WAGNER CONE PROSTHESIS

K032380 · Centerpulse Orthopedics, Inc. · LZO · Sep 22, 2003 · Orthopedic

Device Facts

Record IDK032380
Device NameWAGNER CONE PROSTHESIS
ApplicantCenterpulse Orthopedics, Inc.
Product CodeLZO · Orthopedic
Decision DateSep 22, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Wagner Cone Prosthesis is intended for prosthetic replacement of the proximal portion of the femur in total hip arthroplasty. This device is intended for cementless application where primary fixation occurs predominantly in the distal region. Specific diagnostic indications for use of the Wagner Cone Prosthesis are: - patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis; - those patients with failed previous surgery where pain, deformity, or dysfunction persists; - revision of previously failed hip arthroplasty.

Device Story

Wagner Cone Prosthesis is a titanium-aluminum-niobium alloy hip stem designed for cementless proximal femoral replacement. Device features conical configuration with circular cross-section and eight longitudinal flutes to provide rotational stability; stem diameter tapers from proximal to distal end. Intended for use in total hip arthroplasty where primary fixation occurs in distal femoral region. Implanted by orthopedic surgeons in clinical/hospital settings. Device replaces diseased or failed proximal femur to restore joint function, reduce pain, and correct deformity. Substantial equivalence established via performance testing and design comparisons to predicate hip stems.

Clinical Evidence

No clinical data provided; substantial equivalence is supported by bench testing and design/functional analysis.

Technological Characteristics

Material: Titanium-aluminum-niobium alloy. Design: Conical stem with circular cross-section and eight longitudinal flutes for rotational stability. Fixation: Cementless, distal-fixation. Form factor: Tapered proximal-to-distal geometry.

Indications for Use

Indicated for patients requiring proximal femoral replacement during total hip arthroplasty, including those with noninflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis), inflammatory joint disease (e.g., rheumatoid arthritis), failed previous surgery, or revision of failed hip arthroplasty.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) SUMMARY | SPONSOR NAME: | Centerpulse Orthopedics, Inc.<br>9900 Spectrum Drive<br>Austin, TX 78717 | |-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) CONTACT: | Robert M. Wolfarth<br>Phone: (512) 432-9324<br>E-Mail: Robert.Wolfarth@Centerpulse.com | | TRADE NAME: | Wagner Cone Prosthesis | | COMMON NAME: | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis | | CLASSIFICATION: | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prostheses (Product Code 87 LZO) are Class<br>II per 21 CFR §888.3353, reviewed by the Orthopedic Devices panel. | #### PREDICATE DEVICES: - MP™ Reconstruction Hip Stem by Waldemar Link GmbH & Co KG. (K955296) ● - Restoration™ Modular System by Howmedica Osteonics Corp. (K022549) ● - Wagner Revision Stem (K960588) . - Precedent Revision Hip Stem (K972637) . ## DEVICE DESCRIPTION: This stem is a titanium-aluminum-niobium alloy conical configuration with a taper. The design features a circular cross-section with eight equally spaced conical longitudinal flutes down the long axis of the stem for rotational stability. The stem's core diameter tapers gradually from the proximal end to the distal end. ## INTENDED USE: The Wagner Cone Prosthesis is intended for prosthetic replacement of the proximal portion of the femur in total hip arthroplasty. This device is intended for cementless application where primary fixation occurs predominantly in the distal region. Specific diagnostic indications for use of the Wagner Cone Prosthesis are: - patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., avascular . necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis; - those patients with failed previous surgery where pain, deformity, or dysfunction persists; - . revision of previously failed hip arthroplasty. ## BASIS OF SUBSTANTIAL EQUIVALENCE: Performance tests, design comparisons, and functional analyses conducted on the Wagner Cone Prosthesis demonstrate that it is substantially equivalent to the predicate devices. Page 1 of 1 {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 2 2003 Mr. Robert M. Wolfarth Regulatory Affairs Programs Manager Centerpulse Orthopedics, Inc. 9900 Spectrum Drive Austin, TX 78717 Re: K032380 Trade/Device Name: Wagner Cone Prosthesis Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: LZO Dated: July 30, 2003 Received: August 1, 2003 Dear Mr. Wolfarth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - Mr. Robert M. Wolfarth This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark M Millham Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K032380 #### Indications For Use 510(K) Number. Wagner Cone Prosthesis Device Name: #### Indications for Use: The Wagner Cone Prosthesis is intended for prosthetic replacement of the proximal portion of the femur in total hip arthroplasty. This device is intended for cementless application where primary fixation occurs predominantly in the distal region. Specific diagnostic indications for use of the Wagner Cone Prosthesis are: - patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., . avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis; - those patients with failed previous surgery where pain, deformity, or dysfunction . persists; - revision of previously failed hip arthroplasty. ● (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) (Division Sign-Off) Page 1 of 1 (Division Sign-Off) D. ision of General, Restorative and Neurological Devices 510(k) Number. K032380
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...