The NeoMed Gastrointestinal Tube and Accessories is intended for short-term use < 30 days in neonatal and pediatric patients to provide nutrition via nasal or oral gastric placement.
Device Story
NeoMed Gastrointestinal Tube and Accessories is a PVC single-lumen catheter for delivering liquid nutritional media to neonatal and pediatric patients. The device features a one-piece molded enteral-only connector hub with an integral tethered closure plug and a closed tip with side holes for flow and aspiration. An orange radiopaque barium sulfate stripe is embedded in the tubing wall for X-ray visualization of tip placement. The device is used in clinical settings for short-term gastric access. It is designed to connect with NeoMed extension tubing and oral syringes. By providing a secure, enteral-only connection, the device facilitates safe nutritional delivery, reducing the risk of misconnection and ensuring accurate placement via radiographic confirmation.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verification included biocompatibility testing per ISO 10993-1, tensile strength, dimensional analysis, non-interconnectability, and priming volume measurements.
Technological Characteristics
PVC single-lumen catheter; barium sulfate radiopaque stripe; one-piece molded enteral-only connector hub; tethered closure plug; closed tip with side holes. Biocompatibility per ISO 10993-1. Non-powered, mechanical device.
Indications for Use
Indicated for neonatal and pediatric patients requiring short-term (< 30 days) nutritional delivery via nasal or oral gastric placement.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
K120/82
Page 1 of 3
APR 1 9 2012
# NeoMe
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
DATE:
April 11, 2012
APPLICANT:
NeoMed, Inc. 100 Londonderry Ct Suite 112 Woodstock, GA 30188 Tony Lair, President Tel: 770-516-2225 email: lair1@concentric.net
Official Correspondent: Penny Northcutt, RAC, CQA Regulatory Consultant for NeoMed, Inc. REGSolutions, LLC Tel: 678-428-6978 Fax: 678-513-0937 email: pennynorthcutt@theregsolutions.com
Trade Name:
NeoMed Gastrointestinal Tube and Accessories
Classification Tubes, Gastrointestinal (and accessories) Name:
| DEVICE | Class II per 21 CFR §876.5980 |
|--------------------|-------------------------------|
| CLASSIFICATION AND | Gastroenterology/Urology |
| PRODUCT CODE: | Product Code: KNT |
| PREDICATE DEVICE | Vygon Nutrisafe 2, KO60944 |
|------------------|-----------------------------------------------------|
| NAME: | Kendall Argyle PVC Feeding Tube, K820441 (Sherwood) |
{1}------------------------------------------------
K12 0182
Page 2 of 3
### SUBSTANTIAL EQUIVALENCE:
The NeoMed Gastrointestinal Tube and Accessories is substantially equivalent to the legally marketed Vygon Nutrisafe 2 (K060944) and Kendall Argyle PVC Feeding Tube (K820441). In addition, the NeoMed PVC feeding tube device design is the same as previously cleared NeoMed Silicone (K072881) and Polyurethane (K082238) Feeding Tubes.
### TECNOLOGICAL CHARACTERISTICS:
The NeoMed Gastrointestinal Tube and Accessories have similar technological characteristics when compared to the predicate devices. They have similar indications for use, materials, product design, and principles of operation. Bench testing has demonstrated that the NeoMed Gastrointestinal Tube and Accessories is functionally equivalent to the proposed predicate feeding tubes. Any minor differences do not affect safety or effectiveness of the NeoMed Gastrointestinal Tube and Accessories.
#### DESCRIPTION OF THE DEVICE:
The NeoMed Gastrointestinal Tube and Accessories is a PVC single lumen catheter that is used to deliver liquid nutritional media. This NeoMed PVC feeding tube is designed to connect with other NeoMed accessories such as the NeoMed Extension tubing under FDA clearance K100288 and NeoMed oral syringe under FDA clearance K092908.
The device consists of the following main components: a feeding tube single lumen catheter and a one-piece molded enteral only connector hub with integral tethered connection closure plug. The catheter tubing has an orange radiopaque stripe of barium sulfate embedded in the tubing wall which can be visualized on x-ray, for exact placement of the tip. The tip is closed. The single lumen catheter tubing has side holes for better flow and to provide multiple openings for aspiration.
#### INTENDED USE/INDICATIONS FOR USE:
The NeoMed Gastrointestinal Tube and Accessories is intended for short-term use < 30 days in neonatal and pediatric patients to provide nutrition via nasal or oral gastric placement.
### PERFORMANCE DATA:
The NeoMed Gastrointestinal Tube and Accessories materials have a long history of use in feeding tube manufacturing and are biocompatible and have been evaluated in accordance with ISO 10993-1: Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing and these tests pass.
Design verification functional test results including tensile strength, dimensions, noninterconnectability, and priming volume demonstrate that the NeoMed Gastrointestinal Tube and Accessories performs its intended use and is substantially equivalent to the predicate devices by Kendall and Vygon.
{2}------------------------------------------------
K120182
## CONCLUSION:
Based on the performance testing, it can be concluded that the NeoMed Gastrointestinal Tube and Accessories is equivalent to the Vygon Nutrisafe 2 (KQ60944) and Kendall Argyle PVC Feeding Tube (K820441) predicts with respect to intended use, materials, design, and technological characteristics.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ly
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, overlaid on a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NeoMed, Inc. % Ms. Penny Northcutt President/CEO REGSolutions, LLC 717 Lakeglen Drive SUWANNEE GA 30024
APR 1 9 2012
Re: K120182
Trade/Device Name: NeoMed Gastrointestinal Tube and Accessories Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: January 19, 2012 Received: January 20, 2012
Dear Ms. Northcutt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2-
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please 1 note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin R. Fisher, Ph.D.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: NeoMed Gastrointestinal Tube and Accessories
Indications For Use:
This product is intended for short-term use (< 30 days) in neonatal and pediatric patients to provide nutrition via nasal or oral gastric placement.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Huld Reimer
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K120182
Page 1 of _ 1_____________________________________________________________________________________________________________________________________________________________________________
Traditional 510(k) NeoMed, Inc. 1
NeoMed Gastrointestinal Tube and Accessories
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.