The MODULUS stems are indicated for use in partial or total hip arthroplasty and are intended for press-fit (uncemented) use. When used in total hip arthroplasty, the Modulus Stems are intended for use with Co-Cr-Mo femoral heads and a compatible acetabular cup. When used in partial hip arthroplasty, the Modulus Stems are intended for use with the Lock Bipolar Heads. Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia; - rheumatoid arthritis; - osteoarthritis after femoral heads fractures; - correction of functional deformity; - revisions in cases of good remaining femoral bone stock. The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction.
Device Story
Modular femoral hip system for partial or total hip arthroplasty. Components include femoral stem, femoral neck, CoCrMo femoral heads, cemented UHMWPE cups, and Lock Bipolar heads. Stem and neck made of Ti6Al4V; stem features 5° conicity and round finned section. Modular components joined via Morse taper locking mechanism with safety screw. Used in clinical orthopedic settings by surgeons. Provides structural replacement for hip joint to reduce pain and restore function. Radiographic evaluation enabled by AISI 316-L stainless steel ring in cemented cups. System allows for press-fit (uncemented) femoral fixation and cemented acetabular reconstruction.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated through bench testing, including fatigue testing of femoral stems and modular connections, pull-out testing for bipolar heads, and range-of-motion simulations.
Technological Characteristics
Materials: Ti6Al4V (ASTM F1472, ISO 5832-3), CoCrMo (ASTM F1537), UHMWPE (ASTM F648, ISO 5834-2), AISI 316-L stainless steel. Design: Modular femoral stem with 5° conicity, Morse taper locking mechanism with safety screw, 12/14 neck-head taper. Cemented cups feature grooves for rotational stability. Radiopaque marker ring included. Non-sterile or sterile (implied by orthopedic standard).
Indications for Use
Indicated for skeletally mature patients requiring partial or total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, hip dysplasia), rheumatoid arthritis, osteoarthritis following femoral head fractures, functional deformity correction, or revision surgery with sufficient femoral bone stock.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Foundation (FMP) porous coated spiked acetabular system (K072154)
Submission Summary (Full Text)
{0}------------------------------------------------
K112158
# Summary of Safety and Effectiveness
Date: April 26, 2012
Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 – Villanova di San Daniele Udine - Italy
U.S. Contact Person: Cheryl Hastings Principal Consultant Phone: 574-527-4220
| Product | Product Code | Regulation and Classification Name |
|-------------------------------|--------------|--------------------------------------------------------------------------------------------------------------|
| MODULUS Femoral<br>Hip System | JDI | Hip joint metal/polymer semi-constrained<br>cemented prosthesis per 21 CFR 888.3350 |
| | KWY | Hip joint femoral (hemi-hip) metal/polymer<br>cemented or uncemented prosthesis per 21 CFR<br>888.3390 |
| | LPH | Hip joint metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis per 21<br>CFR 888.3358 |
| | KWZ | Hip joint metal/polymer constrained cemented<br>or uncemented prosthesis per 21 CFR 888.3310 |
### Description:
The Modulus femoral hip system consists of modular femoral hip prostheses, modular CoCrMo heads, cemented UHWMPE cups (for use in total hip arthroplasty) and bipolar heads (for use in partial or hemi-hip arthroplasty).
The Modulus femoral hip prosthesis consists of a femoral stem component and a femoral neck component combined by a Morse taper locking mechanism. A locking screw is provided to improve the security of fixation. The MODULUS femoral hip prosthesis is intended for cementless use in hip joint arthroplasty.
The femoral stem has a 5° conicity with a round finned section. It is made of Ti6A14V (ISO 5832-3, ASTM F1472) and is sandblasted. The stem is available in 11 sizes with diameters ranging from 16 to 26 mm with a 1 mm increment.
The femoral neck is available in two cervico-diaphyseal configurations (125° and 135°) and two lengths (42 and 48 mm). It is made of Ti6Al4V (ISO 5832-3, ASTM F1472). The proximal portion of the neck is highly polished while the distal portion is macroroughened.
Traditional 510(k) - MODULUS Femoral Hip System April 26, 2012
p. 1 of 4
{1}------------------------------------------------
K11 2158
Modular heads are manufactured from CoCrMo (ASTM F1537). They are characterized by a spherical shape and are used with an acetabular cup in total hip arthroplasty. The femoral heads are coupled with the MODULUS hip prosthesis by means of a 12/14 Morse taper. They are available in diameters of 28, 32 and 36 mm and in offset sizes S. M, L, XL.
Cemented cups are manufactured entirely of UHMWPE (ASTM F648, ISO 5834-2).
The acctabular cups are available in two different versions: standard and protruded. The standard version is hemispherical and is designed for use in deep acetabula. The protruded version has a lateral portion designed to resist dislocation.
An AISI 316-L stainless steel ring is inserted in the equatorial region to allow the evaluation of the device position through radiography.
The following table shows the availability of the cemented cup diameters for each type of cup.
| Cup Type | Inner diameter | Range of Outer Diameter |
|----------------|----------------|----------------------------------|
| Standard Cups | 28 mm | 44-58 mm with increments of 2 mm |
| | 32 mm | 50-58 mm with increments of 2 mm |
| Protruded Cups | 28 mm | 44-58 mm with increments of 2 mm |
| | 32 mm | 50-58 mm with increments of 2 mm |
The Modulus femoral prostheses and modular CoCrMo heads are also compatible with the DJO FMP acetabular cups, previously cleared in K974093, K974095, K973119, K023794 and K072154.
Lock Bipolar Heads are modular prostheses characterized by an external shell made from CoCrMo (ASTM F1537) and by a conical liner made from UHMWPE (ASTM F648, ISO 5834-2). A retentive ring allows the shell to assemble with modular heads. Lock Bipolar heads are available in 13 sizes from 45 to 57 mm with a 1 mm increment.
Intended Use: The Modulus femoral hip prosthesis is indicated for use in partial or total hip arthroplasty and is intended for press-fit (uncemented) use. When used in total hip arthroplasty, the MODULUS hip prosthesis is intended for use with a Co-Cr-Mo femoral head and a compatible acetabular cup. When used in partial hip arthroplasty, the MODULUS hip prosthesis is intended for use with a Lock Bipolar Head.
The Modulus femoral hip system is indicated for use in the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular . necrosis and hip dysplasia;
- . rheumatoid arthritis;
- . osteoarthritis after femoral heads fractures;
- correction of functional deformity; .
- revisions in cases of good remaining femoral bone stock. .
Traditional 510(k) - MODULUS Femoral Hip System April 26, 2012
p. 2 of 4
{2}------------------------------------------------
K112158
The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction.
The DJO FMP acetabular cups are intended for press-fit (uncemented) use.
### Predicate Devices:
Keystone Hip (Encore Medical, L.P., K000521) PROFEMUR R femoral system (Wright Mcdical Technology, K041114-K091423) WAGNER CONE stem (Centerpulse orthopaedics, K032380). Modular Revision Hip Stem (Encore Medical, L.P., K092331) Articul/eze femoral heads (Depuy, K883460-K980513) PE-Plus acetabular cups (Plus Orthopaedics, K992153) ZCA All Poly Acetabular Cup (K030153) Global Bipolar system (Smith & Nephew, K023743) Bipolar (Encore Orthopedics Inc. K953510) Foundation porous coated hemispherical shell (Encore Orthopedics, Inc., K974093) Foundation porous coated flared rim shell (Encore Orthopedics, Inc., K974095) Foundation porous acetabular system (Encore Orthopedics, Inc., K973119) Constrained liner (Encore Medical, L.P.,K023794) Foundation (FMP) porous coated spiked acetabular system (Encore Medical, L.P., K072154)
Comparable Features to Predicate Devices: the Modulus femoral hip system is similar to the predicate devices in terms of intended use, indications, design and materials.
The Modulus femoral prosthesis and the predicate femoral prostheses are indicated for use in hip arthroplasty and for press-fit (uncemented) use. The CoCrMo heads and the predicate heads are all indicated for use in total hip arthroplasty, while the Lock Bipolar heads and the predicate bipolar heads are all indicated for use in partial hip arthroplasty. The cemented cups and the predicate acetabular cups are intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction.
Like the Keystone Hip, the Profemur R femoral system and the Modular Revision Hip system, the Modulus femoral prosthesis is a modular prosthesis made up of a cylindrical stem and a proximal body that are joined through a taper connection stabilized by a safety screw. As with the predicates devices, the Modulus prosthesis is characterized by a 12/14 neck-head taper. The ccmented cups and the predicate cups are all monolithic and are characterized by several grooves in the outer surface to enhance fixation and rotational stability of the cup in bone cement. Like the Global Bipolar System, the Lock Bipolar heads are modular devices composed of an external shell, an inner liner and a locking mechanism.
The components of the Modulus femoral hip system are manufactured from the same materials as the predicate devices.
Traditional 510(k) - MODULUS Femoral Hip System April 26, 2012
p. 3 of 4
{3}------------------------------------------------
K112158
Non-Clinical Testing: The Modulus femoral hip system has undergone fatigue testing to demonstrate the strength of the femoral stem and the strength of the modular connections. A pull-out test was performed to test the force necessary to disassemble the Lock bipolar head. A simulation of the Range of Motion has been performed to ensure the device design does not overly limit range of motion. All mechanical testing was done on worst case components or constructs. Where possible, standard test methods were used to allow comparison to testing of the predicate devices. The testing results demonstrated the system's ability to perform under expected clinical conditions.
Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the Modulus femoral hip system to the predicate devices.
Traditional 510(k) – MODULUS Femoral Hip System April 26, 2012
., \$4.4
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or other bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
LimaCorporate S.p.A. % Ms. Cheryl Hastings Principal Consultant P.O. Box 696 Winona Lake, Indiana 46590
#### 1 0 2012 A AY
Re: K112158
Trade/Device Name: Modulus stems, necks, heads, and cups Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
Regulatory Class: II Product Code: LPH, JDI, KWY, KWZ Dated: March 30, 2012 Received: April 4, 2012
Dear Ms. Hastings:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny receir sthe Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Ms. Cheryl Hastings
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
KILL 58 510(k) Number (if known): Unichown
Device Name: MODULUS Femoral Hip System
Indications for Use:
## MODULUS Femoral Hip System Indications for Use
The MODULUS stems are indicated for use in partial or total hip arthroplasty and are intended for press-fit (uncemented) use. When used in total hip arthroplasty, the Modulus Stems are intended for use with Co-Cr-Mo femoral heads and a compatible acetabular cup. When used in partial hip arthroplasty, the Modulus Stems are intended for use with the Lock Bipolar Heads.
Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular . necrosis and hip dysplasia;
- . rheumatoid arthritis;
- osteoarthritis after femoral heads fractures; .
- correction of functional deformity; ●
- revisions in cases of good remaining femoral bone stock. ●
The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
A.J.R
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number_K 112158
Page 1 of 1
Traditional 510(k) - MODULUS Femoral Hip System April 23, 2012
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.