For nasogastric/oralgastric enteral feeding, incorporating safety connectors which eliminates the risk of IV administration through the feeding tube.
Device Story
Nutrisafe 2 is a system of enteral feeding tubes, syringes, and accessories designed to prevent accidental IV administration. The device features a unique, non-luer locking connection mechanism that prevents involuntary disconnection and eliminates the risk of misconnection to IV systems. The system is used for nasogastric or oralgastric feeding in clinical settings. It does not require specialized training as it maintains standard end-user techniques. The locking hub is secured by screwing and released by unscrewing. The device benefits patients by mitigating risks associated with medication/nutrition delivery errors.
Clinical Evidence
Bench testing only. Biocompatibility testing confirmed materials are non-irritant and non-toxic. Performance testing verified that design changes do not adversely affect safety or efficacy. Risk assessment performed per ISO 14971.
Technological Characteristics
Materials: DEHP-free PVC, Polyurethane, and Silicone. Connection: Unique non-luer locking hub. Form factor: Feeding tubes, syringes, and accessories. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for patients requiring nasogastric or oralgastric enteral feeding.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
## 510K Summary
SEP 1 5 2006
### Date of Preparation: March 28, 2006
| Applicant: | Vygon Corporation<br>2495 General Armistead Ave.<br>Norristown, PA 19403 |
|------------------------|--------------------------------------------------------------------------|
| Contact Individual: | Courtney Smith, Regulatory Affairs Manager<br>610-539-9300 Ext. 110 |
| Trade Name: | Nutrisafe 2 |
| Common Name: | Feeding tube |
| Regulation Number: | 876.5980 |
| Product Code: | FPD |
| Classification Name: | Gastrointestinal tube and accessories devices |
| Classification: | Class II |
| Predicate Device Name: | Nutrisafe (K991918/ K003311), Vygon Enteral Nutrition Tube (K820176) |
The Nutrisafe 2 is a complete system of feeding tubes and accessories. Device Description: The feeding tubes are available in several configurations and in three materials: DEHP-free PVC, Polyurethane, and Silicone. A recent report in the FDA Patient Safety News: Show #46, December 2005, discussed the risks associated with health care personnel mistakenly connecting the wrong devices to the wrong luer connection. Often, this kind of error can result in fatal results. One of the benefits of the Nutrisafe 2 is that the connection does not incorporate a luer, and thus eliminating the risk of inadvertently connecting the system to an IV system. The Nutrisafe 2 feeding tube features a locking connection which eliminates the risk of involuntary disconnection; voluntary disconnection is achieved by simply unscrewing the hub connections. Another benefit of the Nutrisafe 2 is that it does not change the technique of the end-user, and therefore it does not require any special training.
{1}------------------------------------------------
K060944
page 2 of 2
The Nutrisafe 2 consists of the following:
| Description | Reference |
|-------------------------------------|------------------------------------------------------------------------------|
| Feeding Tubes | 361.xxx, 362.xxx, 363.xxx, 1361.xxx, 1362.xxx,<br>1363.xxx, 2332.xxx2395.xxx |
| Syringes and Syringe<br>Accessories | 1015.xxx, 818.xxx, 828.xxx |
| Accessories | 368.xxx, 5802.xx |
| Sampling Devices | 821.xxx, 817.xxx |
For nasogastric/oralgastric enteral feeding, incorporating safety connectors Intended Use: which eliminates the risk of IV administration through the feeding tube
Technology Characteristics: The Nutrisafe 2 feeding tubes are available in several configurations and in three materials: DEHP-free PVC, Polyurethane, and Silicone. Additional the feeding system contains a unique connection does not incorporate a luer. The locking connection eliminates the risk of involuntary disconnection; voluntary disconnection is achieved by simply unscrewing the hub connections.
### Summary of Design Control Activities:
Biocompatibility testing of the material demonstrates that it is non-irritant and non-toxic. Performance testing demonstrates that the changes do not affect safety or efficacy. Risk Assessment was conducted in compliance with ISO 14971.
Conclusion:
The only changes between the predicate device (Nutrisafe K003311 / K991918) and the Nutrisafe 2 is the use of DEHP-free PVC and the unique non-luer connection. The only difference between the predicate device (Enteral Nutrition Tube K820176) is the additional sizes and the unique non-luer connection. Biocompatibility testing, performance testing and risk assessment demonstrate that the Nutrisafe 2 meets manufacturers specifications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Ms. Courtney Smith Regulatory Affairs Manager Vygon Corporation 2495 General Armistead Avenue NORRISTOWN PA 19403
SEP 1 6 2006
Re: K060944
Trade/Device Name: Nutrisafe 2 Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Dated: September 8, 2006 Received: September 8, 2006
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows the FDA Centennial logo, which includes the years 1906-2006. Below the logo is the text "Protecting and Promoting Public Health". The logo is circular in shape and contains the letters FDA in the center. The text is written in a decorative font.
{3}------------------------------------------------
Page 2 –
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
1f you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/edrh/industry/support/index.html.
Sincerely yours.
Nancy Chogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K060944
Device Name:
Nutrisafe 2
Indications For Use:
For nasogastric/oralgastric enteral feeding, incorporating safety connectors which eliminates the risk of IV administration through the feeding tube.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rat. A. Pearson
(Division Sign Off)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page 1 of _ 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.