External evaluation with fresh and stock clinical isolates and challenge strains.
—
Indications for Use
VITEK® 2 Streptococcus Penicillin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK® 2 and VITEK® 2 COMPACT Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Streptococcus Penicillin is a quantitative test. Penicillin has been shown to be active against the microorganisms listed below, according to the FDA label for this antimicrobial. Active both in vitro and in clinical infections: Beta hemolytic Streptococci groups C and G Streptococcus pyogenes Streptococcus agalactiae Streptococcus viridans group Streptococcus pneumoniae The VITEK® 2 Streptococcus Penicillin Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.
Device Story
VITEK 2 AST-Streptococcus Penicillin is an in vitro diagnostic test for determining antimicrobial susceptibility of Streptococcus species. The device uses 64-well cards containing premeasured penicillin and culture media. Clinical laboratory technicians prepare a saline suspension of the bacterial isolate, which is used to rehydrate the card. The VITEK 2 system automatically fills, seals, and incubates the card; the VITEK 2 Compact system requires manual filling/sealing. The system monitors well growth over time. Output includes MIC values and interpretive category results. Results assist clinicians in selecting appropriate antibiotic therapy for patients with streptococcal infections.
Clinical Evidence
Performance evaluated via external study comparing VITEK 2 Streptococcus Penicillin to CLSI broth microdilution reference method. Testing included fresh clinical isolates, stock isolates, and challenge strains. Results showed high Essential Agreement (97.7%–100%) and Category Agreement (89.7%–100%) across tested Streptococcus species. Reproducibility and quality control were acceptable.
Technological Characteristics
Microdilution MIC technique; 64-well plastic test cards containing premeasured antibiotic and culture media; automated incubation and optical growth monitoring; compatible with VITEK 2 and VITEK 2 Compact systems; uses discriminant analysis algorithms.
Indications for Use
Indicated for antimicrobial susceptibility testing of Streptococcus species (S. pneumoniae, beta-hemolytic Streptococcus groups C and G, S. pyogenes, S. agalactiae, S. viridans group) in clinical laboratories. Quantitative test for penicillin susceptibility.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K232201
B Applicant
bioMérieux, Inc
C Proprietary and Established Names
VITEK 2 AST-Streptococcus Penicillin (≤0.06 - ≥8 μg/mL), VITEK 2 Streptococcus Penicillin (≤0.06 - ≥8 μg/mL), VITEK 2 Streptococcus Penicillin
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LON | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology |
| LTW | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
| LTT | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
To update VITEK 2 AST-Streptococcus Penicillin device labeling to include FDA-recognized oral penicillin (Penicillin V) breakpoints for *Streptococcus pneumoniae*. Performance data originally obtained and evaluated using parenteral (injection) penicillin breakpoints for *S. pneumoniae* (meningitis and non-meningitis) in K112000 was reanalyzed with oral penicillin breakpoints in this submission.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
Breakpoints and performance data for β-hemolytic Streptococci and Viridians species remain unchanged since the original clearance (K112000).
Previously obtained QC and reproducibility data are applicable to this reevaluation.
## B Measurand:
Penicillin concentrations of 0.06, 0.12, 0.5, and 2 µg/mL. The MIC result range for the VITEK 2 AST-Streptococcus Penicillin test is ≤0.06 - ≥8 µg/mL
## C Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test for penicillin.
## III Intended Use/Indications for Use:
### A Intended Use(s):
The VITEK 2 Streptococcus Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.
### Indication(s) for Use:
VITEK 2 Streptococcus Penicillin is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Streptococcus Penicillin is a quantitative test. Penicillin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active both in vitro and in clinical infections:
- Beta hemolytic streptococci groups C and G
- Streptococcus pyogenes
- Streptococcus agalactiae
- Streptococcus viridans group
- Streptococcus pneumoniae
The VITEK 2 Streptococcus Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.
### B Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
The ability of the AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available or an insufficient number were encountered at the
K232201 - Page 2 of 11
{3}
time of comparative testing: Benzylpenicillin (p01n): Streptococcus pneumoniae (When applying breakpoints of ≤2 (S), 4 (I), ≥8 (R)), Beta-Hemolytic Streptococcus
## C Special Instrument Requirements:
For use with the VITEK 2 and VITEK 2 Compact Systems. VITEK 2 and VITEK 2 Compact Systems (VITEK 2 Systems version 5.02 was used in K11200. New breakpoints will be incorporated into a future VITEK 2 Systems software release.)
## IV Device/System Characteristics:
### A Device Description:
The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well which contains only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically (or allow operator to manually) dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then, the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing, and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. The data are used to determine the minimum inhibitory concentration or "MIC" values for the antimicrobial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
The VITEK 2 AST-Streptococcus Penicillin has the following concentrations in the card: 0.06, 0.12, 0.5, and 2 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The MIC result range for the VITEK 2 AST-Streptococcus Penicillin test is ≤0.06 - ≥8 µg/mL. For all species, the MIC result range indicates that the VITEK 2 system is capable of producing the following MIC results ≤0.06, 0.125, 0.25, 0.5, 1, 2, 4 and ≥8 for AST-Streptococcus Penicillin test. This means the VITEK 2 systems does not provide results lower than 0.06µg/mL or greater than 8 µg/mL for the AST-Streptococcus Penicillin test.
### B Principle of Operation:
The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems use visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well determine organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.
K232201 - Page 3 of 11
{4}
V Substantial Equivalence Information:
A Predicate Device Name(s):
VITEK 2 Streptococcus Penicillin
B Predicate 510(k) Number(s):
K112000
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device K232201 | Predicate K112000 |
| --- | --- | --- |
| Device Trade Name | VITEK 2 AST-Streptococcus Penicillin (≤0.06 - ≥8 μg/mL) | VITEK 2 AST-ST Penicillin (≤0.06 - ≥8 μg/mL) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications for Use | VITEK 2 Streptococcus Penicillin is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Streptococcus Penicillin is a quantitative test.
Penicillin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active both in vitro and in clinical infections:
Beta hemolytic Streptococci groups C and G
Streptococcus pyogenes
Streptococcus agalactiae
Streptococcus viridans group
Streptococcus pneumoniae | Same |
| Test Methodology | Automated quantitative antimicrobial | Same |
K232201 - Page 4 of 11
{5}
| Device & Predicate Device(s): | Device K232201 | Predicate K112000 |
| --- | --- | --- |
| | susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of microorganisms | |
| Antimicrobial Agent | Penicillin | Same |
| Inoculum | Saline suspension of organism | Same |
| Test Card | Streptococcus (AST-ST) Susceptibility Card | Same |
| Analysis Algorithms | Discriminant Analysis | Same |
| Concentrations | 0.06, 0.12, 0.5, 2 | Same |
| General Device Characteristic Differences | | |
| Breakpoints (S/I/R) | • S. pneumoniae (non-meningitis) [≤2 / 4 / ≥8]
• S. pneumoniae (meningitis) [≤0.06/ - / ≥0.12]
• Viridans Streptococci [≤0.12/ 0.25-2 / ≥4]
• Beta-hemolytic Streptococcus [≤0.12/ - / -]
• S. pneumoniae (oral/non-meningitis) [≤0.06/ 0.12-1/ ≥2] | • S. pneumoniae (non-meningitis) [≤2 / 4 / ≥8]
• S. pneumoniae (meningitis) [≤0.06/ - / ≥0.12]
• Viridans Streptococci [≤0.12/ 0.25-2 / ≥4]
• Beta-hemolytic Streptococcus [≤0.12/ - / -]
• S. pneumoniae (oral) [none] |
VI Standards/Guidance Documents Referenced:
1. FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
2. CLSI M7-A5, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard - Fifth Edition", (January 2000)
3. CLSI M07, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard – Seventh Edition", (January 2006)
4. CLSI M07, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard - Eleventh Edition", (January 2018)
5. CLSI M100, "Performance Standards for Antimicrobial Susceptibility Testing; 33rd Edition", (March 2023)
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
K232201 - Page 5 of 11
{6}
1. Precision/Reproducibility:
Reproducibility testing for the VITEK 2 AST-Streptococcus Penicillin was previously evaluated during review of K112000 and was determined to be acceptable.
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Assay Reportable Range:
Not applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Quality Control (QC) testing, Inoculum and Growth failure for the VITEK 2 AST-Streptococcus Penicillin were performed in support of clearance of K112000 and were determined acceptable. QC testing specific to the additional oral penicillin breakpoints was not performed in the current submission since the penicillin QC range of $0.25 - 1\mu \mathrm{g / mL}$ used in K112000 was unchanged for S. pneumoniae ATCC 49619. The QC results with the unchanged QC range are shown in Table 1.
In summary, the QC organism recommended by both the FDA and CLSI, namely S. pneumoniae ATCC 49619 was tested using the reference broth microdilution (BMD) method as well as the VITEK 2 AST-Streptococcus Penicillin by the automatic and the manual dilution methods during the clinical study a minimum 20 times/site. QC results for the VITEK 2 AST-Streptococcus Penicillin were within the expected results range $>95\%$ of the time for both automatic and manual dilution modes of the VITEK 2 (Table 1), which is acceptable. In addition, a similar QC study conducted to evaluate the VITEK 2 Compact System demonstrated QC results within the expected results range $>95\%$ of the time, which is acceptable.
Table 1. Quality Control Result Frequencies for VITEK 2 (Auto-Dilution and Manual Dilution Methods)
| Organism | VITEK 2 Result Range (μg/mL) | BMD Result Range (μg/mL) | VITEK 2 Auto-Dilution | BMD | VITEK 2 Manual Dilution | BMD |
| --- | --- | --- | --- | --- | --- | --- |
| S. pneumoniae ATCC 49619 | | ≤0.03 | | | | |
| | ≤0.06 | 0.06 | | | | |
| | 0.12 | 0.12 | | | | |
| Expected Result 0.25 – 1 μg/mL | 0.25 | 0.25 | 22 | 79 | 12 | 78 |
| | 0.5 | 0.5 | 154 | 86 | 154 | 94 |
| | 1 | 1 | 2 | 3 | 9 | 3 |
| | 2 | 2 | | | | |
| | 4 | 4 | | | | |
K232201 - Page 6 of 11
{7}
| Organism | VITEK 2 Result Range (μg/mL) | BMD Result Range (μg/mL) | VITEK 2 Auto-Dilution | BMD | VITEK 2 Manual Dilution | BMD |
| --- | --- | --- | --- | --- | --- | --- |
| | ≥8 | 8 | | | | |
| | | 16 | | | | |
| | | ≥32 | | | | |
6. Detection Limit:
Not applicable
7. Assay Cut-Off:
Not applicable
B Comparison Studies:
1. Method Comparison with Predicate Device:
The VITEK 2 AST-ST Penicillin was originally cleared in premarket submission K112000 and included indications for Beta-hemolytic streptococci groups C and G, *Streptococcus pyogenes*, *Streptococcus agalactiae*, *Streptococcus viridans* group, and *Streptococcus pneumoniae* for testing with the VITEK 2 auto dilution, the VITEK 2 manual dilution, and the VITEK 2 Compact manual dilution. VITEK 2 AST-ST Penicillin results were compared to the CLSI broth microdilution reference method.
The performance of the VITEK 2 AST-Streptococcus Penicillin with *S. pneumoniae* using the current FDA-recognized oral penicillin breakpoints was evaluated by reanalyzing the data obtained for *S. pneumoniae* in support of K112000. *S. pneumoniae* performance when using both the oral penicillin breakpoints as well as the unchanged parenteral (injection) penicillin (meningitis and non-meningitis) breakpoints are summarized in this submission. Breakpoints for β-hemolytic Streptococci and Viridians species also remain unchanged; performance data are not shown here.
Since there was no change in the design or the penicillin dilution range of the VITEK 2 AST card, the performance evaluation of the VITEK 2 AST-Streptococcus Penicillin with *S. pneumoniae* for the oral penicillin breakpoints was achieved by reanalyzing the penicillin MIC data of the original 510(k) submission (K112000). For this review, the current penicillin (oral and parenteral) breakpoints are applied to *S. pneumoniae* according to the FDA STIC webpage.
A total of 350 (clinical and challenge) *S. pneumoniae* isolates were tested using the VITEK 2 with the auto dilution. *S. pneumoniae* results of the original data using the unchanged parenteral breakpoints and re-analyzed data using the oral breakpoints are summarized in Table 2.
The Essential Agreement (EA) and Categorical Agreement (CA) when using each of the parenteral breakpoints is >90%, which is acceptable, as described in the “Class II Special Control Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance
K232201 - Page 7 of 11
{8}
for Industry and FDA, August 2009" (AST guidance). When using the oral breakpoints, the EA is >90% and the CA is 89.1%, which is acceptable since the majority of errors is minor and the EA of evaluable results is very good, as described in the AST guidance. There were no major or very major errors observed with the oral and parenteral non-meningitis breakpoints.
When analyzing the performance data with the parenteral non-meningitis penicillin breakpoints, the Essential Agreement (EA) and Categorical Agreement (CA) is >90%, which is acceptable, as described in the "Class II Special Control Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, August 2009" (AST guidance). There were no major or very major errors observed.
When analyzing the performance data with the parental meningitis penicillin breakpoints, the EA and CA is >90%, which is acceptable, as described in the AST guidance. As previously reported in K112000, there were 6 major errors (3.2% major error rate) and 2 very major errors (1.2% very major error rate). The major error rate was considered acceptable given the very high overall CA agreement and lack of an intermediate meningitis interpretive criterion for penicillin. During this review, further analysis of the categorical errors was performed and adjustments made by considering the MIC values where the errors occurred. Five of the six major errors and one of the two very major errors was one doubling dilution from the reference and thus in essential agreement. Therefore, the adjusted major error rate is 0.5% (1/189) and the adjusted very major error rate is 0.6% (1/161), which is acceptable. To address the adjustment of the major error, the following performance footnote was included in the device labeling:
"The overall major error rate for penicillin when testing S. pneumoniae and using the meningitis breakpoints is 3.2% (6/189). Based on the essential agreement and lack of an intermediate meningitis breakpoint for penicillin, the overall adjusted major error rate for S. pneumoniae clinical and challenge isolates is 0.5% (1/189)."
When analyzing the performance data with the oral non-meningitis penicillin breakpoints, the EA is >90% and the CA is 89.1%, which is acceptable since the majority of errors are minor and the EA of evaluable results is very good, as described in the AST guidance. There were no major or very major errors observed.
Table 2. Analysis of the Original Performance for Penicillin with Streptococcus pneumoniae Using Parenteral Non-meningitis, Parenteral Meningitis and Oral Non-meningitis Breakpoints with the VITEK 2 (Auto Dilution)
| | Total | #EA | %EA | Eval EA Total | Eval EA # | Eval EA % | #CA | %CA | #R | #S | min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Streptococcus pneumoniae Parenteral Non-meningitis Breakpoints ≤2 (S), 4 (I), ≥8 (R) | | | | | | | | | | | | | |
| Clinical | 300 | 293 | 97.7 | 117 | 110 | 94.0 | 276 | 92.0 | 0 | 266 | 24 | 0 | 0 |
| Challenge | 50 | 49 | 98 | 49 | 48 | 98 | 47 | 94.0 | 0 | 47 | 3 | 0 | 0 |
| Combined | 350 | 342 | 97.7 | 166 | 158 | 95.2 | 323 | 92.3 | 0 | 303 | 27 | 0 | 0 |
| Streptococcus pneumoniae Parenteral Meningitis Breakpoints ≤0.06 (S), ≥0.12 (R) | | | | | | | | | | | | | |
| Clinical | 300 | 293 | 97.7 | 117 | 110 | 94.0 | 292 | 97.3 | 112 | 188 | N/A | 6 | 2 |
| Challenge | 50 | 49 | 98 | 49 | 48 | 98 | 50 | 100 | 49 | 1 | N/A | 0 | 0 |
| Combined | 350 | 342 | 97.7 | 166 | 158 | 95.2 | 342 | 97.7 | 161 | 189 | N/A | 6 | 2 |
| Streptococcus pneumoniae Oral Non-meningitis Breakpoints ≤0.06 (S), 0.12-1 (I), ≥2 (R) | | | | | | | | | | | | | |
K232201 - Page 8 of 11
{9}
| | Total | #EA | %EA | Eval EA Total | Eval EA # | Eval EA % | #CA | %CA | #R | #S | min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical | 300 | 293 | 97.7 | 117 | 110 | 94.0 | 276 | 92.0 | 61 | 188 | 24 | 0 | 0 |
| Challenge | 50 | 49 | 98 | 49 | 48 | 98 | 38 | 76 | 31 | 1 | 14 | 0 | 0 |
| Combined | 350 | 342 | 97.7 | 166 | 158 | 95.2 | 314 | 89.7 | 92 | 189 | 38 | 0 | 0 |
CA - Category Agreement
Eval - Evaluable isolates
R - Resistant isolates
S - Susceptible isolates
min - minor errors
maj - major errors
vmj - very major errors
Essential Agreement (EA) occurs when there is agreement between the MIC result of the reference method and that of the VITEK 2 test card within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the VITEK 2 test card and the reference method or those in which an off scale result is at least two doubling dilutions from the on scale result. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the VITEK 2 test card.
Performance of the VITEK 2 and the VITEK 2 Compact was also evaluated with the same 50 challenge organisms using the manual dilution method. Comparisons to the reference method are shown below in Table 3.
Table 3. Analysis of the Original Performance for Penicillin with Streptococcus pneumoniae Using Parenteral Non-meningitis, Parenteral Meningitis and Oral Non-meningitis Breakpoints with the VITEK 2 Systems (Manual Dilution)
| | Total | #EA | %EA | Eval EA Total | Eval EA # | Eval EA % | #CA | %CA | #R | #S | min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Streptococcus pneumoniae Parenteral Non-meningitis Breakpoints ≤2 (S), 4 (I), ≥8 (R) | | | | | | | | | | | | | |
| VITEK 2 | 50 | 48 | 96 | 48 | 46 | 95.8 | 46 | 92 | 0 | 40 | 4 | 0 | 0 |
| VITEK 2 Compact | 50 | 48 | 96 | 49 | 47 | 95.9 | 47 | 97 | 0 | 39 | 3 | 0 | 0 |
| Streptococcus pneumoniae Parenteral Meningitis Breakpoints ≤0.06 (S), ≥0.12 (R) | | | | | | | | | | | | | |
| VITEK 2 | 50 | 48 | 96 | 48 | 46 | 95.8 | 50 | 100 | 49 | 1 | N/A | 0 | 0 |
| VITEK 2 Compact | 50 | 48 | 96 | 49 | 47 | 95.9 | 50 | 100 | 49 | 1 | N/A | 0 | 0 |
| Streptococcus pneumoniae Oral Non-meningitis Breakpoints ≤0.06 (S), 0.12-1 (I), ≥2 (R) | | | | | | | | | | | | | |
| VITEK 2 | 50 | 48 | 96 | 48 | 46 | 95.8 | 38 | 76 | 31 | 1 | 12 | 0 | 0 |
| VITEK 2 Compact | 50 | 48 | 96 | 49 | 47 | 95.9 | 40 | 80 | 33 | 1 | 10 | 0 | 0 |
## Resistant Organisms
The ability of the AST card to detect resistance with the following combination(s) is unknown because resistant strains were either not available or an insufficient number were encountered at the time of comparative testing:
- Benzylpenicillin (p01n): Streptococcus pneumoniae (When applying breakpoints of ≤2 (S), 4 (I), ≥8 (R)), Beta-Hemolytic Streptococcus
## MIC Trending Analysis
A trending reanalysis was performed using the combined (clinical and challenge) data obtained in K112000 to align with current trending analyses. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower, at least one dilution higher or in
K232201 - Page 9 of 11
{10}
exact agreement with the CLSI reference method are not considered in the trending analysis. Organisms in which the difference between the percentage of isolates with higher vs. lower MIC values was ≥ 30% and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that showed higher or lower MIC values compared to the reference is addressed in the labeling. No trending was observed.
2. Matrix Comparison:
Not applicable
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
The FDA and CLSI susceptibility interpretive criteria for penicillin when tested against streptococci are listed in Table 4.
Table 4. FDA Recognized Interpretive Criteria for Penicillin a
| Pathogen | Minimum Inhibitory Concentrations (μg/mL) | | |
| --- | --- | --- | --- |
| | S | I | R |
| S. pneumoniae (oral, non-meningitis) | ≤0.06 | 0.12 - 1 | ≥2 |
| S. pneumoniae (non-meningitis) | ≤2 | 4 | ≥8 |
| S. pneumoniae (meningitis) | ≤0.06 | - | ≥0.12 |
| Viridans Streptococci | ≤0.12 | 0.25 - 2 | ≥4 |
| Beta hemolytic Streptococcus | ≤0.12 | - | - |
S = Susceptible; I = Intermediate; R = Resistant
a According to FDA STIC Website
VIII Proposed Labeling:
K232201 - Page 10 of 11
{11}
The labeling was updated to include:
- Addition of oral penicillin non-meningitis breakpoints for testing *S. pneumoniae*.
- Performance data using the unchanged FDA-recognized parenteral penicillin breakpoints for *S. pneumoniae* (meningitis and non-meningitis) and additional performance data using the FDA-recognized oral penicillin breakpoints
- Inclusion of a footnote to the performance table to describe the adjusted major error rate when using parental meningitis penicillin breakpoints.
The labeling supports the finding of substantial equivalence for this device when evaluated with the current FDA-recognized penicillin breakpoints for tests with *Streptococcus pneumoniae*.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage
(https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that bioMérieux intends to use to evaluate the VITEK 2 System with penicillin when revised breakpoints for penicillin are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, bioMérieux will update the penicillin device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K232201 - Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.