The CADD® Solis ambulatory infusion pump is indicated for intravenous, intraarterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion, patient controlled PCA doses, or both. The CADD®-Solis Medication Safety Software – Administrator allows you to establish a therapy-based protocol library that will be used by the CADD®-Solis Ambulatory Infusion Pump or by the CADD® -Solis Medication Safety Software -Point of Care to program the CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher).
Device Story
CADD-Solis Ambulatory Infusion Pump delivers measured drug therapy (continuous, PCA, clinician bolus) in hospital settings. Solis Medication Safety Software-Administrator is a server-based application for creating therapy-based protocol libraries. Clinicians use the software to define protocols, which are then loaded into the pump or the Point of Care (POC) software to program CADD-Solis or CADD-Prizm PCS II pumps. System prevents duplicative drug, protocol, or user entries. Output allows clinicians to standardize medication delivery, reducing manual programming errors and enhancing patient safety during infusion therapy.
Clinical Evidence
No clinical data. Safety and effectiveness established through bench-level verification and validation testing.
Technological Characteristics
Ambulatory infusion pump for various delivery routes. Server-based medication safety software for protocol management. Connectivity via protocol library transfer to pumps. Software-based protocol enforcement.
Indications for Use
Indicated for patients requiring intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, intraoperative, epidural, or subarachnoid infusion therapy. Supports continuous infusion, patient-controlled demand doses (PCA), or both. Prescription use only.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
CADD-Prizm® PCS II Model 6101 Ambulatory Infusion System
CADD-Sentry Pro® Medication Safety
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary
# MAR - 7 ZUUn
## GENERAL INFORMATION
| Applicant's Name and Address: | Smiths Medical MD, Inc.<br>1265 Grey Fox Road<br>St. Paul, MN 55112 |
|--------------------------------|---------------------------------------------------------------------------------------------------|
| Contact Person: | Phil Neururer<br>Sr. Regulatory Affairs Specialist |
| Common/Usual Name: | Ambulatory Infusion Pump<br>Pump Communications System |
| Proprietary Name: | CADD®-Solis Ambulatory Infusion Pump<br>CADD®-Solis Medication Safety Software |
| Equivalence Device Comparison: | CADD-Prizm® PCS II Model 6101 Ambulatory Infusion<br>System<br>CADD-Sentry Pro® Medication Safety |
#### II. DEVICE DESCRIPTION
### CADD®-Solis Ambulatory Infusion Pump
The CADD® -Solis ambulatory infusion pump ("Solis Pump") provides measured drug therapy to patients in the hospital setting. The pump can be programmed to deliver medication at a continuous rate, patient controlled analgesia (PCA), clinician bolus, continuous rate plus PCA, and continuous rate plus clinician bolus.
### CADD®-Solis Medication Safety Software – Administrator
The Smiths Medical MD, Inc. CADD® -Solis Medication Safety Software -Administrator ("Solis Safety Software - Administrator"), a server based software program that operates on commercially available computers, is designed to establish user defined therapy based protocols for use by the Solis Pump or for use by use CADD®-Solis Medication Safety Software – Point of Care (POC) to program the CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher). The Solis Safety Software - Administrator does not allow duplicative Drug, Protocol or User identification entries.
072144
{1}------------------------------------------------
#### INTENDED USE OF THE DEVICE III.
## CADD®-Solis Ambulatory Infusion Pump
The CADD® Solis ambulatory infusion pump is indicated for intravenous, intraarterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion, patient controlled PCA doses, or both.
## CADD®-Solis Medication Safety Software – Administrator
The CADD®-Solis Medication Safety Software – Administrator allows you to establish a therapy-based protocol library that will be used by the CADD®-Solis Ambulatory Infusion Pump or by the CADD® -Solis Medication Safety Software -Point of Care to program the CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher).
#### SUMMARY OF STUDIES V.
## A. Functional Testing
The Solis Pump and Solis Safety Software – Administrator were subjected to verification and validation testing. The verification and validation testing performed demonstrate the devices function as intended and perform to specification.
### B. Clinical Studies
Human clinical studies were deemed not necessary to evaluate the safety or effectiveness of the CADD-Sentry Pro™ Medication Safety Software.
### C. Conclusions Drawn from the Studies
Based upon the information provided, the CADD-Sentry Pro™ Medication Safety Software is safe, effective and performs to established specifications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAR - 7 2008
Mr. Phil Neururer Senior Regulatory Affairs Specialist Smiths Medical MD, Incorporated 1265 Grey Fox Road Saint Paul, Minnesota 55112
Re: K072144
Trade/Device Name: CADD® -Solis Ambulatory Infusion Pump CADD® - Solis Medication Safety Software - Administrator Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEA Dated: February 20, 2008 Received: February 21, 2008
Dear Mr. Neururer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Mr. Neururer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chih Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# SMITHS MEDICAL MD, INC. 510(k) Premarket Notification
CADD®-Solis Ambulatory Infusion Pump CADD®-Solis Medication Safety Software
## Indications for Use
510(k) Number: K072144
## Device Name: CADD®-Solis Ambulatory Infusion Pump
Indications for Use:
"The CADD -Solis ambulatory infusion pump is indicated for intravenous, intraarterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both (such as patientcontrolled analgesia)."
Prescription Use
Over-The Counter Use ---------
Per 21 CFR 801.109)
## Device Name: CADD®-Solis Medication Safety Software - Administrator
OR
Indications for Use:
"The CADD®-Solis Medication Safety Software consists of the Administrator and the Point of Care software applications. This software allows you to create a set of standard pump Protocols to be used with the CADD®-Solis Ambulatory Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump. The Point of Care software application is used for the CADD-Prizm® PCS II Ambulatory Infusion pump (with software revision H or higher) only."
Prescription Use _____________________________________________________________________________________________________________________________________________________________
Over-The Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony D. Morton
OR
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K\$\phi\$7a,44
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.