The Acu-Sinch Repair System is intended to be used in conjunction with a clavicle plate of the Congruent Bone Plate System to provide fixation during the healing of clavicle fractures. The Acu-Sinch Repair System also may be used as a stand-alone system for treatment of acromioclavicular and/or coracoclavicular ligament disruption. The Acumed Suture Anchor is intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow in the following procedures: Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction; Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair; Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Iliotibial Band Tenodesis; Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction; Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. Acumed's Locking Clavicle Plating System is designed to provide fixation during fractures, fusions, or osteotomies of the clavicle.
Device Story
Acu-Sinch Repair System and Acumed Suture Anchor provide orthopedic fixation; components include suture anchors, clavicle plates, screws, and sutures. System used for clavicle fracture healing (adjunct to Congruent Bone Plate System) or as stand-alone for acromioclavicular/coracoclavicular ligament disruption. Suture anchors facilitate suture-to-bone fixation across shoulder, foot/ankle, knee, hand/wrist, and elbow. Operated by surgeons in clinical/OR settings. Device constructs provide mechanical stability to bone/ligament structures during healing. Benefits include secure fixation for various orthopedic reconstructions and fracture management.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence consists of load-to-failure testing, fatigue testing, pull-out comparison testing for 2.3mm screws, and engineering analysis of plates to demonstrate substantial equivalence.
Technological Characteristics
Metallic bone fixation system; components include suture anchors, plates, screws, and sutures. Materials and strength characteristics are equivalent to predicate devices. Mechanical fixation principle; non-powered; non-software device.
Indications for Use
Indicated for patients requiring bone fixation or ligament repair in the shoulder, foot/ankle, knee, hand/wrist, or elbow. Specific procedures include rotator cuff, Bankart, SLAP, biceps, deltoid, capsular, ligamentous, and tendon repairs/reconstructions, as well as clavicle fracture fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device (K082095)
{0}------------------------------------------------
# NOV 1 4 2011
510(k) Summary
| Device Trade Name: | Acu-Sinch Repair System and Acumed Suture Anchor |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Acumed, LLC<br>5885 NW Cornelius Pass Road<br>Hillsboro, OR 97124 |
| Contact: | Ms. Lino Tsai<br>Regulatory Specialist<br>Phone: (503) 627-9957 |
| Prepared by: | Musculoskeletal Clinical Regulatory Advisers, LLC<br>1331 H Street NW, 12th Floor<br>Washington, DC 20005<br>Phone: (202) 552-5800<br>Fax: (202) 552-5798 |
| Date Prepared: | October 28, 2011 |
| Classifications: | 21 CFR 888.3030, Single/multiple component metallic<br>bone fixation appliances and accessories |
| | and |
| | 21 CFR 888.3040, Smooth or threaded metallic bone<br>fixation fasteners |
| Class: | II |
| Product Codes: | HTN, HWC, MBI, and HRS |
#### Indications For Use:
The Acu-Sinch Repair System is intended to be used in conjunction with a clavicle plate of the Congruent Bone Plate System to provide fixation during the healing of clavicle fractures.
The Acu-Sinch Repair System also may be used as a stand-alone system for treatment of acromioclavicular and/or coracoclavicular ligament disruption.
The Acumed Suture Anchor is intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow in the following procedures:
- Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Shoulder: . Shoulder: " Rolator "Our repairs, aration Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
{1}------------------------------------------------
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, . Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair
- Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, . Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Iliotibial Band Tenodesis
- Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament . Hand/Wrist: Reconstruction, Radial Collateral Ligament Reconstruction
- Biceps Tendon Reattachment. Ulnar or Radial Collateral Ligament . Elbow: Reconstruction
Acumed's Locking Clavicle Plating System is designed to provide fixation during fractures, fusions, or osteotomies of the clavicle.
### Device Description:
The Acu-Sinch Repair System consists of several components (Suture Anchor, Clavicle Fracture Plate, Acu-Sinch plate, Suture Retainer, Screws, and Suture) which can be combined to create various device constructs to treat the indications listed above. The Suture Anchor can be used for fixation of suture to bone in the anatomic locations listed above.
#### Substantial Equivalence
The components of the Acu-Sinch Repair System are substantially equivalent to the following devices: Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device (K082095); and Arthrex TightRope™ Acromioclavicular (AC) Device (K052776) with respect to indications, materials, and technological characteristics. The suture to be used with the Acu-Sinch Repair System was cleared in K063778. When the Acu-Sinch Repair System is used as an adjunct in fracture repair, the clavicle plates are substantially equivalent to plates cleared in K012655 with respect to materials and plate strength.
The Acumed Suture Anchor is substantially equivalent to the Arthrex Corkscrew FT (K061665) with respect to indications, materials, and technological characteristics. The suture to be used with the Acumed Suture Anchor was cleared in K063778.
### Preclinical Testing:
The Acu-Sinch Repair System was subjected to load to failure and fatigue testing. A pull-out comparison testing was performed to support the use of 2.3mm screws with the Distal Clavicle Plates. In addition, an engineering analysis of the plates was completed. The results demonstrate that the components are substantially equivalent to the predicate.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a double helix-like design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Acumed LLC % Ms. Lino Tsai Regulatory Specialist 5885 NW Cornelius Pass Road Hillsboro, Oregon 97124-9432
NOV 1 4 2011
Re: K112111 Trade/Device Name: Acu-Sinch Repair System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN, HWC, MBI, HRS Dated: October 10, 2011 Received: October 11, 2011
Dear Ms. Tsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
#### Page 2 - Ms. Lino Tsai
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use 4.
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Acu-Sinch Repair System
The Acu-Sinch Repair System is intended to be used in conjunction with a clavicle plate of the Congruent Bone Plate System to provide fixation during the healing of clavicle fractures.
The Acu-Sinch Repair System also may be used as a stand-alone system for treatment of acromioclavicular and/or coracoclavicular ligament disruption.
V Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 2
(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic, Division of Surgices Division of Surger
and Restorative Devices and Restorative Devices
K//2///
510(k) Number
{5}------------------------------------------------
# KII 2111
The Acumed Suture Anchor is intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow in the following procedures:
- Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, . Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux . Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair
- Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral . Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Iliotibial Band Tenodesis
- Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament . Hand/Wrist: Reconstruction, Radial Collateral Ligament Reconstruction
- Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament . Elbow: Reconstruction
Acumed's Locking Clavicle Plating System is designed to provide fixation during fractures, fusions, or osteotomies of the clavicle.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sigh-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number `/
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.