VITEK 2 STRPTOCOCCUS AMPICILLIN

K112075 · bioMerieux, Inc. · LON · Sep 15, 2011 · Microbiology

Device Facts

Record IDK112075
Device NameVITEK 2 STRPTOCOCCUS AMPICILLIN
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateSep 15, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

VITEK® 2 Streptococcus Ampicillin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 Streptococcus Ampicillin is a quantitative test intended for use with the VITEK® 2 and the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Ampicillin has been shown to be activity against the microorganisms listed below: - Beta-hemolytic group Streptococcus species - Viridans group Streptococcus species The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

Device Story

VITEK® 2 AST-ST Ampicillin is a miniaturized, automated antimicrobial susceptibility test (AST) card used with VITEK® 2 and VITEK® 2 Compact systems. Bacterial isolates are diluted in saline, used to rehydrate antimicrobial medium in the card, and processed by the system. The VITEK® 2 system automatically fills, seals, and incubates the card, monitoring growth in each well over time. The VITEK® 2 Compact utilizes manual filling and sealing. The system generates a report providing the Minimum Inhibitory Concentration (MIC) and interpretive category results. Used in clinical laboratories by technicians to assist healthcare providers in selecting appropriate antimicrobial therapy for Streptococcus infections.

Clinical Evidence

Performance evaluated via clinical and challenge studies (n=1272 combined). Compared to CLSI broth microdilution reference method. Results: 99.1% Essential Agreement (EA) and 97.0% Category Agreement (CA). Minor discrepancies observed; no major or very major discrepancies. Reproducibility >95%. Bench testing only.

Technological Characteristics

Miniaturized AST card utilizing doubling dilution methodology. Automated incubation and optical growth monitoring. Compatible with VITEK® 2 and VITEK® 2 Compact instrument platforms. Quantitative MIC determination.

Indications for Use

Indicated for antimicrobial susceptibility testing of isolated colonies of Beta-hemolytic group Streptococcus species and Viridans group Streptococcus species to determine in vitro susceptibility to Ampicillin.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K112075 SEP 1 5 2011 # 510(k) SUMMARY ### VITEK® 2 AST-ST Ampicillin ### 510(k) Submission Information: Contact Person: Fax Number: | Submitter's Name: | bioMérieux, Inc. | |-------------------|----------------------------------------| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | Jolyn Tenllado Director, Regulatory Affairs Phone Number: 314 -731-8386 314-731-8689 September 2011 #### B. Device Name: Formal/Trade Name: Classification Name: Date of Preparation: VITEK® 2 Streptococcus Ampicillin VITEK® 2 AST-ST Ampicillin Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Device, 21 CFR 866.1645 Product Code LON Common Name: C. Predicate Device: VITEK® 2 Gram Positive Amoxicillin for Streptococcus pneumoniae (K063597) #### D. 510(k) Summary: VITEK® 2 AST-ST Ampicillin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 AST-ST Ampicillin is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Ampicillin has been shown to be active against the microorganisms listed below: Beta-hemolytic group Streptococcus species Viridans group Streptococcus species The antimicrobial agent presented in VITEK® 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK® 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology. 00/638-4835 Fax: 314/731-8700 {1}------------------------------------------------ The bacterial isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 36 hours for yeast). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card. VITEK® 2 AST-ST Ampicillin demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued August. 28, 2009. The Premarket Notification (510[k]) presents data in support of VITEK® 2 AST-ST Ampicillin. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 AST-ST Ampicillin by comparing its performance with the CLSI broth microdilution reference method. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms. VITEK 2 AST-ST Ampicillin demonstrated acceptable performance of 99.1% overall essential agreement and 97.0% overall category agreement with the reference method. Reproducibility and Quality Control demonstrated acceptable results using both the VITEK® 2 and VITEK® 2 Compact instrument systems. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The bird is positioned to the right of a circular arrangement of text that spells out "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 bioMérieux, Inc. c/o Jolyn Tenllado Director, Regulatory Affairs 595 Anglum Rd. Hazelwood, Missouri 63042 SEP 15 2011 Re: K112075 Trade/Device Name: VITEK® 2 Streptococcus Ampicillin Regulation Number: 21 CFRS 866.1645 Regulation Name: Short-Term Antimicrobial Susceptibility Test System Regulatory Class: Class II Product Code: LON Dated: July 19. 2011 Received: July 20, 2011 Dear Ms. Tenllado: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. - If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning vour device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice {3}------------------------------------------------ Page 2 - Jolyn Tenllado requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at us toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industrv/support/index.html. Sincerely yours, Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K112075 Device Name: VITEK® 2 Streptococcus Ampicillin ( ≤ 0.25 – ≥ 16 µg/mL) Indications For Use: VITEK® 2 Streptococcus Ampicillin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 Streptococcus Ampicillin is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Ampicillin has been shown to be active against the microorganisms listed below: Beta-hemolytic group Streptococcus species Viridans group Streptococcus species The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Freddie M. Poole Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K112075
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...