K030710 · Biomet, Inc. · KWT · Jun 4, 2003 · Orthopedic
Device Facts
Record ID
K030710
Device Name
BIO-MODULAR SHOULDER SYSTEM
Applicant
Biomet, Inc.
Product Code
KWT · Orthopedic
Decision Date
Jun 4, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The components of Bio-Modular® Shoulder System included in this submission are intended for total shoulder joint arthroplasty. Indications for use include: 1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis 2) Rheumatoid arthritis 3) Revision where other devices or treatments have failed 4) Correction of functional deformity 5) Fractures of the proximal humerus, where other methods of treatment are deemed inadequate 6) Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate Devices with surface coatings are indicated for cemented or uncemented biological fixation application. Polyethylene glenoid components not attached to a metal back are indicated for cemented application only.
Device Story
Bio-Modular Shoulder System is a modular total shoulder replacement prosthesis. Components include humeral stem, modular humeral head, and glenoid component (metal-backed or all-polyethylene). Humeral stem features biplaner taper for stress transfer, lateral fin for rotational stability, and collar to minimize subsidence. Proximal stems are circumferentially porous coated for biological fixation. Reverse Morse taper allows attachment of modular, offset, or bi-polar humeral heads. Metal-backed glenoid components use central peg and optional screws for fixation; all-polyethylene components are for cemented use. Device is implanted by orthopedic surgeons in a clinical/OR setting to replace the shoulder joint, restore function, and alleviate pain from degenerative disease, arthritis, or fractures.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing demonstrating mechanical performance under expected clinical conditions and characterization of the porous surface.
Technological Characteristics
Modular shoulder prosthesis consisting of humeral stem, modular heads, and glenoid components. Materials include metal and polymer. Features include reverse Morse taper, biplaner taper, lateral fin, and circumferential porous coating for biological fixation. Cemented or uncemented application depending on component type. No software or electronic components.
Indications for Use
Indicated for patients requiring total shoulder joint arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, revision of failed prior treatments, functional deformity, proximal humerus fractures, or complex conditions like cuff arthropathy.
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
K030710
p. 1/2
JUN - 4 2003
Image /page/0/Picture/3 description: The image shows the logo for Biomet Orthopedics, Inc. The word "BIOMET" is in a bold, sans-serif font, with the letters connected. Below the word "BIOMET" is the phrase "ORTHOPEDICS, INC." in a smaller, serif font.
## Summary of Safety and Effectiveness
| Applicant or Sponsor: | Biomet Orthopedics, Inc.<br>P.O. Box 587<br>Warsaw, Indiana 46581-0588 |
|-----------------------|---------------------------------------------------------------------------------------------------------|
| Contact Person: | Patricia Sandborn Beres<br>Senior Regulatory Specialist<br>Phone: (574) 267-6639<br>FAX: (574) 372-1683 |
Trade Name: Bio-Modular® Shoulder System
Common Name: Total shoulder replacement components
#### Classification Name:
- 1. Prosthesis, Shoulder, Non-constrained, Metal/Polymer, Cemented (21 CFR Section 888.3650)
- 2. Prosthesis, Shoulder, Semi-constrained, Metal/Polymer, Cemented (21 CFR Section 888.3660)
- 3. Prosthesis, Shoulder, Semi-constrained, Metal/Polymer, Uncemented (21 CFR Section 888,3670)
- 4. Prosthesis, Shoulder, Homi-, Humeral, Metallic, Cemented or Uncemented (21 CFR Section 888.3690)
## Legally Marketed Device to which Substantial Equivalence is claimed:
- The Bio-Modular® Shoulder System cleared by 510(k)s K872454, K915596, ● K992119, K992899 and K002998.
- The Gobal® Advantage Shoulder System cleared by 510(k)s K914695, ● K943300, K974044, K981487 and K011047.
Device Description: The Bio-Modular® Shoulder System consists of a humeral stem that is used in conjunction with a modular head and a metal backed or all polyethylene glenoid component. The Humeral Component is available in various lengths and porous coating levels. The Bio-Modular® Humeral stern features a biplaner taper to promote stress transfer down the length of the stem. Each stem has a lateral fin to provide rotational stability. Holes in the fin provide the surgeon with the option of suture attachment during fracture reconstruction. A collar minimizes subsidence. The stems are circumferentially porous coated proximally to provide fixation by tissue ingrowth.
| MAILING ADDRESS | P.O. Box 587<br>Warsaw, IN 46581-0587 |
|-----------------|---------------------------------------|
|-----------------|---------------------------------------|
| SHIPPING ADDRESS | Airport Industrial Park<br>Warsaw, IN 46580 |
|------------------|---------------------------------------------|
|------------------|---------------------------------------------|
| OFFICE | 219.267.6639 |
|--------|-------------------|
| FAX | 219.267.8137 |
| E-MAIL | biomet@biomet.com |
{1}------------------------------------------------
p. 2/2
The proximal portions of the humeral stems feature a reverse Morse taper for attachment of the modular humeral head. Offset and Bi-Polar humeral heads are also available.
Modular, metal backed glenoid components are available for total shoulder replacement. These devices feature a central peg and the option of using screws for immediate fixation and porous coating to provide biological fixation when used without bone cement. All polyethylene glenoid components are available for cemented application.
Intended Use: The components of Bio-Modular® Shoulder System included in this submission are intended for total shoulder joint arthroplasty. Indications for use include:
- 1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- 2) Rheumatoid arthritis
- 3) Revision where other devices or treatments have failed
- 4) Correction of functional deformity
- 5) Fractures of the proximal humerus, where other methods of treatment are deemed inadequate
- 6) Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate
Devices with surface coatings are indicated for cemented or uncemented biological fixation application. Polyethylene glenoid components not attached to a metal back are indicated for cemented application only.
Summary of Technologles: The materials, surface finishes and processing of the Bio-Modular® Shoulder System are similar to the predicate device.
Non-Clinical Testing: Mechanical testing has demonstrated the device's ability to perform under expected clinical conditions. A full characterization of the porous surface has been provided.
Clinical Testing: None provided.
All trademarks are property of Biomet, Inc. except for the following: Gobal is a trademark of DePuv. Inc.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 4 2003
Ms. Patricia Sandborn Beres Senior Regulatory Specialist Biomet Orthopedics, Inc. 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587
Re: K030710
Trade/Device Name: Bio-Modular® Shoulder System Regulation Number: 21 CFR 888.3650, 888.3660, 888.3670, 888.3690 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis; Shoulder joint metal/polymer semi-constrained cemented prosthesis; Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis; and Shoulder joint humeral (hemishoulder) metallic uncemented prosthesis Regulatory Class: II
Product Code: KWT, KWS, MBF, and HSD Dated: March 4, 2003 Received: March 6, 2003
Dear Ms. Beres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
## Page 2 - Ms. Patricia Sandborn Beres
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
L. Merle A. Millican
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
# 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
#### Device Name: Bio-Modular® Shoulder System
#### Indications For Use:
- 1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- 2) Rheumatoid arthritis
- 3) Revision where other devices or treatments have failed
- 4) Correction of functional deformity
- 5) Fractures of the proximal humerus, where other methods of treatment are deemed inadequate
- 6) Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate
Devices with surface coatings are indicated for cemented or uncemented biological fixation application. Polyethylene glenoid components not attached to a metal back are indicated for cemented application only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96) Division of General, Restorative
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.