The Innova™ Nasal Non-Vented Mask is intended to be used as an accessory to CPAP/bi-level positive pressure systems containing exhalation valves. This product is indicated for use with CPAP/Bi-level systems that have adequate alarins for positive pressure delivery failure. The mask is ingle patient, multi-use in the home, hospital or institutional environment on adult patients (>30kg/66 lbs.).
Device Story
Innova Nasal Non-Vented Mask; patient interface for CPAP/bi-level positive pressure systems. Device provides seal over nose to deliver pressurized air from non-invasive ventilator; requires external exhalation port in circuit for CO2 flushing. Passive device; no active components. Used in home or hospital/institutional environments by adult patients. Healthcare provider selects mask; patient wears mask to receive prescribed therapy. Output is delivery of pressurized air to patient airway; improves patient compliance/therapy efficacy. Device features rigid shell, gel bladder, non-vented elbow, and headgear. Available in two sizes (Small/Medium, Medium/Large).
Clinical Evidence
Bench testing only. Testing included pressure drop, unintentional leak, internal volume/dead space, mechanical drop test, cleaning validation, and shelf-life (ASTM F1980:16). Biocompatibility evaluated per ISO 10993-1:2018, ISO 18562-1:2017, ISO 10993-5:2009, and ISO 10993-10:2010. Performance characteristics (dead space, resistance to flow, leak rates) compared to predicate.
Technological Characteristics
Passive nasal mask; rigid shell, gel bladder, non-vented elbow, headgear. Standard 22mm connection for ventilator circuits. Non-sterile. Biocompatible materials per ISO 10993/ISO 18562. Shelf life 5 years. No software or electronics.
Indications for Use
Indicated for adult patients (>30kg) requiring non-invasive positive airway pressure therapy via CPAP or bi-level systems equipped with exhalation valves and adequate alarms for pressure delivery failure.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 30, 2021
Sleepnet Corporation % Paul Dryden President ProMedic. LLC 131 Bay Point Dr NE St. Petersburg, Florida 33704
Re: K203601
Trade/Device Name: Innova Nasal Non-Vented Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: March 30, 2021 Received: March 31, 2021
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203601
Device Name Innova™ Nasal Non-Vented Mask
#### Indications for Use (Describe)
The Innova™ Nasal Non-Vented Mask is intended to be used as an accessory to CPAP/bi-level positive pressure systems containing exhalation valves. This product is indicated for use with CPAP/Bi-level systems that have adequate alarins for positive pressure delivery failure. The mask is ingle patient, multi-use in the home, hospital or institutional environment on adult patients (>30kg/66 lbs.).
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summarv Page 1 of 7 30-Apr-21
#### Sponsor:
| Sleepnet Corporation | |
|----------------------------|-------------------------------------------------------|
| 5 Merrill Industrial Drive | Tel - 603-758-6625 |
| Hampton, NH 03842 | Fax - 603-758-6699 |
| Sponsor Contact: | Jennifer Kennedy - Director of Regulatory and Quality |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC |
| Proprietary or Trade Name: | Innova™ Nasal Non-Vented mask |
| Common/Usual Name: | Patient interface for CPAP |
| Regulation Number: | 21CFR 868.5905 |
| Regulation Code: | Non-continuous ventilator (IPPB) |
| Product Code: | BZD |
| Regulatory Class: | II |
| Device: | Innova™ Nasal Non-Vented mask |
| Predicate Device: | K102317 - Sleepnet iQ Ventilation Nasal mask |
#### Device Description:
The Sleepnet Innova™ Non-Vented Nasal mask is a patient interface for use with positive pressure equipment.
There are 2 labeled versions:
- Single patient, multi-use (home or hospital/institutional) up to 6 months .
- Single patient, multi-use up to 7 days (hospital/institutional) .
#### Indications for Use:
The Innova™ Nasal Non-Vented Mask is intended to be used as an accessory to CPAP/bi-level positive pressure systems containing exhalation valves. This product is indicated for use with CPAP/BI-LEVEL systems that have adequate alarms for positive pressure delivery failure. The mask is intended for single patient, multi-use in the home, hospital or institutional environment on adult patients (>30kg/66 lbs.).
Patient Population: For adults (>30 kg)
Environment of Use: Home or hospital / institutional environments
#### Substantial Equivalence Discussion:
We discuss the major attributes for demonstrating substantial equivalence below. These refer to the table below this discussion.
The Sleepnet Innova™ Nasal Non-Vented mask is viewed as substantially equivalent to the predicate device because:
{4}------------------------------------------------
#### 510(k) Summary Page 2 of 7 30-Apr-21
#### Indications -
- The masks are to be used on adult patients (>30 kg) for whom positive airway pressure . therapy has been prescribed.
- Similar to the predicate. ●
#### Patient Population -
- . The masks are to be used on adult patients (>30 kg) for whom positive airway pressure therapy has been prescribed.
- Similar to the predicate. ●
#### Environment of Use -
- The masks are intended for use in the home or hospital/institutional environment.
- Similar to the predicate. .
#### Technological Characteristics -
- . Similar technology to the predicate.
#### Non-clinical testing
#### Biocompatibility -
- . With the exception of headgear, the mask materials which are in patient contact are identical to the Sleepnet reference device.
- . Material Certification provided to support biocompatibility evaluation in accordance with ISO 10993-1:2018 and ISO 18562-1:2017
- The material of the headgear has been tested per ISO 10993-5:2009 and ISO 10993-. 10:2010
#### Bench testing
- Pressure drop .
- Unintentional leak .
- Internal Volume / Dead space .
- Mechanical drop test .
- . Cleaning validation
- . Shelf-life via Accelerated aging per ASTM F1980:16
- . Testing following ISO 17510:2015
We have demonstrated that the proposed device is equivalent to the predicate, K102317.
{5}------------------------------------------------
# 510(k) Summary Page 3 of 7
30-Apr-21
K203601
## Table of Comparison to Predicate
| Attributes | Subject | Predicate | Explanation of Differences |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Innova nasal non-vented mask | iQ Ventilation nasal mask | |
| 510(k) | K203601 | K102317 | - |
| Product Classification<br>CFR | BZD<br>CFR 868.5905 | BZD<br>CFR 868.5905 | Similar |
| Indications for Use | The Innova™ Nasal Non-Vented Mask is<br>intended to be used as an accessory to<br>CPAP/bi-level positive pressure systems<br>containing exhalation valves. This product is<br>indicated for use with CPAP/BI-LEVEL<br>systems that have adequate alarms for<br>positive pressure delivery failure. The mask<br>is intended for single patient, multi-use in the<br>home, hospital or institutional environment<br>on adult patients (>30kg/66 lbs.). | The iQ Ventilation Nasal Mask is to be used<br>as an accessory to CPAP/bi-level positive<br>pressure systems that have adequate alarms<br>for positive pressure delivery failure. Use of<br>this product is indicated for use with<br>CPAP/BI-LEVEL POSITIVE PRESSURE<br>SYSTEMS CONTAINING EXHALATION<br>VALVES.<br>The iQ Ventilation Nasal Mask is intended<br>for single patient multi-use in the home<br>environment and multiple patients multi-use<br>in the hospital/institutional environment. | Similar.<br>• Both devices intended for use with<br>positive pressure devices that have<br>exhalation valves.<br>• Both devices intended for use with<br>devices that have adequate alarms and<br>safety systems in case of failure.<br>• The predicate device was cleared for<br>multi-patient use in the<br>hospital/institutional environment. The<br>subject device will be single patient use<br>only. |
| Patient Population | Adult (>30 kg) | Adult (>30 kg) | Similar |
| Patient type | Patients who are appropriate candidates for<br>non-invasive ventilation | Patients who are appropriate candidates for<br>non-invasive ventilation | Similar |
| Prescriptive | Yes | Yes | Similar |
| Principle of Operation | Provides a seal over the nose to allow for<br>delivery of pressurized air from a non-invasive ventilator. Mask is used with<br>circuits that include an exhalation port for<br>flushing out exhaled CO2. | Provides a seal over the nose to allow for<br>delivery of pressurized air from a non-invasive ventilator. Mask is used with<br>circuits that include an exhalation port for<br>flushing out exhaled CO2. | Similar |
| | The device is passive until connected to the<br>ventilator. | The device is passive until connected to the<br>ventilator. | |
| Attributes | Subject | Predicate | Explanation of Differences |
| | Innova nasal non-vented mask | iQ Ventilation nasal mask | |
| Therapy Pressure Range | 4 cm H2O - 20 cm H2O.<br>Typically determined by the equipment to<br>which it is attached. | Greater than 3 cm H2O.<br>Typically determined by the equipment to<br>which it is attached. | Similar<br>This rating is dependent on the equipment<br>to which the device is attached. The<br>subject device has a minimum pressure of<br>4 cm H2O. Most new positive pressure<br>devices have a minimum pressure setting<br>of 4 cm H2O. |
| Anatomical site | Face (seals around nose) | Face (seals around nose) | Similar |
| User Interface to administer<br>therapy | Masks have a standard 22mm connection<br>that connects to 22mm ventilator circuits. | Mask incorporates a tapered tubing, which<br>has a standard 22mm connection that<br>connects to 22mm ventilator circuits. | Similar |
| Contraindications | None | None | Similar |
| Environment of Use | The masks are intended for use in the home<br>or hospital/institutional environment. | The masks are intended for use in the home<br>or hospital/institutional environment. | Similar |
| Use life | Innova nasal non-vented mask<br>Home use -<br>Single patient, multi-use up to 6 months<br>Innova hospital nasal non-vented mask<br>Single patient, multi-use up to 6 months<br>Disposable, Single patient use up to 7 days | Single patient use - 6 months<br>Multi patient use - 6 months | Similar.<br>· Removed multi-patient use for the<br>subject device.<br>· Added Innova Hospital nasal non-vented<br>mask for short-term single patient use<br>(up to 7 days) and single patient. multi-<br>use for up to 6 months in the<br>hospital/institutional environment. |
| Shelf life | 5 years | No shelf life stated | Similar<br>There is no shelf life for both subject and<br>predicate. 5 year shelf life has been<br>validated for the subject device |
| Non-sterile | Yes | Yes | Identical |
| Attributes | Subject | Predicate | Explanation of Differences |
| | Innova nasal non-vented mask | iQ Ventilation nasal mask | |
| Cleaning methods | Innova nasal non-vented mask -<br>• Mild Soap (such as Ivory) and water<br><br>Innova hospital nasal non-vented mask -<br>• Mild Soap (such as Ivory) and water<br>• 70% Isopropyl alcohol | Mild Soap/dish detergent and water | Similar<br>70% isopropyl alcohol has been added for<br>the short term hospital use mask. This<br>method has been validated. This method is<br>similar to Sleepnet Veraseal 3 full face<br>non-vented mask - K190533. |
| Features | | | |
| Available sizes | 2 - Small/Medium & Medium/Large | 1 | Similar<br>The Medium/Large size is similar to the<br>predicate device. The Small/Medium size<br>is smaller. The dead space of both sizes of<br>the subject device is lower than the<br>predicate device but does not raise new<br>safety concerns. |
| Shape | Similar | Similar | Similar |
| Incorporates an Exhaust<br>elbow | No | No | Similar |
| Components of the mask | • Mask shell<br>• Gel bladder<br>• Non-vented elbow<br>• Headgear | • Mask shell<br>• Gel bladder<br>• Non-vented elbow<br>• Headgear<br>• Tubing | Similar |
| Shell design | Rigid | Soft | Similar<br>Shell geometry is similar for both devices.<br>Soft shell incorporated in the predicate<br>device for comfort feature, but the<br>functionality is the same. Rigid shell in the<br>subject device uses the same material as<br>the Sleepnet V3 full face vented mask -<br>K190254, which has a similar environment<br>and duration of use. |
{6}------------------------------------------------
K203601
# 510(k) Summary
{7}------------------------------------------------
K203601
# 510(k) Summary Page 5 of 7
30-Apr-21
{8}------------------------------------------------
### K203601
# 510(k) Summary Page 6 of 7
30-Apr-21
| Attributes | | Subject<br>Predicate | | Explanation of Differences | |
|--------------------------|----------------------------------|-------------------------------------------|----------------------------------|-------------------------------------------|---------------------------------------------|
| | | Innova nasal non-vented mask | iQ Ventilation nasal mask | | |
| Patient Contact per | Surface contact with intact skin | | Surface contact with intact skin | | Similar safety profile |
| ISO 10993-1 | | Externally Communicating with tissue (gas | | Externally Communicating with tissue (gas | |
| | pathway) | | pathway) | | |
| | Permanent contact | | Permanent contact | | |
| | Performance Characteristics | | | | |
| Dead space (ml) | Small/Medium size – 78 ml | | 120 ml | | The dead space of both sizes of the subject |
| | Medium/Large size - 100 ml | | | | device is lower than the predicate device. |
| Resistance to flow | (a) 30 lpm - 0.10 cm H2O | | (@ 30 lpm - 0.31 cm H2O | | The subject device has a lower pressure |
| (pressure drop) (cm H2O) | (@50 lpm - 0.33 cm H2O | | @ 60 lpm - 1.28 cm H2O | | drop compared to the predicate. There are |
| | (a) 60 lpm - 0.46 cm H2O | | | | no performance requirements for pressure |
| | (@100 lpm - 1.42 cm H2O | | | | drop. The pressure drop for a device is |
| | | | | | disclosed in labeling consistent with the |
| | | | | | ISO 17510:2015 requirements. |
| Unintentional leak (lpm) | Pressure | Unintentional leak (Ipm) | Pressure | Unintentional leak (Ipm) | The subject device has a higher |
| | (cm H2O) | | (cm H2O) | | unintentional leak than the predicate. |
| | 5 | 4.86 | 5 | 3.2 | Positive pressure ventilators have leak |
| | 10 | 9.20 | 10 | 5.6 | compensation to account for this |
| | ા ર | 11.43 | ા ર | 9.0 | unintentional leak. Leak compensation |
| | 20 | 14.70 | 20 | 10.8 | found on these devices is 50 lpm or |
| | 25 | 17.53 | 25 | 12.5 | greater. So long as the unintentional leak |
| | 30 | 20.33 | 30 | 14.3 | of the mask is lower than 50 lpm, this leak |
| | | | | | does not cause any loss of therapy. The |
| | | | | | maximum leak is below this threshold. |
{9}------------------------------------------------
#### 510(k) Summary Page 7 of 7 30-Apr-21
#### Substantial Equivalence Conclusion:
The Innova™ Nasal Non-Vented mask has similar indications, technological characteristics and principles of operation and performance to the predicate and performance testing demonstrates that the proposed device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.