K093775 · Romedex International Srl · LJS · Jul 15, 2010 · General Hospital
Device Facts
Record ID
K093775
Device Name
EVGUIDE TIP LOCATION SYSTEM
Applicant
Romedex International Srl
Product Code
LJS · General Hospital
Decision Date
Jul 15, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K093775 · Jul 15, 2010
EVGUIDE TIP LOCATION SYSTEM
Romedex International Srl
Hospital clinical records; Routine clinical practice data
Clinical data from 350 patients in European hospital centers were used to demonstrate the safety and efficacy of the Sapiens TLS in routine clinical settings, including various catheter types and clinical procedures.
Clinical performance; Routine clinical practice; European hospital centers
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
European Clinical Experience; Retrospective/Observational (Routine clinical use)
350 adult patients (ages 19-96) undergoing central venous catheter placement; Sample Size: 350; Number of Sites: 5
Not applicable for this study
Catheter tip placement confirmation
Indications for Use
The Sapiens™ Tip Location System (TLS) is indicated for guidance and positioning of central venous catheters such as PICCs. CVCs, implantable ports, and hemodialysis catheters. The Sapiens™ TLS provides real-time catheter tip location information by using the patient's cardiac electrical activity. Sapiens™ TLS is indicated for use as an alternative method to chest X-ray and fluoroscopy for central venous catheter tip placement confirmation in adult patients. Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P wave as in atrial fibrillation, atrial flutter, severe tachycardia, and pacemaker driven rhythm, and in central venous catheterization procedures performed through femoral or saphaneous vein access which change the presentation of the P wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Device Story
Sapiens™ TLS assists in central venous catheter (CVC) placement by tracking the catheter tip via cardiac electrical activity. System inputs include skin ECG signals and endovascular electrograms acquired via a sterile electrical adaptor connected to a catheter stylet, guidewire, or saline-filled lumen. The system processes these signals to display real-time ECG waveforms on a laptop, allowing clinicians (physicians/nurses) in ORs, clinics, or bedside settings to guide the catheter tip. Output waveforms provide visual confirmation of tip location, potentially reducing reliance on chest X-ray or fluoroscopy. Clinicians use the displayed P-wave morphology to determine optimal tip positioning. The system includes a sterile electrical adaptor, ECG module/cable, and laptop software; an optional label printer documents the procedure.
Clinical Evidence
Clinical data from 362 adult patients (ages 19-96) across five hospital centers. Study included a prospective, multicenter, non-controlled study (n=332) and a human factors/usability study (n=30). Primary endpoint was confirmation of catheter tip placement at the desired location, verified by chest X-ray or fluoroscopy. Results showed 97% confirmation rate. No adverse events or complications reported. Data supports safety and efficacy for PICCs, CVCs, implantable ports, and hemodialysis catheters.
Technological Characteristics
System components: sterile electrical adaptor, ECG module/cable, laptop, optional label printer. Adaptor features insulated, conductive wire with distal alligator clip. Sensing principle: cardiac electrical activity (ECG) via direct stylet/guidewire connection or saline column method. Connectivity: laptop-based display. Software: proprietary Sapiens™ TLS software. Sterilization: sterile adaptor.
Indications for Use
Indicated for guidance and positioning of central venous catheters (PICCs, CVCs, implantable ports, hemodialysis catheters) in adult patients. Contraindicated/limited use in patients with cardiac rhythm alterations (atrial fibrillation, flutter, tachycardia, pacemaker-driven rhythm) or femoral/saphenous vein access, where P-wave presentation is altered; requires additional confirmation methods in these cases.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
.
## 510(k) Summary
| Proprietary Name: | Sapiens™ Tip Location System also known as evGuide™ Tip Location System |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade name: | Sapiens™ Tip Location System (TLS) |
| Product<br>Classification: | Class II, 21 CFR §880.5970<br>LJS - Accessory to Percutaneous, Implanted, Long-Term Intravascular Catheters<br>General Hospital |
| Applicant name: | Romedex International Srl 58 Aleea Arubium, Bucharest, 022944 Romania,<br>+1.650.209.4838, info@romedex.com, www.romedex.com |
| Contact person: | Sorin Grunwald Ph.D., MBA, 175 Colorado Ave, Palo Alto, CA 94303,<br>sorin@romedex.com, Tel: +1.650.209.4838, Fax: +1.650.887.0348 |
| Preparation Date: | June 4, 2010 |
| Predicate Devices: | K091324 - Sherlock 3CG Tip Positioning System<br>K032613 - Transvenous Pacemaker Placement Assist Device<br>K973371 - Conduction Anesthesia Kit<br>K843263 - Arrow-Johans ECG Adaptor |
| Indications for Use: | The Sapiens™ Tip Location System (TLS) is indicated for guidance and positioning of<br>central venous catheters such as PICCs. CVCs, implantable ports, and hemodialysis<br>catheters. The Sapiens™ TLS provides real-time catheter tip location information by<br>using the patient's cardiac electrical activity. Sapiens™ TLS is indicated for use as an<br>alternative method to chest X-ray and fluoroscopy for central venous catheter tip<br>placement confirmation in adult patients. |
| | Limiting but not contraindicated situations for this technique are in patients where<br>alterations of cardiac rhythm change the presentation of the P wave as in atrial<br>fibrillation, atrial flutter, severe tachycardia, and pacemaker driven rhythm, and in central<br>venous catheterization procedures performed through femoral or saphaneous vein access<br>which change the presentation of the P wave. In such patients, who are easily identifiable<br>prior to central venous catheter insertion, the use of an additional method is required to<br>confirm catheter tip location. |
| Device Description: | The Sapiens™ TLS consists of the following clements: sterile electrical adaptor, ECG<br>module and cable, laptop running Sapiens™ TLS software, label printer (optional), and<br>remote control (optional). A stylet or a guidewire inserted in a central venous catheter can<br>be connected to the Sapiens™ TLS system via the Sapiens™ TLS Electrical Adaptor<br>establishing a direct electrical connection to the catheter distal tip for ECG signal<br>measurement. A different ECG signal measurement method - the column of saline<br>method - can be used by connecting the Sapiens™ TLS Electrical Adaptor to the Arrow-<br>Johans Adaptor, by connecting the Arrow-Johans Adaptor to any central venous catheter<br>and by injecting saline solution into the catheter lumen through the Arrow-Johans<br>Adaptor, thus establishing electrical conductivity to the distal tip of the catheter. When<br>the central venous catheter or its associated stylet or guidewire is connected to Sapiens™<br>TLS, the Sapiens™ TLS laptop screen displays skin ECG signals and endovascular<br>electrograms acquired at the location of the distal tip of the catheter. The waveforms<br>provided by Sapiens™ TLS can be used for guiding and positioning of the central venous<br>catheter. These ECG waveforms can be printed using an optional label printer to<br>document the catheter tip location for the patient's file. |
| Bench Top Safety &<br>Performance Tests: | Verification and validation tests have been performed in accordance with Design<br>Controls per 21 CFR §820.30. |
| | Bench top testing has been performed side-by-side with available predicate devices<br>which has demonstrated substantial equivalence of Sapiens™ TLS to the respective<br>predicate devices. The following tests were performed: a) electrical impedance testing<br>and b) ECG waveform accuracy tests. |
| Summary of<br>Non-clinical Data: | Non-clinical studies were performed which have demonstrated safety and efficacy of<br>Sapiens™ TLS using Electrical Adaptor, good correlation between bench top and in-vivo<br>data, and substantial equivalence with predicate devices. The following tests were<br>performed: a) ECG waveform accuracy comparison with a commercially available ECG<br>monitor, b) ECG waveform accuracy comparison with the Conduction Anesthesia Kit<br>(K973371) and c) system usability and validation testing. |
| Summary of<br>Clinical Data: | To date, Sapiens™ TLS has been used in Europe for central venous catheter<br>guidance and positioning in five major hospital centers on more than 350 adult patients<br>(ages 19-96) of both genders for placing several types of central venous catheters: PICCs,<br>CVCs, implantable ports, hemodialysis catheters, and tunneled catheters. Several types of<br>users including nurses and physicians have used Sapiens™ TLS for different clinical<br>procedures, e.g., oncology, anesthesia, patient monitoring in the ICU, hemodialysis in<br>different clinical settings: in the operating room, in outpatient clinics and at the bedside. Side-by-side comparisons with available predicate devices were performed which demonstrated substantial equivalence of Sapiens™ TLS to the respective predicate devices. In the two analyzed subsets of 362 patients (332 patients from a prospective, multicenter, non-controlled study and 30 patients form a human factors/ usability study), the catheter tip placement using Sapiens™ TLS at the desired location was confirmed with chest X-ray or fluoroscopy in 97% of the cases. No adverse events or complications have occurred. |
| Summary of<br>Technological<br>Characteristics<br>Compared to<br>Predicate Devices | The subject Sapiens™ TLS Electrical Adaptor combines design features, materials and<br>technological characteristics of marketed predicate devices including the Transvenous<br>Pacemaker Placement Assist Device (K032613) and Conduction Anesthesia Kit<br>(K973371) such as: a) a sterile, insulated, electrically conductive wire of very low<br>electrical resistance; b) a distal end alligator clip to connect to stylets and guidewires;<br>c) a proximal end connector which can be connected to an ECG cable or to an ECG<br>connection switch. When compared to the Conduction Anesthesia Kit, the Sapiens™ TLS<br>Electrical Adaptor has simplified the design of the device by removing the switch box.<br>When compared to Transvenous Pacemaker Placement Assist Device, the Sapiens™ TLS<br>Electrical Adaptor has simplified the design of the device by removing the connection<br>box. The Sapiens™ TLS Electrical Adaptor may be connected to the ECG pin of the predicate Arrow-Johans Adaptor (K843263) using an ECG cable which allows for the saline conduction method of ECG measurement. Use of the Sapiens™ TLS Electrical Adaptor with the Arrow-Johans Adaptor does not require any modifications of design features, materials, or technological characteristics of the marketed predicate device. Additionally, the subject Sapiens™ TLS System combines design features, components and technological characteristics of the predicate device Sherlock 3CG Tip Positioning System (K091324) but uses only cardiac electrical signal detection to provide real-time catheter tip location information. The subject Sapiens™ System does not use a passive |
4
{1}------------------------------------------------
.
{2}------------------------------------------------
Any differences between technological features of the subject and predicate devices do not raise new questions of safety or efficacy of the subject Sapiens™ TLS device.
Summary of Substantial Equivalence: The Sapiens™ TLS has the same intended use and similar indications for use as the commercially available Sherlock 3CG Tip Positioning System (K091324), Transvenous Pacemaker Placement Assist Device (K032613). Conduction Anesthesia Kit (K973371), and Arrow-Johans ECG Adaptor (K843263). Additionally. clinical and non-clinical performance testing has demonstrated that any differences in technological characteristics do not raise new issues of safety or effectiveness when compared to the aforementioned predicate devices. Therefore, the Sapiens™ TLS meets the requirements for substantial equivalence to the referenced predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white, giving it a formal and official appearance.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
JUL 28 2010
Dr. Sorin Grunwald, Ph.D. US FDA Agent for Romedex International, Srl 175 Colorado Avenue Palo Alto, California 94303
Re: K093775
Trade/Device Name: Sapiens™ TLS Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: June 5, 2010 Received: June 7, 2010
Dear Dr. Grunwald:
This letter corrects our substantially equivalent letter of July 15, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
{4}------------------------------------------------
Page - Dr. Grunwald
and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## SECTION 4: INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K093775
Not known at this time
Sapiens TM TLS Device Name: _
Indications for Use:
The Sapiens™ Tip Location System (TLS) is indicated for guidance and positioning of central venous catheters such as PICCs, CVCs, implantable ports, and hemodialysis catheters. The Sapiens TM TLS provides real-time catheter tip location information by using the patient's cardiac electrical activity. Sapiens TLS is indicated for use as an alternative method to chest X-ray and fluoroscopy for central venous catheter tip placement confirmation in adult patients.
Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P wave as in atrial fibrillation, atrial flutter, severe tachycardia, and pacemaker driven rhythm, and in central venous catheterization procedures performed through femoral or saphaneous vein access which change the presentation of the P wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page of (Posted November 13, 2003)
RUC. Ch 7/20/10
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093775
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.