VascuChek® Kit, VascuChek® Transceiver, VascuChek® Clinical Probe, VascuChek® Charger: The VascuChek® device is intended for the non-invasive transcutaneous evaluation of blood flow in Peripheral Vasculature. VascuChek® Kit, VascuChek® Transceiver, VascuChek® Surgical Probe, VascuChek® Charger: The VascuChek® device is intended for the intraoperative and transcutaneous evaluation of blood flow in the following clinical applications: - Intraoperative (Microvascular and Vascular) - Intraoperative Neurological - Transrectal and Peripheral Vascular.
Device Story
Cardiovascular blood flowmeter; utilizes ultrasonic Doppler technology. System components: reusable transceiver, sterile probe with sheath, charger. Operation: transmitter drives ultrasonic crystal at probe tip; narrow beam waves reflect off blood flow; piezoelectric transducer converts reflected waves to high-frequency electronic signals. Signal processing: amplification, frequency comparison against transmitted signal, conversion to audible Doppler-shifted tone via speaker. Used in clinical/surgical settings by healthcare providers. Output: audible blood flow signals. Benefit: real-time, non-invasive assessment of blood flow to support clinical decision-making during surgery or routine vascular evaluation.
Clinical Evidence
Bench testing only. Evaluations included electrical safety (ANSI AAMI ES60601-1), EMC (IEC 60601-1-2, IEC 60601-2-37), firmware verification, and biocompatibility testing for housing material changes per FDA guidance.
Indicated for patients requiring non-invasive transcutaneous or intraoperative evaluation of blood flow in peripheral, microvascular, vascular, neurological, and transrectal applications.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 30, 2024
Remington Medical, Inc. % Justin Gracyalny Regulatory Affairs Manager Secure BioMed Evaluations 7828 Hickory Flat Hwy, Suite 120 Woodstock, Georgia 30188
Re: K241583
Trade/Device Name: VascuChek® Clinical and VascuChek® Surgical Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW Dated: June 3, 2024 Received: June 3, 2024
Dear Justin Gracyalny:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K241583
Device Name VascuChek® Clinical and VascuChek® Surgical
Indications for Use (Describe)
VascuChek® Kit, VascuChek® Transceiver, VascuChek® Clinical Probe, VascuChek® Charger: The VascuChek® device is intended for the non-invasive transcutaneous evaluation of blood flow in Peripheral Vasculature.
VascuChek® Kit, VascuChek® Transceiver, VascuChek® Surgical Probe, VascuChek® Charger:
The VascuChek® device is intended for the intraoperative and transcutaneous evaluation of blood flow in the following clinical applications:
- Intraoperative (Microvascular and Vascular)
- Intraoperative Neurological
- Transrectal and Peripheral Vascular.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## K241583
#
| Date Prepared | August 30, 2024 |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor | Remington Medical, Inc. |
| | 6830 Meadowridge Court<br>Alpharetta, GA 30005<br>470-719-1084 |
| 510(k) Contact | Secure BioMed Evaluations |
| | Justin Gracyalny, MSE |
| | Linda Braddon, Ph.D. |
| | 7828 Hickory Flat Highway, Suite 120<br>Woodstock, GA 30188<br>770-837-2681<br>Regulatory@SecureBME.com |
| Trade Name | VascuChek® Clinical and VascuChek® Surgical |
| Common Name | Flowmeter, Blood, Cardiovascular |
| Classification<br>Name | 21 CFR §870.2100 Cardiovascular blood flowmeter<br>Product Code: DPW |
| Device Class | II |
| Predicate<br>Devices | K211589 Remington Medical, Inc. VascuChek Kit, VascuChek Transceiver,<br>VascuChek Surgical Probe, VascuChek Charger |
| | K212065 Remington Medical, Inc. VascuChek Kit, VascuChek Transceiver,<br>VascuChek Clinical Probe, VascuChek Charger |
| Device<br>Description | VascuChek® Surgical is a cardiovascular blood flowmeter comprised of two<br>components: a sterile VascuChek™ Surgical Probe with sheath which<br>connects to the reusable, nonsterile VascuChek™ Transceiver. The sheath is<br>deployed over, and encapsulates the transceiver, allowing it to be used within<br>the sterile field.<br><br>VascuChek® Clinical is a cardiovascular blood flowmeter comprised of two<br>components: a sterile VascuChek™ Surgical Probe with sheath which<br>connects to the reusable, nonsterile VascuChek™ Transceiver. The sheath is<br>deployed over, and encapsulates the transceiver, allowing it to be used within<br>the sterile field.<br><br>The VascuChek® device follows Track 1.<br><br>A transmitter in the transceiver drives the ultrasonic transmitting crystal<br>located at the tip of the probe component. The ultrasonic waves generated by<br>the sensor travel through the tissue just under the probe tip in a narrow beam.<br>The reflected ultrasonic waves are received by the transducer and are<br>converted via the piezoelectric effect into a high frequency electronic signal. |
| | band audio Doppler shifted signal which is filtered and converted to audio |
| | via a speaker. During the intervals when the unit is not transmitting, the<br>device passes any reflected signals that it receives to a receiving circuit. This |
| | circuit amplifies the returning echoes, compares their frequency to that of the |
| | transmitted signal and converts any frequency differences into an audible |
| | tone. |
| Indications for<br>Use Statement | VascuChek® Kit, VascuChek® Transceiver, VascuChek® Clinical Probe,<br>VascuChek® Charger: |
| | The VascuChek® device is intended for the non-invasive transcutaneous<br>evaluation of blood flow in Peripheral Vasculature. |
| | VascuChek® Kit, VascuChek® Transceiver, VascuChek® Surgical Probe,<br>VascuChek® Charger: |
| | The VascuChek® device is intended for the intraoperative and<br>transcutaneous evaluation of blood flow in the following clinical<br>applications: |
| | - Intraoperative (Microvascular and Vascular) |
| | - Intraoperative Neurological |
| | - Transrectal and Peripheral Vascular. |
{6}------------------------------------------------
### Comparison of Technological Characteristics
The subject device is substantially equivalent to its predicate devices (K211589 Remington Medical, Inc. VascuChek Kit, VascuChek Transceiver, VascuChek Surgical Probe, VascuChek Charger and K212065 Remington Medical, Inc. VascuChek Kit, VascuChek Transceiver, VascuChek Clinical Probe, VascuChek Charger).
### Similarities to Predicate
The subject device has the same intended use and technological characteristics as the identified predicate devices. The system employs identical technology as the identified predicates. The principles of operation and technological characteristics are identical to the predicate devices. The system has similar or identical technological characteristics including:
- . Identical system and kit components including use of single use sterile (for surgical use) and non-sterile (for clinical use) probes and a multi-use transceiver
- Identical patient population
- Identical intended use ●
- Identical principles of operation ●
- . Identical firmware requirements and performance
### Differences to Predicate
The only differences between the subject and predicate device relate to minor hardware, firmware, and housing material differences in the subject device transceiver. All differences have been addressed through risk analysis and supporting verification / validation testing.
{7}------------------------------------------------
### Non-Clinical Performance Testing Summary
All necessary testing has been performed to assure substantial equivalence to the predicate devices and to demonstrate the subject device performs as intended. All testing was performed in accordance with relevant performance standards or internal test methods. The following evaluations were conducted:
- Electrical Safety Testing per ANSI AAMI ES60601-1 .
- EMC Testing per IEC 60601-1-2, IEC 60601-2-37 ●
- Firmware Verification Testing ●
- Biocompatibility information was provided in accordance with Attachment G of the FDA ● guidance for biocompatibility to support the transceiver housing material change
### Conclusions
Based on the similarities of the intended use / indications for use, technological and functional characteristics, and the results of the non-clinical performance testing, the subject device is substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.