K123209 · Delphinus Medical Technologies, Inc. · IYO · Dec 19, 2013 · Radiology
Device Facts
Record ID
K123209
Device Name
SOFTVUE
Applicant
Delphinus Medical Technologies, Inc.
Product Code
IYO · Radiology
Decision Date
Dec 19, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Intended Use
SoftVue™ is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast when used with an automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening mammography.
Device Story
SoftVue is an automated tomographic B-mode diagnostic ultrasound system used as an adjunct to mammography. The patient lies prone with the breast submerged in a warm water imaging chamber. A built-in curvilinear transducer ring performs automated scanning. A camera provides live video for positioning. The system collects ultrasound data, which is processed by a reconstruction engine into B-mode image slices that can be stacked into a volumetric image. The system includes a water control system for de-gassing and warming, a computer control system, and a touchscreen console for operator input. Patient data is entered via touchscreen or barcode reader. Output consists of images stored on the system or transferred via DICOM to a PACS environment for clinical review. The device is operated by clinical users in a diagnostic setting to assist in breast imaging.
Clinical Evidence
No clinical data. Evidence consists of non-clinical design verification and validation testing, including system performance evaluations, acoustic output testing, and simulated use tests using an anthropomorphic breast phantom seeded with 18 inclusions to measure resolution and contrast.
Technological Characteristics
Automated tomographic B-mode ultrasound system. Components: curvilinear transducer, water control system (de-gassing/warming), computer control, image reconstruction engine, data acquisition system. Connectivity: DICOM/PACS. Safety standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37.
Indications for Use
Indicated for B-mode ultrasonic imaging of the breast in patients. Not for use as a replacement for screening mammography.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K172256 — SoftVue · Delphinus Medical Technologies · Mar 2, 2018
P200040 — SoftVue Automated Whole Breast Ultrasound System with Sequr Breast Interface Assembly · Delphinus Medical Technologies · Oct 6, 2021
K142517 — SoftVue · Delphinus Medical Technologies, Inc. · Oct 31, 2014
K162372 — QT Ultrasound Breast Scanner-1 · Qt Ultrasound, LLC · Jun 6, 2017
K190314 — Advanced Intelligent Volumetric Ultrasound System, Model AiVUS 1000 · Mednovel Technology, Ltd. · Oct 9, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a document with handwritten text at the top and printed text at the bottom. The handwritten text reads "K123209 Page 1 of 4". The printed text at the bottom reads "Section 03 - 510(k) S". The document appears to be a page from a larger document, possibly related to a 510(k) submission.
Delphinus Medical Technologies. Inc.
Traditional 510(k) - SoftVue
DEC 1 9 2013
#### Submitter Information 1.0
- 】.【 This Premarket Notification is submitted by:
Delphinus Medical Technologies, Inc. 46701 commerce Center Drive Plymouth, Michigan 48170
- l.2 Contact Information:
| Contact Name: | Andrea N. Wallen |
|---------------|-------------------------|
| Telephone: | 734-306-1069 |
| Fax: | 734-207-3165 |
| E-mail: | awallen@delphinusmt.com |
- 1.3 October 2, 2012 Date:
#### 2.0 Device Name
- 2.1 Trade/ Proprietary Name: SoftVue
- 2.2 Common Name:
- 沙 System, Imaging, Pulsed Echo Ultrasonic
- ﺗﺮ Transducer. Ultrasonic. Diagnostic
- 2.3 Classification Name:
- 21 CFR § 892.1560: Ultrasonic pulsed echo imaging system : A
- 21 CFR § 892-1570; Diagnostic ultrasonic transducer ; )
#### Predicate Device 3.0
The predicate device is identified as the Sofia™ (ATUS) Imaging Device manufactured (ATUS) by iVu Imaging Corporation. The Sofia™ (ATUS) Imaging Device received market clearance under 510(k) number K080555.
#### 4.0 Device Description
Soft Vue is an automated tomographic B-mode diagnostic ultrasound system to be used as an adjunct to mammography for imaging a patient's breast. SoftVue is comprised of the following subsystems: the Transducer. Table/Housing. Water Control System, Computer Control System, Image Reconstruction System, Power System, and the Data Acquisition System.
{1}------------------------------------------------
SoftVue has a built-in curvilinear transducer that is used to acquire ultrasound images. I mages are acquired from a patient lying prone on the table with their breast submerged in an imaging chamber filled with warm (body temperature) water. The breast is positioned in the center of the transducer. A camera, located at the bottom of the imaging chamber provides a live video feed to the system operator to aid in positioning the patient's breast. Once the scan is initiated. the transducer collects data that are processed to produce a series of ultrasound image slices that can be stacked to vield a volumetric ultrasound image of the breast,
The Water Control System is used to de-gas and warm the water that is used as the image acquisition medium. The Computer Control System controls all of the functionality of the other subsystems. The reconstruction engine processes the image data acquired by the transducer ring into B-mode ultrasound images.
The system includes a barcode reader, touchscreen console (user interface), and display monitor. The touchscreen console and monitor allow the clinical user to perform an imaging procedure on a patient. Patient information is entered into the system using either the touchscreen console or the barcode reader. Device errors and warnings are displayed on the console and/or display.
Soft Vue outputs the image Storage Medium that allows the images to be stored until they are reviewed on a workstation or transferred via DICOM to a PACS environment.
#### 5.0 Intended Use
Soft Vue™ is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast when used with an automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening mammography.
#### Predicate Device Comparison 6.0
Delphinus Medical Technologies claims that the SoftVue device is substantially equivalent to the Sofia (ATUS) Imaging Device cleared by the FDA in K080555. Delphinus Medical Technologies claims substantial equivalence because the proposed device has an equivalent intended use. manufacturing materials, operating principles, physical and operational specifications as compared to the predicate device.
The SoftVue device and the predicate device utilize B-mode gravscale ultrasound images to achieve their intended use. Both the Sofia (ATUS) and Soft Vue are table-top systems and have automatic scanning transducers to image breast tissue.
The specific details regarding similarities and differences between SoftVue and iVu Sofia have been identified and explained in the Comparison Table section provided in Section 10 of this submission. A brief summary of the similarities and differences between Soft Vue™ and the Sofia (ATUS) Imaging Device is included below. The differences noted between SoftVue™ and the Sofia (ATUS) do not present any new issues related to safety and effectiveness.
{2}------------------------------------------------
# Similarities:
- Similar to the Sofia™ (ATUS), Soft Vue uses an automated transducer to acquire 2D ;众 images of a patient's breast.
- // Soft Vue™ and the Sofia™ (ATUS) use broadband transducers.
- نز Both systems acquire and process B-mode grayscale images of a patient's breast.
- Soft Vue™ and the Sofia™ (ATUS) position the patient in a prone position lying on 公 their examination table with the patient's in a pendulous position within an imaging chamber.
- Both the SoftVue™ and Sofia™ (ATUS) position the patient's breast in a fluid ﺎﺭ environment to eliminate the need for breast compression and facilitate the transmission of ultrasound waves.
# Differences:
- The Sofia™ (ATUS) utilizes a commercially available 510(k) approved linear transducer manufactured by GE Medical Systems. SoftVue™ uses a custom curvilinear transducer manufactured for Delphinus Medical Technologies.
- 公 The Sofia™ (ATUS) utilizes a frequency range of 5-13 MHz, whereas SoftVue™ has a single operating frequency of 3 MHz.
#### Summary of Testing 7.0
The function and performance of SoftVue™ will be evaluated through non-clinical design verification and validation testing. Testing includes system performance and simulated use tests. The results of the evaluation tests will demonstrate that SoftVue™ successfully meets the requirements of its intended use.
Delphinus Medical Technologies conducted performance evaluations to verify that Soft Vue 's subsystems successfully meet predetermined specifications and product performance requirements. Results of the testing performed demonstrate that SoftVue's subsystems meet the system and performance requirements necessary for its intended use. A brief list of some of the testing performed is included below.
- > Acoustic output testing Meets the Global Maximum Output
- 产 Sub-system Verification Testing - All requirements were met.
System Verification and Validation Testing will be performed using an anthropomorphic phantom breast designed for measuring resolution and contrast. The breast phantom was seeded with 18 inclusions located in specific locations within the phantom.
In addition to the tests listed above. SoftVue™ will also undergo testing to the safety standards listed in Table 3-1 prior to market release.
{3}------------------------------------------------
| Standard # and Date | Standard Title |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2005 | Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988;<br>Amendment 1, 1991-11, Amendment 2, 1995 |
| IEC 60601-1-1:2000 | Medical electrical equipment -- Part 1-1: General requirements for safety -<br>Collateral standard: Safety requirements for medical electrical systems. |
| IEC 60601-1-2:2007 | Medical Electrical Equipment - Part 1-2: General Requirements for Safety -<br>Collateral standard: Electromagnetic Compatibility - Requirements and Tests<br>(Edition 2:2001 with Amendment 1:2004) |
| IEC 60601-1-4:2000 | Medical electrical equipment - Part 1-4: General requirements for safety -<br>Collateral standard: Programmable electrical medical systems. edition 1.1 |
| IEC 60601-2-37:2007 | Medical electrical equipment - Part 2-37: Particular requirements for the basic<br>safety and essential performance of ultrasonic medical diagnostic and<br>monitoring equipment. |
Table 3 - 1: Safety Testing to be Performed on SoftVue
#### 8.0 Conclusion
SoftVue performs as intended and is substantially equivalent to the Sofia (ATUS) Imaging Device with respect to intended use, design, principles of operation, technology, materials, and performance. Any noted differences do not raise any new issues of safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2013
Delphinus Medical Technologies, Inc. % Ms Andrea Wallen Director of Quality and Regulatory Affairs 46701 Commerce Center Drive PLYMOUTH MI 48170
Re: K123209
Trade/Device Name: SoftVue Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO Dated: November 27, 2013 Received: December 02, 2013
Dear Ms. Wallen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Ms. Wallen
"If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approvad: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
# 510(k) Number (if known)
K123209
Device Name
Soft Vue
Indications for Use (Describe)
SoftVue™ is indicated for use as a B-mode ultrasonic imaging of a patient's breast when used with as automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening manmography.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7)
FORM FDA 3881 (9/13)
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, Including the lime to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of Information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Offica of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."
. . . . . . . . . .
{8}------------------------------------------------
# System: SoftVue
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | |
| Other | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Breast) | N | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
Additional Comments: SoftVue is intended for ultrasonic breast examinations
Concurrence of Center for Devices and Radiological Health (CDRH)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.