The SuperSonic Imagine AIXPLORER® ultrasound system and transducer are intended for general purpose pulse echo ultrasound imaging and Doppler fluid flow analysis of the human body.
Device Story
Cart-based ultrasound system for non-invasive diagnostic imaging; utilizes linear, curved, micro-convex, and motorized linear array transducers. Inputs: acoustic signals processed via scan converter. Outputs: B-Mode, M-Mode, Doppler, 3D imaging, and ShearWave™ elastography displayed on LCD monitor. Operation: clinician-controlled via panel/touchscreen; supports image capture, measurement, and reporting. Elastography mode features a graduated numeric scale for dynamic range selection. Used in clinical settings for general diagnostic ultrasound and fluid flow analysis; aids in clinical evaluation and diagnosis through image post-processing and anatomical measurements.
Clinical Evidence
Bench testing only. Performance evaluated using CIRS 049A elasticity phantom to assess shearwave velocity estimation bias and precision across various target stiffnesses (I-IV) and sizes. Depth dependence of velocity estimation also verified. Results demonstrated satisfactory agreement and high precision for all transducers and depths.
Technological Characteristics
Cart-based ultrasound system; complies with IEC 60601-1 and IEC 60601-2-37. Acoustic output complies with AIUM/NEMA UD-3 (Track 3). Transducers: linear, curved, micro-convex, motorized linear arrays. Imaging modes: B-Mode, M-Mode, Doppler, 3D, ShearWave™ elastography. Connectivity: digital memory, external printer support.
Indications for Use
Indicated for abdominal, small organs, musculoskeletal, superficial musculoskeletal, vascular, peripheral vascular, GYN, pelvic, pediatric, urology, trans-rectal, and trans-vaginal applications. Provides measurement of anatomical structures in these areas.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
## AUG 28 2012
This summary of safety and effectiveness information is submitted in accordance with 21 CFR §807.92.
1) Submitter's name, address, telephone number, contact person Submitted by:
SuperSonic Imagine, S.A. Les Jardins de la Duranne - Bât. E & F 510, rue René Descartes 13857 Aix-en-Provence Cedex France Telephone: 011 33 442 99 24 24
Distributed by: SuperSonic Imagine, Inc. 11714 North Creek Parkway N Suite 150 Bothell, WA 98011 North America Telephone: +1(425) 686 6380
Corresponding Official:
Jacques Souquet Chief Executive Officer Telephone: 011 33 442 99 24 35
Date: August 24, 2012
## 2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, If known:
Common/Usual Name: Diagnostic Ultrasound System with Accessories Proprietary Name: Aixplorer®
Classification:
Regulatory Class: II
| Classification Name: | 21 CFR Section | Product Code |
|------------------------------------------|----------------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
#### Substantially Equivalent/Predicate Devices 3)
AIXPLORER® Ultrasound Imaging System (K091970), cleared August 12, 2009 AIXPLORER® Ultrasound Imaging System (K102041), cleared October 18, 2010
{1}------------------------------------------------
### 4) Description of Device
The SuperSonic Imagine AIXPLORER® system is a cart based ultrasound imaging system used to perform non-invasive diagnostic general purpose ultrasound imaging studies. The system contains a scan converter and can be coupled to a veriety of linear, curved, micro-convex, and motorized linear array intensducers to produce images that are displayed on a LCD monitor. An adjustable control panel with integrated touch screen allows the user to perform an ultrasound exam quickly and efficiently in accordance with ALARA principles. The system also allows the user to perform measurements, capture images to digital memory or to an external device (such as a printer), and review diagnostics studies in the form of a report. The system functions in a manner identical to the predicate devices in the the imaging modes: B-Mode, Calleno in a mannel lucinical to the president for the imaging, Amplitude Doppler, 3D imaging and for ShearWave™ elastography.
#### 5) Intended Use
The SuperSonic Imagine AIXPLORER® ultrasound system and transducer are intended for general purpose pulse echo ultrasound imaging and Doppler fluid flow analysis of the human body.
#### 6) Indication for Use
The SuperSonic Imagine AIXPLORER® ultrasound system is indicated for use in the following applications: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vasculoskeletal, Vascular, Peripheral Vascular, GYN, Pelvic, Pediatric, Urology, Trans-rectal and Trans-vaginal.
The system also provides the ability to measure anatomical structures (Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Peripheral Vascular, GYN, Pelvic, Peditatin, Uniology, Transrectal and Trans-vaginal).
#### 7) Safety Considerations
As a Track 3 ultrasound device, the SuperSonic Imagine AIXPLORER® ultrasound system is designed to comply with the "Standard For Real Time Display Of Thermal And Mechanical Acoustic Output Indices On Diagnostic Ultrasound Equipment" AIUM/NEMA 2004a published by the National Electrical Manufacturers Association as UD -3. With respect to limits on acoustic outputs, the SuperSonic Imagine AIXELORER® ultrasound system complies with the FDA guideline limits set in the September 0, 2008, 510(k) diagnostic ultrasound guidance.
With regard to general safety, the SuperSonic Imagine AIXPLORER® ultrasound system scanner is designed to comply with IEC 60101 -1 (2005) Medical Electrical Equipment, Part 1: General Requirements for Basic Safety and Essential Performance, and IEC 60601 - 2-37 (2007): Particular Requirements For The Safety Of Ultrasonic Medical Diagnostic And Monitoring Equipment.
| Mechanical Index | 1.9 (Maximum) |
|------------------|-------------------|
| TIS/TIB | 0.1 - 4.0 (Range) |
| ISPTA (d) | 720 mW/cm2 |
| ISPPA (d) | 0 - 700 W/cm2 |
The device's acoustic output limits are:
The limits are the same as predicate Track 3 devices. These considerations apply to all modes the system offers.
### 8) Comparison to Predicate Devices
The SuperSonic Imagine AIXPLORER® system and transducers are identical to the predicate devices with regard to intended use, imaging capabilities, safety and effectiveness.
・
{2}------------------------------------------------
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- The systems have the same clinical indications for use. .
- The systems have the same B-Mode (grayscale imaging) and Doppler capabilities. .
- The systems have similar capability in terms of harmonic imaging, spatial compond imaging, . elastography imaging and other image post-processing features to improve the image quality and aid in clinical evaluation and diagnosis.
- . The transducers are identical in materials, manufacture and clinical capability.
- The systems are manufactured with materials which have been evaluated and found to be safe . for the intended use of the device.
- . The systems have acoustic power levels which are below the applicable FDA limits.
- The systems have identical capability in terms of performing measurements, capturing digital . images, reviewing and reporting studies.
- The systems have been found to be manufactured in compliance with approved electrical and . physical safety standards.
The SuperSonic Imagine AlXPLORER® system and transducers are identical to the predicate device . identified in K102041; only the display of the device is changed in the elastography mode. A gradiated and adjustable numeric scale, with units, will be added next to the color bar, so that users will be able to change the dynamic range displayed and return repeatedly and consistently to the selected range for the particular patient examination.
#### SUPERSONIC IMAGINE
{3}------------------------------------------------
### 9) Nonclinical Performance Data
The SuperSonic Imagine AIXPLORER® system and transducers are identical to the predicate device identified in K102041; only the display of the device is changed in the preofically one provide device additional non clinical performance data was submitted for determination of substantial equivalnee.
The following non clinical performance tests were conducted to demonstrate the performance of the device in the elastography mode.
### 9.1 Bias and Precision of Shearwave Velocity Estimation
This test was conducted using a CIRS 049A Quality Assurance elasticity phantom. This phantom contains stepped-cylinder targets of four stiffness types I – IV corresponding to nominal shear wave speeds of 1.6, 2.2, 3.9 and 5.2 m/s of different diameters, embedded in uniform-elasticity background (nominal shear wave speeds of 2.9 m/s).
Shearwave velocity estimation bias was derived as the difference between the mean of five independent shear wave velocity measurements and the nominal shear wave velocity, normalized by the nominal shear wave velocity and expressed as a percentage.
Shearwave velocity estimation precision was derived as the standard deviation of five independent shear wave velocity measurements normalized by the mean of the five independent measurements, and expressed as a percentage.
The tables below document the estimation bias and estimation precision as a function of target size and target stiffness for all transducers where the Elastography mode is available.
| | | | | SLV16-5 | | | | |
|-------------------------|--|---------------------------|----------------|------------------|-----------------|------------------|----------------|------------------|
| | | TargetType: I | TargetType: II | | TargetType: III | | TargetType: IV | |
| Target Diameter<br>(mm) | | BIAS PRECISION<br>(%) (%) | BIAS | PRECISION<br>(%) | BIAS | PRECISION<br>(%) | BIAS | PRECISION<br>(%) |
| 4.1 | | 21.9 3.9 | -7.0 | 2.2 | 19.2 | 2.3 | 24.1 | 1.9 |
| 6.5 | | 7.6 0.2 | -7.8 | 2.2 | -8.7 | 1.3 | 12.5 | 2.2 |
| 10.4 | | 6.4 2.2 | -8.6 | 1.8 | -3.2 | 1.3 | 3.7 | 0.9 |
| 16.7 | | 1.6 0.2 | -8.6 | 1.8 | -0.2 | 1.0 | 9.5 | 0.9 |
| SL15-4 | | | | | | | | | |
|-------------------------|---------------|----------|----------------|----------|-----------------|----------|----------------|----------|---------------|
| | TargetType: I | | TargetType: II | | TargetType: III | | TargetType: IV | | |
| Target Diameter<br>(mm) | | BIAS (%) | PRECISION (%) | BIAS (%) | PRECISION (%) | BIAS (%) | PRECISION (%) | BIAS (%) | PRECISION (%) |
| 4.1 | | 19.5 | 0.2 | -5.3 | 0.3 | - | - | 24.5 | 2.3 |
| 6.5 | | 7.6 | 0.2 | -7.8 | 2.2 | -9.7 | 0.4 | 12.5 | 0.9 |
| 10.4 | | 1.6 | 0.2 | -9.4 | 0.3 | -3.7 | 1.3 | -0.9 | 1.5 |
| 16.7 | | 1.6 | 0.2 | -8.6 | 1.8 | 5.3 | 0.4 | 8.8 | 1.3 |
page 5 of 7
{4}------------------------------------------------
| SL10-2 | | | | | | | | | | | | |
|----------------------|---------------|----------|---------------|----------------|----------|---------------|-----------------|----------|---------------|----------------|----------|---------------|
| Target Diameter (mm) | TargetType: I | BIAS (%) | PRECISION (%) | TargetType: II | BIAS (%) | PRECISION (%) | TargetType: III | BIAS (%) | PRECISION (%) | TargetType: IV | BIAS (%) | PRECISION (%) |
| 4.1 | | 25.5 | 0.2 | | -4.5 | 1.7 | | 21.8 | 1.3 | | 22.2 | 1.0 |
| 6.5 | | 13.5 | 0.2 | | -5.3 | 0.3 | | 10.2 | 1.1 | | -9.8 | 1.7 |
| 10.4 | | 7.6 | 0.2 | | -5.3 | 0.3 | | -0.2 | 1.0 | | -0.1 | 2.0 |
| 16.7 | | 1.6 | 0.2 | | -5.3 | 0.3 | | 6.3 | 1.2 | | 7.2 | 0.9 |
| SE12-3 | | | | | | | | |
|-------------------------|---------------|------------------|----------------|------------------|-----------------|------------------|----------------|------------------|
| | TargetType: I | | TargetType: II | | TargetType: III | | TargetType: IV | |
| Target Diameter<br>(mm) | BIAS<br>(%) | PRECISION<br>(%) | BIAS<br>(%) | PRECISION<br>(%) | BIAS<br>(%) | PRECISION<br>(%) | BIAS<br>(%) | PRECISION<br>(%) |
| 6.5 | | | | | - | | - | |
| | 26.7 | 1.9 | -2.8 | 2.1 | 13.7 | 1.4 | 17.9 | 1.9 |
| 10.4 | 7.6 | 0.2 | -6.1 | 1.8 | -5.7 | 1.3 | -0.9 | 0.8 |
| 16.7 | 1.6 | 0.2 | -5.3 | 0.3 | 1.3 | 1.2 | 9.9 | 0.7 |
| SC6-1 | | | | | | | | |
|-------------------------|---------------|------------------|----------------|------------------|-----------------|------------------|----------------|------------------|
| Target Diameter<br>(mm) | TargetType: I | | TargetType: II | | TargetType: III | | TargetType: IV | |
| | BIAS<br>(%) | PRECISION<br>(%) | BIAS<br>(%) | PRECISION<br>(%) | BIAS<br>(%) | PRECISION<br>(%) | BIAS<br>(%) | PRECISION<br>(%) |
| 6.5 | 43.4 | 0.2 | 2.1 | 1.6 | 27.3 | 0.4 | 39.6 | 1.3 |
| 10.4 | 21.9 | 2.4 | -2.0 | 1.7 | 14.2 | 1.2 | 22.2 | 1.0 |
| 16.7 | 5.2 | 2.8 | -7.0 | 2.2 | -7.2 | 0.0 | 11.0 | 3.4 |
The results of this test demonstrate that the shearwave velocity estimates exhibit high precision for all target types and sizes considered, and are accurate with target sizes compatible to a transidered sis elastography spatial resolution.
## 9.2 Depth Dependence of Elastography Velocity Estimation
The Depth Dependence Test was conducted using the same CIRS049A phantom as in the Bias and Precision Test described above.
Shearwave velocity estimates were obtained through the uniform-elasticity background material of the CIRS 049A phantom at multiple depths up to the penetration depth of each transducer. Five indent measurements were performed per depth and were used to calculate the sheanwave velocity average and standard deviation values per depth.
| The table below provides the results of this test: | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | |
| SLV15-6 | SL15-4 |
|---------|--------|
| SL10-2 | SE12-3 |
| SC6-1 | |
SUPERSONIC IMAGINE
page 6 of 7
{5}------------------------------------------------
| DEPTH<br>(mm) | SWV<br>AVG<br>(m/s) | SWV<br>STD<br>(m/s) | SWV<br>STD<br>(m/s) | SWV<br>STD<br>(m/s) | SWV<br>AVG<br>(m/s) | SWV<br>STD<br>(m/s) | SWV<br>AVG<br>(m/s) | SWV<br>STD<br>(m/s) | SWV<br>AVG<br>(m/s) | SWV<br>STD<br>(m/s) |
|---------------|---------------------|---------------------|---------------------|---------------------|---------------------|---------------------|---------------------|---------------------|---------------------|---------------------|
| 5 | 2.86 | 0.05 | 2.78 | 0.04 | 2.80 | 0.01 | 2.90 | 0.01 | | |
| 10 | 2.80 | 0.01 | 2.80 | 0.01 | 2.72 | 0.04 | 2.80 | 0.01 | | |
| 15 | 2.70 | 0.01 | 2.68 | 0.04 | 2.70 | 0.01 | 2.70 | 0.01 | | |
| 20 | 2.72 | 0.04 | 2.70 | 0.01 | 2.70 | 0.01 | 2.68 | 0.01 | | |
| 25 | 2.70 | 0.01 | 2.70 | 0.01 | 2.68 | 0.04 | 2.64 | 0.05 | | |
| 30 | | | | | 2.70 | 0.01 | | | 2.90 | 0.01 |
| 35 | | | | | 2.70 | 0.01 | | | NA | NA |
| 40 | | | | | 2.80 | 0.01 | | | 2.70 | 0.01 |
| 50 | | | | | | | | 2.70 | 0.01 | |
| 60 | | | | | | | | 2.60 | 0.01 | |
| 70 | | | | | | | | 2.60 | 0.01 | |
The results of this test demonstrate that the shearwave velocity estimates exhibit satisfactory agreement for all transducers and depths considered.
#### 10) Conclusion
The documentation provided demonstrates that:
- 1) The system and transducers are substantially equivalent to the predicate devices.
- 2) There are no new questions of safety and effectiveness concerning the SuperSonic Imagine AIXPLORER® ultrasound system and transducers.
- 3) The ultrasound device has been scientifically evaluated and demonstrated to be as safe and effective as the predicate devices identified in item 3.
The system's acoustic power levels are below the applicable FDA limits.
#### SUPERSONIC IMAGINE
page 7 of 7
{6}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure.
#### Food and Drug Administration
10903 New Hampshire Avenue Silver Spring, MD 20993
AUG 28 2012
Mr. Jacques Souquet Chief Executive Officer SuperSonic Image, S.A. Les Jardins de la Duranne - Bat. E & F 510, rue Rene Descartes AIX-EN-PROVENCE CEDEX 13857 FRANCE
Re: K112255
Trade/Device Name: AIXPLORER® Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: June 26, 2012 Received: June 26, 2012
Dear Mr. Souquet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the AIXPLORER® Ultrasound System, as described in your premarket notification:
## Transducer Model Number
| SL 15-4 (1D linear array) | SE 12-3 (endocavitary) |
|---------------------------|---------------------------|
| SC6-1 (curved array) | SLV16-5 (motorized linear |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can
{7}------------------------------------------------
be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Brendan O'Leary at (301) 796-6898.
Sincerely Yours.
Amherst D. Ohm for
Janine M. Morris Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{8}------------------------------------------------
510(k) number (if known):
Device Name: AIXPLORER® Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
#### Indications for Use:
The SuperSonic Imagine AIXPLORER® ultrasound system is indicated for use in the following applications: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, GYN, Pelvic, Pediatric, Urology, Trans-rectal and Trans-vaginal.
The system also provides the ability to measure anatomical structures (Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Peripheral Vascular, GYN, Pelvic, Pediatric, Urology, Transrectal and Trans-vaginal).
Prescription Use XX
(Part 21 CFR·801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 11
(Division Sign-Off)
Division of Radiological Devices
510k
{9}------------------------------------------------
510(k) number (if known):
Device Name: AIXPLORER® Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | |
|---------------------------|---------------------------------------------------------------------|--|-------------------|---|-----|-----|---------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | | B | M | PWD | CWD | Color Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | | |
| | Abdominal (including urolology) | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7<br>8 |
| | Small Organ (Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Trans-vaginal | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Musculo-skeletal (Superficial) | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Intravascular | | | | | | | | |
| | GYN | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7,<br>8 |
| | Pelvic | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7,<br>8 |
| | Other (Specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| Vessel | Other (Specify) | | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
P = previously cleared by FDA (K102041)
Additional Comments:
510k
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
h
(Division Sign Off)
ഗ്ഗം
(Division Sign-Off) Olvision of Radiological Devices
{10}------------------------------------------------
3: Combined modes include: B+ Pulsed Wave
4: Harmonic Imaging
5: Spatial Compounding
്
# 7: Imaging Guidance for Biopsies
8: 3D Imaging
Prescription Use __XX______ OR Over-The-Counter Use __________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 3 of 11
signature
112255
{11}------------------------------------------------
510(k) Number (if known):
Device Name: SL15-4 transducer (1D Linear Array Transducer)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | | |
|---------------------------|-------------------------------------------------------------------------|-------------------|---|---|---|---|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | P | W | D | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | | | |
| | Abdominal | P | | P | | | | P | P 1, 2, 3 | P 4, 5, 6, 7 |
| | Intra-operative (Specify) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | P | | P | | | | P | P, 1, 2, 3 | P 4, 5, 6, 7 |
| | Small Organ (for example Breast,<br>Thyroid, Testicle, Prostate, Penis) | P | | P | | | | P | P, 1, 2, 3 | P 4, 5, 6, 7 |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skeletal (Conventional) | P | | P | | | | P | P, 1, 2, 3 | P 4, 5, 6, 7 |
| | Musculo-skeletal (Superficial) | P | | P | | | | P | P, 1, 2, 3 | P 4, 5, 6, 7 |
| | Intravascular | | | | | | | | | |
| | GYN | | | | | | | | | |
| | Pelvic | | | | | | | | | |
| | Other (Specify) | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (Specify) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | | P | | | | P | P, 1, 2, 3 | P 4, 5, 6, 7 |
| | Other (Specify) | | | | | | | | | |
P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
1: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
4: Harmonic Imaging
21 %
5. Spatial Compounding
b: Shearwave Elastography
5: Spatial Compounding
6: ShearWave™ Elastography
7: Imaging Guidance for Biopsies
Jmull D. O'R
OIVD
(Division Sign-Off) ર્જિતિ Division of Radiological Devices
{12}------------------------------------------------
Prescription Use __XX________ OR Over-The-Counter Use _
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 5 of 11 .
- Michael D. O'Brien
k112255
{13}------------------------------------------------
510(k) Number (if known):
Device Name: SC6-1 transducer (curved array transducer)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|---------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal (including urology) | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Small Organ (Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Musculo-skeletal (Superficial) | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Intravascular | | | | | | | |
| | GYN | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Pelvic | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
| | Other (Specify) | P | P | | P | | P, 1, 2, 3 | P, 4, 5, 6, 7 |
で
OR Over-The-Counter Use
ت تري
P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
1: Combined modes include: D + Color Friow
2: Combined modes include: B+ ShearWave™ Elastography
XX
3: Combined modes include: B+ Pulsed Wave
4: Harmonic Imaging
5: Spatial Compounding
TM
5: Spatial Compounding
6: ShearWave™ Elastography
7: Imaging Guidance for Biopsies
Prescription Use
Michael D. Okerlund
(Division Sign-Off) of Radiological Devices
210k
{14}------------------------------------------------
## (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) of the may Page 7 of 11 ' 412255
{15}------------------------------------------------
510(k) Number (if known):
Device Name: SE12-3 transducer (endocavitary transducer)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (for example Breast,<br>Thyroid, Testicle, Prostate, penis,<br>etc...) | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7 |
| | Trans-vaginal | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial). | | | | | | | |
| | Intravascular | | | | | | | |
| | GYN | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7 |
| | Pelvic | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7 |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7 |
P = previously cleared by FDA (K102041)
Additional Comments:
1: Combined modes include: B+ Color Flow
1: Combined modes include: B+ ShearWave™ Elastography
Combined modes include B+ Pulsed Wave
(Division Sign-Off) on of Radiological Devices OTVO
{16}------------------------------------------------
Page 9 of 11
4: Harmonic Imaging 5: Spatial Compounding
6: ShearWave™ Elastography 7: Imaging Guidance for Biopsies
Prescription Use __XX_______ OR Over-The-Counter Use _
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Murld D'Ohm
F112255
{17}------------------------------------------------
510(k) Number (if known):
Device Name: SLV16-5 transducer (motorized linear transducer)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Small Organ (for example Breast,<br>Thyroid, Testicle, Prostate, penis,<br>etc...) | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Musculo-skeletal (Superficial) | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| | Intravascular | | | | | | | |
| | GYN | | | | | | | |
| | Pelvic | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | P | P | | | P | P, 1, 2, 3 | P, 4, 5, 6,<br>7,8 |
| Vessel | Other (Specify) | | | | | | | |
P = previously cleared by FDA (K102041)
Additional Comments:
510k
1: Combined modes include: B+ Color Flow
2: Combing a modes include: B ShearWave™ Elastography
స్ట్ ਾ ਕਿ
3.70pmbined modes include: B+P Ised Wave
(Division Sign-Off)
Division of Radiological Devices OTVE
{18}------------------------------------------------
Page 11 of 11
4: Harmonic Imaging 5: Spatial Compounding
6: ShearWave™ Elastography 7: Imaging Guidance for Biopsies 8: 3D imaging
Prescription Use ___XX________ OR Over-The-Counter Use _______________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Michael D Othen
611225-5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.