K080999 · Philips Medical Systems · DPS · May 9, 2008 · Cardiovascular
Device Facts
Record ID
K080999
Device Name
PAGEWRITER CARDIOGRAPH
Applicant
Philips Medical Systems
Product Code
DPS · Cardiovascular
Decision Date
May 9, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Intended Use
To analyze multi-channel ECG signals from adult and pediatric patients with algorithms that provide measurements, data presentations, graphical presentations and interpretations for review by the user. The interpreted ECG with measurements and interpretive statements is offered to the clinician on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the patient, the results of the physical examination, the ECG tracings, and other clinical findings. A qualified physician is asked to over read and validate (or change) the computer generated ECG interpretation.
Device Story
Philips PageWriter Cardiograph (Model 860315) acquires multi-channel ECG signals from adult and pediatric patients; utilizes interpretive algorithms to provide measurements, graphical presentations, and clinical interpretations. Device features improved human factors engineering with a patient interface module and 802.11g wireless connectivity. Operated by clinicians in clinical settings; output serves as advisory information for physician review. Physician must over-read and validate computer-generated interpretations before clinical decision-making. Benefits include streamlined ECG acquisition (3-step process) and enhanced diagnostic support through automated analysis.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, integration tests, and verification of risk mitigations identified in hazard analysis to confirm performance and reliability of the new algorithm relative to the predicate.
Technological Characteristics
Electrocardiograph (21 CFR 870.2340); includes patient interface module; 802.11g wireless connectivity; interpretive ECG algorithm; system-level integration; Class II device.
Indications for Use
Indicated for adult and pediatric patients requiring ECG evaluation for diagnosis, treatment planning, treatment effectiveness assessment, or symptom rule-out.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
K031422 — PHILIPS PAGE WRITER TRIM SERIES CARDIOGRAPH INCLUDING I, II, III, MODELS 860286, 860287, 860288, 860289, 860290, 860291 · Philips Medical Systems · Jul 3, 2003
K173830 — MAC VU360 Resting ECG Analysis System · Ge Medical Systems Information Technologies, Inc. · Sep 18, 2018
K020708 — PHILIPS PAGEWRITER TOUCH CARDIOGRAPH, MODEL M5000A SERIES · Philips Medical Systems, Inc. · Apr 25, 2002
K251670 — MAC 7 Resting ECG Analysis System · Ge Medical Systems Information Technologies, Inc. · Jan 26, 2026
K101115 — ELI 150 ELECTROCARDIOGRAPH MODEL ELI 150 · Mortara Instrument, Inc. · Jun 18, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
#### 8.0 510(k) Summary
K080999
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- 1. The submitter of this pre-market notification is Philips Medical Systems
This summary was prepared on 4 April 2008
- 2. The name of this device is the Philips PageWriter Cardiograph. Classification is as follows:
| 1. 2012<br>Classification<br>2000 2000 1000 2000 2000 2000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 | Pro Code<br>. | CONTRACTOR PROPERTY CALL PROPERTY CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACT<br>Description Description<br>11 2015<br>2018 23:33 2017<br>a do the state the state of the states<br>一个人的一次的人的人的人 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 870.2340. I | DPS<br>বে ব | r 1<br>Electrocardiograph |
- The new device is substantially equivalent to the previously cleared Philips 3. cardiograph cleared under K020708.
- 4. The modification provides the following additional capabilities: improved ease of use ( 3 step ECG), improved human factors engineering with patient interface module, 802.11 g wireless capability, compatibility with extended lead ECG interpretive algorithm.
- 5. The new device has the same Indications for Use as the legally marketed predicate device.
- 6. The new device has the same technological characteristics as the legally marketed predicate device.
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new algorithm with respect to the predicate. Testing involved use of applicable consensus standards, system level tests, integration tests, and verification of risk mitigation identified in the hazard analysis.
page 1 of 1
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem consists of a symbol that resembles an abstract caduceus or a stylized representation of human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
**MAY - 9 2008**
Philips Medical Systems c/o Mr. Paul Schrader Senior Manager Regulatory Affairs Specialist Ultrasound and Monitoring Systems 3000 Minuteman Road Andover, MA 01810-1099
Re: K080999
PageWriter Cardiograph, Model 860315 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: April 4, 2008 Received: April 8, 2008
Dear Mr. Schrader:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Paul Schrader
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
ch
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### 2.0 Indications for Use
# 510(k) Number (if known): K080999
Device Name: PageWriter Electrocardiograph (Model 860315)
# Intended Use
To analyze multi-channel ECG signals from adult and pediatric patients with algorithms that provide measurements, data presentations, graphical presentations and interpretations for review by the user.
The interpreted ECG with measurements and interpretive statements is offered to the clinician on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the patient, the results of the physical examination, the ECG tracings, and other clinical findings. A qualified physician is asked to over read and validate (or change) the computer generated ECG interpretation.
## Indications for Use
Where the clinician decides to evaluate the electrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule-out causes for symptoms.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use N/A (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 14 of 62
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | 1680949 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.