The Cook-Swartz Doppler Flow Probe is for monitoring blood flow in vessels intraoperatively, and following reconstructive micro-vascular procedures, re-implantation, and free-flap transfers.
Device Story
The Cook-Swartz Doppler Flow Probe is a diagnostic ultrasonic transducer used to monitor blood flow in vessels. It consists of a 1 mm diameter piezoelectric crystal transducer (20 MHz) attached to a silicone cuff, which is secured around a blood vessel. The probe connects via wires to the Swartz Doppler Monitoring System. During intraoperative or post-operative monitoring, the transducer detects blood flow signals, which are processed by the compatible monitor. Healthcare providers use the auditory or visual output from the monitor to assess vessel patency and blood flow status. This real-time monitoring helps clinicians detect vascular compromise early, potentially improving outcomes for reconstructive micro-vascular procedures, re-implantations, and free-flap transfers.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility assessments (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity, implantation, chronic toxicity, carcinogenicity) and mechanical performance testing (cuff and suture pad detachment force).
Technological Characteristics
Piezoelectric crystal transducer (20 MHz); 1 mm diameter disc. Materials: NuSil MED-4750 silicone cuff, NuSil MED-1137 silicone adhesive, Master Bond EP42HT or Hysol EE0079/HD0070 epoxy, Neoflon NP-101 FEP-coated wire, NuSil MED-4050 silicone suture pads. Sterilization: Ethylene oxide. Connectivity: Wired connection to Cook Model DP-M350 monitor.
Indications for Use
Indicated for monitoring blood flow in vessels intraoperatively and post-operatively following reconstructive micro-vascular procedures, re-implantation, and free-flap transfers.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
July 17, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Cook Incorporated % Mr. Thomas Kardos Vice-President, Regulatory Affairs 750 Daniels Way, P.O. Box 489 BLOOMINGTON IN 47402
Re: K171272
Trade/Device Name: Cook-Swartz Doppler Flow Probe Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX Dated: July 12, 2018 Received: July 13, 2018
Dear Mr. Kardos:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
(1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171272
Device Name Cook-Swartz Doppler Flow Probe
Indications for Use (Describe)
The Cook-Swartz Doppler Flow Probe is for monitoring blood flow in vessels intraoperatively, and following reconstructive micro-vascular procedures, re-implantation, and free-flap transfers.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
Device: Cook-Swartz Doppler Flow Probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| Fetal<br>Imaging<br>& Other | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | P | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | | Laparoscopic | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | | Other (Specify) | | | | | | |
| | Peripheral | Peripheral vessel | | | | | | |
| Vessel | Other (Specify) | | | P | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
The Cook-Swartz Doppler Flow Probe is intended for use with the Swartz Doppler Monitoring System for monitoring blood flow in vessels intraoperatively, and following reconstructive micro-vascular procedures, re-implantation, and free-flap transfers.
Prescription Use Only (Part 21 CFR 801 Subpart D)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif letters on a gray background. Below the word "COOK" is the word "MEDICAL" in smaller, white, sans-serif letters on a black background. The black background is in the shape of a trapezoid.
# 2.0 510(k) SUMMARY
## Cook-Swartz Doppler Flow Probe As required by 21 CFR 807.92 Date Prepared: April 27, 2017
#### Submitted Bv:
| Submission: | Traditional 510(k) Premarket Notification |
|-----------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact: | Tom Kardos |
| Email: | RegSubmissions@CookMedical.com |
| Contact Phone Number: | 724-845-8621 |
| Contact Fax Number: | 724-845-2848 |
| Device Information: | |
| Trade Name: | Cook-Swartz Doppler Flow Probe |
| Common Name: | Doppler Probe |
| Classification Name: | Diagnostic ultrasonic transducer |
21 CFR 892.1570
ITX
Radiology
II
#### Predicate Device:
Classification Panel:
Regulation:
Product Code:
Device Class:
The Cook-Swartz Doppler Flow Probe, is substantially equivalent to the predicate device, the previously-submitted Cook-Swartz Doppler Flow Probe as described in K022649.
#### Device Description
The Cook-Swartz Doppler Flow Probe contains a piezoelectric crystal transducer assembly at its distal end. The transducer is a 1 mm diameter disc and operates at 20 MHz. The transducer is attached to a silicone cuff (either 5 mm × 32 mm or 5 mm × 17.4 mm) that is designed to be secured around a blood vessel. Proximal to the cuff, braided wires connect the transducer to suture pads and to a proximal polyurethanecovered wire. The polyurethane-covered wire ends in a two-pin plug connector.
{5}------------------------------------------------
Cook Incorporated - Traditional 510(k) Cook-Swartz Doppler Flow Probe April 27, 2017
The cuff is designed to be wrapped around the vessel to be monitored. The length of the cuff represents the circumference of the vessel that can be covered. The cuff is 5 mm wide regardless of length. The width of the cuff represents the length of the vessel that can be covered with the cuff.
#### Intended Use
The Cook-Swartz Doppler Flow Probe is for monitoring blood flow in vessels intraoperatively, and following reconstructive micro-vascular procedures, reimplantation, and free-flap transfers.
#### Comparison to Predicate Device
The Cook-Swartz Doppler Flow Probe is substantially equivalent to the predicate device, the previously-submitted Cook-Swartz Doppler Flow Probe as described in K022649. The table below presents the similarities and differences between the two products for substantial equivalence purposes. The differences between the subject device and the predicate device do not raise any new issues of safety and effectiveness. Performance data are available to support substantial equivalence and were developed using acceptable scientific methods for evaluation.
| | Cook-Swartz Doppler Flow Probe<br>(K022649) | Cook-Swartz Doppler Flow Probe<br>(Subject of this submission) |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Indications for Use | The Cook-Swartz Doppler Flow Probe<br>is for monitoring blood flow in vessels<br>intraoperatively, and following<br>reconstructive micro-vascular<br>procedures, re-implantation, and free-<br>flap transfers. | Same |
| Transducer | | |
| Type | Piezoelectric crystal | Same |
| Signal frequency | 20 MHz | Same |
| Signal amplitude | 15 V (peak-to-peak) into 50 Ω | Same |
| Transmitter pulse width &<br>repetition frequency | 0.4 μs; 78.1 KHz | Same |
| Cuff | | |
| Cuff sizes | 27 mm × 25 mm | 37 mm (L) × 5 mm (W)<br>17.4 mm (L) × 5 mm (W) |
| Material | NuSil MED-4850 silicone | NuSil MED-4750 silicone |
{6}------------------------------------------------
| | Cook-Swartz Doppler Flow Probe<br>(K022649) | Cook-Swartz Doppler Flow Probe<br>(Subject of this submission) |
|---------------------------------|---------------------------------------------|---------------------------------------------------------------------------------------|
| Other patient-contact materials | | |
| Adhesive | NuSil MED-2000 silicone | NuSil MED-1137 silicone |
| Epoxy | Master Bond Polymer System<br>EP42HT epoxy | Master Bond Polymer System<br>EP42HT epoxy or Hysol EE0079 &<br>HD0070 two-part epoxy |
| Wire | Neoflon NP-101 FEP-coated wire | Same |
| Suture pads | NuSil MED-4950 or MED-4850<br>silicone | NuSil MED-4050 Silicone |
| Other characteristics | | |
| Sterilization | Ethylene oxide | Same |
| Compatible Monitor | Cook Model DP-M350 | Same |
#### Technological Characteristics
The transducer, electrical components, and the polyurethane-covered wires and connector of the Cook-Swartz Doppler Flow Probe are unchanged compared to the predicate device; therefore, it was not necessary to retest acoustic and electrical characteristics of the device.
A biocompatibility assessment was made, including assessment of the following biological effects:
- 1. Cytotoxicity
- 2. Sensitization
- 3. Irritation or intracutaneous reactivity
- 4. Acute systemic toxicity
- 5. Material-mediated pyrogenicity
- 6. Subacute/subchronic toxicity
- 7. Genotoxicity
- 8. Implantation
- 9. Chronic toxicity
- 10. Carcinogenicity
The assessment concluded that the materials of the device provide a reasonable assurance of safety with respect to biocompatibility.
{7}------------------------------------------------
Cook Incorporated - Traditional 510(k) Cook-Swartz Doppler Flow Probe April 27, 2017
The following tests were conducted and demonstrate the effective performance of the device:
- Cuff detachment force ●
- Suture pad attachment force
## Conclusion
The results of these tests support the conclusion that the Cook-Swartz Doppler Flow Probe performs acceptably and does not raise new questions of safety and effectiveness, thus supporting a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.