Application | | Mode of Operation | | | | | |
|---------------------------|---------------------------------------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal (including urolology) | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, S |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, S |
| | Small Organ (Breast, Thyroid, Testicle,<br>Prostate, penis, etc...) | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, |
| | Musculo-skeletal (Superficial) | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, |
| | Intravascular | | | | | | | |
| | GYN | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Pelvic | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, |
| Vessel | Other (Specify) | P | P | | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, N, |
N = new indication; P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
/ Idaltonal Sommons:
1: Combined modes include: B+ ShearWave™ Elastography
2: Combined modes include: B+ ShearWave™ Elastography
11: Combined modes include: B+ ShearWave™
3: Combined modes include: B+ Pulsed Wave
5: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6: Spatial Compounding
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
dioiogical Devices
9: Panoramic Imaging
10: 3D Imaging
510k
OR Over-The-Counter Use Prescription Use __XX_
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{10}------------------------------------------------
510(k) Number (if known):
Device Name: SE12-3 transducer (endocavitary transducer)
Device Name: OE IL o transcover (maging or fluid flow analysis of the human body as follows:
| | Clinical Application | | | | | | Mode of Operation | | | | |
|---------------------------|---------------------------------------------------------------------------------|--|---|---|---|---|-------------------|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | | B | M | P | W | D | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| | Ophthalmic | | | | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | | | | |
| | Abdominal | | | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | | |
| | Laparoscopic | | | | | | | | | | |
| | Pediatric | | | | | | | | | | |
| | Small Organ (for example Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | | P | | P | | W | D | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Neonatal Cephalic | | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | | |
| | Trans-rectal | | | P | | | W | D | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Trans-vaginal | | P | P | | | W | D | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Trans-urethral | | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | | | |
| | Muscuio-skeletal (Superficial) | | | | | | | | | | |
| | Intravascular | | | | | | | | | | |
| | GYN | | | P | | | W | D | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Pelvic | | P | P | | | W | D | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Other (Specify) | | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | | |
| | Other (Specify) | | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | P | P | | | W | D | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8 |
| | Other (Specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
# 6: Spatial Compounding
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
(Division Sign-Off)
vision of Radiblogical Devices 0
10: 3D Imaging
the
210k
Prescription Use __XX_________ OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{11}------------------------------------------------
510(k) Number (if known):
Device Name: SLV16-5 transducer (motorized linear transducer)
Device Name: OLV ro-o tranddaoor (motorial a filter analysis of the human body as follows:
| | Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|------------------------|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) | |
| | Ophthalmic | | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | | |
| | Abdominal | P | P | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, 10, N, 9 | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | P | P | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, 10, N, 9 | |
| | Small Organ (for example Breast,<br>Thyroid, Testicle, Prostate, penis,<br>etc...) | P | P | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, 10, N, 9 | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, 10, N, 9 | |
| | Musculo-skeletal (Superficial) | P | P | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, 10, N, 9 | |
| | Intravascular | | | | | | | | |
| | GYN | | | | | | | | |
| | Pelvic | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | P 1, 2, 3, N, 4 | P 5, 6, 7, 8, 10, N, 9 | |
| | Other (Specify) | | | | | | | | |
N = new indication; P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
Auditional Oommonts:
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
2: Combined modes include: B+ ShearWave™
3: Combined modes include: B+ Pulsed Wave
5. Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6. Spatial Compounding
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
iplogical Devices
8. Imaging Guidance for
9. Panoramic Imaging
9: Panoramic Imaging
10: 3D Imaging
510k
OR Over-The-Counter Use Prescription Use _____________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{12}------------------------------------------------
510(k) Number (if known):
Device Name: SL10-2 transducer (linear transducer)
Device Name: SE 10-2 Iranoducer (mour itanodios)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | | | | | Mode of Operation | | |
|---------------------------|---------------------------------------------------------------------------------|---|---|-----|-----|-------------------|-----------------------|------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Small Organ (for example Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Neonatal Cephalic | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 9 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Musculo-skeletal (Superficial) | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Intravascular | | | | | | | |
| | GYN | | | | | | | |
| | Pelvic | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Other (Specify) | N | N | | | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
N = new indication; P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
/ Latitional Somments
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6: Spatial Compounding
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
an Sign-Off ision of Radiological Devices
10: 3D Imaging
ામ Divisi
C
510k
__ OR Over-The-Counter Use Prescription Use _____________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(Please do not write below this line-continue on another Page if needed)
{13}------------------------------------------------
510(k) Number (if known):
Device Name: SMC12-3 transducer (micro-curved transducer)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | | Mode of Operation | | | | | |
|---------------------------|---------------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal<br>Abdominal | N | N | N | N | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | N | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Small Organ (for example Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | N | N | N | N | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Neonatal Cephalic | N | N | N | N | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 9 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | N | N | N | N | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Musculo-skeletal (Superficial) | N | N | N | N | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
| | Intravascular | | | | | | | |
| | GYN | | | | | | | |
| | Pelvic | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | N | N | N | N | N | N, 1, 2, 3, 4 | N, 5, 6, 7, 8, 9 |
N = new indication; P = previously cleared by FDA (K102041)
Additional Comments:
1: **Combined modes include:** B+ Color Flow
Additional Commont:
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6: Spatial Compounding
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
18 M
(Division Sign-Off) of Radiological Devices
9: Panoramic Imaging
10: 3D Imaging
ﻟﻠ
510
OR Over-The-Counter Use Prescription Use ___XX_
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.