DOPPLER MONITOR

K110347 · Cook Vascular, Inc. · DPW · Mar 16, 2011 · Cardiovascular

Device Facts

Record IDK110347
Device NameDOPPLER MONITOR
ApplicantCook Vascular, Inc.
Product CodeDPW · Cardiovascular
Decision DateMar 16, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2100
Device ClassClass 2

Intended Use

The Cook Vascular Blood Flow Monitor - Model DP-M350 is intended for monitoring blood flow in vessels intraoperatively, and following micro-vascular procedures, re-implantation, and free-flap transfers. The Swartz Doppler Flow Probe is supplied sterile and is intended for one-time use.

Device Story

Monitor unit generates 20 MHz signal transmitted to attached piezo-electric crystal probe; probe transmits/receives Doppler frequency sound waves from blood vessels. Returning signal amplified and projected as audible output; monitor includes two input channels and visual LED indicator. Used intraoperatively and post-procedure by clinicians to assess blood flow. Device features on/off switch, channel selector, volume control, and rechargeable battery. Output provides real-time auditory and visual feedback to assist clinicians in monitoring vessel patency and blood flow status, aiding clinical decision-making during and after vascular procedures.

Clinical Evidence

Bench testing only. Testing included electrical safety, battery testing, EMC, software validation, shock/vibration testing, mechanical validation, and packaging testing. No clinical data provided.

Technological Characteristics

20 MHz Doppler signal generation; piezo-electric crystal probe; two input channels; LED visual indicator; rechargeable battery; software-controlled visual/auditory output. Device is a monitor unit designed for use with Cook-Swartz Doppler Flow Probes.

Indications for Use

Indicated for monitoring blood flow in vessels intraoperatively and post-procedure, including micro-vascular procedures, re-implantation, and free-flap transfers.

Regulatory Classification

Identification

A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K110347 Page 1/2 MAR 1 6 2011 ## 510(k) Summary Submitted by: Jim Ferguson Quality Systems Manager Cook Vascular, Incorporated 1186 Montgomery Lane Vandergrift, Pa 15690 724-845-8621, XT 2227 April 10, 2006 Device: Cook Vascular, Inc. Blood Flow Monitor - Model DP-M350 Trade name: Cardiovascular Blood Flow Meter, 870.2100 Proposed Classification: Predicate Devices: The Cook Vascular, Inc. Blood Flow Monitor - Model DP-M350 is the same in terms of intended use, materials of construction and technology characteristics to the predicate devices that have been found substantially equivalent. ## Device Description: The proposed device is monitor unit that generates a (20 MHz) signal and conducts it to an attached probe that contains a piezo-electric crystal which transmits and then receives Doppler frequency sound waves. The returning signal is amplified and projected via the proposed monitoring unit as an audible output. The monitor unit has two input channels and also provides an adjunct visible (LED) indicator that is used as a complimentary information source to the primary audible indicator. The predicate device was submitted as a system, the proposed device is only the monitoring unit. ## Substantial Equivalence: This device will be manufactured to specified process controls and a Quality Assurance program. This device will undergo validated packaging operations indicative of devices currently manufactured by Cook Vascular. Being the same as the predicate devices in respect to indication for use and no changes in the fundamental scientific technology, this device meets the requirements for section 510(k) substantial equivalence. It is to be noted that this change is the roquitements or evil be used with the currently marketed Cook- Swartz Doppler Flow Probes. The only changes to the device are an added visual display, software controls for visual indicator, on/off switch, channel selector and volume. Additionally a chargeable battery system has been added. {1}------------------------------------------------ K 110347 Page 2/2 ## Test Data: The Cook Vascular, Inc. Blood Flow Monitor – Model DP-M350 was subjected to the flowing tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of: - Electrical Safety Testing 1. - 2. Battery Testing - EMC Testing ന് - Software Validation 4. - 5. Shock/Vibration Testing - 6. Mechanical Validation - 7. Packaging Testing The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a Blood Flow Monitor. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or other bird with outstretched wings. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 Cook Vascular, Inc. % Mr. Jim Ferguson, Jr. Quality Systems Manager 1187 Montgomery Lane Vandergrift, Pennsylvania 15690 MAR 1 6 20:1 Re: K110347 Trade/Device Name: Cook Vascular, Inc. Blood Flow Monitor - Model DP-M350 Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular blood flowmeter Regulatory Class: II Product Code: DPW, ITX, JOP Dated: February 3, 2011 Received: February 16, 2011 Dear Mr. Ferguson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in'the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations.administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Mr. Jim Ferguson, Jr. CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html far the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR, Part 807.97). For questions regarding the reporting of adverse events under the MDR regultive (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Aty B. Rh for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use Statement K110347 510(k) Number: Device Name: Cook Vascular, Inc. Blood Flow Monitor - Model DP-M350 Indications for Use: The Cook Vascular Blood Flow Monitor - Model DP-M350 is intended for monitoring blood flow in vessels intraoperatively, and following micro-vascular procedures, re-implantation, and free-flap transfers. The Swartz Doppler Flow Probe is supplied sterile and is intended for one-time use. Prescription Use (Part 21 CFR 801 Subpart D) ✓ · I AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) . Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Knape for WKM (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K110347 i Page 1 of
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%