K243023 · Wellell, Inc. · BZD · May 23, 2025 · Anesthesiology
Device Facts
Record ID
K243023
Device Name
WiZARD 520 Full Face Mask
Applicant
Wellell, Inc.
Product Code
BZD · Anesthesiology
Decision Date
May 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bi-level therapy. The Mask is intended for single-patient reuse in the home and multi-patient multi-use in the hospital environment. The Mask is to be used by adults weighing ≥ 30 kgs whom positive airway pressure (CPAP or bi-level system) has been prescribed.
Device Story
Full face mask interface for CPAP or bi-level positive airway pressure therapy; used in home or hospital settings by adult patients (≥ 30 kg). Device consists of cushion, frame, and 3D-shaped headgear; connects to standard 22 mm breathing tubes. Elbow features integrated exhalation vents to purge CO₂ and maintain continuous airflow. Provides secure facial seal to deliver prescribed pressure (4–40 cmH₂O). Designed for clear line of sight; available in multiple sizes. Healthcare providers or patients use the mask to facilitate respiratory therapy; output is the delivery of pressurized air to the patient's airway. Benefits include effective treatment of Obstructive Sleep Apnea via stable, comfortable interface.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 17510:2015, including CO₂ rebreathing, pressure-flow characteristics, resistance to flow, anti-asphyxia valve operation, and sound levels. Biocompatibility testing performed per ISO 10993 series and ISO 18562 series. Reliability testing included shipping, drop, temperature, and disinfection/cleaning validation.
Technological Characteristics
Full face mask with cushion, frame, and headgear. Connects to 22 mm tubing. Pressure range 4–40 cmH₂O. Materials evaluated per ISO 10993-1 and ISO 18562. Complies with ISO 17510:2015 (mask performance), ISO 5356-1:2015 (conical connectors), and IEC 62366-1:2015 (usability). Non-sterile; suitable for single-patient home use or multi-patient hospital use with disinfection.
Indications for Use
Indicated for adults weighing ≥ 30 kg prescribed CPAP or bi-level positive airway pressure therapy for the treatment of Obstructive Sleep Apnea.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 23, 2025
Wellell Inc.
Chieh Yang
Quality Engineering Manager
No. 9, Min Sheng St.,
Tu-Cheng, New Tapei City, Taiwan
New Taipei City, 236044
Taiwan
Re: K243023
Trade/Device Name: WiZARD 520 Full Face Mask
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous Ventilator (IPPB)
Regulatory Class: Class II
Product Code: BZD
Dated: September 27, 2024
Received: April 21, 2025
Dear Chieh Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243023 - Chieh Yang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243023 - Chieh Yang
Page 3
Sincerely,
Rachana Visaria -S
Rachana Visaria, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243023 | |
| Device Name WiZARD 520 Full Face Mask | |
| Indications for Use (Describe) The Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bi-level therapy. The Mask is intended for single-patient reuse in the home and multi-patient multi-use in the hospital environment. The Mask is to be used by adults weighing ≥ 30 kg whom positive airway pressure (CPAP or bi-level system) has been prescribed. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Wellell Inc.
WiZARD 520 Full Face Mask
K243023
# 510(k) Summary
1. Type of submission: Traditional
2. Date of summary: May 22, 2025
3. Submitter: Wellell Inc.
Address: No. 9, Min Sheng St., Tu-Cheng, New Taipei City, 236044, Taiwan
Phone: +886-2-2268-5568
Fax: +886-2-2268-9662
Contact: Chieh Yang (meow.yang@wellell.com)
Job title: Quality Engineering Manager
4. Identification of the device:
- Proprietary/Trade name: WiZARD 520 Full Face Mask
- Classification product Code: BZD
- Regulation number: 21 CFR 868.5905
- Regulation description: Noncontinuous ventilator
- Review panel: Anesthesiology
- Device class: II
5. Identification of the Predicate Device:
- Predicate device name: WiZARD 310/320 Series CPAP Mask
- Manufacturer: Wellell Inc.
- Classification product code: BZD
- Regulation number: 21 CFR 868.5905
- Device class: II
- 510(k) number: K182394
Page 1 of 8
{5}
Wellell Inc.
WiZARD 520 Full Face Mask
K243023
## 6. Device description
WiZARD 520 Full Face Mask’s cushion and frame provide a secure interface between the mask and the patient's face and the flexible cushion improves seal. A series of vents on the elbow, serve as an exhalation vent to purge the exhaled carbon dioxide from the mask. The mask is designed to maintain a continuous airflow to minimize the amount of CO₂ rebreathed by the patient. The mask features a wider visibility design that ensures a clear line of sight. It also includes a fabric frame equipped with 3D-shaped headgear for enhanced comfort and stability.
WiZARD 520 Full Face Mask is intended for connection to a positive air pressure source such as CPAP or bi-level for the treatment of Obstructive Sleep Apnea.
WiZARD 520 Full Face Mask can be linked to the CPAP or bi-level system's standard 22 mm breathing tube using a swivel hose, flexible tube, and tubing connector. Various sizes (small, medium, and large) are offered to ensure adequate fit over the extended patient population.
## 7. Indication for use
The Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bi-level therapy. The Mask is intended for single-patient reuse in the home and multi-patient multi-use in the hospital environment.
The Mask is to be used by adults weighing ≥ 30kgs whom positive airway pressure (CPAP or bi-level system) has been prescribed.
Page 2 of 8
{6}
Wellell Inc.
WiZARD 520 Full Face Mask
K243023
8. Comparison of technological characteristics with the predicate device
| Characteristic | Predicate Device
WiZARD 320 CPAP Mask | Subject Device
WiZARD 520 Full Face Mask | Comments |
| --- | --- | --- | --- |
| Product Code | BZD | BZD | Equivalent |
| Regulation Number | 868.5905 | 868.5905 | Equivalent |
| Indications for Use | WIZARD 310/320 series CPAP Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bi-level therapy. These masks are intended for single-patient reuse in the home and multi-patient, multi-use in the hospital environment. These masks are to be used on patients greater than 30 kg for whom positive airway pressure (CPAP or bi-level system) has been prescribed. | The Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bi-level therapy. The Mask is intended for single-patient reuse in the home and multi-patient multi-use in the hospital environment.
The Mask is to be used by adults weighing ≥ 30 kgs whom positive airway pressure (CPAP or bi-level system) has been prescribed. | Equivalent |
| Patient Use Type | Adult ≥ 30 kg | Adult ≥ 30 kg | Equivalent |
| Indication | Obstructive sleep apnea | Obstructive sleep apnea | Equivalent |
| Environment | Home, hospital | Home, hospital | Equivalent |
Page 3 of 8
{7}
Wellell Inc.
WiZARD 520 Full Face Mask
K243023
| Characteristic | Predicate Device
WiZARD 320 CPAP Mask | Subject Device
WiZARD 520 Full Face Mask | Comments |
| --- | --- | --- | --- |
| Reprocessing | Single patient reuse or multi-patient reuse | Single patient reuse or multi-patient reuse | Equivalent |
| Patient Support System | CPAP or bi-level system | CPAP or bi-level system | Equivalent |
| Principle of Operation | Mask provides an interface such that airflow from a positive pressure source is directed to the patient's nostril and is held in place with adjustable headgear that straps the mask to the face. | Mask provides an interface such that airflow from a positive pressure source is directed to the patient's nostril and is held in place with adjustable headgear that straps the mask to the face. | Equivalent |
| Use Life | 6 months after unpacking | 6 months after unpacking (or 30 cycles under disinfection) | The subject device claims to incorporate disinfection considerations. |
| Mask Size | L/M/S | L/M/S | Equivalent |
| Mask Dead Space | L: 328 ml
M: 284 ml
S: 223.8 ml | L: 271 ml
M: 228 ml
S: 211 ml | Measured dead space is disclosed in labelling in accordance with ISO 17510:2015. |
| Sterility | Non-sterile | Non-sterile | Equivalent |
| Pressure Range | 4~40 cmH₂O | 4~40 cmH₂O | Equivalent |
| Hose Connection | 22 mm | 22 mm | Equivalent |
| Operation Range | +5°C to +35°C
(+41°F to +95°F)
15% to 95% R.H (non-condensing) | +5°C to +35°C
(+41°F to +95°F)
15% to 95% R.H (non-condensing) | Equivalent |
Page 4 of 8
{8}
Wellell Inc.
WiZARD 520 Full Face Mask
K243023
| Characteristic | Predicate Device
WiZARD 320 CPAP Mask | | Subject Device
WiZARD 520 Full Face Mask | | Comments |
| --- | --- | --- | --- | --- | --- |
| Storage and Transport | -15°C to +60°C
(+5°F to +140°F)
10% to 90% R.H (non-condensing) | | -15°C to +60°C
(+5°F to +140°F)
10% to 90% R.H (non-condensing) | | Equivalent |
| Exhaust Flow | Pressure (cmH₂O) | Flow (l/min) | Pressure (cmH₂O) | Flow (l/min) | Testing and reporting have been conducted in line with ISO17510: 2015. |
| | 4 | 18 | 4 | 18 | |
| | 8 | 26 | 13 | 35 | |
| | 12 | 32 | 22 | 48 | |
| | 16 | 38 | 31 | 59 | |
| | 20 | 45 | 40 | 69 | |
| | 24 | 49 | | | |
| | 28 | 53 | | | |
| | 30 | 57 | | | |
| | 32 | 59 | | | |
| | 36 | 63 | | | |
| | 40 | 66 | | | |
| CO₂ Rebreathing Normal condition | Pressure (cmH₂O) | Relative CO₂ increase | Pressure (cmH₂O) | Relative CO₂ increase | Equivalent
Complies with ISO 17510:2015 CO₂ requirements (<20%) |
| | 4 | 16% | 4 | 6% | |
| | 5 | 14% | 5 | 5% | |
| | 10 | 6% | 10 | 1% | |
| CO₂ Rebreathing Single Fault use | Fault | Relative CO₂ increase | Fault | Relative CO₂ increase | Equivalent
Complies with ISO 17510:2015 CO₂ requirements (<60%) |
| | Fault 1 | 46% | Fault 1 | 41% | |
| | Fault 2 | 50% | Fault 2 | 39% | |
Page 5 of 8
{9}
Wellell Inc.
WiZARD 520 Full Face Mask
K243023
| Characteristic | Predicate Device
WiZARD 320 CPAP Mask | | Subject Device
WiZARD 520 Full Face Mask | | Comments |
| --- | --- | --- | --- | --- | --- |
| Resistance
(Pressure Drop) | Flow (l/min) | Pressure (cmH₂O) | Flow (l/min) | Pressure (cmH₂O) | Equivalent |
| | 50 | 1.0 | 50 | ≤ 1.0 | |
| | 100 | 2.0 | 100 | ≤ 2.0 | |
| Anti-asphyxia
Valve Pressure (cmH₂O) | open-to-atmosphere | 0.54 | open-to-atmosphere | < 1.5 | The AAV pressure are
complied with
ISO17510: 2015. |
| | closed-to-atmosphere | 0.81 | closed-to-atmosphere | < 2.5 | |
| Single-Fault
Inspiration/
Specification
of Expiration
Resistance | Inspiration | 0.9 | Inspiration | ≤ 1.5 | The resistance is
complied with
ISO17510: 2015. |
| | Expiration | 1.0 | Expiration | ≤ 2.2 | |
| Sound | ≤ 30dB | | Sound Pressure Level
≤ 26 dBA
Sound Power Level
≤ 35 dBA | | Measured sound power
level is disclosed in
labelling in accordance
with ISO 17510:2015. |
Page 6 of 8
{10}
Wellell Inc.
WiZARD 520 Full Face Mask
K243023
## 9. Non-clinical testing
The following tests were conducted to evaluate the safety and effectiveness of WiZARD 520 Full Face Mask in accordance with relevant FDA consensus standards including ISO 17510:2015, ISO 10993-1: 2018, ISO 18562-1:2017 and ISO 5356-1:2015.
- Performance test
- CO₂ Rebreathing
- Pressure-Flow characteristics
- Resistance to flow
- Anti-Asphyxia valve operating pressures
- Sound power level test
- Sound pressure level test
- Dead space test
- Biocompatibility Test
- Cytotoxicity Test
- Skin irritation Test
- Skin sensitization Test
- Leachable and extractable Test
- Particulate matter (PM) Test
- Volatile organic compounds (VOC) Test
- Reliability Test
- Shipping test
- Drop test
- Temperature operation test
- Sit test
- Shelf life test report
- Disinfection validation
- Cleaning validation
{11}
Wellell Inc.
WiZARD 520 Full Face Mask
K243023
| Standards | Title |
| --- | --- |
| ISO 17510:2015 | Medical devices -- Sleep apnoea breathing therapy -- Masks and application accessories |
| ISO 10993-1:2018 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process |
| ISO 10993-5:2009 | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity |
| ISO 10993-10:2010 | Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization |
| ISO 10993-11:2017 | Biological evaluation of medical devices - Part 11: Tests for systemic toxicity |
| ISO 10993-17:2023 | Biological evaluation of medical devices - Part 17: Toxicological risk assessment of medical device constituents |
| ISO 10993-18:2020 | Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process. |
| ISO 10993-23:2021 | Biological evaluation of medical devices - Part 23: Tests for irritation |
| ISO 18562-1: 2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process |
| ISO 18562-2: 2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter |
| ISO 18562-3: 2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds |
| IEC 62366-1:2015+AMD1:2020 | Medical devices Part 1: Application of usability engineering to medical devices including Amendment 1 |
| ISO 14971:2019 | Medical devices — Application of risk management to medical devices |
| ISO 5356-1:2015 | Anaesthetic and respiratory equipment — Conical connectors Part 1: Cones and sockets |
## 10. Conclusion
Based on the comparison of characteristics and analysis of non-clinical performance data, we conclude that the WiZARD 520 Full Face Mask is substantially equivalent to the legally marketed predicate device.
Page 8 of 8
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.