Puffin Lite Infant Resuscitation System

K182956 · International Biomedical · BTL · Jan 18, 2019 · Anesthesiology

Device Facts

Record IDK182956
Device NamePuffin Lite Infant Resuscitation System
ApplicantInternational Biomedical
Product CodeBTL · Anesthesiology
Decision DateJan 18, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5925
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Puffin Lite Infant Resuscitator is intended to provide the basic equipment required for pulmonary resuscitation of neonatal infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the neonatal infant.

Device Story

Gas-powered emergency resuscitation system for neonatal infants (<10 kg); used in hospitals, delivery suites, nurseries, and ICUs by trained medical professionals. Device inputs include wall gas or cylinder air/oxygen; internal components include medical blender, flowmeters, airway pressure manometer, PIP control, and PEEP control. Device transforms inputs into precise FIO2 and manual ventilation via T-piece circuit. Output is delivered to patient via face mask or ET tube. Healthcare providers monitor airway pressure via manometer and adjust PIP/PEEP/FIO2 controls to provide ventilation. Benefits include precise, guideline-compliant resuscitation support for neonates. Device is a simplified version of predicate with venturi suction and redundant gas inputs removed to reduce footprint.

Clinical Evidence

Bench testing only. Testing verified valve function after vomitus, inspiratory/expiratory resistance, dead space (<7 mL), FIO2 accuracy (within 5%), flow valve accuracy, airway manometer accuracy, VOC levels, and particulate analysis. All tests passed.

Technological Characteristics

Gas-powered manual resuscitator. Components: medical blender, flowmeters, airway pressure manometer, PIP control, PEEP control. Connectivity: None (standalone). Energy source: Pneumatic (wall gas or cylinder). Form factor: Reduced footprint compared to predicate. Materials: Equivalent to predicate (no new materials).

Indications for Use

Indicated for neonatal infants < 10 kg requiring pulmonary resuscitation, including airway establishment, oxygen/air-oxygen delivery, and manual ventilation.

Regulatory Classification

Identification

A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue. January 18, 2018 International Biomedical Amy Pieper Director of Regulatory Affairs 8206 Cross Park Drive Austin, Texas 78754 Re: K182956 Trade/Device Name: Puffin Lite Infant Resuscitation System Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: Class II Product Code: BTL Dated: December 19, 2018 Received: December 19, 2018 Dear Amy Pieper: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, James J. Lee -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182956 #### Device Name Puffin Lite Infant Resuscitation System #### Indications for Use (Describe) The Puffin Lite Infant Resuscitator is intended to provide the basic equipment required for pulmonary resuscitation of neonatal infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the neonatal infant. Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |---------------------------------------------------------------------------------------------------------------------------| | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY | | Submitter Information: | Regulatory Affairs Contact: | |-------------------|------------------------------------------------------------------|--------------------------------| | | International Biomedical | Amy Pieper | | | 8206 Cross Park Drive | Director of Regulatory Affairs | | | Austin, TX 78754 | (512) 873-0033 - phone | | | U.S.A. | (512) 873-9090 - fax | | | Date Summary Prepared: January 8, 2019 | | | | Device Identification: | | | | Trade Name: Puffin Lite Infant Resuscitation System | | | | Common Name: Ventilator, Emergency Gas Powered (Resuscitator) | | | | Regulatory Class: II | | | | Regulation: 868.5925 | | | | Product Code: BTL | | | | Panel: Anesthesiology | | | Predicate Device: | | | | | LifeBorne Infant Resuscitator. International Biomedical, k140707 | | # Device Description: The resuscitation system provides the basic equipment required for pulmonary resuscitation of neonatal infants. The Puffin Lite Infant Resuscitator is a gas powered emergency resuscitation system. It is intended to be used inside the hospital by trained medical professionals to provide precise FIO2 delivery, and manual ventilation as established by resuscitation guidelines to neonatal infants weighing less than 10 kg (22 lb). {4}------------------------------------------------ The resuscitation system includes two medical gas flowmeters, an integrated oxygen blender, airway pressure manometer, peak inspiratory pressure (PIP) control, positive end expiratory pressure (PEEP) control, a gas supply pressure alarm and T-piece resuscitator. # Modifications from the predicate device were: - The integrated venturi suction was removed from the LifeBorne Infant Resuscitation System to create the Puffin Lite Infant Resuscitation System. - . The overall form-factor was reduced (based on the removal of the suction function and the redundant air and oxygen gas inputs) - Changes made to the device did not impact delivery of flow, pressure or waveform delivered to the patient ### Indications for Use: The Puffin Lite Infant Resuscitator is intended to provide the basic equipment required for pulmonary resuscitation of neonatal infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the neonatal infant. ## Functional Description and Technological Characteristics: The Puffin Lite Infant Resuscitation System incorporates the following components for neonatal resuscitation, all of which are unchanged from the predicate device: - . Medical blender to mix air and oxygen to a precise FIO2 (K883038 or K925982, also part of K140707 submission) - Flowmeters for the delivery of oxygen or air/oxygen mixtures (included in K140707 submission) - Peak Inspiratory Pressure (PIP) control to set the maximum pressure delivered during an inspiratory phase of a manual breath (included in K140707 submission) - Peak End Expiratory Pressure (PEEP) control located on a provided T-piece circuit to set the maximum pressure during the expiratory phase of a manual breath (K093913, also part of K140707 submission) - . An airway pressure manometer to monitor both PIP and PEEP airway pressure (included in K140707 submission) # Substantial Equivalence: {5}------------------------------------------------ The Puffin Lite Infant Resuscitation System described in this submission is substantially equivalent to the predicate device, in regard to intended use. The intended use of the Puffin Lite Infant Resuscitation System is unchanged from the predicate device. Changes made to the Puffin Lite versus the Puffin did not impact delivery of flow, pressure or waveform delivered to the patient. The Puffin Lite differs from the Puffin in that the venturi suction and the redundant air and oxygen inputs were eliminated. These changes allow for a smaller footprint that is easier to integrate. The airway portion of the Puffin Lite is identical to the predicate Puffin Infant Resuscitator. Calibration of the Puffin Lite airway is identical to the predicate Puffin device as all output parameters such as flow, peak inspiratory pressure (PIP) and fractional of inspired oxygen concertation (FIO2) are the same. Because these output parameters have not changed, when the interface circuit is attached, flow, PIP and FIO2 delivered to the patient does not change. Additionally, all pressure, flow and volume waveforms for the Puffin Lite are identical to its predicate device given the same compliance and resistance of the lung impacted. | | Proposed<br>Puffin Lite Infant Resuscitator | LifeBorne Infant Resuscitator<br>K140707 | |-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The Puffin Lite Infant Resuscitator is<br>intended to provide the basic<br>equipment required for pulmonary<br>resuscitation of neonatal infants.<br>Pulmonary resuscitation includes<br>practices necessary to establish a clear<br>airway and provide oxygen or<br>air/oxygen mixtures and/or manual<br>ventilation to the neonatal infant. | The LifeBorne Infant Resuscitator is<br>intended to provide the basic<br>equipment required for pulmonary<br>resuscitation of neonatal infants.<br>Pulmonary resuscitation includes<br>practices necessary to establish a clear<br>airway and provide oxygen or air/oxygen<br>mixtures and/or manual ventilation to<br>the neonatal infant. | | Environment for Use | Hospital, delivery suites, nursery, ICU | Hospital, delivery suites, nursery, ICU | | Patient Population | Neonatal Infant < 10 kg | Infant < 10 kg | | Patient Connection | Face mask; ET tube | Face mask; ET tube | | Air/Oxygen Mixture | 21-100% | 21-100% | | Gas Flow Source | Wall gas or cylinder | Wall gas or cylinder | | Manometer Range | -10 to 80 cm H2O | -10 to 80 cm H2O | | Peak Inspiratory<br>Pressure (PIP) | Max 45 +/- 5 cm H2O | Max 45 +/- 5 cm H2O | | Positive End<br>Expiratory Pressure | 0-6 cm H2O | 0-6 cm H2O | {6}------------------------------------------------ | (PEEP) | | | |------------------------------------------|------------------|---------------------------------| | Vacuum Pressure Range | N/A - No Suction | 0 - 150 mmHg, negative pressure | | Maximum gas flow rate | 15 LPM | 15 LPM | | Maximum pressure relief | 55 cm H2O | 55 cm H2O | | Features: Venturi Vacuum Device | N/A - No Suction | Present | | Features: Vacuum gauge | N/A - No Suction | Present | | Features: Integrated Medical Blender | Present | Present | | Features: Flowmeters for delivery of gas | Present | Present | | Features: PIP control | Present | Present | | Features: PEEP control | Present | Present | | Features: Airway pressure manometer | Present | Present | # Performance Testing: Conformance of the Puffin Lite Infant Resuscitation System to performance specifications has been established through bench testing. | TEST | TEST REQUIREMENTS | SUMMARY OF<br>RESULTS | |---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------| | Valve Function<br>after Vomitus | The proper function of the circuit shall be verified within<br>20 seconds of becoming disabled by vomitus. Function is<br>verified by verifying flow valve accuracy. | Passed. | | Inspiratory<br>Resistance | Pressure generated at the patient connection port during<br>expiration should not exceed -5 cmH2O with inspiratory<br>airflow set to 6 L/min. | Passed. | | Expiratory<br>Resistance | Pressure generated at the patient connection port during<br>expiration should not exceed 5 cmH2O with expiratory<br>airflow set to 6 L/min. | Passed. | | Dead Space | The deadspace volume of the T-Piece circuit should be less | Passed. | {7}------------------------------------------------ | | than 7 mL. | | |----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | FIO2 accuracy | The proper function of the FIO2 adjustment knob shall be verified by comparing the FIO2 setting value with the output oxygen concentration. Values shall be within 5%. | Passed. | | Primary and<br>Secondary Flow<br>Valve - Peak Flow | The proper function of the primary and secondary flow valves shall be verified by comparing the flow setting with the actual measured output flow. | Passed. | | Airway Manometer<br>Accuracy | The proper function of the airway manometer shall be verified by comparing the pressure readings with the actual measured output pressure. | Passed. | | VOC Testing | The device should not add volatile organic compounds (VOCs) to the output gas delivered to the patient. | Passed. | | Particulate Analysis | The output of particulate matter sizes 2.5 microns or less are no more than 12 micrograms/cubic meter of air at one atmospheric pressure. | Passed. | The Puffin Lite Infant Resuscitator met all the performance requirements as outlined above and thus can be found to be substantially equivalent to the predicate device. ### Conclusion: The modified Puffin Lite Infant Resuscitator has the following similarities to the previous LifeBorne Infant Resuscitator that already has 510(k) clearance: - The same intended use ● - Use the same operating principle - Incorporate the same basic design - Incorporate equivalent materials (new materials were not introduced with the modification) - No new features were added to the device - The following features were removed from the predicate device to make the new, smaller, device: Integrated Venturi Suction, Redundant air/oxy inputs The overall technology characteristics and performance are unchanged from the predicate device. Evaluation of the risks and performance data based on the differences between the subject and the predicate does not raise any new issues or concerns related to functionality of the device. Therefore, based on the submitted information in this premarket notification, the candidate device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...