VITEK® 2 Gram Negative Ceftazidime/Avibactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ceftazidime/ Avibactam is a quantitative test. Ceftazidime/Avibactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active both in vitro and in clinical infections: Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis Pseudomonas aeruginosa The following in vitro data are available, but clinical significance is unknown: Citrobacter koseri Enterobacter aerogenes Morganella morganii Providencia stuartii Serratia marcescens The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Streptococcus spp., Streptococcus spp. and clinically significant yeast.
Device Story
VITEK 2 AST-GN Ceftazidime/Avibactam is an automated, miniaturized, growth-based antimicrobial susceptibility test (AST) system. It utilizes 64-well test cards containing premeasured antimicrobial concentrations. Input consists of isolated bacterial colonies diluted in saline. The system automatically (VITEK 2) or manually (VITEK 2 Compact) fills, seals, and incubates cards. An optical scanner measures light attenuation every 15 minutes to monitor bacterial growth. The system calculates the Minimum Inhibitory Concentration (MIC) based on growth kinetics. Results are reported as MIC values and interpretive categories (Susceptible/Resistant). Used in clinical laboratories to guide antibiotic therapy for Gram-negative infections. The device provides rapid, standardized susceptibility data, aiding clinicians in selecting appropriate antimicrobial treatments and improving patient outcomes by identifying effective therapies.
Clinical Evidence
Performance evaluated via clinical and challenge studies (n=1073 total results). Compared to CLSI broth microdilution reference method. Overall Essential Agreement (EA) was 94.5% and Category Agreement (CA) was 98.7%. Major discrepancy rate was 1.4% (adjusted to 0.5% due to lack of intermediate category). P. aeruginosa showed a trend toward higher MICs; a limitation was added for repeat testing when MIC is ≥16 µg/mL.
Technological Characteristics
Miniaturized broth microdilution card containing antimicrobial concentrations. Automated incubation and optical growth monitoring system. Connectivity via VITEK 2/Compact instrument platforms. Software-based analysis of growth kinetics to determine MIC.
Indications for Use
Indicated for antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic Gram-negative bacilli, including Citrobacter, Enterobacter, Escherichia, Klebsiella, Proteus, Pseudomonas, Morganella, Providencia, and Serratia species.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
K161217 — VITEK 2 Gram Negative Ceftriaxone (<=0.25->=64 ug/mL) · bioMerieux, Inc. · Jan 12, 2017
K091899 — VITEK 2 GRAM NEGATIVE MEROPENEM · bioMerieux, Inc. · Aug 14, 2009
K211630 — VITEK 2 AST-Gram Negative Piperacillin / Tazobactam (<=4 - =>128 µg/mL) · bioMerieux, Inc. · Aug 30, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 19, 2017
bioMerieux, Inc. Ms. Cherece Jones Staff Regulatory Affairs Specialist 595 Anglum Rd. Hazelwood, Missouri 63042
Re: K172944
Trade/Device Name: VITEK 2 AST-GN Ceftazidime/Avibactam (<0.12 ->16 ug/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system Regulatory Class: Class II Product Code: LON, LTW, LTT Dated: September 25, 2017 Received: September 26, 2017
Dear Ms. Cherece Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ribhi Shawar -S For
Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172944
#### Device Name
VITEK® 2 AST-GN Ceftazidime/Avibactam (≤0.12 ->16 µg/mL)
#### Indications for Use (Describe)
VITEK® 2 Gram Negative Ceftazidime/Avibactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ceftazidime/ Avibactam is a quantitative test. Ceftazidime/Avibactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
### Active both in vitro and in clinical infections:
Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis Pseudomonas aeruginosa
### The following in vitro data are available, but clinical significance is unknown:
Citrobacter koseri Enterobacter aerogenes Morganella morganii Providencia stuartii Serratia marcescens
The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Streptococcus spp., Streptococcus spp. and clinically significant yeast.
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of a stylized globe with horizontal lines on the left side and a solid black fill on the right side. A curved line extends from the top of the globe, arcing over it. Below the globe is the company name, "BIOMÉRIEUX," in a sans-serif font.
# 510(k) SUMMARY
# VITEK® 2 AST-GN Ceftazidime/Avibactam
### 510(k) Submission Information:
| Submitter's Name: | bioMérieux, Inc. |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| Contact Person: | Cherece L. Jones<br>Staff Regulatory Affairs Specialist |
| Phone Number: | 314-731-8684 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | September 25, 2017 |
| <b>B. Device Name:</b> | |
| Formal/Trade Name: | VITEK® 2 AST-GN Ceftazidime/Avibactam (≤0.12 - ≥ 16 µg/mL) |
| Classification Name: | 21 CFR 866.1645<br>Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility System<br>Product Code: LON |
| Common Name: | VITEK® 2 AST-GN Ceftazidime/Avibactam |
| <b>C. Predicate Device:</b> | |
| | VITEK® 2 AST-GN Ceftolozane/Tazobactam (≤0.25 - ≥ 32 ug/ml) (K161510) |
## D. 510(k) Summary:
VITEK® 2 Gram Negative Ceftazidime/Avibactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ceftazidime/Avibactam is a quantitative test. Ceftazidime/Avibactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
## Active both in vitro and in clinical infections:
{5}------------------------------------------------
Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis Pseudomonas aeruginosa
## The following in vitro data are available, but clinical significance is unknown:
Citrobacter koseri Enterobacter aerogenes Morganella morganii Providencia stuartii Serratia marcescens
The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.
The antimicrobial presented in VITEK® 2 AST-GN Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK® 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK® 2 AST-GN Ceftazidime/Avibactam demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems: Guidance for Industry and FDA (Issued August 28, 2009).
The Premarket Notification (510[k]) presents data in support of VITEK® 2 AST-GN Ceftazidime/Avibactam. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm
{6}------------------------------------------------
the acceptability of VITEK® 2 AST-GN Ceftazidime/Avibactam by comparing its performance with the CLSI broth microdilution reference method incubated at 16-20 hrs. The data is representative of performance on both the VITEK® 2 Compact instrument platforms.
The VITEK® 2 AST-GN Ceftazidime/Avibactam (≤ 0.12 – ≥ 16 µg/mL) demonstrated acceptable performance as presented in Table 1 below:
# Table 1: Performance Characteristics for VITEK® 2 AST-GN Ceftazidime/Avibactam (ڪ 0.12 – ≥ 16 µg/mL)
| | % Essential Agreement | % Category Agreement* |
|----------------------------------------------------|-----------------------|-----------------------|
| | (EA) | (CA) |
| Overall Performance (with the<br>reference method) | 94.5 | 98.7 |
* The overall categorical major error rate for Entervase and Pseudomonas 1.4% (14998). Nine (9) major errors were one dilution apart from the reference method and as such fall within essential agreement, and the lack of an intermediate breakpoint for Ceftazidime/Avibactam, the adjusted categorical major error rate is 0.5% (5/998).
Reproducibility and Quality Control demonstrated acceptable results.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.