CADD SOLIS VIP AMBULATORY INFUSION PUMP AND CADD SOLIS MEDICATION SAFETY SOFTWARE
K111275 · Smiths Medical Asd, Inc. · FRN · Feb 1, 2013 · General Hospital
Device Facts
Record ID
K111275
Device Name
CADD SOLIS VIP AMBULATORY INFUSION PUMP AND CADD SOLIS MEDICATION SAFETY SOFTWARE
Applicant
Smiths Medical Asd, Inc.
Product Code
FRN · General Hospital
Decision Date
Feb 1, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CADD®-Solis VIP Ambulatory Infusion Pump is indicated for intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion. PCA (patient-controlled analgesia) delivery is used for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both, such as patientcontrolled analgesia. Continuous delivery allows the infusion of drug/fluid at a constant, programmed rate. Intermittent delivery allows the infusion of a specific volume of drug/fluid at a regular, programmed interval. Step delivery allows an incremental increase in infusion rate to a specified maximum infusion rate for a specified total infusion volume. Taper delivery allows a plateau rate of infusion with the option of tapering at the beginning and/or end and has a programmable KVO rate at the end of the infusion. The CADD™ Solis Medication Safety Software – Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD -Solis VIP Ambulatory Infusion Pump, CADD®-Solis Ambulatory Infusion Pump or CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher). The CADD™ Solis Medication Safety Software-Point of Care allows use of a computer to send therapy-based protocols developed by the CADD™-Solis Medication Safety Software-Administrator to the CADD® Solis Ambulatory Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher).
Device Story
Ambulatory infusion pump with linear peristaltic mechanism; microprocessor-controlled. Inputs: user-programmed therapy parameters via color LCD/keypad. Outputs: controlled drug/fluid delivery. Used in hospital, homecare, and outpatient settings by clinicians or patients. Software modules (Administrator/POC) manage therapy-based protocol libraries; allow protocol creation, editing, and download to pumps via USB. Enhances medication safety by enforcing programmed limits and protocols. Benefits: precise, programmable drug delivery across multiple clinical routes; reduces medication errors via standardized protocols.
Clinical Evidence
No clinical data; bench testing only. Verification and validation included electrical/mechanical safety (IEC 60601-1, IEC 60601-2-24), software validation, usability/human factors, and biocompatibility testing. All testing met pre-established specifications.
Technological Characteristics
Linear peristaltic pump; microprocessor-controlled. Powered by 4 AA batteries, AC adapter, or rechargeable battery pack. Features color LCD (320x320 pixels), USB connectivity, and real-time clock. Software-based protocol management. Materials are standard medical-grade housing/components. Sterilization not specified; infusion disposables are biocompatible.
Indications for Use
Indicated for patients requiring intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural, or subarachnoid infusion therapy. Supports continuous, intermittent, step, taper, and patient-controlled analgesia (PCA) delivery modes.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
# KIII 275
#### 510(k) Summary
# FEB 0 1 2013
| DATE PREPARED | January 29, 2013 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| I. GENERAL INFORMATION | |
| Applicant's Name and<br>Address: | Smiths Medical ASD, Inc.<br>1265 Grey Fox Road<br>St. Paul, MN 55112 |
| Contact Person: | Julie Tapper<br>Senior Regulatory Affairs Specialist |
| Common/Usual Name: | Ambulatory Infusion Pump<br>Pump Communications System |
| Proprietary Name: | CADD®-Solis VIP Ambulatory Infusion Pump<br>CADD™-Solis Medication Safety Software |
| Predicate Devices: | K072144, CADD®-Solis Ambulatory Infusion Pump<br>K960826, CADD-Prizm® VIP Ambulatory Infusion Pump<br>K082783, CADD™-Solis Medication Safety Software-Point of Care<br>K072144, CADD™-Solis Medication Safety Software-Administrator |
# II. DEVICE DESCRIPTION
#### CADD®-Solis VIP Ambulatory Infusion Pump
The CADD®-Solis VIP Ambulatory Infusion Pump ("Solis VIP Pump") has a microprocessor and linear peristaltic pumping mechanism, similar in design to the Smiths Medical ASD, Inc. CADD®-Solis Ambulatory Infusion Pump (K072144) and CADD-Prizm® VIP Ambulatory Infusion Pump (K960826). The user activates the Solis VIP Pump via a color LCD screen and keypad user interface. Commands are issued to the microprocessor by activating the user interface. Microprocessor actions are controlled by a program, which is contained in the pump's memory.
The Solis VIP Pump consists of components, such as the user interface, sensors, communication ports, power ports, structural (housing) components, electronics, pumping mechanism, watchdog timer, pump battery and circuitry, real time clock, on-board memory, and pump log. The Solis VIP Pump exterior surface components include the pump housing, LCD lens, labels, and keypad, and the materials of construction for these components are widely used in the medical industry. The Solis VIP Pump is designed to be used with an infusion disposable, such as medication cassette reservoir.
The Solis VIP Pump provides measured drug therapy to patients in the hospital, homecare, and outpatient settings. The Solis VIP Pump is intended for therapies that require a Continuous. Intermittent. Tapering, or Step rates of infusion, and Patient Controlled Analgesia (including Clinician Bolus).
{1}------------------------------------------------
## CADDTM-Solis Medication Safety Software-Administrator
The CADD™-Solis Medication Safety Software-Administrator ("Administrator") module allows the user to create, edit, and save therapy-based protocols and pump settings within user-defined protocol libraries. The system administrator user determines user access and library editing capabilities. Other Administrator module features include barcode printing, reports, and sending and receiving pump protocol information.
# CADD™-Solis Medication Safety Software-Point of Care
The CADD™-Solis Medication Safety Software-Point of Care ("POC") module allows the user to download therapy based protocols to the Solis Pump, and Prizm Pump (software revision H or higher) and send and receive pump settings via USB connection.
# III. INTENDED USE OF THE DEVICE
# CADD®-Solis VIP Ambulatory Infusion Pump
The CADD®-Solis VIP Ambulatory Infusion Pump is indicated for intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
PCA (patient-controlled analgesia) delivery is used for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both, such as patientcontrolled analgesia.
Continuous delivery allows the infusion of drug/fluid at a constant, programmed rate.
Intermittent delivery allows the infusion of a specific volume of drug/fluid at a regular, programmed interval.
Step delivery allows an incremental increase in infusion rate to a specified maximum infusion rate for a specified total infusion volume.
Taper delivery allows a plateau rate of infusion with the option of tapering at the beginning and/or end and has a programmable KVO rate at the end of the infusion.
#### CADD™-Solis Medication Safety Software-Administrator
The CADD™ Solis Medication Safety Software – Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD -Solis VIP Ambulatory Infusion Pump, CADD®-Solis Ambulatory Infusion Pump or CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher).
## CADD"-Solis Medication Safety Software-Point of Care:
The CADDIM Solis Medication Safety Software-Point of Care allows use of a computer to send therapy-based protocols developed by the CADD™-Solis Medication Safety Software-Administrator to the CADD® Solis Ambulatory
{2}------------------------------------------------
Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher).
## VI. SUMMARY OF STUDIES
#### Performance Testing
Smiths Medical performed extensive verification and validation testing on the Solis VIP Pump and Administrator and POC software.
Testing was completed in accordance with FDA's draft guidance, Total Product Life Cycle-Infusion Pump - Premarket Notification [510(k)] Submissions, issued on April 23, 2010; and the General Principles of Software Validation; Final Guidance for Industry and FDA Staff, issued on January 11, 2002.
Also, testing was completed in accordance with FDA consensus standards, IEC 60601-1:1988 Medical electrical equipment - Part 1: General reguirements for safety, plus Amendment 1:1991. Amendment 2:1995 and Amendment 2 Corrigendum: 2005, and IEC 60601-2-24:1998 - Medical electrical equipment - Part 2-24: Particular requirements for the safety of infusion pumps and controllers.
An overall summary of the performance testing was, as follows:
| Electrical and mechanical safety tests | Electromagnetic compatibility tests |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Solis VIP Pump software validation | Administrator and POC software validation |
| Usability/human factors validations for<br>the Solis VIP Pump and Administrator<br>and POC Software | Infusion disposable biocompatibility<br>testing, including extractables and<br>leachables |
#### Clinical Studies
Human clinical studies were deemed unnecessary to evaluate the safety or effectiveness of the Solis VIP Pump, and Administrator and POC software.
# Testing Conclusion
All testing met pre-established specifications. and successfully demonstrated that the devices performed as intended. The testing results allowed for a conclusion to be made that the Solis VIP Pump, and Administrator and POC software were as safe and effective as the predicate devices.
#### VII. STATEMENT OF EQUIVALENCE
The Solis VIP Pump, and Administrator and POC software are substantially equivalent to the predicate devices, based on comparisons of the device classifications. intended use, and technological characteristics. Verification and validation tests confirmed the suitability of the devices for their intended uses. The test results did not raise new safety or performance questions, and confirmed that the Solis VIP Pump, and Administrator and POC software devices are substantially equivalent to the predicate devices.
{3}------------------------------------------------
| ADD®-Solis VIP Ambulatory Infusion Pu | | | |
|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| evice name | CADD®- | ADD®-Solis Ambulatory Infusio Pump | CADD-Prizm® VIP Ambulator Smiths Mesical ASS |
| | | | |
| | | Smiths Medical ASD, K072144 (March 7, 2008) | |
| | | | |
| | | | |
| | | | |
| Tanufacturer Notification numbr Substantial equivalence Substantial Eguivalence Subscription EDATURES Pump Lym FEAT UREST Display | - Linear Peristaltic display with 320 x 320 pixe<br>Color | Linear Peristaltic olor display with 320 x 320 pixel | 32<br>Linear Peristalic Tonchromatics displayatriis Pixol, Dixplandisplanding Mariity Program Pixel, Mistrials EPS Portalian Provinsian Provinsia EPS Power Power Power Provinsian P |
| | | | |
| ower requiremen | | | |
| | 4AA Batteries AC Adapter echargeable Battery Pac<br>d | 4AA Batteries | |
| | No | No | Yes |
| | No | No | No |
| rinting capabilities Remote communications capabilities USB port PROGRAMMING FUNCTIONS | Yes | Yes | No |
| | | | |
| Step | Yes | No | No |
| ntermitten | Yes | No | Yes |
| aper | Yes | No | Yes (TPN |
| | Yes | Y es | Yes |
| ontinuous infusion ontinuous infusion with bolus (PC) | Yes | ਿੰਡ ਨੇੜੇ ਵਿੱਚ ਮੈਡ | Yes |
| ecurity | Yes | | Yes |
| | Yes | Y es | Y es |
| | Yes | Y es | Yes |
| | Yes | Yes | Yes |
| | Y es | Yes | Yes |
| | Yes | No | Yes |
| | ﺮ ،<br>Yes | Yes | Yes |
| | Yes | Yes | Y es |
| | | | |
| | Yes | Yes | Yes |
| | Yes | Yes | Yes |
| | Yes | Yes | Yes |
| | | | |
| CADD®-Solis VIP Ambulatory Infusion Pump | | | |
| Device name | CADD®-Solis VIP Ambulatory<br>Infusion Pump | CADD®-Solis Ambulatory Infusion<br>Pump | CADD-Prizm® VIP Ambulatory<br>Infusion Pump |
| No-battery alert | Yes | Yes | Yes |
| Pump in stop mode | Yes | Yes | Yes |
| High pressure | Yes | Yes | Yes |
| Power up fault | Yes | Yes | Yes |
| Low volume in medication reservoir | Yes | Yes | Yes |
| Cassette detachment | Yes | Yes | Yes |
| Upstream occlusion | Yes | Yes | Yes |
| Downstream occlusion | Yes | Yes | Yes |
| Air-in-line | Yes | Yes | Yes |
| Key stuck | Yes | Yes | Yes |
| <b>INFUSION SPECIFICATIONS</b> | | | |
| Minimum continuous delivery rate | 0 mL/hr | 0 mL/hr | 0 mL/hr |
| Maximum continuous delivery rate | 100 mL/hr in PCA<br>500 mL/hr in Continuous,<br>Intermittent, Taper, Step | 30 mL/hr | 30 mL/hr in PCA<br>350 mL/hr in Continuous,<br>Intermittent, TPN |
| Maximum patient bolus | 50 mL | 20 mL | 9.9 mL |
| Maximum clinician bolus | 50 mL | 20 mL | 20 mL |
| Reservoir volume | 1 to 9999 mL | 1 to 9999 mL | 1 to 9999 mL |
| Patient controlled access PCA (dosing) | Yes | Yes | Yes |
| Dose lockout time | Yes | Yes | Yes |
| Doses per hour limit | Yes | Yes | Yes |
| Delivery limit | Yes | Yes | Yes |
| Clinician bolus | Yes | Yes | Yes |
| Programmable maximum delivery rate<br>(Continuous rate + bolus) | Yes | Yes | Yes |
# COMPARISION OF TECHNOLOGICAL CHARACTERISTICS v.
.
Page 4 of 6
-
.
K111275 510(k) Summar
Smiths Medical ASD, Inc
{4}------------------------------------------------
:
:
·
1
: :
:
·
: 上一篇:
K111275 510(k) Summary
.
Smiths Medical ASD, Inc.
.
.
Page 5 of 6
: 上一
{5}------------------------------------------------
| CADD TM-Solis Medication Safety<br>Software Administrator and POC<br>Device Name | CADDTM-Solis Medication Safety<br>Software (software update) | CADDTM-Solis Medication Safety<br>Software | |
|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Manufacturer | Smiths Medical ASD, Inc. | Smiths Medical ASD, Inc. | |
| 10(k) notification number<br>(substantial equivalence date) | K111275<br>(TBD) | K082783<br>(December 17, 2008) | K072144<br>(March 7, 2008) |
| <b>COMPATIBILITY</b> | | | |
| Pump compatibility | CADD®-Solis VIP Ambulatory<br>Infusion Pump, CADD®-Solis<br>Ambulatory Infusion Pump and<br>CADD-Prizm® PCS II 6101 (rev H<br>and higher) | CADD®-Solis Ambulatory Infusion<br>Pump and CADD-Prizm® PCS II 6101<br>(rev H and higher) | CADD®-Solis Ambulatory Infusion<br>Pump and CADD-Prizm® PCS II 6101<br>(rev H and higher) |
| Accessory compatibility | USB cable for the CADD®-Solis VIP<br>pump, and CADD®-Solis pump<br>21-6144 Interface cable/null modem<br>cable for CADD-Prizm® pump | USB cable for the CADD®-Solis<br>pump<br>21-6144 Interface cable/null modem<br>cable for CADD-Prizm® pump | USB cable for the CADD®-Solis<br>pump<br>21-6144 Interface cable/null modem<br>cable for CADD-Prizm® pump |
| PC software compatibility<br>Computer equipment | Windows 2000, XP, VISTA<br>RS232 serial port, USB, and CD-<br>ROM | Windows 2000, XP, VISTA<br>RS232 serial port, USB, and CD-<br>ROM | Windows 2000, XP, VISTA<br>RS232 serial port, USB, and CD-<br>ROM |
KI11275 510(k) Summary
Smiths Medical ASD, Inc
Page 6 of 6
:
. .
.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### Febauary 1, 2013
Ms. Julie Tapper Senior Regulatory Affairs Specialist Smiths Medical ASD, Incorporated 1265 Grey Fox Road ST. PAUL, MN 55112
Re: K111275
Trade/Device Name: CADD® -Solis VIP Ambulatory Infusion Pump and CADD™-Solis Medication Safety Software Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: January 21, 2013 Received: January 22, 2013
Dear Ms. Tapper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Ms. Tapper
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours;
hor
Theth
Harshita.D.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and
Dental Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# SMITHS MEDICAL ASD, INC. 510(k) Premarket Notification
CADD®-Solis VIP Ambulatory Infusion Pump CADD"-Solis Medication Safety Software
#### Indications for Use --
510(k) Number: K 111275
TAB 5 - Page 1 of 2
1 389
# Device Name: CADD®-Solis VIP Ambulatory Infusion Pump
Indications for Use:
"The CADD®-Solis VIP Ambulatory Infusion Pump is indicated for intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
PCA (patient-controlled analgesia) delivery is used for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both, such as patientcontrolled analgesia.
Continuous delivery allows the infusion of drug/fluid at a constant, programmed rate.
Intermittent delivery allows the infusion of a specific volume of drug/fluid at a regular, programmed interval.
Step delivery allows an incremental increase in infusion rate to a specified maximum infusion rate for a specified total infusion volume.
Taper delivery allows a plateau rate of infusion with the option of tapering at the beginning and/or end and has a programmable KVO rate at the end of the infusion."
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The Counter Use _ OR Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by Richard C. Chapman Date: 2013.02.01 13:20:30 -05'00'
#### (Division Sign-Off)
: vision of Anesthesiology, General Hespital rection Control, Dental Devices
510(k) Number:
{9}------------------------------------------------
# SMITHS MEDICAL ASD, INC. 510(k) Premarket Notification CADD -Solis VIP Ambulatory Infusion Pump CADD™-Solis Medication Safety Software
## Indications for Use --
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
#### Device Name: CADD" Solis Medication Safety Software - Administrator
Indications for Use:
"The CADD" Solis Medication Safety Software - Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD -Solis VIP Ambulatory Infusion Pump, CADD -Solis Ambulatory Infusion Pump or CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher)."
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The Counter Use _________________________________________________________________________________________________________________________________________________________ OR Per 21 CFR 801.109
# Device Name: CADD" Solis Medication Safety Software - Point of Care
Indications for Use:
"The CADD" Solis Medication Safety Software -- Point of Care allows use of a computer to send therapy-based protocols developed by the CADD"-Solis Medication Safety Software - Administrator to the CADD® Solis Ambulatory Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher)."
Prescription Use _____________________________________________________________________________________________________________________________________________________________ OR Over-The Counter Use _________________________________________________________________________________________________________________________________________________________ Per 21 CFR 801.109
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by Richard C. Chapman Date: 2013.01.30 10:21:19 -05'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _
TAB 5 - Page 2 of 2
1390
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.