The Neo-Tee™ T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with neonates and infants weighing less than 10kg (22 lb).
Device Story
Gas-powered manual resuscitator for neonates/infants < 10 kg; provides respiratory support via face mask or endotracheal tube. Device features T-piece configuration, integrated manometer, adjustable Peak Inspiratory Pressure (PIP), adjustable Positive End-Expiratory Pressure (PEEP), and pressure relief valve. Operated by trained clinicians in hospitals, delivery suites, ICUs, and nurseries. Clinician connects device to wall gas or cylinder; manually controls breath delivery by occluding/releasing T-piece port. Manometer provides real-time pressure feedback to clinician, allowing adjustment of PIP/PEEP to meet patient needs. Facilitates established resuscitation standards; helps maintain airway and oxygenation for infants failing to establish effective breathing.
Clinical Evidence
Bench testing only. Tests included vomitus resistance, water immersion, oxygen concentration, inspiratory/expiratory resistance, PEEP, delivered volume, pressure limitation, and durability (drop test). Device met all requirements of ISO 10651:2006 or demonstrated equivalence to predicate performance.
Technological Characteristics
Gas-powered T-piece resuscitator. Components: manometer (up to 60 cmH2O), adjustable PIP/PEEP valves, pressure relief valve. Connectivity: wall gas or cylinder (1-15 LPM). Patient interface: face mask or ET tube. Materials: patient-contact materials identical to predicate devices. No software or electronic components.
Indications for Use
Indicated for emergency respiratory support in neonates and infants weighing < 10 kg. Used via face mask or endotracheal tube in hospital, delivery suite, ICU, and nursery settings by trained clinicians.
Regulatory Classification
Identification
A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
{0}------------------------------------------------
K093913
## 510(k) Summary Page 1 of 5 30-Mar-10
APR - 3 2010
| Mercury Medical, Inc. | |
|----------------------------|-------------------------------------------------------------------------|
| 11300 - 49th St. North | Tel - (727) 573-0088 |
| Clearwater, FL 33762-4807 | Fax - (727) 571-3922 |
| Official Contact: | Jeff Ratner - VP Engineering and Quality Assurance |
| Proprietary or Trade Name: | Mercury Medical T-Piece Resuscitator (Neo-Tee™) |
| Common/Usual Name: | Infant T-Piece Resuscitator |
| Classification Name/Code: | BTL - powered emergency ventilator<br>CFR 868.5925 |
| Device: | Mercury Medical T-Piece Resuscitator |
| Predicate Devices: | GE - Giraffe and Panda - K070210<br>Fisher & Paykel - NeoPuff - K892885 |
#### Device Description:
The Mercury Medical T-Piece Resuscitator (Neo-Tee™) is a gas power resuscitator for use with neonates / infants < 10 kg (22 lb).
It is a simple T-piece, with a manometer and the ability to adjust Peak Inspiratory Pressure (PIP) and Positive End-Expiratory Pressure (PEEP). It incorporates a pressure relief valve for excessive pressure.
The Mercury Medical T-Piece Resuscitator (Neo-Tee™) can be connected to the patient via a face mask or endotracheal tube.
#### Indications for Use:
The (Neo-Tee™) T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with neonates and infants weighing less than 10kg (22 1b).
Environment of Use: Hospital, delivery suites, ICU and Nursery settings
{1}------------------------------------------------
510(k) Summary
Summary of substantial equivalence
.
| | Indications for Use | | |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Proposed<br>Mercury Medical T-Piece<br>Resuscitator (Neo-TeeTM) | Predicate<br>K070210 GE Giraffe and Panda | Predicate<br>K892885 F&P NeoPuff |
| | The Neo-TeeTM T-Piece Resuscitator is<br>a gas powered emergency resuscitator<br>intended to provide emergency<br>respiratory support by means of a face<br>mask or a tube inserted into a patient's<br>airway. It is intended for use with<br>neonates and infants weighing less than<br>10kg (22 lb). | Provides the basic equipment required<br>for pulmonary resuscitation of infants.<br>Pulmonary resuscitation includes<br>practices necessary to establish a clear<br>airway and provide oxygen and<br>air/oxygen mixtures and/or manual<br>ventilation to the infant. These are<br>clinical practices that represent the<br>established standard or care.*<br><br>Resuscitation may be required whenever<br>an infant fails to establish effective,<br>adequate breathing patterns necessary to<br>meet tissue oxygen demands and/or to<br>rid the body of carbon dioxide.<br><br>* As stated in collaborative guidelines<br>written by the American Heart<br>Association (AHA) and the American<br>Academy of Pediatrics (AAP) in the<br>Textbook of Neonatal Resuscitation, 5th<br>Edition. | Manually operated, gas powered<br>resuscitator which provides controlled<br>and accurate resuscitation of newborn<br>babies |
| Environment of use | Hospital, delivery suites, nursery, ICU | Hospital, delivery suites, nursery, ICU | Hospital, delivery suites, nursery, ICU |
| Patient Population | Infant and neonates < 10 kg | Infant < 10 kg | Infant < 10 kg |
| Prescriptive | Persons trained in infant / neonate<br>resuscitation | For professional use only, by trained<br>clinicians. | Persons trained in infant resuscitation |
| | Proposed<br>Mercury Medical T-Piece<br>Resuscitator (Neo-Tee™) | Predicate<br>K070210 GE Giraffe and Panda | Predicate<br>K892885 F&P NeoPuff |
| Patient connection | Face mask<br>ET tube | Face mask<br>ET tube | Face mask<br>ET tube |
| Features and Performance Characteristics | | | |
| Gas flow provided by<br>Components | Wall gas or cylinder<br>T-piece configuration<br>Manometer<br>Adjustable PIP<br>Adjustable PEEP<br>Masks | Wall gas or cylinder<br>T-piece configuration<br>Manometer<br>Adjustable PIP<br>Adjustable PEEP<br>Masks | Wall gas or cylinder<br>T-piece configuration<br>Manometer<br>Adjustable PIP<br>Adjustable PEEP<br>Masks |
| Manometer range | Up to 60 cmH2O<br>Cleared under K032593 | -10 to 80 cm H2O | -20 to 80 cm H2O |
| Manometer accuracy | +/- 3 cm H2O up to 15 cm H₂O<br>+/- 5 cm H2O > 15 cm H₂O | +/- 5% of full scale | Not specified |
| Peak Inspiratory Pressure<br>(PIP) | 0-40 cm H2O @ 15 Lpm | Maximum PIP 45 +/- 5 cm H2O | @ 8 Lpm – 2 to 73 cm H2O<br>@ 10 Lpm – 2 to 80 cm H2O |
| Positive End-Expiratory<br>Pressure (PEEP) | @ 5 Lpm - $<2.1$ cm H2O<br>@ 8 Lpm - $<5.0$ cm H2O<br>@ 10 Lpm - $<7.2$ cm H2O<br>@ 15 Lpm - $<14.6$ cm H2O | @ 5 Lpm - $<5$ cm H2O<br>@ 8 Lpm - $<5$ cm H₂O<br>@ 10 Lpm - $<5$ cm H2O<br>@ 15 Lpm - $<6$ cm H2O | @ 5 Lpm – 1 to 5 cm H2O<br>@ 8 Lpm – 1 to 9 cm H2O<br>@ 10 Lpm – 2 to 15 cm H₂O<br>@ 15 Lpm – 3 to 25 cm H2O |
| Operating time based on<br>e-cylinder | @ 15 Lpm ~ 44 minutes | @ 5 Lpm - 121 minutes<br>@ 10 Lpm - 77 minutes<br>@ 15 Lpm - 54 minutes | 5 to 50 Lpm – 50 minutes |
| Maximum pressure relief | 40 cm H2O @ 15 Lpm | | 40 cm H2O |
| Air / Oxygen mixture | Up to 100% O2 | 21 to 100% O2 | Not specified |
| Maximum gas flow rate | 15 Lpm | 15 Lpm | 15 Lpm |
| Warnings | Proposed<br>Mercury Medical T-Piece<br>Resuscitator (Neo-TeeTM) | Predicate<br>K070210 GE Giraffe and Panda | Predicate<br>K892885 F&P NeoPuff |
| | 1. Incorrect operation of this device can be hazardous.<br>2. Do not use in the presence of flammable materials and ensure that no sources of ignition are present while the device is in use. Fire hazards are possible in oxygen enriched environments.<br>3. Users should release the T-piece (manual control) as soon as the breath is delivered. Failure to do so will result in an extended breath, which may prevent exhalation.<br>4. Do not leave patient unattended when device is in use.<br>5. The recommended gas flow range is from 1-15 LPM. DO NOT EXCEED 15 LPM. | Not available | 1. Ensure no smoking, naked flames or sources of ignition are present while the unit is in use.<br>2. Recommended operating gas flow range 5 to 15 Lpm. Do not attempt to use a higher flow than 15 Lpm. |
Page S1.17
{2}------------------------------------------------
510(k) Summary
.
Page S1.18
{3}------------------------------------------------
510(k) Summary
It is our view that there are no significant differences that affect the safety or effectiveness of the intended device as compared to the
predicate device.
Page S1.19
{4}------------------------------------------------
# 510(k) Summary Page 5 of 5 30-Mar-10
The Mercury Medical T-Piece Resuscitator (Neo-Tee™) is viewed as substantially equivalent to the predicate devices because:
#### Indications -
- Similar to predicate F&P NeoPuff K892885 .
### Technology -
- Similar technology and design -.
- o GE Giraffe K070210 and F&P NeoPuff K892885
#### Materials -
- The materials in patient contact are identical to predicate devices manufactured by . Mercury Medical.
### Environment of Use -
- . Identical to predicate -
- 0 GE Giraffe K070210 and F&P NeoPuff K892885
### Performance Testing -
The following performance / bench tests were performed and shown to demonstrate that the Mercury Medical T-piece Resuscitator (Neo-Tee™) either met the requirements of industry standard, ISO 10651:2006 or were substantially equivalent to the predicate devices.
| Test | Test requirements |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Vomitus Resistance | Must subsequently pass Oxygen Concentration, Inspiratory Resistance,<br>Expiratory Resistance, PEEP, Delivered Volume and Pressure Limitation tests |
| Water Immersion | Resuscitator must continue to function within the tolerances for normal use |
| Oxygen<br>Concentration | Volumetric concentration of oxygen must be > 85% |
| Inspiratory<br>Resistance | Pressure at patient connection port must be > -6 cm H2O with inspiratory airflow of 6<br>L/min |
| Expiratory<br>Resistance | Pressure at patient connection port must be < 6 cm H2O with expiratory airflow of 6<br>L/min |
| PEEP Test | PEEP must be < 2 cm H2O during Oxygen Concentration Test |
| Delivered Volume | Delivered volume must be <20 mL @ C1, R200 and > 3.75 mL @ C10, R20 |
| Pressure Limitation | Pressure must not exceed 60 cm H2O @ 60 L/min |
| Storage/Operating<br>Conditions | Must subsequently pass Delivered Volume and Pressure Limitation tests |
| Drop | Resuscitator must continue to function within the tolerances for normal use |
The (Neo-Tee™) T-piece resuscitator met all the performance requirements as outline above and thus can be found to either comply with ISO 10651:2006 or is equivalent to the predicates.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mercury Medical, Incorporated C/O Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
APR - 3 2010
Re: K093913
Trade/Device Name: Mercury Medical T-Piece Resuscitator (Neo-Tee™) Regulation Number: 21CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: II Product Code: BTL Dated: December 19, 2009 Received: December 22, 2009
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
## Page 2- Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
h11p://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm forthe Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{7}------------------------------------------------
## Indications for Use Statement
Page 1 of 1
K093913 510(k) Number:
Device Name:
Mercury Medical T-Piece Resuscitator (Neo-Tee™)
Indications for Use:
The Neo-Tee™ T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with neonates and infants weighing less than 10kg (22 lb).
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultze
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093913
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.