K142096 · Mercury Medical · BTL · Oct 17, 2014 · Anesthesiology
Device Facts
Record ID
K142096
Device Name
T-PIECE RESUSCITATOR
Applicant
Mercury Medical
Product Code
BTL · Anesthesiology
Decision Date
Oct 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5925
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22lb).
Device Story
Gas-powered, manually operated emergency resuscitator; provides respiratory support via face mask, supraglottic airway, or endotracheal tube. Device features T-piece configuration, integrated manometer, and adjustable Peak Inspiratory Pressure (PIP) and Positive End-Expiratory Pressure (PEEP). Includes pressure relief valve for safety. Operated by trained personnel in hospital, sub-acute, or pre-hospital (EMS) environments. User manually delivers ventilation (up to 60 BPM) using wall gas or cylinder source. Output allows healthcare providers to monitor airway pressure and deliver controlled respiratory support, potentially stabilizing patients with respiratory distress. Benefits include portable, adjustable pressure delivery for patients >10kg.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 10651-5 standards, including vomitus resistance, water immersion, oxygen concentration, inspiratory/expiratory resistance, PEEP testing, delivered volume, pressure limitation, storage/operating conditions, and drop testing.
Technological Characteristics
Gas-powered T-piece resuscitator. Materials identical to predicate NeoTee. Features integrated manometer (K954486), pressure relief valve, and adjustable PIP/PEEP. Patient connector: 15mm ID / 22mm OD. Standards: ISO 10651-5. Operational gas flow: 0-35 lpm. Pressure relief: 60 cm H2O (factory set 40 cm H2O).
Indications for Use
Indicated for emergency respiratory support in patients weighing >10kg (>22 lbs) via face mask, supraglottic airway, or endotracheal tube. Intended for use in hospital, sub-acute facilities, and pre-hospital (EMS) settings by trained personnel.
Regulatory Classification
Identification
A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 17, 2014
Mercury Medical, Incorporated C/O Mr. Paul Dryden President ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134-2958
Re: K142096
Trade/Device Name: T-Piece Resuscitator Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: Class II Product Code: BTL Dated: August 14, 2014 Received: August 15, 2014
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Tejashri Purohit-Sheth, M.D." in a stylized font. The text is black and appears to be centered. The letters are thick and have a slight curve to them, giving the text a formal appearance.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K142096
Device Name
T-piece Resuscitator Indications for Use (Describe)
> The T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22lb).
Environment of Use - Hospital, sub-acute facilities, and pre-hospital (EMS)
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
PSC Publishing Services (301) 443-6740 EF
Page 1 of 2
{3}------------------------------------------------
#### 510(k) Summary Page 1 of 6 14-Oct-14
| Mercury Medical, Inc.<br>11300 - 49th St. North<br>Clearwater, FL 33762-4807 | Tel – (727) 573-0088<br>Fax – (727) 571-3922 |
|------------------------------------------------------------------------------|------------------------------------------------------------|
| Official Contact: | Jeff Ratner - VP Engineering and Quality Assurance |
| Proprietary or Trade Name: | T-Piece Resuscitator |
| Common/Usual Name: | Powered emergency ventilator |
| Classification Code /Name: | BTL - powered emergency ventilator<br>CFR 868.5925 |
| Device: | T-Piece Resuscitator |
| Predicate Devices: | NeoForce – ISPIRA - K092085<br>Mercury - NeoTee™ - K093913 |
#### Device Description:
The Mercury T-Piece Resuscitator is manually operated, gas powered resuscitator for use with patients greater than 10 kg (>22 lb).
It is a simple T-piece, with a manometer and the ability to adjust Peak Inspiratory Pressure (PIP) and Positive End-Expiratory Pressure (PEEP). It incorporates a pressure relief valve for excessive pressure.
The T-Piece Resuscitator can be connected to the patient via a face mask, Supraglottic airway or endotracheal tube.
#### Indications for Use:
The T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22 lbs.).
Environment of Use: Hospital, sub-acute facilities, and pre-hospital (EMS)
#### Substantial Equivalence:
The Mercury T-Piece Resuscitator is a similar design T-piece resuscitation system to our own reference device, Mercury NeoTee™ (K093913) and with the NeoForce ISPIRA (K092085) for the indications for use and patient population. We will provide information which demonstrates that the proposed device is substantially equivalent to the predicates.
Table 5.1 lists the similarities and differences of the predicate. NeoForce ISPIRA (K092085) and the proposed device. While Table 5.2 compares the reference device, Mercury NeoTeeTM (K093913) which has the identical design of patient circuit but is indicated for neonate / infant populations and a more limited environment of use.
{4}------------------------------------------------
#### 510(k) Summary Page 2 of 6 14-Oct-14
#### Summary of substantial equivalence Table 5.1 - Table of the Similarities and Differences of Proposed Device vs. the Predicate
| | Proposed<br>Mercury Medical T-Piece<br>Resuscitator | Predicate<br>NeoForce ISPIRA<br>K092085 |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The T-Piece Resuscitator is a powered<br>emergency ventilator intended to<br>provide emergency respiratory support<br>by means of a face mask or a tube<br>inserted into a patient's airway. It is<br>intended for use with patients greater<br>than 10kg (>22 lbs.). | The NeoForce ISPIRA Resuscitation<br>System is a manually operated, gas<br>powered resuscitator intended for<br>controlled and accurate pulmonary<br>resuscitation and emergency respiratory<br>support of pediatric and adult patients<br>with a body weight of more than 22 lbs.<br>(10 kg) in the hospital, pre-hospital<br>(EMS), and sub-acute / alternate site<br>facility environments via face mask,<br>laryngeal mask or endotracheal tube.<br>The device is also intended to provide<br>CPAP to spontaneously breathing<br>patients in the hospital, pre-hospital<br>(EMS) and sub-acute / alternate site<br>facility environments via face mask,<br>laryngeal mask or endotracheal tube. |
| Environment of use | Hospital, sub-acute facilities, and pre-<br>hospital (EMS) | Hospital, pre-hospital (EMS), and sub-<br>acute / alternate site facility<br>environments |
| Patient Population | Patients greater than 10 Kg (>22 lbs.) | Patients greater than 10 Kg (>22 lbs.) |
| Prescriptive | Persons trained in resuscitation | Persons trained in resuscitation |
| Patient connection | Face mask<br>Supraglottic airway<br>Endotracheal Tube | Face mask<br>Supraglottic airway (LMA)<br>Endotracheal Tube |
| Features and Performance Characteristics | | |
| Gas flow provided by | Wall gas or cylinder | Wall gas or cylinder |
| Components | Hand-piece which is a T-piece configuration with Manometer Corrugated tubing and connector Control Panel which allows for PIP and PEEP adjustments Oxygen tubing Masks (optional) | T-piece configuration Manometer part of controller Corrugated tubing and connectors Controller which allows for PIP, PEEP, and CPAP adjustments Tubing to monitor pressure Oxygen delivered via corrugated hose Mask has some CPAP features |
| Ventilation Frequency | Manually delivered by the user up to<br>60 BPM | Manually delivered by the user up to<br>60 BPM |
| Maximum pressure relief | 60 cm H2O<br>Factory set at 40 cm H2O | 5-80 cm H2O.<br>Factory set at 40 cm H2O |
| | Proposed<br>Mercury Medical T-Piece Resuscitator | Predicate<br>NeoForce ISPIRA |
| | | K092085 |
| Delivered Pressure | Up to 60 cm H2O | 5 to 80 cm H2O |
| CPAP | Not offered but offers PEEP<br>Identical to K093913 – Mercury NeoTee™ | 5 -15 cm H2O@ 8-12 lpm flow |
| Delivered Volume | 60 - 700 ml with flow rates between<br>5 - 35 lpm | 190 - 675 ml with flow rates between<br>4 - 36 lpm |
| Inspiratory resistance | 2 cm H2O at minimum PEEP setting<br>@ 60 lpm | Less than - 5 cm H2O |
| Expiratory resistance | 2.4 cm H2O at minimum PEEP setting<br>@ 60 lpm | Less than + 5 cm H2O |
| Oxygen concentration with<br>optional blender | 21 - 98% based upon blender setting | 21 - 100% based upon blender setting |
| Manometer range | Up to 60 cmH2O<br>Cleared under K954486 | -20 to 80 cm H2O |
| Manometer accuracy | +/- 3 cm H2O up to 15 cm H2O<br>+/- 5 cm H2O > 15 cm H2O | +/- 2 cm H2O |
| Peak Inspiratory Pressure<br>(PIP) | 0-60 cm H2O | 5-80 cm H2O |
| Positive End-Expiratory<br>Pressure (PEEP) | 0 to 60 cm H2O | Not offered. Use CPAP instead |
| Operational gas flow rate | 0 - 35 lpm | 0 - 60 lpm |
| Operational time with 400 L<br>cylinder | @ 35 lpm - 11 minutes | @ 36 lpm - 10 minutes |
| Dead space of circuit | < 7.5 ml | < 15 ml |
| Operating Temperature range | -18 to 60°C | 10 to 40°C |
| Storage Temperature | -40 to 50°C | -20 to 60°C |
| Immersion resistance | Temporary submersion of circuit does not<br>affect functionality | Temporary submersion of circuit does<br>not affect functionality |
| Reusable | No | Yes |
| Standards | ISO 10651.5 | ISO 10651.5 |
{5}------------------------------------------------
## 510(k) Summary
Page 3 of 6
14-Oct-14
{6}------------------------------------------------
### 510(k) Summary Page 4 of 6
14-Oct-14
# Table 5.2 - Table of the Similarities and Differences of Proposed
vs.
| | Proposed<br>Mercury Medical T-Piece<br>Resuscitator | Reference<br>Mercury NeoTee™<br>K093913 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The T-Piece Resuscitator is a powered<br>emergency ventilator intended to<br>provide emergency respiratory support<br>by means of a face mask or a tube<br>inserted into a patient's airway. It is<br>intended for use with patients greater<br>than 10kg (>22 lbs.). | The Mercury Medical T-Piece<br>Resuscitator is a powered emergency<br>ventilator intended to provide<br>emergency respiratory support by means<br>of a face mask or a tube inserted into a<br>patient's airway. It is intended for use<br>with neonates and infants weighing less<br>than 10kg (<22 lbs.). |
| Environment of use | Hospital, sub-acute facilities, and pre-<br>hospital (EMS) | Hospital, delivery suites, nursery, ICU |
| Patient Population | Patients greater than 10 Kg (>22 lbs.) | Patients less than 10 Kg (<22 lbs.) |
| Prescriptive | Persons trained in resuscitation | Persons trained in infant / neonate<br>resuscitation |
| Patient connection | Face mask<br>Supraglottic airway<br>Endotracheal Tube | Face mask<br>Endotracheal Tube |
| Features and Performance Characteristics | | |
| Gas flow provided by | Wall gas or cylinder | Wall gas or cylinder |
| Components | Hand-piece which is a T-piece<br>configuration with Manometer Corrugated tubing and connector Control Panel which allows for PIP<br>and PEEP adjustments Oxygen tubing Masks (optional) | Hand-piece which is a T-piece<br>configuration with Manometer Corrugated tubing and connector Control Panel which allows for PIP<br>and PEEP adjustments Oxygen tubing Masks (optional) |
| Maximum pressure relief | 60 cm H2O | 40 cm H2O |
| Delivered Pressure | Up to 60 cm H2O | Up to 40 cm H2O |
| Delivered Volume | 60 - 700 ml with flow rates between<br>5 - 35 lpm | 15-200 ml |
| Manometer range | Up to 60 cmH2O (K954486) | Up to 60 cmH2O ( K954486) |
| Manometer accuracy | +/- 3 cm H2O up to 15 cm H2O<br>+/- 5 cm H2O > 15 cm H2O | +/- 3 cm H2O up to 15 cm H2O<br>+/- 5 cm H2O > 15 cm H2O |
| Inspiratory resistance | 2 cm H2O at minimum PEEP setting<br>@ 60 lpm | 0.4 cm H2O at minimum PEEP setting<br>@ 6 lpm |
| Expiratory resistance | 2.4 cm H2O at minimum PEEP setting<br>@ 60 lpm | 0.5 cm H2O at minimum PEEP setting<br>@ 6 lpm |
| Peak Inspiratory Pressure<br> | 0-60 cm H2O | 0-40 cm H2O @ 15 Lpm |
| (PIP) | | |
| | Proposed<br>Mercury Medical T-Piece<br>Resuscitator | Reference<br>Mercury NeoTee™<br>K093913 |
| Positive End-Expiratory | 0-60 cm H2O | 2 cmH₂O at 2.5 lpm up to |
| Pressure (PEEP) | | 15 cm H2O @ 15 lpm |
| Operational time with 400 L | @ 35 lpm – 11 minutes w/ 400 L | @ 15 lpm – 26 minutes w/ 400L |
| cylinder | @ 35 lpm ~ 19 min with 660 L cylinder | @ 15 lpm - 44 minutes w/ 660L |
| Operational gas flow rate | 0 – 35 lpm | 0 – 15 lpm |
| Dead space of circuit | < 7.5 ml | < 3.48 ml |
{7}------------------------------------------------
#### 510(k) Summary Page 5 of 6 14-Oct-14
#### Substantial Equivalence Rationale
The Mercury T-Piece Resuscitator is viewed as substantially equivalent to the predicate devices because:
Indications - Identical indications for use to our reference device, NeoTee™ (K093913). Similar to the predicate NeoForce (K092085), which are is a gas powered emergency ventilator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. The environments of use are different as the proposed device as the predicate was specifically designed for neonate use.
Discussion - The differences in indications for use between the reference device, NeoTeeTM (K093913), patient population and environment differences are covered with the NeoForce™ (K092085).
Technology - The technology is identical to the reference NeoTee™ (K093913) which includes all the features and components.
Discussion -The only difference between the proposed T-piece and the reference NeoTee™ is the size of the PEEP cap to accommodate high gas flow for adults. The patient connector for the proposed device is 15mm ID / 22mm OD to allow for connection to standard face mask with 22 mm ID fittings as well as endotracheal tubes with 15 mm OD fittings vs. the reference NeoTee™ (K093913) only has a 15mm ID connector which allows it to connect to endotracheal tubes and face mask with 15 mm OD fittings. These differences do not raise any new safety concerns.
Materials - The materials in patient contact are identical to the NeoTee™ device manufactured by Mercury Medical.
Discussion - The materials are identical and the patient contact is also identical to the reference NeoTee™ (K093913). The manufacturing, handling and processing are identical to the reference device and we can make a certification claim.
#### Non-clinical Testing
We performed a number of tests to demonstrate the performance of the proposed device to the reference or predicate devices.
{8}------------------------------------------------
#### 510(k) Summary Page 6 of 6 14-Oct-14
Bench testing - We performed applicable testing per ISO 10651-5 for gas powered resuscitators. These tests included:
- Vomitus Resistance
- . Water Immersion
- Oxygen Concentration
- Inspiratory Resistance .
- Expiratory Resistance
- PEEP Test ●
- Delivered Volume
- Pressure Limitation ●
- Storage/Operating Conditions ●
- . Drop
Discussion - We have demonstrated that the T-Piece Resuscitator meets the performance and design specifications requirements as outlined in requirements and specifications as outlined in ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators.
#### Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.