K113144 · Philips Medical Systems · DPS · Apr 3, 2012 · Cardiovascular
Device Facts
Record ID
K113144
Device Name
PAGEWRITER TC 20, 30, 50, 70
Applicant
Philips Medical Systems
Product Code
DPS · Cardiovascular
Decision Date
Apr 3, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
To acquire multi-channel ECG signals from adult and pediatric patients from body surface ECG electrodes and to record, display, analyze and store these ECG signals for review by the user. To be used in healthcare facilities by trained healthcare professionals. Analysis of the ECG signals is accomplished with algorithms that provide measurements, data presentations, graphical presentations and interpretations for review by the user. The interpreted ECG with measurements and interpretive statements is offered to the clinician on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the patient, the results of the physical examination, the ECG tracings, and other clinical findings. A qualified physician is asked to over read and validate (or change) the computer generated ECG interpretation.
Device Story
The PageWriter TC20 is a diagnostic electrocardiograph used in healthcare facilities by trained professionals. It acquires multi-channel ECG signals via body surface electrodes; samples at 8000 Hz; performs 12-bit or 16-bit A/D conversion. The device processes signals using internal algorithms to provide measurements, graphical presentations, and interpretive statements. Output is displayed on a 6.5-inch color touch screen or printed via a high-resolution thermal-array printer. Data can be stored internally or on USB media and transmitted via LAN/wireless connectivity. The device assists clinicians by providing advisory interpretations; a qualified physician must review and validate all computer-generated findings. The TC20 integrates the ECG circuit directly onto the main board, differing from the predicate TC30 which uses a separate Patient Interface Module (PIM), though both maintain identical isolation and signal processing principles.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing against FDA-recognized standards (AAMI-EC11, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-25). The 12-lead ECG interpretation algorithm was previously cleared in prior submissions.
Indicated for adult and pediatric patients requiring ECG evaluation for diagnosis, treatment planning, treatment effectiveness monitoring, or symptom investigation. For use by trained healthcare professionals in healthcare facilities.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
## 510(k) Summary
## abbreviated 510(k) Premarket Notification
For
Philips Medical Systems
Image /page/0/Picture/6 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The text is black and appears to be slightly pixelated or distressed, giving it a textured look. The background is white, providing a high contrast that makes the word easily readable despite the pixelation.
Philips Electrocardiograph
PageWriter TC20, 30, 50, 70
1. Submitter's Name, Address, Telephone Number:
Name: Philips Medical Systems
Address: 3000 Minuteman Road
Andover, MA 01810, U.S.A.
Phone Number: (978) 659-2404
### Contact Person:
Paul Schrader Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 Tel. 978-659-2404 Fax 978-659-3610 Paul.Schrader@philips.com
## The Date the Summary was Prepared:
18 October, 2011
{1}------------------------------------------------
### 2. The Name of the Device:
PageWriter TC20/860332 Electrocardiograph
- Classification Name: Electrocardiograph.
CFR: \$870.2340
Product Code: DPS
II Class:
Panel: Cardiovascular
## 3. The identification of the Legally Marketed Device (s) to Which the Submitter Claims Equivalence:
PageWriter TC30/860306
### 4. Description of the Device that is the Subject of the 510(k)
The TC series of cardiographs all provide the same basic functionality and utilize the same software code to provide different features. For example the TC 70 cardiograph has a digital patient interface module, the largest display, has the complete set of clinical features and the latest ECG data management capabilities.
The TC 20 cardiograph is the newest addition to the TC family of cardiographs. It provides the following basic feature set for diagnostic cardiographs: 12 lead ECG acquisition and analysis, rhythm printing of ECG, 12 lead ECG report, battery and line operated power, ECG storage and LAN / wireless connectivity.
A detailed comparison with the existing TC 30 cardiograph (predicate device) can be found in section 6 of this 510(K) summary.
| Item | Part number |
|-----------------------------------------|--------------|
| PW6L CBL 10-Lead Patient Cable IEC | 989803175891 |
| PW6L CBL 10-Lead Patient Cable AHA | 989803175901 |
| PW6L CBL Long 10-Lead Patient Cable IEC | 989803175911 |
| PW6L CBL Long 10-Lead Patient Cable AHA | 989803175921 |
### New accessories for analog front end:
{2}------------------------------------------------
## 5. Statement of the Indications for Use of the Device That is the Subject of the 510(k):
To acquire multi-channel ECG signals from adult and pediatric patients from body surface ECG electrodes and to record, display, analyze and store these ECG signals for review by the user. To be used in healthcare facilities by trained healthcare professionals. Analysis of the ECG signals is accomplished with algorithms that provide measurements, data presentations, graphical presentations and interpretations for review by the user.
The interpreted ECG with measurements and interpretive statements is offered to the clinician on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the patient, the results of the physical examination, the ECG tracings, and other clinical findings. A qualified physician is asked to over read and validate (or change) the computer generated ECG interpretation.
## 6. Summary of the Technological characteristics of the New Device in Comparison to Those of the Predicate Device
PageWriter TC20 has the same technological characteristics as the predicate device PageWriter TC30.
Following table provides the Technological characteristics of the New Device in Comparison to Those of the Predicate Device.
| | Philips | Philips | |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Feature/<br>Model | PageWriter TC20<br>(New Device) | PageWriter TC30<br>(Predicate)<br>K080999 | Comparison |
| | Philips | Philips | |
| Feature/<br>Model | PageWriter TC20<br>(New Device) | PageWriter TC30<br>(Predicate)<br>K080999 | Comparison |
| Intended use | To acquire multi-channel ECG<br>signals from adult and pediatric<br>patients from body surface ECG<br>electrodes and to record, display,<br>analyze and store these ECG<br>signals for review by the user. To<br>be used in healthcare facilities by<br>trained healthcare professionals.<br>Analysis of the ECG signals is<br>accomplished with algorithms<br>that provide measurements, data<br>presentations, graphical<br>presentations and interpretations<br>for review by the user. | To acquire multi-channel ECG<br>signals from adult and pediatric<br>patients from body surface ECG<br>electrodes and to record, display,<br>analyze and store these ECG<br>signals for review by the user. To<br>be used in healthcare facilities by<br>trained healthcare professionals.<br>Analysis of the ECG signals is<br>accomplished with algorithms that<br>provide measurements, data<br>presentations, graphical<br>presentations and interpretations<br>for review by the user. | Same |
| | The interpreted ECG with<br>measurements and interpretive<br>statements is offered to the<br>clinician on an advisory basis<br>only. It is to be used in<br>conjunction with the clinician's<br>knowledge of the patient, the<br>results of the physical<br>examination, the ECG tracings,<br>and other clinical findings. A<br>qualified physician is asked to<br>over read and validate (or change)<br>the computer generated ECG<br>interpretation. | The interpreted ECG with<br>measurements and interpretive<br>statements is offered to the<br>clinician on an advisory basis only.<br>It is to be used in conjunction with<br>the clinician's knowledge of the<br>patient, the results of the physical<br>examination, the ECG tracings,<br>and other clinical findings. A<br>qualified physician is asked to over<br>read and validate (or change) the<br>computer generated ECG<br>interpretation. | |
| Indications of<br>use | Where the clinician decides to<br>evaluate the electrocardiogram of<br>adult and pediatric patients as part<br>of decisions regarding possible<br>diagnosis, potential treatment,<br>effectiveness of treatment or to<br>rule-out causes for symptoms. | Where the clinician decides to<br>evaluate the electrocardiogram of<br>adult and pediatric patients as part<br>of decisions regarding possible<br>diagnosis, potential treatment,<br>effectiveness of treatment or to<br>rule-out causes for symptoms. | Same |
| | Philips | Philips | |
| Feature/<br>Model | PageWriter TC20<br>(New Device) | PageWriter TC30<br>(Predicate)<br>K080999 | Comparison |
| Power | AC or battery | AC or battery | Same |
| Battery<br>Chemistry | Lithium Ion battery | Lithium Ion battery (identical<br>battery) | Same |
| ECG<br>Acquisition | AUTO (12 leads) RHYTHM (up to 12 leads) DISCLOSE (1 to 12 leads) | AUTO (12 leads) RHYTHM (up to 12 leads) DISCLOSE (1 to 12 leads) | same |
| Keyboard | ¾-size qwerty keyboard | ¾-size qwerty keyboard | same |
| Touch screen<br>display | 640 x 480 pixel resolution 13.3 cm x 9.9 cm (6.5-<br>inch diagonal) color LCD<br>with Touch Screen | 640 x 480 pixel resolution 13.3 cm x 9.9 cm (6.5-inch<br>diagonal) color LCD with<br>Touch Screen | same |
| Raw data<br>acquisition | 8000 samples per seconds<br>on individual leads for 12<br>lead ECG 8000 samples per seconds<br>on individual leads for<br>SAECG | 8000 samples per seconds<br>on individual leads for 12<br>lead ECG 8000 samples per seconds<br>on individual leads for<br>SAECG | same |
| Sampling rate<br>on cardiograph | 500, 1000 and 2000<br>samples per second per<br>electrode/lead. 12 bit and 16 bit A/D<br>conversion provides 5µV,<br>2.5µV and 1µV<br>resolution. | 500, 1000 and 2000<br>samples per second per<br>electrode/lead. 12 bit and 16 bit A/D<br>conversion provides 5µV,<br>2.5µV and 1µV resolution. | same |
| Filters | AC noise Baseline wander Artifact | AC noise Baseline wander Artifact | same |
| Printer | | | |
| | Philips | Philips | |
| Feature/<br>Model | PageWriter TC20<br>(New Device) | PageWriter TC30<br>(Predicate)<br>K080999 | Comparison |
| Printer<br>resolution | High-resolution, digital-<br>array printer using<br>thermal-sensitive paper200 dpi (voltage axis) by<br>500 dpi (time axis) | High-resolution, digital-<br>array printer using thermal-<br>sensitive paper200 dpi (voltage axis) by500 dpi (time axis) | same |
| Battery<br>Operation | | | |
| Capacity | Typically 50 ECGs and<br>copies on a single charge<br>or 40 minutes of<br>continuous rhythm<br>recordingFully charged 1 battery is<br>to last up to 5 hours under<br>normal usage | Typically 50 ECGs and<br>copies on a single charge<br>or 40 minutes of<br>continuous rhythm<br>recordingFully charged 1 battery is<br>to last up to 5 hours under<br>normal usage | same |
| Networking | | | |
| Networking<br>connection | 10/100 Base-T IEEE<br>802.3 Ethernet via RJ45<br>connector (standard)SDIO wireless LAN<br>connection supporting<br>802.11 b/g standards | 10/100 Base-T IEEE<br>802.3 Ethernet via RJ45<br>connector (standard)Compact Flash card<br>wireless LAN connection<br>supporting 802.11 a/b/g<br>standards | Substantially<br>Equivalent |
| ECG storage | XML File Format<br>(Schema 1.04)Up to 200 ECGs to<br>internal flash memoryUp to 200 ECGs per USB<br>Memory Stick | XML File Format (Schema<br>1.04)Up to 200 ECGs to internal<br>flash memoryUp to 200 ECGs per USB<br>Memory Stick | same |
| Orders | Receive Orders from<br>TraceMaster via the<br>NetworkUp to 200 Orders stored<br>in internal database | Receive Orders from<br>TraceMaster via the<br>NetworkUp to 200 Orders stored<br>in internal database | same |
| ECG file<br>formats | XML (Schema 1.03 and<br>1.04, 1.04.01, 1.04.02) | XML (Schema 1.03 and<br>1.04. 1.04.01, 1.04.02) | same |
| ECG leads<br>and<br>interpretation | | | |
| ECG<br>acquisition | 12-Lead ECG<br>acquisition | 12-Lead ECG acquisition | same |
| Feature/<br>Model | Philips<br>PageWriter TC20<br>(New Device) | Philips<br>PageWriter TC30<br>(Predicate)<br>K080999 | Comparison |
| Algorithm | • 12-Lead ECG Algorithm | • 12-Lead ECG Algorithm | same |
| Patient<br>information<br>module (PIM) | No | Yes | See analysis<br>on difference |
| Patient cable | yes | No | See analysis<br>on difference |
{3}------------------------------------------------
{4}------------------------------------------------
.
{5}------------------------------------------------
.
## 510(K) Summary for Philips PageWriter TC20, 30, 50, 70 Cardiograph
{6}------------------------------------------------
#### Analysis of difference:
For TC20, the ECG circuit is on the main PCA board of the cardiograph; for TC30, the ECG circuit is on the PIM. The TC20 ECG circuit and TC30 PIM have similar basic functions which are to collect the analog signals from the electrode-lead wires, convert the signals to digital signals and send the digital signals to FPGA/processor for processing. The TC30 PIM and TC20 ECG circuit provide isolation between the mains and the patient circuit using the same scientific principles and technical design.
The above table and the analysis demonstrate that the ECG circuit of the TC20 is substantially equivalent with PIM of TC30, from both effectiveness and safety point of view.
Using the FDA "510(k) SUBSTANTIAL EQUIVALENCE DECISIO-MAKING PROCESS", we conclude that the new device, Philips PageWriter TC20 w/ A.05 SW, is substantially equivalent with the predicate, Philips PageWriter TC30.
{7}------------------------------------------------
### 7. Brief Discussion of the Nonclinical Tests
Tests were conducted to Philips PageWriter TC20 cardiograph and the product meet specific acceptance criteria of the following FDA recognized standards:
| Standard number | Description of Standard |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI-EC11:1991 | Diagnostic Electrocardiographic Device |
| IEC 60601-1: 1990+A1 1991+A2 1995 | Medical electrical equipment -<br>Part 1: General requirements for safety, including<br>Amendment 1 (1991) and Amendment 2 (1995). |
| IEC 60601-1-2: 2001+A1 2004 | Medical Electrical Equipment -<br>Part 1: General Requirements for Safety;<br>Electromagnetic Compatibility Requirements and<br>Tests including Amendment 1: 2004 |
| IEC 60601-2-25: 1993+A1:1999 | Medical Electrical Equipment-<br>Part 2: Particular Requirements for Safety and<br>Electrocardiographs |
### 8. Clinical Test
The PageWriter TC20 cardiograph is not required to do clinical test for determination of substantial equivalence. The ECG 12 lead algorithm used for interpretive statements was cleared in a previous 510(K) submittal.
### 9. Conclusion that PageWriter TC20 is as safe and effective as TC30
The PageWriter TC20 cardiograph has been tested to FDA recognized standards and it complies with the standards. We conclude that TC20 is as safe, as effective, and performs at least as safely and effectively as Page Writer TC30 cardiograph.
{8}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR - 3 2012
Philips Medical Systems c/o Ms. Dawn Tibodeau Senior Reviewer TUV SUD America, Inc. 1775 Old Highway 8 NW New Brighton, MN 55112
Re: K113144 Trade/Device Name: Phillips Electrocardiograph, Page Writer TC20 Regulatory Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: II (two) Product Code: 74 DPS Dated: March 15, 2012 Received: March 16, 2012
Dear Ms. Tibodeau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{9}------------------------------------------------
Page 2 - Ms. Dawn Tibodeau
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# 113144
## Indications for Use Statement
#### 510(k) Number: K113144
## Device Name: Philips Electrocardiograph, PageWriter TC20 (860332)
To acquire multi-channel ECG signals from adult and pediatric patients from body surface ECG 10 acquire multi-chamer DCG signals in and store these ECG signals for review by the user. To electrodes and to record, unsplay, and beathere professionals. Analysis of the ECG signals be used in healthoute tacmises of hat provide measurements, data presentations, graphical presentations and interpretations for review by the user.
The interpreted ECG with measurements and interpretive statements is offered to the clinician on The Interpreted LCG with measurements conjunction with the clinician's knowledge of the all advisory basis only. It is to be assu interest the ECG tracings, and other clinical findings. A patient, the results of the physical enammand validate (or change) the computer generated ECG interpretation.
X Prescription Use
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|------------------------------------|
| Division of Cardiovascular Devices |
| 510(k) Number | K113144 |
|---------------|---------|
|---------------|---------|
Page 2 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.