K172256 · Delphinus Medical Technologies · IYO · Mar 2, 2018 · Radiology
Device Facts
Record ID
K172256
Device Name
SoftVue
Applicant
Delphinus Medical Technologies
Product Code
IYO · Radiology
Decision Date
Mar 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
SoftVue™ is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast when used with an automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening mammography.
Device Story
SoftVue 3.0 is a B-mode ultrasonic imaging system for breast imaging. Patient lies prone on a table with breast submerged in a warm water imaging chamber. A Breast Interface Assembly uses gentle vacuum to center, shape, and stabilize the breast. An automated curvilinear array transducer (single-crystal) scans the breast to acquire ultrasound data. The system processes data into B-mode image slices, which can be stacked for volumetric visualization. It also generates colorized and grayscale relative stiffness maps and combined reflection/sound speed images for radiologist reference. A touchscreen interface allows operator control and patient data entry; DICOM worklist integration is supported. Images are transmitted to an external PACS server for review. The system includes a water conditioning unit for de-gassing and temperature control. The device aids radiologists by providing additional reference information for breast tissue assessment.
Clinical Evidence
No clinical data. Performance was established via non-clinical design verification and validation, including subsystem and full system evaluation, phantom imaging analysis, and acoustic output testing.
Indicated for B-mode ultrasonic imaging of the breast in patients using an automatic scanning curvilinear array transducer. Not for use as a replacement for screening mammography.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Delphinus Medical Technologies % Ms. Jillian M. Reed Regulatory and Clinical Affairs Consultant 45525 Grand River Avenue NOVI MI 48374
March 2, 2018
Re: K172256
Trade/Device Name: SoftVue Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: January 26, 2018 Received: January 31, 2018
Dear Ms. Reed:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
| <b>Indications for Use</b> |
| Form Approved: OMB No. | 0910-0120 |
|------------------------|--------------------------|
| Expiration Date: | January 31, 2017 |
| | See PRA Statement below. |
| 510(k) Number (if known) | K172256 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SoftVue |
| Indications for Use (Describe) | SoftVue™ is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast when used with an automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening mammography. |
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (1/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
{3}------------------------------------------------
# อท/นู้บริจาระหว่าง
swollot se ypod usund att to sissfirms woll binness, in 21,50138.0 1.3.11.2019.11.1
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|---------------------------------|-------------------|---|-----|----------------|--------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD<br>Doppler | Color<br>(Specify) | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging & Other | Fetal<br>Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Breast) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
xipuəddə sidi 1961년 bəbb = ヨ :AQJ vq bərrələ (ISBN) 777-11-2017-2017 19:40 = 14
suoitsuimsin 1585-19 viri 107 bəbdərini si ən Vilos :sənəmməd İstinadlar
{4}------------------------------------------------
# 1.0 Submitter Information
- 1.1 This Premarket Notification is submitted by:
Delphinus Medical Technologies, Inc. 45525 Grand River Avenue Novi, MI 48374
- 1.2 Contact Information:
| Contact Name: | Jillian M. Reed |
|-------------------|-------------------------|
| Office Telephone: | 860-238-7724 |
| E-mail: | jillianreed@charter.net |
- 1.3 Date: February 20, 2018
### Device Name 2.0
- Trade/ Proprietary Name: SoftVue 2.1
- 2.2 Common Name:
- A System, Imaging, Pulsed Echo Ultrasonic
- Transducer, Ultrasonic, Diagnostic A
### 2.3 Classification Name:
- > 21 CFR § 892.1560: Ultrasonic pulsed echo imaging system
- > 21 CFR § 892.1570: Diagnostic ultrasonic transducer
### 2.4 Product Codes
- > IYO Ultrasonic pulsed echo imaging system
- > ITX Diagnostic ultrasound transducer
### 3.0 Predicate Device
The predicate device is identified as SoftVue™ 2.0, manufactured by Delphinus Medical Technologies. SoftVue™ 2.0 received market clearance under 510(k) number K142517.
## 4.0 Device Description
SoftVue™ 3.0 is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast using a curvilinear array transducer that completely surrounds the breast. The SoftVue™ System is comprised of the following subsystems: Transducer, Table/Housing Assembly, Water Conditioning System, Computer Control System, Image Reconstruction System, Power System, and the Data Acquisition System.
{5}------------------------------------------------
The SoftVue™ System has a built-in curvilinear transducer that is used to acquire ultrasound data. Data are acquired from a patient lying prone on the table with their breast submerged in an imaging chamber filled with warm (body temperature) water. The breast is positioned in the center of the transducer. A Breast Interface Assembly connects to the distal end of the breast using a gentle vacuum, drawing the breast downward to center, shape, and stabilize the breast for optimal imaging. Cameras located at the bottom of the imaging chamber provide a live video feed to the system operator to aid in positioning the patient's breast.
Once the scan is initiated, the transducer collects data that are processed to produce a series of B-mode ultrasound image slices that can be stacked to yield a volumetric ultrasound image of the breast. SoftVue™ also outputs colorized and grayscale relative stiffness image stacks that provide the radiologist with additional reference information.
The Water Control System is used to de-gas and warm the water that is used as the image acquisition medium. The Computer Control System controls the functionality of the other subsystems. The reconstruction engine processes the image data acquired by the transducer ring into B-mode ultrasound images.
The system includes a touchscreen display (user interface). The touchscreen display allows the user to control the system and perform an imaging procedure. Patient information is entered into the system using the QWERTY keyboard on the touchscreen display, or the user can import the patient information from a DICOM modality worklist on an externally networked RIS server. Device errors and warnings are displayed on the touchscreen display.
SoftVue™ outputs the images to an externally networked PACS server which allows the images to be stored until they are reviewed on a workstation.
# Intended Use 5.0
SoftVue™ is indicated for use as a B-mode ultrasonic imaging system for imaging of a patient's breast when used with an automatic scanning curvilinear array transducer. The device is not intended to be used as a replacement for screening mammography.
### Predicate Device Comparison 6.0
Delphinus Medical Technologies claims that the SoftVue device is substantially equivalent to the SoftVue system cleared by the FDA in K142517. Delphinus Medical Technologies claims substantial equivalence because the proposed device has an equivalent intended use, manufacturing materials, operating principles, physical and operational specifications as compared to the predicate device.
The SoftVue device and the predicate device utilize B-mode grayscale ultrasound images to achieve their intended use. Both the modified SoftVue device and the predicate SoftVue device are table-top systems that have automatic scanning transducers to image breast tissue.
The specific details regarding similarities and differences between the modified SoftVue 3.0 device and the predicate SoftVue 2.0 device have been identified and explained in the
{6}------------------------------------------------
Comparison Table section of this submission. A summary of the similarities and differences between the modified SoftVue 3.0 device and the predicate SoftVue 2.0 device is included below. The differences noted between the modified SoftVue 3.0 device and the predicate SoftVue 2.0 device do not present any new issues related to safety and effectiveness.
# Similarities:
- A The modified and predicate SoftVue devices use an automated transducer to acquire stacked coronal images of a patient's breast.
- The modified and predicate SoftVue devices use broadband transducers. A
- A The modified and predicate SoftVue devices have a single operating frequency of 3 MHz.
- A Both systems acquire and process B-mode grayscale images of a patient's breast.
- A Both systems position the patient in a prone position lying on the examination table with the patient's breast in a pendulous position within an imaging chamber.
- A Both systems position the patient's breast in a fluid environment to eliminate the need for breast compression and facilitate the transmission of ultrasound waves.
# Differences:
- A The modified SoftVue device includes a Breast Interface Assembly which is used to center, stabilize, and optimize the shape of the breast. The predicate device does not have a Breast Interface Assembly.
- A The modified SoftVue device allows user access to directly view the sound speed images used to create the color relative stiffness maps. Access to these images is for additional reference information only. The predicate device does not allow user access to sound speed images.
- A The modified SoftVue device uses a single-crystal transducer. The predicate device uses a standard PZT crystal.
- A The modified SoftVue device uses a different transducer array firing pattern, as compared to the predicate device.
- >> The modified SoftVue device delivers an update transmit pulse configuration to enhance low frequency content.
- A The modified SoftVue device offers an alternative reflection imaging output that combines reflection and sound speed images, as additional reference information for radiologists. The predicate device does not output an alternative reflection image.
### 7.0 Summary of Non-Clinical Testing
- The function and performance of the modified SoftVue device has been evaluated 7.1 through non-clinical design verification and validation testing, consisting of both subsystem and full system evaluation. Testing includes system performance and image
{7}------------------------------------------------
quality tests. The results of the evaluation tests demonstrate that the modified SoftVue device successfully meets the requirements of its intended use.
- 7.2 The SoftVue system has undergone acoustic output testing and was found to be well within industry-accepted limits.
- 7.3 The SoftVue system has been tested to applicable industry safety standards listed in Table 3, below.
| Standard # and Date | Standard Title |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI/IEC 60601-1:2005;<br>Corr. 1:2006; Corr. 2:2007 | Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988;<br>Amendment 1, 1991-11, Amendment 2, 1995 |
| IEC 60601-1-2:2007 | Medical Electrical Equipment - Part 1-2: General Requirements for Safety -<br>Collateral standard: Electromagnetic Compatibility - Requirements and Tests<br>(Edition 2:2001 with Amendment 1:2004) |
| IEC 62304:2006 | Medical device software - Software life cycle processes |
| IEC 60601-2-37:2007 | Medical electrical equipment - Part 2-37: Particular requirements for the basic<br>safety and essential performance of ultrasonic medical diagnostic and<br>monitoring equipment. |
# Table 1: Safety Testing
### 8.0 Summary of Clinical Testing
- 8.1 No clinical testing is required to be performed to assess the changes to the SoftVue System. However, a phantom imaging analysis was conducted to assess performance against established requirements. A description of the phantom testing and results are contained in this submission.
### 9.0 Conclusion
The modified SoftVue 3.0 device performs as intended and is substantially equivalent to the predicate SoftVue 2.0 device with respect to intended use, design, principles of operation, technology, materials, and performance. Any noted differences do not raise any new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.