The Hard Shell Vented Nasal Mask is intended to provide an interface for application of positive airway pressure therapy, such as CPAP or bi-level. The mask is intended for: - Single use, disposable . - . Single patient, multi-use in the home setting - . Multi-patient, multi- use in the hospital or institutional settings - Single patient, short-term use (up to 7 days) in the hospital or institutional settings ● The masks are to be used on adult patients (>66 lbs. />30 kg) for whom CPAP or bi-level therapy has been prescribed.
Device Story
Innova Nasal Vented Mask (also referred to as Hard Shell Vented Nasal Mask) serves as patient interface for CPAP or bi-level positive airway pressure therapy. Device consists of rigid shell, gel cushion, headgear, and exhaust elbow with ports for CO2 washout. Used in home, hospital, or institutional environments; operated by patients or healthcare providers. Device delivers prescribed pressure (4-20 cm H2O) to patient airway. Healthcare providers use device to facilitate respiratory therapy; output is therapeutic pressure delivery. Benefits include effective airway pressure application for adult patients. Design is nearly identical to SleepNet iQ Nasal mask (K102317) with modifications to exhaust elbow and duration-of-use labeling.
Clinical Evidence
Bench testing only. Comparative performance testing included exhaust flow (pressure vs. flow), pressure drop (resistance to flow), internal deadspace, and CO2 washout per ISO 17510-2. Cleaning validation and environmental/mechanical testing (temperature, drop test) were performed. Results demonstrated equivalence to predicate; CO2 washout performance met ISO 17510-2 criteria.
Technological Characteristics
Rigid shell, gel cushion, headgear, and swivel elbow with exhaust ports. Materials identical to K102317 and K090935. Pressure range 4-20 cm H2O. Connectivity: None (mechanical interface). Sterilization: Not specified (cleaning instructions provided for soap/water or OPA).
Indications for Use
Indicated for adult patients (>66 lbs / >30 kg) requiring CPAP or bi-level positive airway pressure therapy. Intended for single-use disposable, single-patient multi-use (home), or multi-patient multi-use (hospital/institutional) settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
# K121321
# 510(k) Summary Page 1 of 6 19-Nov-12
| SleepNet Corporation<br>5 Merrill Industrial Drive<br>Hampton, NH 03842 | Tel - 603-758-6625<br>Fax - 603-758-6699 |
|-------------------------------------------------------------------------|-------------------------------------------------------------------|
| Official Contact: | Jennifer Kennedy – Director of Quality |
| Proprietary or Trade Name: | Innova Nasal Vented Mask |
| Common/Usual Name: | Patient interface |
| Classification Code/Name: | BZD – non-continuous ventilator (IPPB)<br>21CFR 868.5905, Class 2 |
| Predicate Device: | K092835 - Respironics ComfortGel Blue nasal mask |
#### Device Description:
The SleepNet Innova Nasal Vented Mask is similar to other CPAP mask. The Innova Nasal Vented mask is nearly identical to our iQ Nasal mask, K102317, except for the following changes:
- Slightly different indications for use, but the predicate is Respironics ComfortGel Blue . Nasal mask, K092835.
- . A range of durations of use for marketing purposes only
- o single use, disposable
- single patient, multi-use up to 7 days o
- single patient, multi-use in home setting o
- o multi-patent, multi-use in institutional settings
- . material for shell, hard vs. soft
- a new exhaust elbow design .
#### Indications for Use:
The Innova Nasal Vented Mask is intended to provide an interface for application of positive airway pressure therapy, such as CPAP or bi-level.
- . The mask is intended for:
- Single use, disposable .
- Single patient, multi-use in the home setting .
- Multi-patient, multi- use in the hospital or institutional settings. �
- o Single patient, short-term use (up to 7 days) in the hospital or institutional settings
· The masks are to be used on adult patients (>66 lbs. />30 kg) for whom CPAP or bi-level therapy has been prescribed.
Patient Population: For adults (> 66 lbs. / >30 kg)
# NOV 3 0 2012
{1}------------------------------------------------
# 510(k) Summary Page 2 of 6 19-Nov-12
#### Environment of Use: Home or hospital or institutional environments
## Predicate Device Comparison:
The Innova Nasal Vented Mask as compared to the predicate is viewed as substantially equivalent to the predicate device because:
#### Indications -
- The Innova Nasal Vented Mask is intended to provide a patient interface for application . of positive airway pressure therapy with CPAP or bi-level devices.
The masks are to be used on adult patients (>66)bs. />30 kg) for whom positive airway pressure therapy has been prescribed. Identical to Respironics ComfortGel Blue Nasal (K092835)
Discussion - The indications for use are identical to the predicate Respironics ComfortGel Blue Nasal (K092835)
#### Patient Population -
- The masks are to be used on adult patients (>66 lbs. />30 kg) for whom positive airway . pressure therapy has been prescribed. Identical to Respironics ComfortGel Blue Nasal (K092835)
Discussion - The patient population is identical to the predicate Respironics ComfortGel Blue Nasal (K092835).
#### Technology -
- . Technology, method of manufacture and construction of a gel cushion, hard or soft shell and elbow with exhalation holes are identical to the predicate SleepNet iQ Nasal mask -K 102317 and similar to the predicate Respironics ComfortGel Blue Nasal (K092835).
Discussion - The technology, shape, design, configuration of the mask, head strap as well as the manufacturing methods are identical to the predicate, SleepNet iQ Nasal mask (K102317).
The use of multiple ports in the elbow for exhalation and CO2 washout is substantially equivalent as demonstrated by the comparative Pressure vs. Flow curves and the CO- washout testing vs. the predicate Respironics ComfortGel Blue Nasal (K092835) and the differences are not clinically significant and do not affect safety and effectiveness.
#### Environment of Use -
- The masks are intended for use in the home or hospital or institutional environment. . Identical to predicate - Respironics ComfortGel Blue Nasal (K092835)
Discussion - The environments of use are identical to the predicate.
{2}------------------------------------------------
510(k) Summary Page 3 of 6 19-Nov-12
# Environment of Use -
- The masks are intended for use in the home or hospital or institutional environment. . Identical to predicate - Respironics ComfortGel Blue Nasal (K092835)
Discussion - The environments of use are identical to the predicate.
| | Table of Comparison to Predicate | |
|--|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | will which beliefler under first and the below | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| Attributes | Innova Nasal vented mask | Respironics ComfortGel Blue<br>K092835 |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Innova Nasal Vented Mask is<br>intended to provide an interface for<br>application of positive airway<br>pressure therapy, such as CPAP or<br>bi-level.<br><br>The mask is intended for:<br>• Single use, disposable<br>• Single patient, multi-use in the<br>home setting<br>• Multi-patient, multi- use in the<br>hospital or institutional settings<br>• Single patient, short-term use (up to<br>7 days) in the hospital or institutional<br>settings<br><br>The masks are to be used on adult<br>patients (>66 lbs. / >30 kg) for whom<br>CPAP or bi-level therapy has been<br>prescribed. | The ComfortGel Blue Nasal Mask is<br>intended to provide an interface of<br>CPAP or bi-level therapy to patients.<br>The mask is for single patient use in<br>the home or multi-patient use in the<br>hospital / institutional environment.<br>The mask is to be used on adult<br>patients (>66 lbs. / >30 kg) for whom<br>CPAP or bi-level therapy has been<br>prescribed. |
| Patient Population | Adult (>66 lbs. / >30 kg) | Adult (>66 lbs. / >30 kg) |
| Environment of Use | The masks are intended for use in the<br>home or hospital or institutional<br>environment. | The masks are intended for use in the<br>home or hospital/institutional<br>environment. |
| Prescriptive | Yes | Yes |
| Duration of Use | Single use, disposable<br>Single patient, multi-use<br>Multi-patient, multi- use<br>Single patient, short-term use (up to<br>7 days) | Single patient, multi-use<br>Multi-patient, multi-use |
| Cleaning methods | Soap and water<br>OPA | Soap and water |
| Incorporates an Exhaust port<br>elbow | Yes | Yes |
| Delivered Pressure range | $4 - 20 cm H_2O$ | $4 - 30 cm H_2O$ |
| Available sizes | 3 | 4 |
| Shape | Similar | Similar |
| Shell | Rigid | Rigid |
| Materials | Identical to Minime K090935 or<br>iQ Nasal K102317 | |
| Attributes | Innova Nasal vented mask | Respironics ComfortGel Blue<br>K092835 |
| CO2 washout profile | Pressure | ETCO2% at |
| Tested per ISO 17510-2<br>(largest size tested, which is<br>worst case) | | mask (% increase) |
| | 4 cm H2O<br>6.0 (15%) | 4 cm H2O<br>5.6 (6%) |
| | 5 cm H2O<br>5.9 (14%) | 5 cm H2O<br>5.5 (3%) |
| | 10 cm H2O<br>5.8 (11%) | 10 cm H2O<br>5.4 (2%) |
| | Occluded<br>6.7 (30%) | Occluded<br>6.3 (19%) |
| Deadspace | | |
| Small | 82 ml | 97 ml |
| Medium | 100 ml | 99 ml |
| Large | 108 ml | 118 ml |
| Exhaust - pressure / flow | Pressure (cmH2O) Flow (lpm) | Pressure (cmH2O)<br>Flow (lpm) |
| | 4<br>18.2 | 4<br>17.3 |
| | 10<br>30.4 | 10<br>27.0 |
| | 20<br>43.8 | 20<br>38.7 |
| Pressure Drop | 30 lpm - 0.04 cm H2O | 30 lpm - 0.04 cm H2O |
| | 50 lpm - 0.16 cm H2O | 50 lpm - 0.15 cm H2O |
| | 60 lpm - 0.24 cm H2O | 60 lpm - 0.22 cm H2O |
| | 100 lpm - 0.82 cm H2O | 100 lpm - 0.74 cm H2O |
| Components | Headgear | Headgear |
| | Shell / Cushion | Shell / Cushion |
| | Swivel elbow | Swivel elbow |
{3}------------------------------------------------
# 510(k) Summary Page 4 of 6 19-Nov-12
# Non-clinical Testing Summary
#### Comparative Performance -
We have performed comparative performance testing pre- and post-conditioning that included:
# Exhaust Flow (Pressure vs. Flow)
- Standard test method using CPAP unit at various flow rates and measuring the leak at the . patient interface
- No pass / fail criteria, reportable values only. It should be comparable to the predicate. .
- . Discussion - The proposed device has equivalent "Leak or Flow" rate at all pressure vs. the predicate. The differences do not have any clinical significance
# Pressure Drop (resistance to Flow)
- Standard test method of using a standard flow and measuring pressure drop .
- . No pass / fail criteria, reportable values only
- . Discussion - The proposed device is equivalent to the predicate.
{4}------------------------------------------------
# 510(k) Summary Page 5 of 6 19-Nov-12
## Internal Volume / Deadspace
- Standard test method of measuring the volume of the parts near the patient. .
- . No pass / fail criteria, reportable values only. It should be comparable to the predicate.
- Discussion The proposed device is equivalent and CO2 washout and Pressure vs. Flow . are the key performance features for CPAP mask performance. The difference4s are not clinically significant and do not raise any new safety or effectiveness concerns.
#### CO2 washout
- Tested per ISO 17510-2 .
- Pass / fail criteria per ISO 17510-2 allow for changes in CO2 at various pressures to be < . 20% and < 60% from baseline).
- Discussion the performance of the proposed device is well within the pass / fail criteria . as stated in ISO 17510-2.
# Cleaning validation
- Repeated cleaning was performed per the recommended cleaning instructions .
- Visual and performance testing was performed and compared pre and post- cleaning . and found to be similar
- . Discussion - The proposed mask can be cleaned as intended and meet its performance specifications
## Environmental / Mechanical testing
- . Test method included subjecting samples to high and low temperatures and a drop test
- Pass / fail criteria was that they would meet the performance specifications which was . performed after the cleaning durability
- . Discussion - The proposed device met the performance specifications after conditioning and cleaning.
#### Materials -
- The materials in patient contact are identical to our own predicate devices SleepNet iQ . Nasal (K102317) and MiniMe mask (K090935).
- . Discussion - The materials are identical to the cited predicates.
{5}------------------------------------------------
# 510(k) Summary Page 6 of 6 19-Nov-12
# Materials / Patient Coatact / Duration of Use (prolonged duration of use)
·
| Component | Patient Contact | Predicate for Biocompatibility |
|----------------|-------------------------------------------------------------------------|----------------------------------------------------|
| Gel Bladder | Surface contact Skin<br>External communicating,<br>Tissue / Gas Pathway | Identical - Sleepnet MiniMe Pediatric mask K090935 |
| Bladder film | Surface contact Skin<br>External communicating,<br>Tissue / Gas Pathway | Identical - Sleepnet IQ Ventilation mask K102317 |
| Shell | External communicating,<br>Tissue / Gas Pathway | Identical - Sleepnet MiniMe Pediatric mask K090935 |
| Elbow assembly | External communicating,<br>Tissue / Gas Pathway | Identical - Sleepnet MiniMe Pediatric mask K090935 |
| Headgear | Surface contact Skin<br>External communicating,<br>Tissue / Gas Pathway | Identical - Sleepnet IQ Ventilation mask K102317 |
# Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate can be found to be substantially equivalent.
.
.
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
## November 30, 2012
Sleepnet Corporation C/O Mr. Paul Dryden President Promedic. Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
Re: K121321
Trade/Device Name: Hard Shell Vented Nasal Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: November 19, 2012 Received: November 20, 2012
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Kwame Ø. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital,
Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
510(k) Number:
K121321 (To be assigned)
Device Name:
Hard Shell Vented Nasal Mask
Indications for Use:
The Hard Shell Vented Nasal Mask is intended to provide an interface for application of positive airway pressure therapy, such as CPAP or bi-level.
The mask is intended for:
- Single use, disposable .
- . Single patient, multi-use in the home setting
- . Multi-patient, multi- use in the hospital or institutional settings
- Single patient, short-term use (up to 7 days) in the hospital or institutional settings ●
The masks are to be used on adult patients (>66 lbs. />30 kg) for whom CPAP or bi-level therapy has been prescribed.
Prescription Use XX (Part 21 CFR 801 Subpart D)
or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Lester W. Schultheis Jr 2012.11.30 10:45:42 -05'00'
(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.