Profisil Fluoride Varnish Combi pack (14801), Profisil Fluoride Varnish Normal pack mint (14802), Profisil Fluoride Varnish Normal pack berry (14804), Profisil Fluoride Varnish Normal pack unflavored (14806), Profisil Fluoride Varnish Sample pack (14800)
Applicant
Kettenbach GmbH & Co. KG
Product Code
LBH · Dental
Decision Date
Jul 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3260
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Profisil® Fluoride Varnish is a 5 % sodium fluoride varnish which produces mechanical occlusion of the dentinal tubules in the treatment of tooth hypersensitivity.
Device Story
Profisil Fluoride Varnish is a 5% sodium fluoride suspension in a dimethicone gel; used by dental professionals to treat tooth hypersensitivity. The varnish is applied to the tooth surface using an applicator brush; it adheres to the dentition for several hours. The mechanism of action involves physical occlusion of patent/open dentinal tubules; the sodium fluoride reacts with calcium to form insoluble calcium fluoride, creating a semi-permanent barrier. This process blocks open pores in the tooth structure, alleviating hypersensitivity. The device is provided in single-use unit-dose packages.
Clinical Evidence
No clinical data provided. Bench testing only, including biocompatibility (ISO 10993 series) and physical property testing per EN ISO 17730:2020 (pH, dynamic viscosity, fluoride release, and SEM).
Technological Characteristics
5% sodium fluoride suspended in dimethicone gel. Single-use, unit-dose delivery system. Adheres to tooth surface. Tested per EN ISO 17730:2020. Biocompatibility evaluated per ISO 10993-1, 5, 10, 11, 23.
Indications for Use
Indicated for the treatment of tooth hypersensitivity in patients experiencing dentinal hypersensitivity. Contraindicated in individuals with known intolerance to fluoride.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 24, 2023
Kettenbach GmbH & Co. KG Katja Simon Regulatory Affairs Manager Im Heerfeld 7 Eschenburg, Hessen 35713 GERMANY
Re: K230333
Trade/Device Name: Profisil Fluoride Varnish Combi pack (14801), Profisil Fluoride Varnish Normal pack mint (14802), Profisil Fluoride Varnish Normal pack berry (14804), Profisil Fluoride Varnish Normal pack unflavored (14806), Profisil Fluoride Varnish Sample pack (14800)
Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: LBH Dated: June 22, 2023 Received: June 22, 2023
Dear Katja Simon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K230333
Device Name
Profisil Fluoride Varnish Combi pack (14801);
Profisil Fluoride Varnish Normal pack mint (14802);
Profisil Fluoride Varnish Normal pack berry (14804);
Profisil Fluoride Varnish Normal pack unflavored (14806);
Profisil Fluoride Varnish Sample pack (14800)
Indications for Use (Describe)
Profisil® Fluoride Varnish is a 5 % sodium fluoride varnish which produces mechanical occlusion of the dentinal tubules in the treatment of tooth hypersensitivity.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours be response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for KETTENBACH DENTAL. The logo features a graphic of six circles arranged in a triangular shape, with the company name "KETTENBACH DENTAL" written in a sans-serif font below it. Underneath the company name, the tagline "Simply intelligent" is written in a smaller font.
# 510(k) Summary - K230333
#### l. Submitter:
Kettenbach GmbH & Co KG Im Heerfeld 7 35713 Eschenburg Germany
Phone: +49 2774 705 0 Fax: +49 2774 705 66
## Contact person:
Simon, Katja Regulatory Affairs Manager Im Heerfeld 7 35713 Eschenburg Germany Phone: + 49 277 4705 0 E-mail: katja.simon@kettenbach.com
## Date of preparation:
July 24th, 2023
## II. Device
Name of Device: Common or Usual Name: Classification Name: Regulatory Class: Product Code:
Profisil® Fluoride Varnish Profisil® Fluoride Varnish Dental, Varnish, Cavity (21 CFR 872.3260) = LBH
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for KETTENBACH DENTAL. The logo consists of six blue circles arranged in a pyramid shape above the company name. Below the company name is the tagline "Simply intelligent".
## . Predicate devices:
Vanish™ Varnish (3M Espe), K092141 This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
### IV. Device Description:
Profisil® Fluoride Varnish is designed to securely adhere to the tooth surface for several hours while releasing fluoride ions. The varnish contains 5 % sodium fluoride suspended in a mucosa-friendly, pleasantly flavored dimethicone gel. The varnish is available in following versions: mint, berry and unflavored.
#### V. Indications for Use:
Profisil® Fluoride Varnish is a 5 % sodium fluoride varnish which produces mechanical occlusion of the dentinal tubules in the treatment of tooth hypersensitivity.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image features the logo for KETTENBACH DENTAL. The logo consists of a triangular arrangement of six blue circles above the text "KETTENBACH DENTAL", with the word "DENTAL" in a slightly lighter shade of blue. Below the company name is the tagline "Simply intelligent" in a smaller font size.
## Table 1: COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.
| | Predicate Devices | Substantial Equivalent Device | Conclusion |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product | Vanish Varnish | Profisil® | |
| Manufacturer | 3M Espe | Kettenbach GmbH & Co. KG | |
| Product<br>Description | 3M™ ESPE™ Vanish™ Varnish 5%<br>Sodium Fluoride White Varnish is a<br>22.600 ppm fluoride-containing varnish for<br>application to enamel and dentin for the<br>treatment of hypersensitivity teeth. It will<br>spread on and adhere to moist teeth.<br>Vanish varnish is an alcohol-based<br>solution of modified rosin. The product is<br>sweetened with xylitol. Vanish varnish<br>contains an innovative tri-calcium<br>phosphate (TCP) ingredient exclusively<br>from 3M. The ingredient TCP serves as a<br>source of bioavailable calcium and<br>phosphate ions when the varnish is<br>applied to the teeth. Calcium and<br>phosphate ions are naturally occurring<br>components of saliva long associated with<br>maintaining healthy teeth. | Profisil® Fluoride Varnish is designed to<br>securely adhere to the tooth surface for<br>several hours while releasing fluoride<br>ions. The varnish contains 5 % sodium<br>fluoride suspended in a mucosa-friendly,<br>pleasantly flavored dimethicone gel. The<br>varnish is available in following versions:<br>mint, berry and unflavored. | Treatment of<br>hypersensitivity teeth is the<br>technological principle for<br>both the subject and<br>predicate devices. The<br>content of sodium fluoride<br>is identical.<br>The following differences<br>exist between the subject<br>and the predicate device:<br>Profisil® does not contain<br>TCP |
| Intended purpose | 5% Sodium Fluoride Varnish for the<br>treatment of dentinal hypersensitivity | 5% Sodium Fluoride Varnish indicated<br>for the treatment of dentinal<br>hypersensitivity. | The intended purpose is<br>identical |
| Indication | • Treatment of hypersensitive teeth<br>• Use on exposed dentin and root<br>sensitivity<br>• Under temporary restoratives and<br>cements where post-operative sensitivity<br>is of concern | Profisil® Fluoride Varnish is a 5 %<br>sodium fluoride varnish which produces<br>mechanical occlusion of the dentinal<br>tubules in the treatment of tooth<br>hypersensitivity. | The indication is the same<br>and are intended for the<br>same general population of<br>patients, experiencing<br>dentinal hypersensitivity.<br>Both devices produce an<br>occlusion of the dentinal<br>tubules in the treatment of<br>tooth hypersensitivity. |
| Basic formulation | • Colophony / Polyamide resin<br>• Ethyl alcohol<br>• Flavor<br>• 5% Sodium fluoride | • Silicone Encapsulant<br>• Thickener/ Adhesive<br>• Flavor / Sweetener<br>• 5% Sodium fluoride | The percentage of sodium<br>fluoride is identical.<br>However, Profisil does not<br>contain TCP |
| | Predicate Devices | Substantial Equivalent Device | Conclusion |
| Product | Vanish Varnish | Profisil® | |
| Manufacturer | 3M Espe | Kettenbach GmbH & Co. KG | |
| Mode of action /<br>technology | Dentin tubule occlusion<br>The material acts to physically block or<br>occlude patent/open dentin tubules on the<br>dentition. Vanish Varnish has unit-dose<br>packages delivering up to 0.25 mL to 0.4<br>mL of 5% sodium fluoride.<br>However, this product contains<br>substances that may cause an allergic<br>reaction by skin contact in certain<br>individuals. | Dentin tubule occlusion<br>The material acts to physically block or<br>occlude patent/open dentin tubules on<br>the dentition. | Same mode of action. |
| Single use | Single use only | Single use only | Identical, both are single<br>use devices |
| Method of<br>application | Unit-dose packages delivering up to 0.25<br>ml or 0.4ml of 5% sodium fluoride varnish.<br>Varnish and an applicator brush placed in<br>a molded tray and closed with a seal | Unit-dose packages delivering in 0.5ml<br>of 5% sodium fluoride varnish. Varnish<br>and an applicator brush placed in a<br>molded tray and closed with a seal | The delivery systems and<br>the mode of application are<br>similar. |
| Contraindications /<br>Precautions | This product contains substances that<br>may cause an allergic reaction by skin<br>contact in certain individuals. Avoid use of<br>this product in patients with known<br>colophony allergies. Discontinue patient's<br>use of other prescriptive fluoride<br>preparations for 24 hours following Vanish<br>Varnish 5% sodium fluoride varnish<br>application. Children taking fluoride<br>supplements should discontinue for 1-3<br>days following treatment. | Do not use this product on any individual<br>who has a known intolerance to fluoride.<br>If any signs of inflammation or allergic<br>reaction occur, discontinue use<br>immediately and refrain from further use<br>with this patient.<br>Other prescription products containing<br>fluoride should not be used within 24<br>hours of using the varnish.<br>Patients using regular, systemic<br>fluoridation should stop using these<br>products for three days following use of<br>this varnish | Discontinue patient's use of<br>other prescriptive fluoride<br>preparations for 24 hours /<br>three days is similar |
| Shelf Life | 24 months | 24 months | The shelf life is identical |
| Performance<br>testing | A comparison of the parameters in respect, pH value, dynamic viscosity, fluoride release and electron microscopy<br>(SEM) studies were performed. The results demonstrated the substantial equivalence to the predicate device. | | |
| Summary of<br>Product<br>Description, | The indication of the subject device is the same for the predicate device. Contraindications, application, and side<br>effects for both products are basically the same - there is a slight difference in the wording. | | |
| | Predicate Devices | Substantial Equivalent Device | Conclusion |
| Product | Vanish Varnish | Profisil® | |
| Manufacturer | 3M Espe | Kettenbach GmbH & Co. KG | |
| Intended purpose<br>Indications,<br>Contraindications /<br>Precautions | | | |
| Summary<br>Chemical<br>Composition | The Profisil® formulation has been thoroughly assessed for biocompatibility. | | |
| Summary of<br>Finished<br>Device<br>Specification | According to EN ISO 17730:2020 "Dentistry - fluoride varnishes" apart from the total fluoride, no other properties<br>appear as being relevant for having the device functioning or performing as intended.<br>Specific physical properties have been assessed and verified according to product standard EN ISO 17730:2020 –<br>Dentistry- fluoride varnishes".<br>The acceptance criteria of EN ISO 17730:2020 was fulfilled. | | |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image features the logo for KETTENBACH DENTAL. Above the company name, there is a graphic of six circles arranged in a triangular pattern. Below the company name, the words "Simply intelligent" are written in a smaller font.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image features the logo for KETTENBACH DENTAL. The logo consists of six blue circles arranged in a triangular formation above the text "KETTENBACH DENTAL". Below the company name is the tagline "Simply intelligent" in a smaller font size. The overall design is clean and professional, suggesting a focus on innovation and simplicity.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for KETTENBACH DENTAL. The logo consists of a blue graphic of six circles arranged in a triangular shape above the company name. The text "KETTENBACH DENTAL" is written in blue, with the word "DENTAL" in a slightly lighter shade. Below the company name is the tagline "Simply intelligent" in a smaller, lighter blue font.
### VII. Performance Data:
Biocompatibility:
The following performance data were provided in support of the substantial equivalence determination. The biocompatibility evaluation for Profisil® Fluoride Varnish was conducted in accordance with ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.:
- · Cytotoxicity ISO 10993-5
- · Sensitization ISO 10993-10
- Irritation ISO 10993-23
- Pyrogenicity ISO 10993-11
- · Systemic toxicity ISO 10993-11
To evaluate the biological safety of Kettenbach's Fluoride Varnish, consideration was given to the type of patient contact, the potential hazards associated with the chemicals used in the devices, and historical use of the materials used in the devices. Profisil® Fluoride Varnish is considered to have met the requirements of ISO 10993-1:2018, ISO 7405:2018, FDA's Use of International standard ISO 10993-1, and the European Union Medical Regulation (EU) 2017/745 is an externally communicating device with limited (<24 hours) contact with the patient's tissue/bone/dentin and can be considered safe for use.
# Non-clinical performance:
According to EN ISO 17730:2020 "Dentistry - fluoride varnishes" apart from the total fluoride, no other properties appear as being relevant for having the device functioning or performing as intended.
Specific physical properties have been assessed and verified according to product standard EN ISO 17730:2020 - Dentistry- fluoride varnishes".
The acceptance criteria of EN ISO 17730:2020 was fulfilled.
- pH value
- · Dynamic viscosity
- · Fluoride release
- · Electron microscopy (SEM) study
Summary:
Based on the performance data, Profisil® Fluoride Varnish was found to have a safety and effectiveness profile that is similar to the predicate device.
Clinical Performance Data: Data from clinical studies was not provided.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image features the logo for KETTENBACH DENTAL. The logo consists of six blue circles arranged in a triangular formation above the text "KETTENBACH DENTAL", with "DENTAL" in a slightly lighter shade of blue. Below the company name is the tagline "Simply intelligent" in a smaller font size.
# VIII. Conclusion:
Kettenbach Profisil® Fluoride Varnish is comprised of a highly substantive and highly viscous dimethicone gel carrier containing a suspension of 5% sodium by weight sodium fluoride crystals.
The material acts to physically block or occlude patent/open dentin tubules on the dentition. This is achieved by applying the sodium fluoride dimethicone gel with an applicator brush to the affected dentition. The gel is allowed to penetrate the porosity of the tooth structure and block these open pores. The sodium fluoride reacts with calcium to form insoluble calcium fluoride to create a semi-permanent occluding material to alleviate the effect of dentinal hypersensitivity.
Profisil® Fluoride Varnish is substantially equivalent to the predicate device 3M Vanish Varnish in the indications for use and mechanism of action. Both products are intended for the reduction of dentinal hypersensitivity, and both produce this effect by the mechanical occlusion of the open dentinal tubules upon and within the dentition. Both Kettenbach Profisil® Fluoride Varnish and 3M Vanish Varnish are applied in an identical manner and are intended for the same general dental population.
Laboratory data demonstrate both products are equivalent in efficacy and release fluoride at an equivalent rate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.