NUPRO MODEL 13016901

K103160 · CAO Group, Inc. · LBH · Jan 6, 2011 · Dental

Device Facts

Record IDK103160
Device NameNUPRO MODEL 13016901
ApplicantCAO Group, Inc.
Product CodeLBH · Dental
Decision DateJan 6, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3260
Device ClassClass 2
AttributesTherapeutic

Intended Use

NUPRO® cavity varnish is indicated for the coating or occlusion of exposed dentin tubules under temporary restorations, and for the treatment of hypersensitive teeth.

Device Story

NUPRO® is an ethanol-based, sodium fluoride-containing cavity varnish (5% NaF, 22,600ppm fluoride ion). Applied by dentists to intact tooth surfaces or exposed dentin as a pre-treatment for temporary restorations. Supplied as a white, opaque, viscous liquid in unit-dose preloaded dispensing dishes with a disposable brush. Mechanism involves physical occlusion of dentinal tubules to alleviate hypersensitivity. Benefits include reduced tooth sensitivity and protection of exposed dentin. Used in clinical dental settings.

Clinical Evidence

Bench testing only. Biocompatibility evaluated per ISO 7405:2008. Performance testing confirmed 5% fluoride concentration, adhesion force of 77kPa, and dentin tubule penetration of 3-5 microns. Dye penetration tests confirmed equivalent fluid ingress prevention compared to predicates.

Technological Characteristics

Ethanol-based varnish containing 5% sodium fluoride (22,600ppm). Viscous liquid form factor. Packaged in unit-dose dispensing dishes with disposable brush. Biocompatibility tested per ISO 7405:2008. No electronic or software components.

Indications for Use

Indicated for patients requiring treatment for hypersensitive teeth or coating/occlusion of exposed dentin tubules under temporary dental restorations. For professional use by dentists.

Regulatory Classification

Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K10 3160 # 510(k) Summary of Safety and Effectiveness CAO Group. Inc. 4628 West Skyhawk Drive West Jordan, UT 84084 Tel: 801-256-9282 Fax: 801-256-9287 Prepared By: Robert K. Larsen. Preparation Date: October 18, 2010 | Device Name: | | |-------------------------|-----------------| | Trade Name: | NUPRO® | | Common Name: | Cavity Varnish | | Product Classification: | Varnish, Cavity | ### Legally Marketed Predicate Devices for Substantial Equivalence: Ascent F-Coat, manufactured by CAO Group, Inc. (K100540) તન DuraShield, manufactured by Dentsply International. Inc. (K082198) #### Rationale for Substantial Equivalence: The aforementioned varnishes share similar indications for use on teeth for alleviating hypersensitivity with the submitted device. The predicate devices and submitted device share similar design features including the inclusion of sodium fluoride, the use of carrier solvents. application time and application methods. #### Description of Submitted Device: The NUPRO® cavity varnish is an ethanol based, sodium fluoride containing cavity varnish. The material is applied to tooth surfaces where hypersensitivity is a concern or to surfaces where dentinal tubule occlusion is indicated. The varnish is applied by a dentist to intact tooth surfaces, or to exposed dentin as a pre-treatment prior to application of temporary restorative materials. The material is a white opaque viscous liquid that is prepackaged in unitdose preloaded dispensing dishes with a disposable brush applicator. {1}------------------------------------------------ ### Intended Uses of NUPRO®: NUPRO® cavity varnish is indicated for the coating or occlusion of exposed dentin tubules under temporary restorations, and for the treatment of hypersensitive teeth. ーステース (1) 2 スポート (1) 2017 - 10:10 (1) 2017 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - #### Technological Characteristics and Substantial Equivalence: Ascent F-Coat is a 5% sodium fluoride varnish containing a total 22,600ppm equivalent fluoride ion. The varnish uses an ethanol carrier and makes use of a synthetic polymer as the varnish matrix. The product is intended for use by a dentist for the treatment of hypersensitivity and for covering exposed dentin under temporary dental restorations. The material is packaged in unit-dose blister packs and is applied by a disposable brush. The Durashield Varnish is a 5% sodium fluoride varnish containing a total 22,600ppm equivalent fluoride ion. The varnish uses an ethanol carrier and makes use of natural resin derivatives as the varnish matrix. The product is intended for use by a dentist for the relief of hypersensitivity where exposed dentin or cementum, exist. The material is packaged in unit-dose blister packs and is applied by a disposable brush. #### Conformity to Standards: The NUPRO® cavity varnish was evaluated in terms of the recognized standard ISO 7405:2008 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry. ### Performance Data Laboratory tests on constituents of the device that are unique to the current submission demonstrate that such constituents are biocompatible. Laboratory testing demonstrates that the subject device contains fluoride ion at the labeled concentration of 5% or 22,600ppm and is at least equivalent to the predicate devices which share similar labeled fluoride content. Laboratory testing demonstrates that the subject device provides adhesion to the tooth surface at least equal to that of the predicate devices with an average lateral shear adhesion force of 77kPa. Photomicrographs of the subject device reveal that it is capable of penetrating exposed dentin tubules to a minimum of 3-5 microns. Laboratory dye penetration tests demonstrate that the subject device is substantially equivalent to the predicate devices in preventing the ingress of fluids into the exposed dentin tubules. {2}------------------------------------------------ ## 1999 - 1999 - 1999 - 1999 Conclusion The NUPRO® cavity varnish is substantially equivalent to the listed cavity varnishes without raising any new issues of safety or effectiveness. This device shares similar intended uses, similar formulation, and similar functional and performance characteristics. The device is designed to comply with relevant federal and international safety and performance standards. ー - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JAN 6 2011 Mr. Robert K. Larsen Regulatory Affairs Manager CAO Group, Incorporated 4628 West Skyhawk Drive West Jordan, Utah 84084 Re: K103160 Trade/Device Name: NUPRO® Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: II Product Code: LBH Dated: October 21, 2010 Received: October 26, 2010 Dear Mr. Larsen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Larsen Enclosure Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Nh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ # Indications for Use 103160 510(k) Number (if known): Device Name: Indications For Use: NUPRO® cavity varnish is indicated for the coating or occlusion of exposed dentin tubules under temporary restorations, and for the treatment of hypersensitive teeth. Prescription Use X AND/OR Over-The-Counter Use (Per 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) 4-1 Concurrence of CDRH, Office of Device Evaluation (ODE) Suan Rianen (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K103160 Page 1 of 1
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%