K111746 · Biomet, Inc. · KWT · Dec 15, 2011 · Orthopedic
Device Facts
Record ID
K111746
Device Name
COMPREHENSIVE SEGMENTAL REVISION SYSTEM (SRS)
Applicant
Biomet, Inc.
Product Code
KWT · Orthopedic
Decision Date
Dec 15, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Comprehensive® Segmental Revision System is intended for use with or without bone cement in the proximal shoulder. The Comprehensive® Segmental Revision System is intended for use with bone cement in distal humeral and total humeral applications. Tissue Attachment Augments provide the option for tissue stabilization and attachment.
Device Story
Comprehensive® Segmental Revision System (SRS) is a modular orthopedic prosthesis for proximal, distal, or total humeral replacement in cases of significant bone loss. System components include humeral heads, proximal bodies, intercalary segments, humeral stems, total humeral couplers, distal bodies with modular flanges, and tissue attachment augments. Device is implanted by orthopedic surgeons in a clinical/surgical setting to replace diseased or damaged bone/joint structures. Components are assembled to restore skeletal anatomy and function. Output is a reconstructed humeral structure providing joint stability and mobility. Benefits include restoration of function and structural support in complex revision or oncology cases.
Clinical Evidence
No clinical data submitted. Substantial equivalence determined via non-clinical bench testing, including engineering analysis of construct strength, range of motion, cantilever fatigue testing, cyclic loading, and MR compatibility testing per ASTM F2182-09.
Technological Characteristics
Modular humeral prosthesis constructed from Ti-6Al-4V and Co-Cr-Mo alloys. Features include porous proximal coating, taper head attachments, and modular tissue attachment augments. Dimensions: humeral stems (4-20mm diameter, 75-200mm length), proximal bodies (42-71mm resection), intercalary segments (19mm diameter, 30-120mm length), humeral heads (40-54mm diameter). MR compatibility tested per ASTM F2182-09.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, failed prior treatments, functional deformity, or oncology-related bone loss. Proximal/total humeral replacement indicated for acute/chronic fractures with humeral head involvement unmanageable by other methods. Distal/total humeral replacement indicated for acute/chronic fractures with humeral epicondyle involvement unmanageable by other methods.
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Discovery® - Mosaic® Total Humeral System (K043505)
Mosaic® Non-Modular Proximal Body and EAS Offset Modular Humeral Head (K042321)
Discovery® Elbow - Mosaic® Proximal Humeral Replacement System (K033280)
3-Piece Proximal Humeral Replacement System (K020045)
Short and Long Tissue Attachment Sleeves (K022079)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a blurry, black and white image of what appears to be Arabic text. The text is written in a calligraphic style, with the letters connected to each other. Due to the blurriness of the image, it is difficult to read the text. The background of the image is white.
K111746
DEC 1 5 2011
・・
re
Image /page/0/Picture/3 description: The image shows the logo for BIOMET MANUFACTURING CORP. The word "BIOMET" is in a stylized font with the letters connected. Below the logo, the words "MANUFACTURING CORP." are printed in a simple, sans-serif font. The text is black on a white background.
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| SUBMITTER INFORMATION | | |
|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Name | Biomet Manufacturing Corp. | |
| Address | 56 East Bell Drive<br>Warsaw, IN 46582 | |
| Phone number | (574) 267-6639 | |
| Fax number | (574) 371-1027 | |
| Establishment Registration Number | 1825034 | |
| Name of contact person | Patricia Sandborn Beres<br>Senior Regulatory Specialist | |
| Date prepared | December 5, 2011 | |
| DEVICE INFORMATION | | |
| Name of device | | |
| Trade or proprietary name | Comprehensive® Segmental Revision System (SRS) | |
| Common or usual name | Shoulder, elbow and total humeral replacement prosthesis | |
| Classification name | Shoulder joint metal/polymer non-constrained cemented prosthesis Shoulder joint metal/polymer semi-constrained cemented prosthesis Shoulder joint metal/polymer/metal non-constrained or semi-constrained porous coated uncemented prosthesis Elbow joint metal/polymer constrained cemented prosthesis | |
| Classification panel | Orthopedics | |
| Regulation | 21 CFR§888.3650/3660/3670/3150 | |
| Product Code(s) | KWT, KWS, MBF, JDC | |
| Legally marketed device(s) to<br>which equivalence is claimed | K043505- Discovery® - Mosaic® Total Humeral System K042321- Mosaic® Non-Modular Proximal Body and EAS Offset Modular Humeral Head K033280 -- Discovery® Elbow - Mosaic® Proximal Humeral Replacement System K020045 - 3-Piece Proximal Humeral Replacement System K022079 - Short and Long Tissue Attachment Sleeves K925613 - Proximal Humeral Replacement System K030710 - Bio-Modular® Shoulder System | |
| Reason for 510(k) submission | Update of existing product line | |
| Device description | The Comprehensive® SRS address the needs of the upper extremity, especially in cases where there is marked bone loss. Applications of the system include proximal humeral (shoulder) replacements distal humeral (elbow) replacements and total humeral replacements. Components of the system include humeral heads, proximal humeral bodies, intercalary segments, humeral stems, total humeral couplers, distal bodies with a modular flange, and modular tissue attachment augments. | |
| Indications for use | The Comprehensive® Segmental Revision System is intended for use<br>in cases of:<br>1. Non-inflammatory degenerative joint disease including<br>osteoarthritis and avascular necrosis.<br>2. Rheumatoid arthritis.<br>3. Revision where other devices or treatments have failed.<br>4. Correction of functional deformity.<br>5. Oncology applications including bone loss due to tumor resection.<br><br>When used in a proximal or total humeral replacement, the<br>Comprehensive® Segmental Revision System is also intended for:<br>Treatment of acute or chronic fractures with humeral head<br>(shoulder) involvement, which are unmanageable using other<br>treatment methods.<br><br>When used as a distal or total humeral replacement, the<br>Comprehensive® Segmental Revision System is also intended for:<br>Treatment of acute or chronic fractures with humeral epicondyle<br>(elbow) involvement, which are unmanageable using other<br>treatment methods. | |
| Intended use of the device | The Comprehensive® Segmental Revision System is intended for use<br>with or without bone cement in the proximal shoulder.<br><br>The Comprehensive® Segmental Revision System is intended for use<br>with bone cement in distal humeral and total humeral applications.<br><br>Tissue Attachment Augments provide the option for tissue<br>stabilization and attachment. | |
| SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE<br>COMPARED TO THE PREDICATE | | |
| Characteristic | Comprehensive® SRS | Predicate |
| Materials | Ti-6Al-4V<br>Co-Cr-Mo | K020045<br>K020045 |
| Principal of Operation | Proximal humeral, distal humeral<br>or total humeral replacement | K925613, K020045, K033280,<br>K043505 |
| Proximal Bodies | Resection Levels - 42-71mm | K020045 |
| Humeral Stems | Taper Head Attachment | K030710 |
| | Diameters - 4-20mm | K020045, K033280, K925613 |
| | Lengths - 75-200mm | K020045, K033280, K925613 |
| | Porous Coating Proximally | K020045, K033280, K030710 |
| Intercalary Segments and Couplers | Diameter - 19mm | K020045, K033280, K043505 |
| | Lengths - 30-120mm | K020045, K033280, K043505 |
| Tissue Attachment | Augments | K022079 |
| Humeral Heads | Diameters - 40-54mm | K030710 |
| | Extended Articular Surface | K042321, K030710 |
| Distal Humeral Bodies | Resection Height - 50, 60, 70mm | K033280, K043505 |
{2}------------------------------------------------
111746
p.2/3
510(k) Summary
Comprehensive® Segmental Revision System (SRS)
Page 2 of 3
{3}------------------------------------------------
K111746
\$\rho.^{3}/_{3}\$
510(k) Summary
Comprehensive® Segmental Revision System (SRS)
Page 3 of 3
-
| PERFORMANCE DATA |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary Of Non-Clinical Tests Conducted For Determination Of Substantial Equivalence |
| Engineering analysis to determine weakest point of the construct |
| Engineering analysis to determine range of motion |
| Engineering analysis to justify smaller diameter long stems |
| Cantilever fatigue testing to compare stem strength to predicate |
| Cyclic loading followed by screw torque out to confirm augment stability |
| Engineering analysis for flange loading determination |
| Cyclic fatigue testing of humeral flange |
| Static Axial Separation of SRS taper junction |
| Static Axial Separation of Comprehensive® taper junction |
| Shear testing to determine elbow condyle strength |
| MR Compatibility to ASTM F2182-09 |
| Summary Of Clinical Tests Conducted For Determination Of Substantial Equivalence And/Or Clinical Information |
| No clinical data submitted |
| CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA |
| No clinical data was necessary for a determination of substantial equivalence. |
| The results of analysis and mechanical testing indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices. |
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Biomet Manufacturing Corporation % Ms. Patricia Sandborn Beres Senior Regulatory Specialist 56 East Bell Drive Warsaw, Indiana 46582
DEC 1 5 2011
Re: K111746
Trade/Device Name: Comprehensive® Segmental Revision System (SRS) Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWT, KWS, MBF, JDC Dated: December 12, 2011 Received: December 13, 2011
Dear Ms. Beres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
## Page 2 - Ms. Patricia Sandborn Beres
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincercly yours,
Eunel Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: _Comprehensive® Segmental Revision System (SRS)
İndications For Use:
The Comprehensive® Segmental Revision System is intended for use in cases of:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
. ")
- 3. Revision where other devices or treatments have failed.
- 4. Correction of functional deformity.
- 5. Oncology applications including bone loss due to tumor resection.
When used in a proximal or total humeral replacement, the Comprehensive® Segmental Revision System is also intended for:
Treatment of acute or chronic fractures with humeral head (shoulder) involvement, which are unmanageable using other treatment methods.
When used as a distal or total humeral replacement, the Comprehensive® Segmental Revision System is also intended for:
Treatment of acute or chronic fractures with humeral epicondyle (elbow) involvement, which are unmanageable using other treatment methods.
The Comprehensive® Segmental Revision System is intended for use with or without bone cement in the proximal shoulder.
The Comprehensive® Segmental Revision System is intended for use with bone cement in distal humeral and total humeral applications.
Tissue Attachment Augments provide the option for tissue stabilization and attachment.
AND/OR
Prescription Use __ X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use __ NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K111746
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.