VITEK 2 AST STREPTOCOCUS VANCOMYCIN

K121100 · bioMerieux, Inc. · LON · Nov 19, 2012 · Microbiology

Device Facts

Record IDK121100
Device NameVITEK 2 AST STREPTOCOCUS VANCOMYCIN
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateNov 19, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

VITEK® 2 AST Streptococcus Vancomycin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 AST Streptococcus Vancomycin is a quantitative test intended for use with the VITEK® 2 and ViTEK® 2 Compact Systems as a laboratory aid in th determination of in vitro susceptibility to antimicrobial agents. Vancomycin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Viridans group streptococci In vitro data are available, but their clinical significance is unknown Streptococcus agalactiae Streptococcus pyogenes. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 The VT LN 27 kitimilloredial outdative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

Device Story

VITEK 2 AST Streptococcus Vancomycin card; miniaturized, automated version of doubling dilution technique. Input: bacterial isolate diluted in saline. Process: card rehydrated with isolate; VITEK 2 system monitors growth in 64 micro-wells via transmittance optics (visible light) over 4-16 hours. Algorithm: growth-based threshold calculation; discriminant analysis determines MIC. Output: MIC value (0.125–8 µg/mL) and SIR interpretation. Used in clinical laboratories; operated by technicians. Benefits: automated, rapid determination of in vitro susceptibility to assist clinical decision-making.

Clinical Evidence

External evaluation conducted using fresh clinical isolates and stock challenge strains. Performance compared against CLSI broth microdilution reference method. Results demonstrated 100% overall Category Agreement. Reproducibility and Quality Control testing performed on VITEK 2 and VITEK 2 Compact platforms showed acceptable results.

Technological Characteristics

Miniaturized 64-well test card; microdilution MIC technique. Sensing: visible light transmittance optics. Connectivity: VITEK 2/VITEK 2 Compact systems. Software: growth-based algorithm using discriminant analysis. Sterilization: not specified.

Indications for Use

Indicated for antimicrobial susceptibility testing of Streptococcus species (Viridans group, S. agalactiae, S. pyogenes) using isolated colonies. For professional laboratory use only.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121/000 # RIEUX #### 510(k) SUMMARY ## VITEK® 2 AST Streptococcus Vancomycin NOV 1 9 2012 #### 510(k) Submission Information: | Submitter's Name: | bioMérieux, Inc. | |----------------------|-----------------------------------------------------------------------------------------------------| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | | Contact Person: | Jocelyn Jennings | | | Senior Manager, Regulatory Affairs | | Phone Number: | 919-620-2894 | | Fax Number: | 919-620-2548 | | Date of Preparation: | April 5, 2012 | | B. Device Name: | | | Formal/Trade Name: | VITEK® 2 AST Streptococcus Vancomycin<br>(< 0.125 - 8 ug/ml) | | Classification Name: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 | | Common Name: | VITEK 2 AST-ST Vancomycin | | C. Predicate Device: | VITEK 2 AST-GP Amoxicillin K063597 | #### D. 510(k) Summary: VITEK® 2 AST Streptococcus Vancomycin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 AST Streptococcus Vancomycin is a quantitative test intended for use with the VITEK® 2 and ViTEK® 2 Compact Systems as a laboratory aid in th determination of in vitro susceptibility to antimicrobial agents. Vancomycin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Viridans group streptococci In vitro data are available, but their clinical significance is unknown Streptococcus agalactiae Streptococcus pyogenes bioMérieux, Inc. 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700 {1}------------------------------------------------ The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/m). The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology. The bacterial isolate to be tested is diluted to a standardized concentration in 0.45 - 0 .50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card. VITEK 2 AST Streptococcus Vancomycin demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued August 28, 2009. The Premarket Notification [510(k)] presents data in support of VITEK 2 AST Streptococcus Vancomvoin. An external evaluation was conducted with fresh clinical isolates and stock challence strains. The external evaluations were designed to confirm the acceptability of VITEK 2 AST Streptococcus Vancomycin by comparing its performance with the CLSI broth microdilution reference method. The data is representative of performance on both the VITEK 2 and VITEK 2 Compact instrument platforms. VITEK 2 AST Streptococcus Vancomycin demonstrated acceptable performance of 100% overall Category Agreement. Reproducibility and Quality Control demonstrated acceptable results using both the VITEK 2 Compact instrument systems. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract caduceus, with three vertical lines and a serpentine form at the bottom. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 bioMerieux, Inc. C/O Jocelyn Jennings, Senior Manager, Regulatory Affairs 595 Anglum Road Hazelwood MO 63042 ### NOV 1 9 2012 Re: K121100 VITEK®2 AST Streptococcus Vancomycin Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON Dated: November 8, 2012 Received: November 9, 2012 Dear Ms. Jennings: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - [Ms. Jennings] forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, # Sally A. Hojvat Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known):_k121100 Device Name: VITEK®2 AST Streptococcus Vancomycin (≤ 0.125 - 8 µg/mL) #### Indications For Use: VITEK® 2 AST Streptococcus Vancomycin is designed for antimicrobial susceptibility testing of Streptococcus species. VITEK® 2 AST Streptococcus Vancomycin is a quantitative test intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the for ass with the VTTER 2 and VTER 2 otimicrobial agents. Vancomycin has been shown to be a fan this detentified of in 'ntro's strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Viridans group streptococci In vitro data are available, but their clinical significance is unknown Streptococcus agalactiae Streptococcus pyogenes The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 The VT LN 27 kitimilloredial outdative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) foaia Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K121100
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...